High-sensitivity operation of single-beam optically pumped magnetometer in a kHz frequency range
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
Optically pumped magnetometers (OPM) can be used in various applications, from magnetoencephalography to magnetic resonance imaging and nuclear quadrupole resonance (NQR). OPMs provide high sensitivity and have the significant advantage of non-cryogenic operation. To date, many magnetometers have been demonstrated with sensitivity close to 1 fT, but most devices are not commercialized. Most recently, QuSpin developed a model of OPM that is low cost, high sensitivity, and convenient for users, which operates in a single-beam configuration. Here we developed a theory of single-beam (or parallel two-beam) magnetometers and showed that it is possible to achieve good sensitivity beyond their usual frequency range by tuning the magnetic field. Experimentally we have tested and optimized a QuSpin OPM for operation in the frequency range from DC to 1.7 kHz, and found that the performance was only slightly inferior despite the expected decrease due to deviation from the spin-exchange relaxation-free regime.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“OPMs on the other hand, like the QZFM series by QuSpin (QuSpin Inc., Louisville, CO, USA), represent a promising, emerging technology that combines near-SQUID sensitivity with room-temperature operation while offering scalability towards wearable and multi-subject recording [25,26,44,45].”
From The Heart’s Electromagnetic Field in Emotions, Empathy and Human Connection: Biosensor-Derived Insights into Heart–Brain Axis Mechanisms and a Basis for Novel BioMagnetoTherapies (2026) · cites this paper as [44]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Budker, D., & Romalis, M. (2007) · Nature Physics · doi:10.1038/nphys566 · Cited by 1,916
A subfemtotesla multichannel atomic magnetometer
Kominis, I. K., Kornack, T. W., Allred, J. C., & Romalis, M. V. (2003) · Nature · doi:10.1038/nature01484 · Cited by 1,631
Induction coil sensors—a review
Tumanski, S. (2007) · Measurement Science and Technology · doi:10.1088/0957-0233/18/3/r01 · Cited by 590 · cited 2×
“8 kHz, and becomes less competitive than search coils [12], except for applications in which measurements are performed at small distances, for example, in the detection of NMR signals or neuronal activities.”
BackgroundImproved Limit on the Permanent Electric Dipole Moment ofHg199
Griffith, W. C., Swallows, M. D., Loftus, T. H., Romalis, M. V., Heckel, B. R., & Fortson, E. N. (2009) · Physical Review Letters · doi:10.1103/physrevlett.102.101601 · Cited by 489
Tunable Atomic Magnetometer for Detection of Radio-Frequency Magnetic Fields
Savukov, I. M., Seltzer, S. J., Romalis, M. V., & Sauer, K. L. (2005) · Physical Review Letters · doi:10.1103/physrevlett.95.063004 · Cited by 313
Self-citationSpin-exchange relaxation-free magnetometry using elliptically polarized light
Shah, V., & Romalis, M. V. (2009) · Physical Review A · doi:10.1103/physreva.80.013416 · Cited by 273 · cited 3×
“When the modulation field β β ω = t sin y y 0 is applied in the dynamic regime where the ω is not small, the solution of Bloch equation to a small field δβy gives at the modulation frequency [13]”
BackgroundMethodInfluentialSelf-citationNMR Detection with an Atomic Magnetometer
Savukov, I. M., & Romalis, M. V. (2005) · Physical Review Letters · doi:10.1103/physrevlett.94.123001 · Cited by 250
Self-citationSubfemtotesla radio-frequency atomic magnetometer for detection of nuclear quadrupole resonance
Lee, S.-K., Sauer, K. L., Seltzer, S. J., Alem, O., & Romalis, M. V. (2006) · Applied Physics Letters · doi:10.1063/1.2390643 · Cited by 176 · cited 3×
“Applications in NMR, MRI, nuclear quadrupole resonance (NQR), underground communications, MEG, and MCG can be readily developed.”
