Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Open accessAbstract
Background: We assessed the performance of a modified Dexcom G4 Platinum system with an advanced algorithm, in comparison with frequent venous samples measured on a laboratory reference (YSI) during a clinic session and in comparison to self-monitored blood glucose (SMBG) during home use. Methods: Fifty-one subjects with diabetes were enrolled in a prospective multicenter study. Subjects wore 1 sensor for 7-day use and participated in one 12-hour in-clinic session on day 1, 4, or 7 to collect YSI reference venous glucose every 15 minutes and capillary SMBG test every 30 minutes. Carbohydrate consumption and insulin dosing and timing were manipulated to obtain data in low and high glucose ranges. Results: In comparison with the laboratory reference method (n = 2,263) the system provided a mean and median absolute relative differences (ARD) of 9.0% and 7.0%, respectively. The mean absolute difference for CGM was 6.4 mg/dL when the YSIs were within hypoglycemia ranges (≤ 70 mg/dL). The percentage in the clinically accurate Clarke error grid A zone was 92.4% and in the benign error B zone was 7.1%. Majority of the sensors (73%) had an aggregated MARD in reference to YSI ≤ 10%. The MARD of CGM-SMBG for home use was 11.3%. Conclusions: The study showed that the point and rate accuracy, clinical accuracy, reliability, and consistency over the duration of wear and across glycemic ranges were superior to current commercial real-time CGM systems. The performance of this CGM is reaching that of a self-monitoring blood glucose meter in real use environment.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
CGM utilizes subcutaneous sensors for continuous monitoring of glucose levels with a high degree of accuracy.
“CGM utilizes subcutaneous sensors for continuous monitoring of glucose levels with a high degree of accuracy. 17 , 18”
CGM and intermittently viewed CGM devices for at-home use have accuracy similar to capillary blood glucose monitors.
“continuous glucose monitoring (CGM) and intermittently viewed CGM devices for at-home patient use with minimally invasive devices have become available, which have similar accuracy to capillary blood glucose monitors.91”
From Type 1 diabetes (2018) · cites this paper as [91]
Errors in CGM increase the difficulty of predicting BG values (approximately 9% of mean absolute relative difference for best sensors).
“Errors in CGM also increase the difficulty of predicting BG values (approximately 9% of the mean absolute relative difference for the best sensors70).”
From Artificial intelligence in diabetes management: Advancements, opportunities, and challenges (2023) · cites this paper as [70]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · see citing works
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237 · see citing works
“Turn it Off!”: Diabetes Device Alarm Fatigue Considerations for the Present and the Future
Shivers, J. P., Mackowiak, L., Anhalt, H., & Zisser, H. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700324 · Cited by 159 · see citing works
A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors
Damiano, E. R., McKeon, K., El-Khatib, F. H., Zheng, H., Nathan, D. M., & Russell, S. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532203 · Cited by 139 · see citing works
Christiansen, M. P., Bailey, T. S., Watkins, E., Liljenquist, D. R., Price, D. A., Nakamura, K., Boock, R. J., & Peyser, T. A. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0077 · Cited by 136 · see citing works
Self-citationKropff, J., Bruttomesso, D., Doll, W., Farret, A., Galasso, S., Luijf, Y. M., Mader, J. K., Place, J., Boscari, F., Pieber, T. R., et al. (2014) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12378 · Cited by 132 · see citing works
Šoupal, J., Škrha, J., Fajmon, M., Horová, E., Mráz, M., Škrha, J., & Prázný, M. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0205 · Cited by 120 · see citing works
Mazze, R. S., Strock, E. S., Borgman, S. K., Wesley, D., Stout, P. J., & Racchini, J. R. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0041 · Cited by 116 · see citing works
Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose Sensor
Bailey, T. S., Ahmann, A., Brazg, R., Christiansen, M. P., Garg, S. K., Watkins, E., Welsh, J. B., & Lee, S. W. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0222 · Cited by 100 · see citing works
Self-citationTack, C. J., Pohlmeier, H., Behnke, T., Schmid, V., Grenningloh, M., Forst, T. M., & Pfützner, A. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0170 · Cited by 91 · see citing works
