Continuous Glucose Monitoring in Adults with Type 1 Diabetes: Real-World Data from the German/Austrian Prospective Diabetes Follow-Up Registry
The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
Background: To analyze key indicators of metabolic control in adults with type 1 diabetes (T1D) using real-time or intermittent scanning continuous glucose monitoring (rtCGM/iscCGM) during real-life care, based on the German/Austrian/Swiss Prospective Diabetes Follow-up (DPV) registry. Methods: Cross-sectional analysis including 233 adults with T1D using CGM. We assessed CGM metrics by gender, age group (18 to <30 years vs. ≥30 years), insulin delivery method (multiple daily injections vs. continuous subcutaneous insulin infusion [CSII]) and sensor type (iscCGM vs. rtCGM), working days versus weekends, and daytime versus night-time using multivariable linear regression models (adjusted for demographic variables) or Wilcoxon signed-rank test. Results: Overall, 79/21% of T1D patients used iscCGM/rtCGM. Those aged ≥30 years spent more time in range (TIR [70–180 mg/dL] 54% vs. 49%) and hypoglycemic range <70 mg/dL (7% vs. 5%), less time in hyperglycemic range >180 mg/dL (38% vs. 46%) and had a lower glucose variability (coefficient of variation [CV] 36% vs. 37%) compared with adults aged <30 years. We found no significant differences between genders. Multivariable regression models revealed the highest Time In Range (TIR) and lowest time with sensor glucose >250 mg/dL, CV and daytime–night-time differences in those treated with CSII and rtCGM. Glucose profiles were slightly more favorable on working days. Conclusions: In our real-world data, rtCGM versus iscCGM was associated with a higher percentage of TIR and improved metabolic stability. Differences in ambulatory glucose profiles on working and weekend days may indicate lifestyle habits affecting glycemic stability. Real-life CGM results should be included in benchmarking reports in addition to hemoglobin A1c (HbA1c) and history of hypoglycemia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,237 · cited 2×
“Our definition of ‘‘continuous’’ glucose monitoring as at least 50% of the 24-h period to be documented is more conservative than the international consensus recommendation ( > 70%), 13 but of course somewhat arbitrary.”
Self-citation6. Glycemic Targets: Standards of Medical Care in Diabetes—2020
American Diabetes Association (2020) · Diabetes Care · doi:10.2337/dc20-s006 · Cited by 744
the SWITCH Study Group, Battelino, T., Conget, I., Olsen, B., Schütz-Fuhrmann, I., Hommel, E., Hoogma, R., Schierloh, U., Sulli, N., & Bolinder, J. (2012) · Diabetologia · doi:10.1007/s00125-012-2708-9 · Cited by 494
The relationships between time in range, hyperglycemia metrics, and HbA1c
Beck, R. W., Bergenstal, R. M., Cheng, P., Kollman, C., Carlson, A. L., Johnson, M. L., & Rodbard, D. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818822496 · Cited by 479
Karges, B., Schwandt, A., Heidtmann, B., Kordonouri, O., Binder, E., Schierloh, U., Boettcher, C., Kapellen, T., Rosenbauer, J., & Holl, R. W. (2017) · JAMA · doi:10.1001/jama.2017.13994 · Cited by 452
Self-citationToward Defining the Threshold Between Low and High Glucose Variability in Diabetes
Monnier, L., Colette, C., Wojtusciszyn, A., Dejager, S., Renard, E., Molinari, N., & Owens, D. R. (2017) · Diabetes Care · doi:10.2337/dc16-1769 · Cited by 390
Improving the Clinical Value and Utility of CGM Systems: Issues and Recommendations
Petrie, J. R., Peters, A. L., Bergenstal, R. M., Holl, R. W., Fleming, G. A., & Heinemann, L. (2017) · Diabetes Care · doi:10.2337/dci17-0043 · Cited by 135
Self-citationŠoupal, J., Petruželková, L., Flekač, M., Pelcl, T., Matoulek, M., Daňková, M., Škrha, J., Svačina, Š., & Prázný, M. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0171 · Cited by 126 · cited 1×
“25,26 This refers to a recent meta-analysis 27 of randomized clinical trials comparing CSII versus MDI in T1D pointing out the advantage of CSII being significantly smaller when rapid-acting analogs were used instead of human insulin.”
Beck, R. W., Riddlesworth, T. D., Ruedy, K. J., Kollman, C., Ahmann, A. J., Bergenstal, R. M., Bhargava, A., Bode, B. W., Haller, S., Kruger, D. F., et al. (2017) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(17)30217-6 · Cited by 119
Benkhadra, K., Alahdab, F., Tamhane, S. U., McCoy, R. G., Prokop, L. J., & Murad, M. H. (2017) · Endocrine · doi:10.1007/s12020-016-1039-x · Cited by 113 · cited 1×
“25,26 This refers to a recent meta-analysis 27 of randomized clinical trials comparing CSII versus MDI in T1D pointing out the advantage of CSII being significantly smaller when rapid-acting analogs were used instead of human insulin.”
Pala, L., Dicembrini, I., & Mannucci, E. (2019) · Acta Diabetologica · doi:10.1007/s00592-019-01326-5 · Cited by 112 · cited 2×
“25,26 This refers to a recent meta-analysis 27 of randomized clinical trials comparing CSII versus MDI in T1D pointing out the advantage of CSII being significantly smaller when rapid-acting analogs were used instead of human insulin.”
Frequent monitoring of A1c during pregnancy as a treatment tool to guide therapy
Jovanovič, L., Savas, H., Mehta, M., Trujillo, A., & Pettitt, D. J. (2011) · Diabetes Care · doi:10.2337/dc10-1455 · Cited by 60
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Visser, M. M., Charleer, S., Fieuws, S., De Block, C., Hilbrands, R., Van Huffel, L., Maes, T., Vanhaverbeke, G., Dirinck, E., Myngheer, N., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(21)00789-3 · Cited by 165
Metadata
- Authors
- Sandig, D., Grimsmann, J., Reinauer, C., Melmer, A., Zimny, S., Müller-Korbsch, M., Forestier, N., Zeyfang, A., Bramlage, P., Danne, T., Meissner, T., & Holl, R. W.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2020-08-01
- Volume
- 22
- Issue
- 8
- Pages
- 602-612
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W3033048454
- Cited by
- 43
- References
- 12
- FWCI
- 5.25
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.