Determination of the spring constants of probes for force microscopy/spectroscopy
The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
A direct, accurate and convenient procedure for calibrating the spring constants of probes used for force microscopy/spectroscopy is described. It amounts to deflecting an `unknown' cantilever with a `standard' lever, where the standard lever has been precalibrated. The absolute and relative accuracies of the procedure are - 30% and - 20%, respectively; the former is limited by uncertainties in the determination of the spring constant for the `standard' lever, as well as that for the `unknown'. The method differs from others of the static deflection variety by its exploitation of the routine features of current instruments. The technique is compared with other static and dynamic methods currently being used.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Focused ion beam technology and applications
Melngailis, J. (1987) · Journal of Vacuum Science & Technology B: Microelectronics Processing and Phenomena · doi:10.1116/1.583937 · Cited by 615
Sub-20-nm-wide line fabrication in poly(methylmethacrylate) using a Ga+ microprobe
Kubena, R. L., Stratton, F. P., Ward, J. W., Atkinson, G. M., & Joyce, R. J. (1989) · Journal of Vacuum Science & Technology B Microelectronics Processing and Phenomena · doi:10.1116/1.584460 · Cited by 39
Transport characteristics of AlGaAs/GaAs wires fabricated by focused Ga-ion-beam implantation
Nakata, S., Hirayama, Y., Tarucha, S., & Horíkoshi, Y. (1991) · Journal of Applied Physics · doi:10.1063/1.348510 · Cited by 34
Cited by
Force measurements with the atomic force microscope: Technique, interpretation and applications
Butt, H.-J., Cappella, B., & Kappl, M. (2005) · Surface Science Reports · doi:10.1016/j.surfrep.2005.08.003 · Cited by 4,297
Micromechanical cantilever-based biosensors
Raiteri, R., Grattarola, M., Butt, H.-J., & Skládal, P. (2001) · Sensors and Actuators B: Chemical · doi:10.1016/s0925-4005(01)00856-5 · Cited by 734
Normal and torsional spring constants of atomic force microscope cantilevers
Green, C. P., Lioe, H., Cleveland, J. P., Proksch, R., Mulvaney, P., & Sader, J. E. (2004) · Review of Scientific Instruments · doi:10.1063/1.1753100 · Cited by 551
Comparison of calibration methods for atomic-force microscopy cantilevers
Burnham, N. A., Chen, X., Hodges, C. S., Matei, G. A., Thoreson, E. J., Roberts, C. J., Davies, M. C., & Tendler, S. J. B. (2003) · Nanotechnology · doi:10.1088/0957-4484/14/1/301 · Cited by 458
Stolz, M., Raiteri, R., Daniels, A. U., VanLandingham, M. R., Baschong, W., & Aebi, U. (2004) · Biophysical Journal · doi:10.1016/s0006-3495(04)74375-1 · Cited by 454
Leite, F. L., Bueno, C. C., Da Róz, A. L., Ziemath, E. C., & Oliveira, O. N. (2012) · International Journal of Molecular Sciences · doi:10.3390/ijms131012773 · Cited by 453
Single molecule force spectroscopy in biology using the atomic force microscope
Zlatanova, J., Lindsay, S. M., & Leuba, S. H. (2000) · Progress in Biophysics and Molecular Biology · doi:10.1016/s0079-6107(00)00014-6 · Cited by 375
Lévy, R., & Maaloum, M. (2002) · Nanotechnology · doi:10.1088/0957-4484/13/1/307 · Cited by 366
Spring constant calibration of atomic force microscope cantilevers of arbitrary shape
Sader, J. E., Sanelli, J. A., Adamson, B. D., Monty, J. P., Wei, X., Crawford, S. A., Friend, J. R., Marusic, I., Mulvaney, P., & Bieske, E. J. (2012) · Review of Scientific Instruments · doi:10.1063/1.4757398 · Cited by 283
Antigen binding forces of individually addressed single-chain Fv antibody molecules
Ros, R., Schwesinger, F., Anselmetti, D., Kubon, M., Schäfer, R., Plückthun, A., & Tiefenauer, L. (1998) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.95.13.7402 · Cited by 253
Probing Soft Matter with the Atomic Force Microscopies: Imaging and Force Spectroscopy
McConney, M. E., Singamaneni, S., & Tsukruk, V. V. (2010) · Polymer Reviews · doi:10.1080/15583724.2010.493255 · Cited by 241
Sun, Y., Akhremitchev, B., & Walker, G. C. (2004) · Langmuir · doi:10.1021/la036461q · Cited by 201
Cook, S. M., Lang, K. M., Chynoweth, K. M., Wigton, M., Simmonds, R. W., & Schäffer, T. E. (2006) · Nanotechnology · doi:10.1088/0957-4484/17/9/010 · Cited by 186
Schaer-Zammaretti, P., & Ubbink, J. (2003) · Ultramicroscopy · doi:10.1016/s0304-3991(03)00044-5 · Cited by 166
Norman, M. D. A., Ferreira, S. A., Jowett, G. M., Bozec, L., & Gentleman, E. (2021) · Nature Protocols · doi:10.1038/s41596-021-00495-4 · Cited by 162
Mullin, N., & Hobbs, J. K. (2014) · Review of Scientific Instruments · doi:10.1063/1.4901221 · Cited by 58
Applications of biosensing atomic force microscopy in monitoring drug and nanoparticle delivery
Lamprecht, C., Hinterdorfer, P., & Ebner, A. (2014) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2014.917078 · Cited by 46
“The determination of the spring constant, which is essential when performing AFM force spectroscopy experiments, can be done by a calibrated reference [40], the added mass method [41], the thermal noise method [42] or by the Sadar method [43].”
Method
Metadata
- Authors
- Gibson, C. T., Watson, G. S., & Myhra, S.
- Venue
- Nanotechnology
- Publisher
- IOP Publishing
- Published
- 1996-09-01
- Volume
- 7
- Issue
- 3
- Pages
- 259-262
- ISSN
- 0957-4484 (ISSN-L)
- 1361-6528
- OpenAlex
- W1994079521
- Cited by
- 237
- References
- 3
- FWCI
- 1.68
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.