MethodInfluentialMagnetic resonance imaging with an optical atomic magnetometer
Xu, S., Yashchuk, V. V., Donaldson, M. H., Rochester, S. M., Budker, D., & Pines, A. (2006) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0605396103 · Cited by 168
Detection of NMR signals with a radio-frequency atomic magnetometer
Savukov, I., Seltzer, S. J., & Romalis, M. V. (2006) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2006.12.012 · Cited by 153
Self-citation
Cited by
Tang, J., Zhai, Y., Li, C., Zhang, Y., Li, L., Zhao, B., Zhou, B., Han, B., & Liu, G. (2021) · Optics Express · doi:10.1364/oe.425851 · Cited by 129
Miniaturized Magnetic Sensors for Implantable Magnetomyography
Zuo, S., Heidari, H., Farina, D., & Nazarpour, K. (2020) · Advanced Materials Technologies · doi:10.1002/admt.202000185 · Cited by 122
Parameter Modeling Analysis of a Cylindrical Ferrite Magnetic Shield to Reduce Magnetic Noise
Ma, D., Lu, J., Fang, X., Yang, K., Wang, K., Zhang, N., Han, B., & Ding, M. (2021) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2021.3050351 · Cited by 99
Quantum‐Based Magnetic Field Sensors for Biosensing
Bao, B., Hua, Y., Wang, R., & Li, D. (2023) · Advanced Quantum Technologies · doi:10.1002/qute.202200146 · Cited by 108
Three-axis closed-loop optically pumped magnetometer operated in the SERF regime
Yan, Y., Lu, J., Zhang, S., Lu, F., Yin, K., Wang, K., Zhou, B., & Liu, G. (2022) · Optics Express · doi:10.1364/oe.458367 · Cited by 89
Wang, J., Fan, W., Yin, K., Yan, Y., Zhou, B., & Song, X. (2020) · Physical Review A · doi:10.1103/physreva.101.053427 · Cited by 77
Kim, Y. J., Chu, P.-H., & Savukov, I. (2018) · Physical Review Letters · doi:10.1103/physrevlett.121.091802 · Cited by 67
Self-citationHybrid GMR Sensor Detecting 950 pT/sqrt(Hz) at 1 Hz and Room Temperature
Guedes, A. F. S., Macedo, R. J., Jaramillo, G., Cardoso, S., Freitas, P. P., & Horsley, D. A. (2018) · Sensors · doi:10.3390/s18030790 · Cited by 47
“Since then, further advancements enabling the demonstration of SERF AM for imaging of brain activities [7] and the development of compact optically pumped atomic magnetometers (OPMs) prototypes aiming at commercialization have been reported [8,9].”
BackgroundWang, J., Zhou, B., Liu, X., Wu, W., Chen, L., Han, B., & Fang, J. (2019) · IEEE Access · doi:10.1109/access.2019.2920955 · Cited by 45
Analysis of coil constant of triaxial uniform coils in Mn–Zn ferrite magnetic shields
Ma, D., Lu, J., Fang, X., Wang, K., Wang, J., Zhang, N., Chen, H., Ding, M., & Han, B. (2021) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/abf53c · Cited by 42
Zhou, W., Han, B., Wang, J., Wu, D., Zhao, F., & Li, N. (2021) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2021.3125561 · Cited by 41
Metasurface-integrated elliptically polarized laser-pumped SERF magnetometers
Liang, Z., Hu, J., Zhou, P., Liu, L., Hu, G., Wang, A., & Ye, M. (2024) · Microsystems & Nanoengineering · doi:10.1038/s41378-024-00715-3 · Cited by 41
Kim, Y. J., Chu, P.-H., Savukov, I., & Newman, S. (2019) · Nature Communications · doi:10.1038/s41467-019-10169-1 · Cited by 47
“The SERF atomic magnetometer used in this work operates in a single-beam configuration in which a single circularly polarized laser beam is used to pump and probe the Rb electron spins [39].”
BackgroundMethodInfluentialSelf-citationGuo, Q., Hu, T., Chen, C.-Q., Feng, X., Wu, Z., Zhang, Y., Zhang, M., Chang, Y., & Yang, X. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3102146 · Cited by 38
Study of Shielding Ratio of Cylindrical Ferrite Enclosure With Gaps and Holes
Ma, D., Ding, M., Lu, J., Yao, H., Zhao, J., Yang, K., Cai, J., & Han, B. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2904719 · Cited by 35
“The maximum diameter of the ferrite magnetic shield reported at present is 180 mm [26].”
