CoChat

LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time

Open access
Schaart, Dennis R, Seifert, Stefan, Vinke, Ruud, van Dam, Herman T, Dendooven, Peter, Löhner, Herbert, Beekman, Freek J · 2010 · Physics in Medicine and Biology
Cited by 225

The DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.

Abstract

The use of time-of-flight (TOF) information in positron emission tomography (PET) enables significant improvement in image noise properties and, therefore, lesion detection. Silicon photomultipliers (SiPMs) are solid-state photosensors that have several advantages over photomultiplier tubes (PMTs). SiPMs are small, essentially transparent to 511 keV gamma rays and insensitive to magnetic fields. This enables novel detector designs aimed at e.g. compactness, high resolution, depth-of-interaction (DOI) correction and MRI compatibility. The goal of the present work is to study the timing performance of SiPMs in combination with LaBr(3):Ce(5%), a relatively new scintillator with promising characteristics for TOF-PET. Measurements were performed with two, bare, 3 mm x 3 mm x 5 mm LaBr(3):Ce(5%) crystals, each coupled to a 3 mm x 3 mm SiPM. Using a (22)Na point source placed at various positions in between the two detectors, a coincidence resolving time (CRT) of approximately 100 ps FWHM for 511 keV annihilation photon pairs was achieved, corresponding to a TOF positioning resolution of approximately 15 mm FWHM. At the same time, pulse height spectra with well-resolved full-energy peaks were obtained. To our knowledge this is the best CRT reported for SiPM-based scintillation detectors to date. It is concluded that SiPM-based scintillation detectors can provide timing resolutions at least as good as detectors based on PMTs.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/0031-9155/55/7/n02

Cites

22 of 22 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 22 references

Cited by

See all 225

Metadata

Authors
Schaart, D. R., Seifert, S., Vinke, R., van Dam, H. T., Dendooven, P., Löhner, H., & Beekman, F. J.
Venue
Physics in Medicine and Biology
Published
2010-03-19
Volume
55
Issue
7
Pages
N179-N189
ISSN
  • 0031-9155 (ISSN-L)
  • 1361-6560
OpenAlex
W2151969366
Cited by
225
References
22
FWCI
20.11
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time. https://app.cochat.ai/papers/10.1088/0031-9155/55/7/n02
The paper
Schaart, D. R., Seifert, S., Vinke, R., van Dam, H. T., Dendooven, P., Löhner, H., & Beekman, F. J. (2010). LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/55/7/n02

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.