A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose
Open accessAbstract
OBJECTIVE: The objectives of this study were 1) to construct new error grids (EGs) for blood glucose (BG) self-monitoring by using the expertise of a large panel of clinicians and 2) to use the new EGs to evaluate the accuracy of BG measurements made by patients. RESEARCH DESIGN AND METHODS: To construct new EGs for type 1 and type 2 diabetic patients, a total of 100 experts of diabetes were asked to assign any error in BG measurement to 1 of 5 risk categories. We used these EGs to evaluate the accuracy of self-monitoring of blood glucose (SMBG) levels in 152 diabetic patients. The SMBG data were used to compare the new type 1 diabetes EG with a traditional EG. RESULTS: Both the type 1 and type 2 diabetes EGs divide the risk plane into 8 concentric zones with no discontinuities. The new EGs are similar to each other, but they differ from the traditional EG in several significant ways. When used to evaluate a data set of measurements made by a sample of patients experienced in SMBG, the new type 1 diabetes EG rated 98.6% of their measurements as clinically acceptable, compared with 95% for the traditional EG. CONCLUSIONS: The consensus EGs furnish a new tool for evaluating errors in the measurement of BG for patients with type 1 and type 2 diabetes.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Analytical accuracy is evaluated by comparing the bias between the result provided by the device under test and a gold standard method.
“Analytical accuracy, instead, is evaluated by comparing the bias between the result provided by the device under test and a gold standard method (specified by the manufacturer) (Figure 2D) 57,58.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [57]
The SMBG clinical trial (pp. 2, 5) compares patients’ meter readings with hexokinase assay values and reports their percentage deviations, supporting the comparison with a reference method. The paper does not define analytical accuracy as a measure of bias or say that the reference method is specified by the manufacturer; its measurements also include user error. (LLM, conf=0.86)
Experimental — accuracy not yet measuredIn the clinical trial, 99.1% (209/211) of subject palm results were within Zone A of the Parkes-Consensus Error Grid.
“Furthermore, 99.5% (217/218) of trial staff capillary fingertip results and 99.1% (209/211) of subject palm results were within Zone A, with the remainder (0.5% [1/218] and 0.9% [2/211], respectively) within Zone B (Fig. 3B and C).”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
The Consensus EG section (p. 2) establishes the Parkes grid and defines Zone A, so the source supports the identification of the grid used to classify results. Its own SMBG clinical trial (p. 2) involved 152 subjects and 304 data points; it does not report subject palm results or the 209/211 (99.1%) figure attributed to it. (LLM, conf=0.88)
Experimental — accuracy not yet measuredThe Parkes error grid modified the Clarke grid for type 1 and type 2 diabetes to avoid discontinuity between risk areas.
“In 2000, this grid was further modified by Parkes et al., for type 1 and type 2 diabetes (Figure 2 B, C, respectively), to avoid discontinuity between risk areas, in which small changes in glucose levels could lead to significant changes of clinical impact (Parkes error grid or consensus error grid) 57.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [57]
In the laboratory study, all 600/600 results were within Zone A of the Parkes-Consensus Error Grid.
“Also, all results (600/600) were within Zone A of the Parkes-Consensus Error Grid (data not shown).”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
In the clinical trial, 99.5% (217/218) of trial staff capillary fingertip results were within Zone A of the Parkes-Consensus Error Grid.
“Furthermore, 99.5% (217/218) of trial staff capillary fingertip results and 99.1% (209/211) of subject palm results were within Zone A, with the remainder (0.5% [1/218] and 0.9% [2/211], respectively) within Zone B (Fig. 3B and C).”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
By Parkes-Consensus Error Grid analysis, 100% (218/218) of subject-obtained capillary fingertip results in the clinical trial were within Zone A.
“By Parkes-Consensus Error Grid analysis, 100% of subject capillary fingertip results were within Zone A.”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
The Parkes-Consensus Error Grid has 5 levels of increasing risk, with Zones A and B denoting little or no clinical risk to the patient.
“There are 5 levels of increasing risk, with Zones A and B denoting little or no clinical risk to the patient [15].”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
ISO 15197:2013 section 6.3 criteria also require that ≥99% of individual measured glucose values fall within Zones A and B of the Parkes-Consensus Error Grid for type 1 diabetes.
“ISO 15197:2013 section 6.3 criteria also require that ≥99% of individual measured glucose values shall fall within Zones A and B of the Parkes-Consensus Error Grid [15] for type 1 diabetes; thus, a Parkes-Consensus Error Grid analysis was performed.”