BackgroundA kilohertz bandwidth and sensitive scalar atomic magnetometer using an optical multipass cell
Li, S., Liu, J., Jin, M., Tetteh, A. K., Dai, P., Xu, Z., & Nwodom, T. E.-T. (2022) · Measurement · doi:10.1016/j.measurement.2022.110704 · Cited by 40
Study of Magnetic Noise of a Multi-Annular Ferrite Shield
Lu, J., Ma, D., Yang, K., Quan, W., Zhao, J., Xing, B., Han, B., & Ding, M. (2020) · IEEE Access · doi:10.1109/access.2020.2976785 · Cited by 31
Alignment of magnetic sensing and clinical magnetomyography
Arekhloo, N. G., Parvizi, H., Zuo, S., Wang, H., Nazarpour, K., Marquetand, J. C., & Heidari, H. (2023) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1154572 · Cited by 36
Huang, J., Fan, W., Wang, Z., Yuan, L., Zhang, K., Pei, H., Pang, H., & Quan, W. (2022) · Optics Express · doi:10.1364/oe.451042 · Cited by 30
Nondestructive in-line sub-picomolar detection of magnetic nanoparticles in flowing complex fluids
Bougas, L., Langenegger, L. D., Mora, C. A., Zeltner, M., Stark, W. J., Wickenbrock, A., Blanchard, J. W., & Budker, D. (2018) · Scientific Reports · doi:10.1038/s41598-018-21802-2 · Cited by 32
Fu, Y., Sun, J., Ruan, J., & Quan, W. (2022) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2022.113487 · Cited by 27
Yang, K., Lu, J., Ding, M., Zhao, J., Ma, D., Li, Y., Xing, B., Han, B., & Fang, J. (2019) · IEEE Access · doi:10.1109/access.2019.2935034 · Cited by 26
Detection of ultra-low field NMR signal with a commercial QuSpin single-beam atomic magnetometer
Savukov, I., Kim, Y. J., & Schultz, G. M. (2020) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2020.106780 · Cited by 27
Self-citationHigh-sensitivity pump–probe atomic magnetometer based on single fiber-coupled
Zhao, B., Tang, J., Yang, H., Li, L., Zhang, Y., Liu, Y., & Zhai, Y. (2022) · Optics & Laser Technology · doi:10.1016/j.optlastec.2022.109025 · Cited by 25
Design of Highly Uniform Radial Coils Considering the Coupling Effect of Magnetic Shielding Cylinder
Wang, W., Zhou, X., Zhao, F., Xie, X., & Zhou, W. (2023) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2023.3341108 · Cited by 24
“A TOMIC sensors are widely used in weak magnetic biomedicine [1], [2], aerospace magnetic detection [3], cosmological axion spin precession experiment, and geomagnetic navigation [4] because of their advantages of ultrahigh sensitivity, portability, and small size [5], [6], [7].”
MethodVarious noise reduction techniques of magnetoresistive sensors and their applications: A review
Jiao, Q., Jin, Z., Zhang, C., & Chen, J. (2024) · Measurement · doi:10.1016/j.measurement.2024.116143 · Cited by 25
Effect of gaps on magnetic noise of cylindrical ferrite shield
Lu, J., Sun, C., Ma, D., Yang, K., Zhao, J., Han, B., Quan, W., Zhang, N., & Ding, M. (2021) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/abf169 · Cited by 22
Correlating the microstructure of Mn–Zn ferrite with magnetic noise for magnetic shield applications
Sun, B., Ma, D., Bai, G., Lu, J., Yang, K., Wang, K., Xu, X., Zhai, Y., Quan, W., & Han, B. (2022) · Ceramics International · doi:10.1016/j.ceramint.2022.12.045 · Cited by 23
In Situ Measurement of Nonorthogonal Angles of a Three-Axis Vector Optically Pumped Magnetometer
Li, S., Lu, J., Ma, D., Wang, K., Gao, Y., Sun, C., & Han, B. (2022) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2022.3144737 · Cited by 21
“When the external magnetic field changes, the absorption of the circularly polarized light by the alkali metal atoms will change, which is reflected in the detection output of the PD [24], [27]”
BackgroundMethodOptimising the sensing volume of OPM sensors for MEG source reconstruction
Bezsudnova, Y., Koponen, L. M., Barontini, G., Jensen, O. B., & Kowalczyk, A. (2022) · NeuroImage · doi:10.1016/j.neuroimage.2022.119747 · Cited by 24
“For the optimal pumping rate, the atomic shot noise is given by [3, 56]”
MethodInfluentialSuppression of modulation-magnetic-fields crosstalk for single-beam optically-pumped magnetometers
Suo, Y., Song, X., Jiang, L., Jia, L., Long, T., & Wu, Z. (2023) · Measurement · doi:10.1016/j.measurement.2023.113059 · Cited by 20
Design of locally arranged sensor arrays in wearable OPM-MEG based on sensor volume constraints
Wang, W., Cao, F., An, N., Wen, L., Xu, W., Wang, C., Gao, Z., Yu, D., Gao, Y., & Ning, X. (2024) · Measurement · doi:10.1016/j.measurement.2024.114373 · Cited by 21
Real-Time Biomagnetic Signal Processing for Uncooled Magnetometers in Cardiology
Reermann, J., Elzenheimer, E., & Schmidt, G. U. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2893236 · Cited by 19
“However, the internal sensor setup required for proper operation is still rather complex [14].”
