Physical Performance of the new hybrid PET/CT Discovery-690

Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., Gilardi, M. C.·2011·Medical Physics

Abstract

PURPOSE: The aim of this work was the assessment of the physical performance of the new hybrid PET∕CT system: Discovery-690. METHODS: The Discovery-690 combines a lutetium-yttrium-orthosilicate (LYSO) block detector designed PET tomograph with a 64-slice CT scanner. The system is further characterized by a dedicated powerful computing platform implementing fully 3D-PET iterative reconstruction algorithms. These algorithms can account for time of flight (TOF) information and∕or a 3D model of the PET point spread function (PSF). PET physical performance was measured following NEMA NU-2-2007 procedures. Furthermore, specific tests were used: (i) to measure the energy and timing resolution of the PET system and (ii) to evaluate image quality, by using phantoms representing different clinical conditions (e.g., brain and whole body). Data processing and reconstructions were performed as required by standard procedures. Further reconstructions were carried out to evaluate the performance of the new reconstruction algorithms. In particular, four algorithms were considered for the reconstruction of the PET data: (i) HD = standard configuration, without TOF and PSF, (ii) TOF = HD + TOF, (iii) PSF = HD + PSF, and (iv) TOFPSF = HD + TOF + PSF. RESULTS: The transverse (axial) spatial resolution values were 4.70 (4.74) mm and 5.06 (5.55) mm at 1 cm and 10 cm off axis, respectively. Sensitivity (average between 0 and 10 cm) was 7.5 cps∕kBq. The noise equivalent count rate (NECR) peak was 139.1 kcps at 29.0 kBq∕ml. The scatter fraction at the NECR peak was 37%. The correction accuracy for the dead time losses and random event counts had a maximum absolute error below the NECR peak of 2.09%. The average energy and timing resolution were 12.4% and 544.3 ps, respectively. PET image quality was evaluated with the NEMA IEC Body phantom by using four reconstruction algorithms (HD, TOF, PSF, and TOFPSF), as previously described. The hot contrast (after 3 iterations and for a lesion∕background activity ratio of 4:1) for the spheres of 10, 13, 17, and 22 mm was (HD) 29.8, 45.4, 55.4, and 68.1%; (TOF) 39.9, 53.5, 62.7, and 72.2%; (PSF) 28.3, 47.3, 60.4, and 71.8%; (TOFPSF) 43.8, 62.9, 70.6, and 76.4%. The cold contrast for the spheres of 28 and 37 mm was (HD) 62.4 and 65.2%; (TOF) 77.1 and 81.4%; (PSF) 62.0 and 65.2%; (TOFPSF) 77.3 and 81.6%. Similar hot and cold contrast trends were found during the analyses of other phantoms representing different clinical conditions (brain and whole body). Nevertheless, the authors observed a predominant role of either TOF or PSF, depending on the specific characteristics and dimensions of the phantoms. CONCLUSIONS: Discovery-690 shows very good PET physical performance for all the standard NEMA NU-2-2007 measurements. Furthermore, the new reconstruction algorithms available for PET data (TOF and PSF) allow further improvements of the D-690 image quality performance both qualitatively and quantitatively.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.3635220

Cites

57 of 57 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 57 references

Cited by

See all 380

Metadata

Authors
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C.
Venue
Medical Physics
Publisher
Wiley
Published
2011-09-14
Volume
38
Issue
10
Pages
5394-5411
ISSN
0094-2405 (ISSN-L)
OpenAlex
W1970453159
Cited by
380
References
57
FWCI
17.47
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Physical Performance of the new hybrid PET/CT Discovery-690. https://app.cochat.ai/papers/10.1118/1.3635220
The paper
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011). Physical Performance of the new hybrid PET/CT Discovery-690. Medical Physics. https://doi.org/10.1118/1.3635220

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.