Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation
Open accessAbstract
The addition of accurate system modeling in PET image reconstruction results in images with distinct noise texture and characteristics. In particular, the incorporation of point spread functions (PSF) into the system model has been shown to visually reduce image noise, but the noise properties have not been thoroughly studied. This work offers a systematic evaluation of noise and signal properties in different combinations of reconstruction methods and parameters. We evaluate two fully 3D PET reconstruction algorithms: (1) OSEM with exact scanner line of response modeled (OSEM+LOR), (2) OSEM with line of response and a measured point spread function incorporated (OSEM+LOR+PSF), in combination with the effects of four post-reconstruction filtering parameters and 1-10 iterations, representing a range of clinically acceptable settings. We used a modified NEMA image quality (IQ) phantom, which was filled with 68Ge and consisted of six hot spheres of different sizes with a target/background ratio of 4:1. The phantom was scanned 50 times in 3D mode on a clinical system to provide independent noise realizations. Data were reconstructed with OSEM+LOR and OSEM+LOR+PSF using different reconstruction parameters, and our implementations of the algorithms match the vendor's product algorithms. With access to multiple realizations, background noise characteristics were quantified with four metrics. Image roughness and the standard deviation image measured the pixel-to-pixel variation; background variability and ensemble noise quantified the region-to-region variation. Image roughness is the image noise perceived when viewing an individual image. At matched iterations, the addition of PSF leads to images with less noise defined as image roughness (reduced by 35% for unfiltered data) and as the standard deviation image, while it has no effect on background variability or ensemble noise. In terms of signal to noise performance, PSF-based reconstruction has a 7% improvement in contrast recovery at matched ensemble noise levels and 20% improvement of quantitation SNR in unfiltered data. In addition, the relations between different metrics are studied. A linear correlation is observed between background variability and ensemble noise for all different combinations of reconstruction methods and parameters, suggesting that background variability is a reasonable surrogate for ensemble noise when multiple realizations of scans are not available.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536
Cited by
A Systematic Review of the Factors Affecting Accuracy of SUV Measurements
Adams, M. C., Turkington, T. G., Wilson, J. M., & Wong, T. Z. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.10.4923 · Cited by 662
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 380
Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls
Rahmim, A., Qi, J., & Sossi, V. (2013) · Medical Physics · doi:10.1118/1.4800806 · Cited by 357
Akamatsu, G., Ishikawa, K., Mitsumoto, K., Taniguchi, T., Ohya, N., Baba, S., Abe, K., & Sasaki, M. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.112.103861 · Cited by 206
Rausch, I., Cal-González, J., Dapra, D., Gallowitsch, H. J., Lind, P., Beyer, T., & Minear, G. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0132-1 · Cited by 133
Metadata
- Authors
- Tong, S., Alessio, A. M., & Kinahan, P. E.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2010-02-11
- Volume
- 55
- Issue
- 5
- Pages
- 1453-1473
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2062405372
- Cited by
- 190
- References
- 1
- FWCI
- 12.29
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. https://app.cochat.ai/papers/10.1088/0031-9155/55/5/013
- The paper
- Tong, S., Alessio, A. M., & Kinahan, P. E. (2010). Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/55/5/013
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.