BackgroundCompact high-bandwidth single-beam optically-pumped magnetometer for biomagnetic measurement
Wu, T., Xiao, W., Peng, X., Wu, T., & Guo, H. M. (2024) · Biomedical Optics Express · doi:10.1364/boe.545624 · Cited by 19
“But miniaturized SERF magnetometers typically operate with a bandwidth of around 100 Hz and are not suitable for capturing high-frequency mouse MCG signals [18,19].”
Yin, Y., Zhou, B., Yin, K., Wang, Y., Tang, J., Ye, M., Ning, X., & Han, B. (2021) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/ac19e3 · Cited by 19
Wang, K., Lu, J., Yan, Y., Tang, J., Ye, M., Yin, K., Chai, Z., Zhou, B., Ning, X., & Han, B. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3136213 · Cited by 18
Liu, Y., Gao, H., Ma, L., Quan, J., Fan, W., Xu, X., Fu, Y., Duan, L., & Quan, W. (2022) · Materials · doi:10.3390/ma15228267 · Cited by 17
Zhang, S., Zhou, Y., Chai, Z., Liu, Q., Yin, K., & Han, B. (2022) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2022.3232658 · Cited by 17
“By solving the Bloch equation, which describes the atomic spin polarization of a SERF AM to the magnetic field, the equation between Pz and Bx is given in [35].”
BackgroundEnzymatic Reactions Observed with Zero- and Low-Field Nuclear Magnetic Resonance
Eills, J., Picazo‐Frutos, R., Bondar, O., Cavallari, E., Carrera, C., Barker, S. J., Utz, M., Herrero-Gómez, A., Marco‐Rius, I., Tayler, M. C. D., et al. (2023) · Analytical Chemistry · doi:10.1021/acs.analchem.3c02087 · Cited by 15
Liu, L., Lu, Y., Zhuang, X., Zhang, Q., & Fang, G. (2020) · Applied Sciences · doi:10.3390/app10207172 · Cited by 16
Analysis and Suppression of Thermal Magnetic Noise of Ferrite in the SERF Co-Magnetometer
Pang, H., Liu, F., Fan, W., Wu, J., Yuan, Q., Wu, Z., & Quan, W. (2022) · Materials · doi:10.3390/ma15196971 · Cited by 15
“Introduction Atomic magnetometers and co-magnetometers based on a spin-exchange relaxationfree (SERF) regime have been widely used in frontier science research [1,2], ultra-weak magnetic field measurement [3–5], inertial navigation [6,7] and magnetoencephalography [8,9] because of their ultra-high-sensitivity magnetic…”
MethodEvaluation of optical parameters for a microminiature Rb vapor cell in a dual-beam SERF magnetometer
Wang, Y., Shi, T., Zhou, W., Tang, J., Zhou, B., Jin, G., Han, B., & Zou, S. (2022) · Optics Express · doi:10.1364/oe.458827 · Cited by 16
Suo, Y., Song, X., Jia, L., Long, T., Ye, M., & Zhou, B. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3198502 · Cited by 14
In-Situ Relaxation Rate Measurement in Magnetic Modulated Atomic Magnetometers
Yin, K., Zhou, B., Lu, J., Lu, F., Zhang, S., Yan, Y., & Ye, M. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3131734 · Cited by 15
Temperature characteristics of Rb-N2 single-beam magnetometer with different buffer gas pressures
Zhang, S., Zhou, Y., Lu, F., Yan, Y., Wang, W., Zhou, B., Zhai, Y., Lu, J., & Ye, M. (2023) · Measurement · doi:10.1016/j.measurement.2023.112860 · Cited by 13
Perspective: magnetic quantum sensors for biomedical applications
Wu, K., & He, R. (2025) · Nanotechnology · doi:10.1088/1361-6528/adb635 · Cited by 19
“OPMs are compact, operate at room temperature, and achieve exceptional sensitivity, detecting magnetic fields as weak as ∼ fT/sqrt(Hz) [28–30].”
