Glycemic Outcomes During Early Use of the MiniMed™ 780G Advanced Hybrid Closed-Loop System with Guardian™ 4 Sensor
Open accessThe DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Background: Safety and significant improvement in overall glycated hemoglobin (A1C) and percentage of time spent in (TIR), below (TBR), and above (TAR) glucose range were demonstrated in the pivotal trial of adolescents and adults using the MiniMed™ advanced hybrid closed-loop (AHCL) system with the adjunctive, calibration-required Guardian™ Sensor 3. The present study evaluated early outcomes of continued access study (CAS) participants who transitioned from the pivotal trial investigational system to the approved MiniMed™ 780G system with the non-adjunctive, calibration-free Guardian™ 4 Sensor (MM780G+G4S). Study data were presented alongside those of real-world MM780G+G4S users from Europe, the Middle East, and Africa. Methods: The CAS participants ( N = 109, aged 7–17 years and N = 67, aged >17 years) used the MM780G+G4S for 3 months and data of real-world MM780G+G4S system users ( N = 10,204 aged ≤15 years and N = 26,099 aged >15 years) were uploaded from September 22, 2021 to December 02, 2022. At least 10 days of real-world continuous glucose monitoring (CGM) data were required for analyses. Glycemic metrics, delivered insulin and system use/interactions underwent descriptive analyses. Results: Time in AHCL and CGM use were >90% for all groups. AHCL exits averaged 0.1/day and there were few blood glucose measurements (BGMs) (0.8/day–1.0/day). Adults in both cohorts met most consensus recommendations for glycemic targets. Pediatric groups met recommendations for %TIR and %TBR, although not those for mean glucose variability and %TAR, possibly due to low use of recommended glucose target (100 mg/dL) and active insulin time (2 h) settings (28.4% in the CAS cohort and 9.4% in the real-world cohort). The CAS pediatric and adult A1C were 7.2% ± 0.7% and 6.8% ± 0.7%, respectively, and there were no serious adverse events. Conclusions: Early clinical use of the MM780G+G4S was safe and involved minimal BGMs and AHCL exits. Consistent with real-world pediatric and adult use, outcomes were associated with achievement of recommended glycemic targets. Clinical Trial Registration number: NCT03959423
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,237
Self-citationInternational consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
Self-citationSix-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030
Self-citationGarg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585 · cited 1×
“0% (3710 paired points) in the MiniMed 670G system pivotal trial.(35) The GS3 with calibration-free algorithm (i.”
Self-citationPhillip, M., Nimri, R., Bergenstal, R. M., Barnard-Kelly, K., Danne, T., Hovorka, R., Kovatchev, B. P., Messer, L. H., Parkin, C. G., Ambler-Osborn, L., et al. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac022 · Cited by 443
Self-citation7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
Khunti, K., Alsifri, S., Aronson, R., Cigrovski Berković, M., Enters‐Weijnen, C., Forsén, T., Galstyan, G., Geelhoed‐Duijvestijn, P., Goldfracht, M., Gydesen, H., et al. (2016) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12689 · Cited by 276
One Year Real-World Use of the Control-IQ Advanced Hybrid Closed-Loop Technology
Breton, M. D., & Kovatchev, B. P. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0097 · Cited by 250
Carlson, A. L., Sherr, J. L., Shulman, D. I., Garg, S. K., Pop-Busui, R., Bode, B. W., Lilenquist, D. R., Brazg, R. L., Kaiserman, K. B., Kipnes, M. S., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0319 · Cited by 230
Self-citationReal-World Performance of the MiniMed™ 780G System: First Report of Outcomes from 4120 Users
Silva, J. D., Lepore, G., Battelino, T., Arrieta, A., Castañeda, J., Grossman, B., Shin, J., & Cohen, O. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0203 · Cited by 229
Self-citation
Cited by
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Real-world evidence of automated insulin delivery system use
Considine, E. G., & Sherr, J. L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0442 · Cited by 59
Self-citationAkturk, H. K., American Diabetes Association Diabetes Technology Interest Group, Aloi, J., Shah, V. N., Akturk, H. K., McKee, A. M., Ekhlaspour, L., Kim, S., & Everett, E. (2024) · Clinical Diabetes · doi:10.2337/cd23-0080 · Cited by 16
Metadata
- Authors
- Cordero, T. L., Dai, Z., Arrieta, A., Niu, F., Vella, M., Shin, J., Rhinehart, A. S., McVean, J., Lee, S. W., Slover, R. H., Forlenza, G. P., Shulman, D. I., Pop-Busui, R., Thrasher, J. R., Kipnes, M. S., Christiansen, M. P., Buckingham, B. A., Pihoker, C., Sherr, J. L., Kaiserman, K. B., Vigersky, R. A., & for the MiniMed™ AHCL Study Group
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2023-09-01
- Volume
- 25
- Issue
- 9
- Pages
- 652-658
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W4378746521
- Cited by
- 30
- References
- 16
- FWCI
- 4.96
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Glycemic Outcomes During Early Use of the MiniMed™ 780G Advanced Hybrid Closed-Loop System with Guardian™ 4 Sensor. https://app.cochat.ai/papers/10.1089/dia.2023.0123
- The paper
- Cordero, T. L., Dai, Z., Arrieta, A., Niu, F., Vella, M., Shin, J., Rhinehart, A. S., McVean, J., Lee, S. W., Slover, R. H., Forlenza, G. P., Shulman, D. I., Pop-Busui, R., Thrasher, J. R., Kipnes, M. S., Christiansen, M. P., Buckingham, B. A., Pihoker, C., Sherr, J. L., . . . for the MiniMed™ AHCL Study Group. (2023). Glycemic Outcomes During Early Use of the MiniMed™ 780G Advanced Hybrid Closed-Loop System with Guardian™ 4 Sensor. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2023.0123
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.