Graphical and Numerical Evaluation of Continuous Glucose Sensing Time Lag
Kovatchev, B. P., Shields, D., & Breton, M. D. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0044 · Cited by 91 · see citing works
Matuleviciene, V., Joseph, J. I., Andelin, M., Hirsch, I. B., Attvall, S., Pivodic, A., Dahlqvist, S., Klonoff, D. C., Haraldsson, B. S., & Lind, M. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0238 · Cited by 83 · see citing works
Real-Life Utilization of Real-Time Continuous Glucose Monitoring: The Complete Picture
Ramchandani, N., Arya, S., Ten, S., & Bhandari, S. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500407 · Cited by 80 · see citing works
Brazg, R. L., Klaff, L. J., & Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700117 · Cited by 68 · see citing works
Current Application of Continuous Glucose Monitoring in the Treatment of Diabetes
Hermanides, J., Phillip, M., & DeVries, J. H. (2011) · Diabetes Care · doi:10.2337/dc11-s219 · Cited by 57 · see citing works
Accuracy and Precision Evaluation of Seven Self-Monitoring Blood Glucose Systems
Kuo, C., Hsu, C., Ho, C.-S., Su, T., Wu, M., & Wang, C. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0223 · Cited by 55 · see citing works
Secher, A. L., Madsen, A. B., Ringholm, L., Barfred, C., Stage, E., Andersen, H. U., Damm, P., & Mathiesen, E. R. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03426.x · Cited by 50 · see citing works
Haruta, S., Hanafusa, T., Fukase, H., 宮嶌, 宏., & OKI, T. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503763816409 · Cited by 50 · see citing works
How to Assess and Compare the Accuracy of Continuous Glucose Monitors?
Wentholt, I. M. E., Hart, A. A. M., Hoekstra, J. B. L., & DeVries, J. H. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0216 · Cited by 49 · see citing works
Bay, C., Kristensen, P., Pedersen-Bjergaard, U., Tarnow, L., & Thorsteinsson, B. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0004 · Cited by 49 · see citing works
Dexcom G4AP: An Advanced Continuous Glucose Monitor for the Artificial Pancreas
Garcia, A., Rack-Gomer, A. L., Bhavaraju, N. C., Hampapuram, H., Kamath, A., Peyser, T. A., Facchinetti, A., Zecchin, C., Sparacino, G., & Cobelli, C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700604 · Cited by 47 · see citing works
Pleus, S., Schmid, C., Link, M., Zschornack, E., Klötzer, H.-M., Haug, C., & Freckmann, G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700405 · Cited by 34 · see citing works
Kirkpatrick, P. N. (2002) · Nature Reviews Drug Discovery · doi:10.1038/nrd948 · Cited by 18 · see citing works
(Analytical) Accuracy of Blood Glucose Meters and Patients: How Do They Come Together?
Heinemann, L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700101 · Cited by 5 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [15]15. Brauker JH, Edelman SV. The function of continuous glucose sensors. Review of Endocrinology , 2007; May 20-22.
Cited by
Sorted by citations
DiMeglio, L. A., Evans-Molina, C., & Oram, R. A. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31320-5 · Cited by 1,561 · see citing works
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · see citing works
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 890 · see citing works
Artificial intelligence for diabetes management and decision support: Literature review
Contreras, I., & Vehi, J. (2018) · Journal of Medical Internet Research · doi:10.2196/10775 · Cited by 560 · see citing works
DiMeglio, L. A., Acerini, C. L., Codner, E., Craig, M. E., Hofer, S. E., Pillay, K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12737 · Cited by 604 · see citing works
Cappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477 · see citing works
Polonsky, W. H., Hessler, D., Ruedy, K. J., & Beck, R. W. (2017) · Diabetes Care · doi:10.2337/dc17-0133 · Cited by 346 · see citing works
Aleppo, G., Ruedy, K. J., Riddlesworth, T. D., Kruger, D. F., Peters, A. L., Hirsch, I., Bergenstal, R. M., Toschi, E., Ahmann, A. J., Shah, V. N., et al. (2017) · Diabetes Care · doi:10.2337/dc16-2482 · Cited by 329 · see citing works
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320 · see citing works
Artificial intelligence in diabetes management: Advancements, opportunities, and challenges
Guan, Z., Li, H., Liu, R., Cai, C., Liu, Y., Li, J., Wang, X., Huang, S., Wu, L., Liu, D., et al. (2023) · Cell Reports Medicine · doi:10.1016/j.xcrm.2023.101213 · Cited by 287 · see citing works
“Errors in CGM also increase the difficulty of predicting BG values (approximately 9% of the mean absolute relative difference for the best sensors 70 ).”
BackgroundReddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245 · see citing works
Wearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 242 · see citing works
Outcome measures for artificial pancreas clinical trials: a consensus report
Maahs, D. M., Buckingham, B. A., Castle, J. R., Cinar, A., Damiano, E. R., Dassau, E., DeVries, J. H., Doyle, F. J., Griffen, S. C., Haidar, A., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2716 · Cited by 211 · see citing works
Wadwa, R. P., Laffel, L. M., Shah, V. N., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0150 · Cited by 197 · see citing works
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197 · see citing works
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
Kropff, J., Choudhary, P., Neupane, S., Barnard-Kelly, K. D., Bain, S. C., Kapitza, C., Forst, T. M., Link, M., DEHENNIS, A. D., & DeVries, J. H. (2016) · Diabetes Care · doi:10.2337/dc16-1525 · Cited by 171 · see citing works
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170 · see citing works
Technology in the management of type 1 diabetes mellitus — current status and future prospects
Tauschmann, M., & Hovorka, R. (2018) · Nature Reviews Endocrinology · doi:10.1038/s41574-018-0044-y · Cited by 161 · see citing works
Christiansen, M. P., Klaff, L. J., Brazg, R., Chang, A. R., Levy, C. J., Lam, D. W.-H., Denham, D. S., Atiee, G. J., Bode, B. W., Walters, S. J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0142 · Cited by 146 · see citing works
Self-citationA review of optical methods for continuous glucose monitoring
Jernelv, I. L., Mileńko, K., Fuglerud, S. S., Hjelme, D. R., Ellingsen, R., & Aksnes, A. (2018) · Applied Spectroscopy Reviews · doi:10.1080/05704928.2018.1486324 · Cited by 136 · see citing works
Aberer, F., Hajnsek, M., Rumpler, M., Zenz, S., Baumann, P. M., Elsayed, H., Puffing, A., Treiber, G. M., Pieber, T. R., Sourij, H., et al. (2017) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12907 · Cited by 132 · see citing works
Bailey, T. S., Grunberger, G., Bode, B. W., Handelsman, Y., Hirsch, I. B., Jovanovič, L., Roberts, V. L., Rodbard, D., Tamborlane, W. V., & Walsh, J. S. A. (2016) · Endocrine Practice · doi:10.4158/ep151124.cs · Cited by 126 · see citing works
Self-citationPettus, J., & Edelman, S. V. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663747 · Cited by 120 · see citing works
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2746340 · Cited by 110 · see citing works
HbA1c and Glucose Management Indicator Discordance: A Real-World Analysis
Perlman, J. E., Gooley, T., Mcnulty, B., Meyers, J., & Hirsch, I. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0501 · Cited by 80 · see citing works
Continuous glucose monitoring: Current use in diabetes management and possible future applications
Vettoretti, M., Cappon, G., Acciaroli, G., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818774078 · Cited by 96 · see citing works
Basu, A., Veettil, S., Dyer, R. B., Peyser, T. A., & Basu, R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0410 · Cited by 94 · see citing works
Wearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91 · see citing works
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86 · see citing works
Laffel, L. M. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0380 · Cited by 86 · see citing works
Sadhu, A. R., Serrano, I. A., Xu, J., Nisar, T., Lucier, J., Pandya, A. R., & Patham, B. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964264 · Cited by 84 · see citing works
Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose Swings
Pleus, S., Schoemaker, M., Morgenstern, K., Schmelzeisen‐Redeker, G., Haug, C., Link, M., Zschornack, E., & Freckmann, G. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815578716 · Cited by 84 · see citing works
Asarani, N. A. M., Reynolds, A. N., Styles, S. E., de Bock, M. I., & Wheeler, B. J. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819870849 · Cited by 81 · see citing works
Contreras, I., Oviedo, S., Vettoretti, M., VISENTIN, R., & Vehı́, J. (2017) · PLoS ONE · doi:10.1371/journal.pone.0187754 · Cited by 80 · see citing works
Long-term implantable glucose biosensors
Bobrowski, T., & Schuhmann, W. W. (2018) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2018.05.004 · Cited by 73 · see citing works