Chai, C., Yang, X. M., Zheng, Y., Heyat, M. B. B., Li, Y., Yang, D.-B., Chen, Y.-H., & Sawan, M. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.117321 · Cited by 15
JOM-4S Overhauser Magnetometer and Sensitivity Estimation
Gong, X., Chen, S., & Zhang, S. (2021) · Sensors · doi:10.3390/s21227698 · Cited by 12
Jofre, M., Romeu, J., & Jofre, L. (2023) · New Journal of Physics · doi:10.1088/1367-2630/acb37a · Cited by 16
Everaert, K., Sander, T. H., Körber, R., Löwa, N., Waeyenberge, B. V., Leliaert, J., & Wiekhorst, F. (2023) · Nanoscale Advances · doi:10.1039/d3na00016h · Cited by 13
A modular optically pumped magnetometer system
Coussens, T., Gialopsou, A., Abel, C., Bason, M. G., James, T. M., Evans, W., Woodley, M. T. M., Nightingale, D., Nicolau, D., Page, L., et al. (2024) · Quantum Science and Technology · doi:10.1088/2058-9565/ad4e60 · Cited by 11
“It is also possible to shift the sensitive bandwidth range to higher frequencies by applying DC field offsets [48].”
BackgroundMyers, J. Z., Bullinger, F., Kempf, N., Plaumann, M., Ortmeier, A., Theis, T., Povolni, P., Romanowski, J., Engelmann, J., Scheffler, K., et al. (2024) · Physical review. B./Physical review. B · doi:10.1103/physrevb.109.184443 · Cited by 11
Zhang, J., Meng, X., Han, T., Wei, X., Wang, L., Zhao, Y., Fu, G., Tian, N., Wang, Q., Qin, S., et al. (2023) · IET Nanodielectrics · doi:10.1049/nde2.12049 · Cited by 13
Compact optically pumped magnetometer light source stabilization with regulated feedbacks
Niu, Y., Zou, Z., Cheng, L., & Ye, C. (2023) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2023.114869 · Cited by 11
Bandwidth compensation for ultra-high-sensitivity SERF magnetometers in magnetocardiac sensing
Jia, Y., Pei, H., Yang, Y., Jiang, L., Lu, J., Ning, X., Cui, Y., & Xiang, M. (2024) · Measurement · doi:10.1016/j.measurement.2024.116095 · Cited by 10
Chu, P.-H., Kim, Y. J., & Savukov, I. (2019) · Physical review. D/Physical review. D. · doi:10.1103/physrevd.99.075031 · Cited by 12
Self-citationSpin dynamic response to a time dependent field
Bevilacqua, G., Biancalana, V., Dancheva, Y., Fregosi, A., & Vigilante, A. (2021) · Applied Physics B · doi:10.1007/s00340-021-07673-y · Cited by 11
Flow Rate Estimation Based on Magnetic Particle Detection Using a Miniatured High-Sensitivity OPM
Liu, Y., Huang, B., Li, J., Li, R., & Zhai, Y. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3565069 · Cited by 15
“Bougas et al. [18] introduced a highly sensitive in-line system for trace MP detection using the Quspin OPM [19], which applies to extracorporeal blood purification and industrial wastewater treatment, and beneficial to mitigate environmental contamination and biohazards.”
Silva, M., Silva, J. F., Leitão, D. C., Cardoso, S., & Freitas, P. P. (2019) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2019.2899808 · Cited by 8
“The portfolio of the state-of-the-art low magnetic field detection includes the well-known quantum interference devices (SQUID) [1], atomic magnetometers (AMs) [2], spin-exchange relaxation-free (SERF) AMs [3], and more recently, optical AMs (OPMs) based on magnetooptical properties of atomic samples [4], [5].”