From Accuracy and user performance evaluation of the Contour Next Link 2.4 blood glucose monitoring system (2015) · cites this paper as [15]
The accuracy of continuous glucose monitors in the hypoglycemic range is poor.
“However, the accuracy of CGMs in the hypoglycemic range is poor as demonstrated by error grid analysis (19,20).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [20]
The SMBG clinical trial (p. 5) uses error-grid analysis to assess measurements from patients using their own blood-glucose meters, not continuous glucose monitors. The paper discusses low-glucose error-grid boundaries, but our copy does not report CGM accuracy in the hypoglycemic range (searched: continuous glucose monitoring, hypoglycemia, low glucose accuracy). (LLM, conf=0.78)
Experimental — accuracy not yet measuredRetrospective and real-time CGMs have poor accuracy in the hypoglycemic range as demonstrated by error grid analysis.
“However, the accuracy of CGMs in the hypoglycemic range is poor as demonstrated by error grid analysis (19,20).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [20]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · see citing works
Accuracy of Perceiving Blood Glucose in IDDM
Cox, D. J., Clarke, W. L., Gonder-Frederick, L. A., Pohl, S. L., Hoover, C. W., Snyder, A. L., Zimbelman, L. K., Carter, W. R., Bobbitt, S. A., & Pennebaker, J. W. (1985) · Diabetes Care · doi:10.2337/diacare.8.6.529 · Cited by 169 · see citing works
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119 · see citing works
Accuracy of home blood glucose meters during hypoglycemia
Trajanoski, Z., Brunner, G. A., Gfrerer, R., Wach, P., & Pieber, T. R. (1996) · Diabetes Care · doi:10.2337/diacare.19.12.1412 · Cited by 85 · see citing works
Understanding Error Grid Analysis
Cox, D. J., Gonder-Frederick, L. A., Kovatchev, B. P., Julian, D. M., & Clarke, W. L. (1997) · Diabetes Care · doi:10.2337/diacare.20.6.911 · Cited by 73 · see citing works
Trajanoski, Z., Brunner, G. A., Schaupp, L. A., Ellmerer, M., Wach, P., Pieber, T. R., Kotanko, P., & Skrabal, F. (1997) · Diabetes Care · doi:10.2337/diacare.20.7.1114 · Cited by 66 · see citing works
Dermal Interstitial Glucose as an Indicator of Ambient Glycemia
Service, F. J., O’Brien, P. C., Wise, S., Ness, S. M., & LeBlanc, S. M. (1997) · Diabetes Care · doi:10.2337/diacare.20.9.1426 · Cited by 47 · see citing works
Kabadi, U. M., O'Connell, K. M., Johnson, J., & Kabadi, M. (1994) · Diabetes Care · doi:10.2337/diacare.17.10.1110 · Cited by 44 · see citing works
Reservations on the Use of Error Grid Analysis for the Validation of Blood Glucose Assays
Gough, D., & Botvinick, E. L. (1997) · Diabetes Care · doi:10.2337/diacare.20.6.1034 · Cited by 35 · see citing works
Pohl, S. L., Gonder-Frederick, L. A., Cox, D. J., et al. (1985) · Diabetes Care · doi:10.2337/diacare.8.6.617 · Cited by 29 · see citing works
Accuracy of Blood Glucose Monitoring for Patients: What It Is and How to Achieve It
Fleming, D. R. (1994) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572179402000606 · Cited by 18 · see citing works
How accurate are home blood-glucose meters with special respect to the low glycemic range?
Moberg, E., Lundblad, S., LINS, P.-E. S., & Adamson, U. (1993) · Diabetes Research and Clinical Practice · doi:10.1016/0168-8227(93)90119-p · Cited by 14 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · Diabetes Care · doi:10.2337/dc12-2480 · Cited by 1,389 · see citing works
“However, the accuracy of CGMs in the hypoglycemic range is poor as demonstrated by error grid analysis (19,20).”
BackgroundGlucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584 · see citing works
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2012-4127 · Cited by 412 · see citing works
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes</i> — <i>2023</i>
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301 · see citing works
Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises
Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400632 · Cited by 285 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes—2024</i>
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281 · see citing works
7. Diabetes Technology: <i>Standards of Medical Care in Diabetes—2022</i>
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358 · see citing works
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234 · see citing works
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210 · see citing works
“, for type 1 and type 2 diabetes (Figure 2 B, C, respectively), to avoid discontinuity between risk areas, in which small changes in glucose levels could lead to significant changes of clinical impact (Parkes error grid or consensus error grid) [57].”
