The DOI matches its Crossref record; 32 of the 32 references it cites were found in a registry or an index.

Noise correlation in CBCT projection data and its application for noise reduction in low‐dose CBCT

Abstract

Purpose: To study the noise correlation properties of cone‐beam CT (CBCT) projection data and to incorporate the noise correlation information to a statistics‐based projection restoration algorithm for noise reduction in low‐dose CBCT. Methods: In this study, the authors systematically investigated the noise correlation properties among detector bins of CBCT projection data by analyzing repeated projection measurements. The measurements were performed on a TrueBeam onboard CBCT imaging system with a 4030CB flat panel detector. An anthropomorphic male pelvis phantom was used to acquire 500 repeated projection data at six different dose levels from 0.1 to 1.6 mAs per projection at three fixed angles. To minimize the influence of the lag effect, lag correction was performed on the consecutively acquired projection data. The noise correlation coefficient between detector bin pairs was calculated from the corrected projection data. The noise correlation among CBCT projection data was then incorporated into the covariance matrix of the penalized weighted least‐squares (PWLS) criterion for noise reduction of low‐dose CBCT. Results: The analyses of the repeated measurements show that noise correlation coefficients are nonzero between the nearest neighboring bins of CBCT projection data. The average noise correlation coefficients for the first‐ and second‐order neighbors are 0.20 and 0.06, respectively. The noise correlation coefficients are independent of the dose level. Reconstruction of the pelvis phantom shows that the PWLS criterion with consideration of noise correlation (PWLS‐Cor) results in a lower noise level as compared to the PWLS criterion without considering the noise correlation (PWLS‐Dia) at the matched resolution. At the 2.0 mm resolution level in the axial‐plane noise resolution tradeoff analysis, the noise level of the PWLS‐Cor reconstruction is 6.3% lower than that of the PWLS‐Dia reconstruction. Conclusions: Noise is correlated among nearest neighboring detector bins of CBCT projection data. An accurate noise model of CBCT projection data can improve the performance of the statistics‐based projection restoration algorithm for low‐dose CBCT.

Overview

Integrity signals

Registry

Crossref · 10.1118/1.4865782

Cites

32 of 32 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 32 references

Research works printed but not linked: 1

  1. Hsieh J., 2003, Computer Tomography, Principles, Design, Artifacts, and Recent Advances

Cited by

See all 48

Metadata

Authors
Zhang, H., Ouyang, L., Ma, J., Huang, J., Chen, W., & Wang, J.
Affiliations
  • Southern Medical University (CN)
  • The University of Texas Southwestern Medical Center (US)
Corresponding author
  • Ma, Jianhua
  • Wang, Jing
Venue
Medical Physics
Publisher
Wiley
Published
2014-02-19
Volume
41
Issue
3
Pages
031906-031906
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2014598003
Cited by
47
References
32
Citations by year
2026: 1 · 2025: 3 · 2024: 4 · 2023: 3 · 2022: 2 · 2021: 1 · 2020: 4 · 2019: 6 · 2018: 8 · 2017: 3 · 2016: 4 · 2015: 5 · 2014: 2
FWCI
3.65
Type
journal article
Language
en
Open access
closed
Licence
Locations
  • Medical Physics (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • NNSFC (81371544, 81000613, 81101046)
  • National Natural Science Foundation of China (81371544, 81000613, 81101046)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Image and Signal Denoising Methods
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
Keywords
cone-beam computed tomography, noise correlation, low-dose CT, noise reduction, penalized weighted least squares, covariance matrix, noise models, pelvis phantom
MeSH terms
Algorithms, Anthropometry, Humans, Image Processing, Computer-Assisted, Male, Radiation Dosage, Radiographic Image Enhancement, Radiographic Image Interpretation, Computer-Assisted, Reproducibility of Results, Models, Statistical, Poisson Distribution, Least-Squares Analysis, Phantoms, Imaging, Cone-Beam Computed Tomography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Noise correlation in CBCT projection data and its application for noise reduction in low‐dose CBCT. https://app.cochat.ai/papers/10.1118/1.4865782
The paper
Zhang, H., Ouyang, L., Ma, J., Huang, J., Chen, W., & Wang, J. (2014). Noise correlation in CBCT projection data and its application for noise reduction in low‐dose CBCT. Medical Physics. https://doi.org/10.1118/1.4865782

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.