BackgroundComprehensive influence of modulated and bias magnetic fields on an atomic magnetometer
Yin, Y., Zhou, B., Yin, K., Tang, J., Ning, X., Han, B., & Fang, J. (2021) · Measurement Science and Technology · doi:10.1088/1361-6501/abdb79 · Cited by 11
“The influence of the small modulated magnetic field on the magnetometer can be solved analytically by the Bloch equation [16].”
Wu, K.-Y., Chen, S. Y., Gong, J., Peng, M., & Yan, H. (2022) · Physical review. D/Physical review. D. · doi:10.1103/physrevd.105.055020 · Cited by 9
Sun, Y., Zhou, Y., Wang, X., Bai, D., Wang, Y., & Zhou, Z. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3375422 · Cited by 7
“The magnetometer’s − 3-dB bandwidth is 285 Hz, and this can be improved by increasing the incident light energy [19], using amplitude-modulated light [40], or configuring the magnetometer to operate in closed-loop operation mode [41], [42].”
MethodTolstrup, K., Akhtari, M., Brisinda, D., Meloni, A. M., Siegel, R. J., & Fenici, R. (2024) · American Heart Journal Plus Cardiology Research and Practice · doi:10.1016/j.ahjo.2024.100483 · Cited by 12
Magnetic field sensing based on multi-order resonances of atomic spins
Yang, H., Wang, Q., Zhao, B., Li, L., Zhai, Y., Han, B., & Tang, F. (2022) · Optics Express · doi:10.1364/oe.443679 · Cited by 6
Compact Mz atomic magnetometer with a thick silicon core MEMS vapor cell
Li, Q., Qin, J., Jiang, Y., Jia, S., & Jiang, Z. (2024) · Optics Communications · doi:10.1016/j.optcom.2024.130866 · Cited by 5
All-Optical Parametric-Resonance Magnetometer Based on 4He Atomic Alignment
Wang, B., Peng, X., Wang, H., Xiao, W., & Guo, H. M. (2022) · Sensors · doi:10.3390/s22114184 · Cited by 4
“[11] and Twinleaf LLC [12] have been successfully applied to measure fields from humans [13], plants [14], and pre-polarized nuclear magnetic resonance samples [15].”
MethodNamita, K., Ito, Y., & Kobayashi, T. (2021) · Japanese Journal of Applied Physics · doi:10.35848/1347-4065/ac0ba6 · Cited by 4
OPM Gradiometer for Magnetorelaxometry
Prospero, A. G., Miranda, J. R. D. A., & Baffa, O. (2022) · Springer eBooks · doi:10.1007/978-3-031-05363-4_12 · Cited by 5
Chu, P.-H., (NNSA), U. N. N. S. A., Kim, Y. J., & Savukov, I. (2019) · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · doi:10.2172/1569722 · Cited by 4
Self-citationSimulation Research on Low-Frequency Magnetic Noise in Fe-Based Nanocrystalline Magnetic Shields
Kang, S., Fan, W., Lu, J., & Quan, W. (2025) · Materials · doi:10.3390/ma18020330 · Cited by 3
Wu, H., Zhao, X., Cui, P., Zhang, H., Liu, J., & Wen, T. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3533616 · Cited by 5
Bandwidth Extension Enabled by Modulated VCSEL Pumping for Quantum Sensing Application
Zhou, P., Quan, W., Xu, F., Liang, Z., Hu, J., Liu, L., Hu, G., Xu, Y., & Ye, M. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3568078 · Cited by 3
“The operational bandwidth of a single-beam Qspin magnetometer was increased from 135 to 1700 Hz by switching modulation modes using transverse magnetic fields [18].”
Analysis and Suppression of Pump Beam Alignment Error in SERF Co-Magnetometer
Yuan, Q., Fan, W., Pang, H., Han, X. S., Wang, Z., & Quan, W. (2025) · Photonics · doi:10.3390/photonics12060550 · Cited by 2
Palantzas, A., & Anagnostouli, M. (2026) · Sensors · doi:10.3390/s26051738 · Cited by 1
“OPMs on the other hand, like the QZFM series by QuSpin (QuSpin Inc., Louisville, CO, USA), represent a promising, emerging technology that combines near-SQUID sensitivity with room-temperature operation while offering scalability towards wearable and multi-subject recording [25,26,44,45].”
Li, Y., Zhou, X., & Wang, W. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3559930 · Cited by 2
“The commercial atomic mag-netometers developed by QuSpin in the United States adopt a single-beam structure combined with magnetic field modulation [23], enabling automatic and precise three-axis magnetic field compensation.”
