A Glucose Meter Accuracy and Precision Comparison: The FreeStyle Flash Versus the Accu-Chek Advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
BACKGROUND: This study compared the accuracy and precision of five blood glucose (BG) meters. METHODS: Diabetes patients undergoing venipuncture for glucose testing were randomized to one of two groups consisting of three meters: FreeStyle Flash (Abbott Diabetes Care, Alameda, CA), Accu-Chek Advantage (Roche Diagnostics Corp., Indianapolis, IN), and Accu-Chek Compact Plus (Roche Diagnostics) or FreeStyle Flash, Ascensia Contour (Bayer Healthcare, Diagnostic Division, Tarrytown, NY), and BD Logic (BD Diabetes Care, Franklin Lake, NJ). Within 5 min following venipuncture, duplicate finger BG measurements from three ipsilateral fingers were taken. Finger glucose measurements were compared with laboratory reference values. Accuracy was assessed by a Clarke error grid analysis (EGA) and within 10% of the laboratory value criteria. Meter precision was determined by calculating the absolute mean differences in glucose values between duplicate samples. RESULTS: Finger sticks were obtained from 202 patients. Mean venipuncture BG was 148 mg/dL (SD +/- mg/64 dL; range 25-439 mg/dL). Accuracy by Clarke EGA (Zone A results) was demonstrated in 69% of Advantage samples, 75% of Compact Plus, and 96% of the first group of Flash versus 88% of the Contour, 67% of the Logic, and 91% of the second Flash samples (P < 0.05 for both Flash and Contour). Meter accuracy using the 10% criteria was demonstrated in 30%, 38%, 70%, 46%, 48%, and 68% of the samples, respectively (P < 0.05 for both Flash groups compared to each of the other meters). There were no differences in meter precision. CONCLUSIONS: No statistically significant differences in accuracy were evident using the Clarke EGA criteria (pooled results of Zone A and B), though the more strict 20% accuracy criteria (Zone A results only) found the Flash and Contour to have significantly greater accuracy compared to the Advantage, Compact Plus, and the Logic. Using the 10% accuracy criteria found the Flash to have significantly greater accuracy than each of the other four meters. All five meters demonstrated similar precision.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Ohkubo, Y., Kishikawa, H., Araki, E., Miyata, T., Isami, S., Motoyoshi, S., Kojima, Y., Furuyoshi, N., & Shichiri, M. (1995) · Diabetes Research and Clinical Practice · doi:10.1016/0168-8227(95)01064-k · Cited by 3,079
Standards of Medical Care in Diabetes—2007
American Diabetes Association (2007) · Diabetes Care · doi:10.2337/dc07-s004 · Cited by 1,190
Self-Monitoring of Blood Glucose in Patients With Type 2 Diabetes Who Are Not Using Insulin
Welschen, L. M. C., Bloemendal, E., Nijpels, G., Dekker, J. M., Heine, R. J., Stalman, W. A. B., & Bouter, L. M. (2005) · Diabetes Care · doi:10.2337/diacare.28.6.1510 · Cited by 374
Assessing Glycemia in Diabetes Using Self-monitoring Blood Glucose and Hemoglobin A 1c
Saudek, C. D., Derr, R. L., & Kalyani, R. R. (2006) · JAMA · doi:10.1001/jama.295.14.1688 · Cited by 373
Cited by
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
A history of blood glucose meters and their role in self-monitoring of diabetes mellitus
Clarke, S. F., & Foster, J. R. (2012) · British Journal of Biomedical Science · doi:10.1080/09674845.2012.12002443 · Cited by 456
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Boulet, G., Halpern, E. M., Lovblom, L. E., Weisman, A., Bai, J.-W., Eldelekli, D., Keenan, H. A., Brent, M. H., Paul, N., Bril, V., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0216 · Cited by 26
Metadata
- Authors
- Thomas, L. E., Kane, M. P., Bakst, G., Busch, R. S., Hamilton, R. A., & Abelseth, J. M.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2008-04
- Volume
- 10
- Issue
- 2
- Pages
- 102-110
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W1549956737
- Cited by
- 49
- References
- 4
- FWCI
- 3.05
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.