Welsh, J. B., Gao, P., Derdzinski, M., Puhr, S., Johnson, T. K., Walker, T. C., & Graham, C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0374 · Cited by 72 · see citing works
Optical sensors for continuous glucose monitoring
Yu, Z., Jiang, N., Kazarian, S. G., Taşoğlu, S., & Yetisen, A. K. (2021) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/abe6f8 · Cited by 67 · see citing works
Wood, A. J., O'Neal, D. N., Furler, J., & Ekinci, E. I. (2018) · Internal Medicine Journal · doi:10.1111/imj.13770 · Cited by 65 · see citing works
Basics and use of continuous glucose monitoring (CGM) in diabetes therapy
Freckmann, G. (2020) · Journal of Laboratory Medicine · doi:10.1515/labmed-2019-0189 · Cited by 66 · see citing works
Freckmann, G., Link, M., Pleus, S., Westhoff, A., Kamecke, U., & Haug, C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0105 · Cited by 64 · see citing works
Standardization process of continuous glucose monitoring: Traceability and performance
Freckmann, G., Nichols, J. H., Hinzmann, R., Klonoff, D. C., Ju, Y., Diem, P., Makris, K., & Slingerland, R. J. (2021) · Clinica Chimica Acta · doi:10.1016/j.cca.2020.12.025 · Cited by 62 · see citing works
Cappon, G., Vettoretti, M., Marturano, F., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818759558 · Cited by 62 · see citing works
Clinical Implications of Accuracy Measurements of Continuous Glucose Sensors
Bailey, T. S. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0050 · Cited by 60 · see citing works
Self-citationModeling Transient Disconnections and Compression Artifacts of Continuous Glucose Sensors
Facchinetti, A., Favero, S. D., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0250 · Cited by 59 · see citing works
Nonadjunctive Use of Continuous Glucose Monitoring for Diabetes Treatment Decisions
Castle, J. R., & Jacobs, P. G. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816631569 · Cited by 56 · see citing works
Eichenlaub, M., Pleus, S., Rothenbühler, M., Bailey, T. S., Bally, L., Brazg, R., Bruttomesso, D., Diem, P., Boija, E. E., Fokkert, M. J., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0465 · Cited by 51 · see citing works
Self-citationDeveloping Insulin Delivery Devices with Glucose Responsiveness
Wang, Z., Wang, J., Kahkoska, A. R., Buse, J. B., & Gu, Z. (2020) · Trends in Pharmacological Sciences · doi:10.1016/j.tips.2020.11.002 · Cited by 49 · see citing works
Avari, P., Reddy, M., & Oliver, N. S. (2019) · Diabetic Medicine · doi:10.1111/dme.13942 · Cited by 45 · see citing works
Forlenza, G. P., Argento, N. B., & Laffel, L. M. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0034 · Cited by 47 · see citing works
Li, A., Riddell, M. C., Potashner, D., Brown, R. E., & Aronson, R. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0387 · Cited by 44 · see citing works
Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems
Welsh, J. B., Psavko, S., Zhang, X., Gao, P., & Balo, A. K. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221099879 · Cited by 43 · see citing works
A Review of Safety and Hazards Associated With the Artificial Pancreas
Ramkissoon, C. M., Aufderheide, B., Bequette, B. W., & Vehı́, J. (2017) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2749038 · Cited by 41 · see citing works
Applications and pitfalls of hemoglobin A1C and alternative methods of glycemic monitoring
Lundholm, M. D., Emanuele, M. A., Ashraf, A., & Nadeem, S. (2020) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2020.107585 · Cited by 42 · see citing works
Avari, P., Uduku, C., George, D., Herrero, P., Reddy, M., & Oliver, N. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0276 · Cited by 40 · see citing works
Lupoli, R., Lembo, E., Ciciola, P., Schiavo, L., Pilone, V., & Capaldo, B. (2020) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2020.06.029 · Cited by 40 · see citing works
Lindner, N., Kuwabara, A., & Holt, T. (2021) · Systematic Reviews · doi:10.1186/s13643-021-01644-2 · Cited by 38 · see citing works
Peyser, T. A., Nakamura, K., Price, D. A., Bohnett, L. C., Hirsch, I. B., & Balo, A. K. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0415 · Cited by 37 · see citing works
Thabit, H., Leelarathna, L., Wilinska, M. E., Elleri, D., Allen, J. M., Lubina-Solomon, A., Walkinshaw, E., Stadler, M., Choudhary, P., Mader, J. K., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0062 · Cited by 37 · see citing works