Jiang, W., Li, J., Li, X., & Liu, Y. H. (2026) · Sensors · doi:10.3390/s26092593 · Cited by 1
“In comparison with SQUIDs, SERF AMs provide non-cryogenic operation [22], ultrahigh sensitivity [21], adequate bandwidth [23,24], and promising prospects for wearable and cost-effective biomagnetic systems [9,25–28].”
Yan, Y., Lu, J., Wang, K., Lu, F., Zhang, S., Liu, Z., Li, X., & Zhou, B. (2024) · Chinese Optics Letters · doi:10.3788/col202422.101201 · Cited by 1
“The conventional SERF atomic magnetometer has a limited dynamic range (generally less than ±10 nT) and bandwidth (usually ∼ 100 Hz) [5] .”
A Magnetic Field Gradiometer based on the Laser Helium Optically Pumped Magnetic Field Sensors
(2020) · International Journal of Circuits Systems and Signal Processing · doi:10.46300/9106.2020.14.44
Study on Digital Demodulation Technology of All-Optical Cesium Optically Pumped Magnetometer
Yu, Z. (2022) · Asia Symposium on Signal Processing (ASSP) · doi:10.1109/assp57481.2022.00013
Miniaturized Brain Imaging Apparatus Employing Light, Sound, and Magnetic Fields
Na, S., Zhang, J., & Fan, B. (2024) · Springer eBooks · doi:10.1007/978-3-031-61411-8_18
Magnetic particle detection using a uniformly polarized four-channel SERF atomic magnetometer
Liu, Y., Huang, B., Song, M. K., An, Y., & Zhai, Y. (2025) · Optics & Laser Technology · doi:10.1016/j.optlastec.2025.113851 · Cited by 1
Li, J., Wang, H., Wu, T., Liu, Y., & Guo, H. M. (2026) · Review of Scientific Instruments · doi:10.1063/5.0302200
Kowalczyk, A., Bezsudnova, Y., Barontini, G., & Jensen, O. B. (2026) · University of Birmingham Research Portal (University of Birmingham) · doi:10.25500/edata.bham.00001639
Calculation of Magnetic Noise of Cylindrical Ferrite Shield
Gao, X., & Lu, J. (2026) · Lecture notes in electrical engineering · doi:10.1007/978-981-95-8232-7_42
Applications of carbon nanomaterials in quantum technologies
McDonnell, L., Spencer, J., Signe, J. L., & Vovrosh, J. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-33473-3.00016-4
Measurement bandwidth enhancement of SERF atomic magnetometers using frequency response regulation
Xing, W., Jia, L., Long, T., Zhang, J., Sun, H., Wu, Z. W., Duan, Z., & Song, X. (2026) · Measurement · doi:10.1016/j.measurement.2026.122508
Guo, S., Tian, K., Zhang, X., & Zheng, S. (2026) · Measurement Science and Technology · doi:10.1088/1361-6501/aeaf09
“In contrast, OPM offers sensitivity in the sub-flytesla range under ultra-low magnetic fields and features a more compact form factor, making it suitable for clinical applications [9, 10].”
Metadata
- Authors
- Savukov, I., Kim, Y. J., Shah, V, & Boshier, M. G.
- Affiliations
- Los Alamos National Laboratory (US)
- QuSpin (United States) (US)
- Corresponding author
- Savukov, I
- Venue
- Measurement Science and Technology
- Publisher
- IOP Publishing
- Published
- 2017-02-02
- Volume
- 28
- Issue
- 3
- Pages
- 035104
- ISSN
- 0957-0233 (ISSN-L)
- 1361-6501
- OpenAlex
- W2573282927
- Cited by
- 86
- References
- 21
- Citations by year
- 2026: 7 · 2025: 9 · 2024: 12 · 2023: 10 · 2022: 17 · 2021: 14 · 2020: 6 · 2019: 8 · 2018: 3
- FWCI
- 2.28
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- creativecommons.org/licenses/by/3.0/ (vor)
- Locations
- Measurement Science and Technology (publishedVersion, cc-by)
- OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
- Funders
- U.S. DOE through the LANL/LDRD
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Chemistry › Spectroscopy › Electron Spin Resonance Studies
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Mechanical and Optical Resonators
- Keywords
- optically pumped magnetometers, high sensitivity, magnetic field tuning
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.