A Three-Way Accuracy Comparison of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense Continuous Glucose Monitoring Devices in a Home-Use Study of Subjects with Type 1 Diabetes
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Background: There is a dearth of comparative accuracy studies of continuous glucose monitoring (CGM) devices in the home-use setting, and none with the Eversense implantable CGM. Methods: We evaluated the accuracy of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense during a 6-week free-living home-use bionic pancreas study involving 23 subjects with type 1 diabetes who wore all three devices concurrently. The primary outcome was the mean absolute relative difference (MARD) between CGM readings and point-of-care (POC) plasma-glucose (PG) values obtained approximately twice daily by the subjects. We compared PG values with CGM readings when available from all three CGMs in the 5 min preceding the PG values ( n = 829 sets). Since the Libre Pro records readings every 15 min, we also did a two-way comparison between the G5 and the Eversense with a higher number of comparisons ( n = 2277 sets). Results: All three CGM systems produced higher average MARDs than during in-clinic studies. However, since all three CGM systems were worn by the same individuals and used the same meter for comparator PG measurements, we could directly compare their performances. In the three-way comparison, Eversense achieved the lowest nominal MARD (14.8%) followed by Dexcom G5 (16.3%) and Libre Pro (18.0%) (Eversense vs. Libre Pro P = 0.004, other comparisons P = NS). There was a statistically significant difference ( P = 0.008) in the two-way comparison of the MARDs for Eversense (15.1%) and G5 (16.9%). Conclusions: The point accuracy of the Eversense was better than two other CGMs when compared with POC PG values.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
Ólafsdóttir, A. F., Attvall, S., Sandgren, U., Dahlqvist, S., Pivodic, A., Skrtic, S., Theodorsson, E., & Lind, M. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0392 · Cited by 173
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127 · cited 2×
“14 Therefore, although the POC PG meter is less accurate than reference methods, this should affect the apparent accuracy of all the CGM systems equally since, to our knowledge, factory calibrations and calibrations algorithms of CGM devices use the YSI as the standard.”
Self-citation
Cited by
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Standardization process of continuous glucose monitoring: Traceability and performance
Freckmann, G., Nichols, J. H., Hinzmann, R., Klonoff, D. C., Ju, Y., Diem, P., Makris, K., & Slingerland, R. J. (2021) · Clinica Chimica Acta · doi:10.1016/j.cca.2020.12.025 · Cited by 62
Nørgaard, S. K., Mathiesen, E. R., Nørgaard, K., & Ringholm, L. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0109 · Cited by 38
“7,8 The international consensus recommendations on CGM data reporting recommends to report percentage of CGM readings as time-in-target range (TIR), time-above-target range (TAR), and time-below-target range (TBR).”
Gubitosi-Klug, R. A., Braffett, B. H., Bebu, I., Johnson, M. L., Farrell, K., Kenny, D., Trapani, V. R., Meadema-Mayer, L., Soliman, E. Z., Pop-Busui, R., et al. (2022) · Diabetes Care · doi:10.2337/dc21-0629 · Cited by 31
Metadata
- Authors
- Jafri, R. Z., Balliro, C. A., El-Khatib, F., Maheno, M. M., Hillard, M. A., O'Donovan, A., Selagamsetty, R., Zheng, H., Damiano, E. R., & Russell, S. J.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2020-11-01
- Volume
- 22
- Issue
- 11
- Pages
- 846-852
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W3032319542
- Cited by
- 90
- References
- 4
- FWCI
- 7.80
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.