Physical and clinical performance of the mCT time-of-flight PET/CT scanner
Open accessThe DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
Time-of-flight (TOF) measurement capability promises to improve PET image quality. We characterized the physical and clinical PET performance of the first Biograph mCT TOF PET/CT scanner (Siemens Medical Solutions USA, Inc.) in comparison with its predecessor, the Biograph TruePoint TrueV. In particular, we defined the improvements with TOF. The physical performance was evaluated according to the National Electrical Manufacturers Association (NEMA) NU 2-2007 standard with additional measurements to specifically address the TOF capability. Patient data were analyzed to obtain the clinical performance of the scanner. As expected for the same size crystal detectors, a similar spatial resolution was measured on the mCT as on the TruePoint TrueV. The mCT demonstrated modestly higher sensitivity (increase by 19.7 ± 2.8%) and peak noise equivalent count rate (NECR) (increase by 15.5 ± 5.7%) with similar scatter fractions. The energy, time and spatial resolutions for a varying single count rate of up to 55 Mcps resulted in 11.5 ± 0.2% (FWHM), 527.5 ± 4.9 ps (FWHM) and 4.1 ± 0.0 mm (FWHM), respectively. With the addition of TOF, the mCT also produced substantially higher image contrast recovery and signal-to-noise ratios in a clinically-relevant phantom geometry. The benefits of TOF were clearly demonstrated in representative patient images.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillator
Melcher, C. L., & Schweitzer, J. S. (1992) · IEEE Transactions on Nuclear Science · doi:10.1109/23.159655 · Cited by 846
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Benefit of Time-of-Flight in PET: Experimental and Clinical Results
Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.044834 · Cited by 584
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536 · cited 2×
“In this experiment, the torso phantom was filled with a solution of water and 18F containing an initial background activity concentration of 5.3 ± 10% kBq ml−1.”
MethodInfluentialSelf-citationTime-of-flight positron emission tomography: status relative to conventional PET.
Budinger, T. F. (1983) · PubMed · Cited by 283
An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging
Lois, C., Jakoby, B. W., Long, M. J., Hubner, K. F., Barker, D. W., Casey, M. E., Conti, M., Panin, V. Y., Kadrmas, D. J., & Townsend, D. W. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.068098 · Cited by 265
Self-citationFast reconstruction of 3D PET data with accurate statistical modeling
Comtat, C., Kinahan, P. E., Defrise, M., Michel, C., & Townsend, D. W. (1998) · IEEE Transactions on Nuclear Science · doi:10.1109/23.681983 · Cited by 165
Self-citationComparison of 3-D reconstruction with 3D-OSEM and with FORE+OSEM for PET
Liu, X. E., Comtat, C., Michel, C. J., Kinahan, P. E., Defrise, M., & Townsend, D. W. (2001) · IEEE Transactions on Medical Imaging · doi:10.1109/42.938248 · Cited by 143
Self-citationJakoby, B. W., Bercier, Y., Watson, C. C., Bendriem, B., & Townsend, D. W. (2009) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2009.2015764 · Cited by 110
Self-citationNEMA NU 2 performance tests for scanners with intrinsic radioactivity.
Watson, C. C., Casey, M., Eriksson, L., Mulnix, T., Adams, D., & Bendriem, B. (2004) · PubMed · Cited by 90
Self-citationFeasibility of time-of-flight reconstruction in positron emission tomography.
Mullani, N. A., Markham, J., & Ter‐Pogossian, M. M. (1980) · PubMed · Cited by 68
Instrumentation trends in nuclear medicine
Budinger, T. F. (1977) · Seminars in Nuclear Medicine · doi:10.1016/s0001-2998(77)80037-8 · Cited by 35
Cited by
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 874
Self-citationCherry, S. R., Jones, T., Karp, J. S., Qi, J., Moses, W. W., & Badawi, R. D. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.184028 · Cited by 763
“The mini EXPLORER system is based on the components from a Siemens mCT scanner (26), which have been reconfigured into a scanner that is roughly half the diameter and twice the axial length.”
MethodPerformance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484
Spencer, B. A., Berg, E., Schmall, J. P., Omidvari, N., Leung, E. K., Abdelhafez, Y. G., Tang, S., Deng, Z., Dong, Y., Lv, Y., et al. (2021) · Journal of Nuclear Medicine · doi:10.2967/jnumed.120.250597 · Cited by 403
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 380
Metadata
- Authors
- Jakoby, B. W., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2011-03-22
- Volume
- 56
- Issue
- 8
- Pages
- 2375-2389
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2047284752
- Cited by
- 492
- References
- 12
- FWCI
- 29.88
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.