BackgroundMethod7. Diabetes Technology: <i>Standards of Medical Care in Diabetes—2021</i>
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s007 · Cited by 202 · see citing works
Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals
Gough, D. A., Kumosa, L. S., Routh, T. L., Lin, J. T., & Lucisano, J. Y. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3001148 · Cited by 179 · see citing works
7. Diabetes Technology: <i>Standards of Medical Care in Diabetes—2019</i>
American Diabetes Association (2019) · Diabetes Care · doi:10.2337/dc19-s007 · Cited by 179 · see citing works
Alva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820958754 · Cited by 164 · see citing works
Technical Aspects of the Parkes Error Grid
Pfützner, A., Klonoff, D. C., Pardo, S., & Parkes, J. L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700517 · Cited by 122 · see citing works
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86 · see citing works
Device and method for noninvasive glucose assessment
Segman, Y. (. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818763457 · Cited by 55 · see citing works
Pemberton, J. S., Wilmot, E. G., Barnard‐Kelly, K., Leelarathna, L., Oliver, N., Randell, T., Taplin, C. E., Choudhary, P., & Adolfsson, P. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14962 · Cited by 44 · see citing works
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46 · see citing works
Klaff, L. J., Brazg, R., Hughes, K., Tideman, A. M., Schachner, H. C., Stenger, P., Pardo, S., Dunne, N., & Parkes, J. L. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0069 · Cited by 46 · see citing works
Self-citationThe Diabetes Research In Children, N. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322640971 · Cited by 44 · see citing works
Perera, N. J., Stewart, P. M., Williams, P. F., Chua, E. L., Yue, D. K., & Twigg, S. M. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03362.x · Cited by 33 · see citing works
Evaluation of a new noninvasive glucose monitoring device by means of standardized meal experiments
Pfützner, A., Strobl, S., Demircik, F., Redert, L., Pfützner, J., Pfützner, A. H., & Lier, A. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818758769 · Cited by 33 · see citing works
Ji, L., Guo, L., Zhang, J., Li, Y., & Chen, Z. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211037991 · Cited by 26 · see citing works
Baumstark, A., Jendrike, N., Pleus, S., Haug, C., & Freckmann, G. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0408 · Cited by 22 · see citing works
Performance of a new real‐time continuous glucose monitoring system: A multicenter pilot study
Zhou, J., Zhang, S., Li, L., Wang, Y., Lu, W., Sheng, C., Li, Y., Bao, Y., & Jia, W. (2018) · Journal of Diabetes Investigation · doi:10.1111/jdi.12699 · Cited by 22 · see citing works
Tian, T., Aaron, R. E., Kohn, M. A., & Klonoff, D. C. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231214281 · Cited by 16 · see citing works
Yu-Fei, W., Wei-Ping, J., Ming-Hsun, W., Miao-O, C., Ming-Chang, H., Chi-Pin, W., & Ming-Shih, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817705143 · Cited by 15 · see citing works
Bailey, T., Wallace, J. F., Greene, C., Pardo, S., Brown, D., Pflug, B., & Klaff, L. J. (2015) · Clinica Chimica Acta · doi:10.1016/j.cca.2015.06.001 · Cited by 14 · see citing works
Self-citationFreckmann, G., Baumstark, A., & Pleus, S. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817713220 · Cited by 13 · see citing works
Harrison, B., & Brown, D. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1704253 · Cited by 12 · see citing works
Evaluation of a continuous glucose monitoring system in healthy dairy calves and adult goats
Brobst, M. N., Abi-Nader, B. A., Blasczynski, S. J., & Chigerwe, M. (2024) · American Journal of Veterinary Research · doi:10.2460/ajvr.24.03.0076 · Cited by 6 · see citing works
- Authors
- Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H.
- Affiliations
- Becton Dickinson (United States) (US)
- ORCID
- Ginsberg, B H
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2000-08-01
- Volume
- 23
- Issue
- 8
- Pages
- 1143-1148
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2111352982
- Cited by
- 620
- References
- 13
- Citations by year
- 2026: 36 · 2025: 50 · 2024: 57 · 2023: 32 · 2022: 27 · 2021: 38 · 2020: 36 · 2019: 33 · 2018: 37 · 2017: 34 · 2016: 29 · 2015: 22 · 2014: 20 · 2013: 21 · 2012: 25
- FWCI
- 2.18
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- error grid analysis, self-monitoring of blood glucose, measurement error, type 1 diabetes, type 2 diabetes, diabetes self-management
- MeSH terms
- Adult, Blood Glucose, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Female, Humans, Male, Physicians, Regression Analysis, Software, United States, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review