Acciaroli, G., Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2706974 · Cited by 35 · see citing works
Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison
Freckmann, G., Link, M., Kamecke, U., Haug, C., Baumgartner, B., & Weitgasser, R. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819826965 · Cited by 34 · see citing works
Mariani, H., Layden, B. T., & Aleppo, G. (2017) · Clinical Diabetes · doi:10.2337/cd16-0008 · Cited by 33 · see citing works
Faccioli, S., Favero, S. D., VISENTIN, R., Bonfanti, R., Iafusco, D., Rabbone, I., Marigliano, M., Schiaffini, R., Bruttomesso, D., Cobelli, C., et al. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817706377 · Cited by 32 · see citing works
Unblinded CGM Should Replace Blinded CGM in the Clinical Management of Diabetes
Ahn, D. T., Pettus, J., & Edelman, S. V. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816632241 · Cited by 30 · see citing works
FDA Approval of Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing
Shapiro, A. R. (2017) · JAMA · doi:10.1001/jama.2017.12075 · Cited by 26 · see citing works
Shah, V. N., Snell‐Bergeon, J. K., Demmitt, J. K., Joshee, P., Garcetti, R., Pyle, L. L., & Polsky, S. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0093 · Cited by 25 · see citing works
Tissue Response to Subcutaneous Infusion Catheter
Zhang, E., & Cao, Z. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819837972 · Cited by 24 · see citing works
Link, M., Kamecke, U., Waldenmaier, D., Pleus, S., Garcia, A., Haug, C., & Freckmann, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819895022 · Cited by 23 · see citing works
Performance of a new real‐time continuous glucose monitoring system: A multicenter pilot study
Zhou, J., Zhang, S., Li, L., Wang, Y., Lu, W., Sheng, C., Li, Y., Bao, Y., & Jia, W. (2018) · Journal of Diabetes Investigation · doi:10.1111/jdi.12699 · Cited by 22 · see citing works
Acciaroli, G., Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0088 · Cited by 22 · see citing works
Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Is It Safe?
Shapiro, A. R. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816688303 · Cited by 22 · see citing works
Chaugule, S., & Graham, C. (2017) · Journal of Medical Economics · doi:10.1080/13696998.2017.1360312 · Cited by 22 · see citing works
Schiavon, M., Man, C. D., & Cobelli, C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0248 · Cited by 22 · see citing works
Hypoglycemia Reduction and Accuracy of Continuous Glucose Monitoring
Kovatchev, B. P. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0144 · Cited by 22 · see citing works
Wersäll, J. H., Adolfsson, P., Forsander, G., & Hanas, R. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13377 · Cited by 21 · see citing works
Longato, E., Acciaroli, G., Facchinetti, A., Hakaste, L., Tuomilehto, J., Maran, A., & Sparacino, G. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2018.03.007 · Cited by 19 · see citing works
A Model of Self-Monitoring Blood Glucose Measurement Error
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817698498 · Cited by 18 · see citing works
Joseph, J. I., EISLER, G. J., Diaz, D., Khalf, A., Loeum, C., & Torjman, M. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0106 · Cited by 18 · see citing works
Ghosh, J., Choudhury, S. R., Singh, K., & Koner, S. (2024) · International Journal of Advancement in Life Sciences Research · doi:10.31632/ijalsr.2024.v07i02.004 · Cited by 18 · see citing works
Sharifi, A., Varsavsky, A., Ulloa, J., Horsburgh, J. C., McAuley, S. A., Krishnamurthy, B., Jenkins, A. J., Colman, P. G., Ward, G. M., MacIsaac, R. J., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815612096 · Cited by 16 · see citing works
Camerlingo, N., Vettoretti, M., Sparacino, G., Facchinetti, A., Mader, J. K., Choudhary, P., & Favero, S. D. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14483 · Cited by 16 · see citing works
Methodology for Safe and Secure AI in Diabetes Management
Foppen, R. J. G., Gioia, V. L., Gupta, S., Johnson, C. L., Giantsidis, J., & Papademetris, M. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241304434 · Cited by 16 · see citing works
Hochfellner, D. A., Simic, A., Taucher, M. T., Sailer, L. S., Kopanz, J., Pöttler, T., & Mader, J. K. (2022) · Biosensors · doi:10.3390/bios12020106 · Cited by 15 · see citing works
