Improving the Quality of Self-Monitoring Blood Glucose Measurement: A Study in Reducing Calibration Errors
The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Two independent studies reported that 16% of people who self-monitor blood glucose used incorrectly coded meters. The degree of analytical error, however, was not characterized. Our study objectives were to demonstrate that miscoding can cause analytical errors and to characterize the potential amount of bias that can occur. The impact of calibration error with three selfblood glucose monitoring systems (BGMSs), one of which has an autocoding feature, is reported. METHODS: Fresh capillary fingerstick blood from 50 subjects, 18 men and 32 women ranging in age from 23 to 82 years, was used to measure glucose with three BGMSs. Two BGMSs required manual coding and were purposely miscoded using numbers different from the one recommended for the reagent lot used. Two properly coded meters of each BGMS were included to assess within-system variability. Different reagent lots were used to challenge a third system that had autocoding capability and could not be miscoded. RESULTS: Some within-system comparisons showed deviations of greater than +/-30% when results obtained with miscoded meters were compared with data obtained with ones programmed using the correct code number. Similar erroneous results were found when the miscoded meter results were compared with those obtained with a glucose analyzer. For some miscoded meter and test strip combinations, error grid analysis showed that 90% of results fell into zones indicating altered clinical action. Such inaccuracies were not found with the BGMS having the autocoding feature. CONCLUSIONS: When certain meter code number settings of two BGMSs were used in conjunction with test strips having code numbers that did not match, statistically and clinically inaccurate results were obtained. Coding errors resulted in analytical errors of greater than +/-30% (-31.6 to +60.9%). These results confirm the value of a BGMS with an automatic coding feature.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Skeie, S., Thue, G., Nerhus, K., & Sandberg, S. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.7.994 · Cited by 131
Sell-Monitoring of Blood Glucose
(1996) · Diabetes Care · doi:10.2337/diacare.19.1.s62 · Cited by 89
Kristensen, G. B. B., Nerhus, K., Thue, G., & Sandberg, S. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2004.031575 · Cited by 68
Kabadi, U. M., O'Connell, K. M., Johnson, J., & Kabadi, M. (1994) · Diabetes Care · doi:10.2337/diacare.17.10.1110 · Cited by 44
Cited by
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406
Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
Self-Monitoring of Blood Glucose: The Use of the First or the Second Drop of Blood
Hortensius, J., Slingerland, R. J., Kleefstra, N., Logtenberg, S. J. J., Groenier, K. H., Houweling, S. T., & Bilo, H. J. G. (2011) · Diabetes Care · doi:10.2337/dc10-1694 · Cited by 76
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
Post-analytical errors with portable glucose meters in the hospital setting
Carraro, P., & Plebani, M. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.03.013 · Cited by 37
Metadata
- Authors
- Baum, J. M., Monhaut, N. M., Parker, D. R., & Price, C. P.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2006-06
- Volume
- 8
- Issue
- 3
- Pages
- 347-357
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2043345648
- Cited by
- 42
- References
- 5
- FWCI
- 3.72
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.