The Accuracy of Home Glucose Meters in Hypoglycemia
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Background: Home glucose meters (HGMs) may not be accurate enough to sense hypoglycemia. We evaluated the accuracy and the capillary and venous comparability of five different HGMs (Optium Xceed™ [Abbott Diabetes Care, Alameda, CA, USA], Contour ® TS [Bayer Diabetes Care, Basel, Switzerland], Accu-Chek ® Go [Roche Ltd., Basel, Switzerland], OneTouch ® Select™ [Lifescan, Milpitas, CA, USA], and EZ Smart [Tyson Bioresearch Inc., Chu-Nan, Taiwan]) in an adult population. Methods: The insulin hypoglycemia test was performed to 59 subjects (56 males; 23.6 ± 3.2 years old). Glucose was measured from forearm venous blood and finger capillary samples both before and after regular insulin (0.1 U/kg) was injected. Venous samples were analyzed in the reference laboratory by the hexokinase method. In vitro tests for method comparison and precision analyses were also performed by spiking the glucose-depleted venous blood. Results: All HGMs failed to sense hypoglycemia to some extend. EZ Smart was significantly inferior in critical error Zone D, and OneTouch Select was significantly inferior in the clinically unimportant error Zone B. Accu-Chek Go, Optium Xceed, and Contour TS had similar performances and were significantly better than the other two HGMs according to error grid analysis or International Organization for Standardization criteria. The in vitro tests were consistent with the above clinical data. The capillary and venous consistencies of Accu-Chek Go and OneTouch Select were better than the other HGMs. Conclusions: The present results show that not all the HGMs are accurate enough in low blood glucose levels. The patients and the caregivers should be aware of these restrictions of the HGMs and give more credit to the symptoms of hypoglycemia than the values obtained by the HGMs. Finally, these results indicate that there is a need for the revision of the accuracy standards of HGMs in low blood glucose levels.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Standards of Medical Care in Diabetes—2009
American Diabetes Association (2009) · Diabetes Care · doi:10.2337/dc09-s013 · Cited by 2,167
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Hypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes.
Cryer, P. (2002) · Diabetologia · doi:10.1007/s00125-002-0822-9 · Cited by 794 · cited 2×
“3,4 However, not all the hypoglycemic events are symptomatic, and HGMs may not be sensitive enough to detect hypoglycemia.”
Hypoglycemia in Type 2 Diabetes
Zammitt, N. N., & Frier, B. M. (2005) · Diabetes Care · doi:10.2337/diacare.28.12.2948 · Cited by 435
Hay, L. C., Wilmshurst, E. G., & Fulcher, G. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503763816427 · Cited by 143 · cited 1×
“22 However, HGMs may not be accurate enough to sense hypoglycemia.”
Evolution of analytical performance in portable glucose meters in the last decade
Böhme, P., Floriot, M., Sirveaux, M.-A., Durain, D., Ziegler, O., Drouin, P., & Guerci, B. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.1170 · Cited by 51
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
Cited by
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Performance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59
A review of blood glucose monitor accuracy
King, F., Ahn, D., Hsiao, V., Porco, T., & Klonoff, D. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0232 · Cited by 37
Hasslacher, C., Kulozik, F., & Platten, I. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814522804 · Cited by 22
Metadata
- Authors
- Sonmez, A., Yilmaz, Z., Uckaya, G., Kilic, S., Tapan, S., Taslipinar, A., Aydogdu, A., Yazici, M., Yilmaz, M. I., Serdar, M., Erbil, M. K., & Kutlu, M.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2010-08
- Volume
- 12
- Issue
- 8
- Pages
- 619-626
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2069519004
- Cited by
- 23
- References
- 7
- FWCI
- 1.76
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Accuracy of Home Glucose Meters in Hypoglycemia. https://app.cochat.ai/papers/10.1089/dia.2009.0183
- The paper
- Sonmez, A., Yilmaz, Z., Uckaya, G., Kilic, S., Tapan, S., Taslipinar, A., Aydogdu, A., Yazici, M., Yilmaz, M. I., Serdar, M., Erbil, M. K., & Kutlu, M. (2010). The Accuracy of Home Glucose Meters in Hypoglycemia. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2009.0183
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.