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318868 · Cited by 15 · see citing works
Pettus, J., Gill, J., Paranjape, S., Stewart, J. A., Malla, S., Edelman, S. V., Bergenstal, R. M., & Bode, B. W. (2019) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13751 · Cited by 15 · see citing works
Price, D. A., & Walker, T. C. (2016) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2016.12.01.24 · Cited by 15 · see citing works
Graf, A., McAuley, S. A., Sims, C. M., Ulloa, J., Jenkins, A. J., Voskanyan, G. R., & O'Neal, D. N. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816682762 · Cited by 14 · see citing works
Data Reliability and Quality in Body Area Networks for Diabetes Monitoring
Huzooree, G., Khedo, K. K., & Joonas, N. (2018) · EAI/Springer Innovations in Communication and Computing · doi:10.1007/978-3-030-00865-9_4 · Cited by 13 · see citing works
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 13 · see citing works
Pleus, S., Link, M., Hinzmann, R., Öter, Ş., Eichenlaub, M., Baumstark, A., Haug, C., & Freckmann, G. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0489 · Cited by 12 · see citing works
Intraperitoneal Glucose Sensing is Sometimes Surprisingly Rapid
Fougner, A. L., Kölle, K., Skjærvold, N. K., Elvemo, N.-A. L., Hjelme, D. R., Ellingsen, R., Carlsen, S. M., & Stavdahl, Ø. (2016) · Modeling Identification and Control A Norwegian Research Bulletin · doi:10.4173/mic.2016.2.4 · Cited by 12 · see citing works
Muhs, A., Bobrowski, T., Lielpētere, A., & Schuhmann, W. W. (2022) · Angewandte Chemie International Edition · doi:10.1002/anie.202211559 · Cited by 12 · see citing works
Predicting Insulin Treatment Scenarios with the Net Effect Method: Domain of Validity
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0148 · Cited by 12 · see citing works
Jia, Z., Huang, L., Liu, H., Huang, Y., Li, W., Pi, X., & Zheng, X. (2019) · Applied Biochemistry and Biotechnology · doi:10.1007/s12010-019-03142-7 · Cited by 11 · see citing works
Camerlingo, N., Vettoretti, M., Sparacino, G., Facchinetti, A., Mader, J. K., Choudhary, P., & Favero, S. D. (2021) · Diabetic Medicine · doi:10.1111/dme.14758 · Cited by 10 · see citing works
Continuous Glucose Monitoring Use in Clinical Trials for On-Market Diabetes Drugs
Fox, B. Q., Benjamin, P. F., Aqeel, A., Fitts, E., Flynn, S., Levine, B., Maslak, E., Milner, R. L., Ose, B. M., Poeschla, M., et al. (2021) · Clinical Diabetes · doi:10.2337/cd20-0049 · Cited by 10 · see citing works
McAuley, S. A., Dang, T. T., Horsburgh, J. C., Bansal, A., Ward, G. M., Aroyan, S., Jenkins, A. J., MacIsaac, R. J., Shah, R. V., & O'Neal, D. N. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816629982 · Cited by 10 · see citing works
Continuous Glucose Monitoring for the Internist
Lee, G. S., & Lupsa, B. C. (2021) · Medical Clinics of North America · doi:10.1016/j.mcna.2021.06.004 · Cited by 9 · see citing works
Schiavon, M., Galderisi, A., Basu, A., Kudva, Y. C., Cengiz, E., & Dalla Man, C. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0397 · Cited by 8 · see citing works
Retrofitting Real-Life Dexcom G5 Data
Favero, S. D., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0413 · Cited by 8 · see citing works
- Authors
- Bailey, T. S., Chang, A., & Christiansen, M.
- Affiliations
- AMCR Institute (US)
- John Muir Health (US)
- Diablo Clinical Research (US)
- Corresponding author
- Bailey, Timothy S.
- ORCID
- Bailey, Timothy S.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-11-03
- Volume
- 9
- Issue
- 2
- Pages
- 209-214
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4604574
- OpenAlex
- W2165138941
- Cited by
- 168
- References
- 24
- Citations by year
- 2026: 7 · 2025: 7 · 2024: 9 · 2023: 7 · 2022: 7 · 2021: 14 · 2020: 15 · 2019: 23 · 2018: 21 · 2017: 30 · 2016: 18 · 2015: 10
- FWCI
- 15.54
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Dexcom
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- continuous glucose monitoring, diabetes mellitus, clinical accuracy, hypoglycemia, self-monitoring of blood glucose, Clarke error grid, mean absolute relative difference, sensor reliability, insulin dose
- MeSH terms
- Adult, Aged, Aged, 80 and over, Algorithms, Blood Glucose, Diabetes Mellitus, Female, Humans, Male, Middle Aged, Blood Glucose Self-Monitoring, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review