Measurement Performance of Two Continuous Tissue Glucose Monitoring Systems Intended for Replacement of Blood Glucose MonitoringParts of the data have previously been presented at the 77th Scientific Sessions of the American Diabetes Association in San Diego, CA; June 9–13, 2017 and at the 17th Annual Diabetes Technology Meeting in Bethesda, MD, November 2–4, 2017.Trial number: DRKS00011920; registered at the Deutsches Register Klinischer Studien (German clinical trials register), an approved Primary Register in the WHO International Clinical Trials Registry Platform.

Open access
Freckmann, Guido, Link, Manuela, Pleus, Stefan, Westhoff, Antje, Kamecke, Ulrike, Haug, Cornelia · 2018 · Diabetes Technology & Therapeutics
Cited by 64

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

BACKGROUND: G5 [DG5] and FreeStyle Libre [FL]) are intended to replace blood glucose monitoring (BGM) and, according to manufacturer labeling, are distributed as such in some jurisdictions, including the United States and the European Union. METHODS: The measurement performance of these two systems in comparison with a BGM system was analyzed in a 14-day study with 20 participants comprising study site visits, which included phases of induced rapid glucose changes, and home use phases. Performance analysis was mainly based on deviations between CGM readings and BGM results. Sensor-to-sensor precision was also analyzed. RESULTS: Approximately 25% of DG5 and FL results showed differences from BGM results exceeding 15 mg/dL or 15% (at glucose concentration below or above 100 mg/dL, respectively) at times of therapeutic decisions, and ∼5% of differences exceeded 30 mg/dL or 30%. Performance was different depending on the setting (study site visits, home use phases, and phases of induced rapid glucose changes). In consensus error grid (CEG) analysis, both systems showed >99.5% of results within the clinically acceptable zones A and B. CONCLUSIONS: In this study, both systems showed deviations from blood glucose (BG) measurements, the current standard approach in diabetes therapy. Although a large percentage of results was found in CEG zones A and B, for approximately one in four therapeutic decisions, CGM and BG readings differed by at least 15 mg/dL or 15%. Such deviations should be taken into account when using CGM systems.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2018.0105

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 7 references

Cited by

See all 64

Metadata

Authors
Freckmann, G., Link, M., Pleus, S., Westhoff, A., Kamecke, U., & Haug, C.
Venue
Diabetes Technology & Therapeutics
Published
2018-08
Volume
20
Issue
8
Pages
541-549
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2960340768
Cited by
64
References
7
FWCI
5.63
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Measurement Performance of Two Continuous Tissue Glucose Monitoring Systems Intended for Replacement of Blood Glucose MonitoringParts of the data have previously been presented at the 77th Scientific Sessions of the American Diabetes Association in San Diego, CA; June 9–13, 2017 and at the 17th Annual Diabetes Technology Meeting in Bethesda, MD, November 2–4, 2017.Trial number: DRKS00011920; registered at the Deutsches Register Klinischer Studien (German clinical trials register), an approved Primary Register in the WHO International Clinical Trials Registry Platform. https://app.cochat.ai/papers/10.1089/dia.2018.0105
The paper
Freckmann, G., Link, M., Pleus, S., Westhoff, A., Kamecke, U., & Haug, C. (2018). Measurement Performance of Two Continuous Tissue Glucose Monitoring Systems Intended for Replacement of Blood Glucose MonitoringParts of the data have previously been presented at the 77th Scientific Sessions of the American Diabetes Association in San Diego, CA; June 9–13, 2017 and at the 17th Annual Diabetes Technology Meeting in Bethesda, MD, November 2–4, 2017.Trial number: DRKS00011920; registered at the Deutsches Register Klinischer Studien (German clinical trials register), an approved Primary Register in the WHO International Clinical Trials Registry Platform. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2018.0105

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.