Validity of continuous glucose monitoring for categorizing glycemic responses to diet: implications for use in personalized nutrition

Open access
Merino, Jordi, Linenberg, Inbar, Bermingham, Kate M, Ganesh, Sajaysurya, Bakker, Elco, Delahanty, Linda M, Chan, Andrew T, Capdevila Pujol, Joan +7 more·2022·The American Journal of Clinical Nutrition

Abstract

BACKGROUND: Continuous glucose monitor (CGM) devices enable characterization of individuals' glycemic variation. However, there are concerns about their reliability for categorizing glycemic responses to foods that would limit their potential application in personalized nutrition recommendations. OBJECTIVES: We aimed to evaluate the concordance of 2 simultaneously worn CGM devices in measuring postprandial glycemic responses. METHODS: Within ZOE PREDICT (Personalised Responses to Dietary Composition Trial) 1, 394 participants wore 2 CGM devices simultaneously [n = 360 participants with 2 Abbott Freestyle Libre Pro (FSL) devices; n = 34 participants with both FSL and Dexcom G6] for ≤14 d while consuming standardized (n = 4457) and ad libitum (n = 5738) meals. We examined the CV and correlation of the incremental area under the glucose curve at 2 h (glucoseiAUC0-2 h). Within-subject meal ranking was assessed using Kendall τ rank correlation. Concordance between paired devices in time in range according to the American Diabetes Association cutoffs (TIRADA) and glucose variability (glucose CV) was also investigated. RESULTS: The CV of glucoseiAUC0-2 h for standardized meals was 3.7% (IQR: 1.7%-7.1%) for intrabrand device and 12.5% (IQR: 5.1%-24.8%) for interbrand device comparisons. Similar estimates were observed for ad libitum meals, with intrabrand and interbrand device CVs of glucoseiAUC0-2 h of 4.1% (IQR: 1.8%-7.1%) and 16.6% (IQR: 5.5%-30.7%), respectively. Kendall τ rank correlation showed glucoseiAUC0-2h-derived meal rankings were agreeable between paired CGM devices (intrabrand: 0.9; IQR: 0.8-0.9; interbrand: 0.7; IQR: 0.5-0.8). Paired CGMs also showed strong concordance for TIRADA with a intrabrand device CV of 4.8% (IQR: 1.9%-9.8%) and an interbrand device CV of 3.2% (IQR: 1.1%-6.2%). CONCLUSIONS: Our data demonstrate strong concordance of CGM devices in monitoring glycemic responses and suggest their potential use in personalized nutrition.This trial was registered at clinicaltrials.gov as NCT03479866.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/ajcn/nqac026

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Baseline glucose CV in ITT data was higher than previous reports demonstrating strong reproducibility and concordance between devices within participants of 3.7% median CV (IQR 1.7%-7.1%).

    “Review of the intent-to-treat (ITT) baseline glucose CGM data exhibited higher coefficient of variation (CV) than previous reports demonstrating a strong reproducibility and concordance between devices within participants of 3.7% median CV (IQR 1.7%-7.1%)(20).”

    From Prebiotic soda lowers postprandial glucose compared to traditional soda pop: a randomized controlled trial (2025) · cites this paper as [20]

    The Results (Concordance in glycemic responses) report a median 3.7% CV (IQR 1.7%–7.1%) for agreement between same-brand devices measuring glucose response after standardized meals. That figure is not a baseline glucose CV, and the reported within-person variability for the same meal on different days is substantially higher. The citing paper's ITT baseline comparison is its own finding, not one established by this source. (LLM, conf=0.88)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 23

  1. [1]Zeevi D, Korem T, Zmora N, Israeli D, Rothschild D, Weinberger A, Ben-Yacov O, Lador D, Avnit-Sagi T, Lotan-Pompan M et al. Personalized nutrition by prediction of glycemic responses. Cell. 2015; 163(5): 1079– 94. 26590418
  2. [3]Mendes-Soares H, Raveh-Sadka T, Azulay S, Edens K, Ben-Shlomo Y, Cohen Y, Ofek T, Bachrach D, Stevens J, Colibaseanu D et al. Assessment of a personalized approach to predicting postprandial glycemic responses to food among individuals without diabetes. JAMA Network Open. 2019; 2(2): e188102. 30735238 PMC6484621
  3. [4]Berry SE, Valdes AM, Drew DA, Asnicar F, Mazidi M, Wolf J, Capdevila J, Hadjigeorgiou G, Davies R, Al Khatib H et al. Human postprandial responses to food and potential for precision nutrition. Nat Med. 2020; 26(6): 964– 73. 32528151 PMC8265154
  4. [5]Ordovas JM, Ferguson LR, Tai ES, Mathers JC. Personalised nutrition and health. BMJ. 2018; 361: k2173. PMC6081996 29898881
  5. [6]Wang DD, Hu FB. Precision nutrition for prevention and management of type 2 diabetes. Lancet Diabetes Endocrinol. 2018; 6(5): 416– 26. 29433995
  6. [7]Rodgers GP, Collins FS. Precision nutrition—the answer to “what to eat to stay healthy.”. JAMA. 2020; 324(8): 735– 6. 32766768
  7. [9]Kearney ML, Thyfault JP. Exercise and postprandial glycemic control in type 2 diabetes. Curr Diabetes Rev. 2016; 12(3): 199– 210. 26073704
  8. [11]de Andrade Mesquita L, Pavan Antoniolli L, Cittolin-Santos GF, Gerchman F. Distinct metabolic profile according to the shape of the oral glucose tolerance test curve is related to whole glucose excursion: a cross-sectional study. BMC Endocr Disord. 2018; 18(1): 56. 30115058 PMC6097323
  9. [13]Monnier L, Mas E, Ginet C, Michel F, Villon L, Cristol JP, Colette C. Activation of oxidative stress by acute glucose fluctuations compared with sustained chronic hyperglycemia in patients with type 2 diabetes. JAMA. 2006; 295(14): 1681– 7. 16609090
  10. [14]Monnier L, Colette C, Owens DR. Glycemic variability: the third component of the dysglycemia in diabetes. Is it important? How to measure it?. J Diabetes Sci Technol. 2008; 2(6): 1094– 100. 19885298 PMC2769808
  11. [16]Hilliard ME, Levy W, Anderson BJ, Whitehouse AL, Commissariat PV, Harrington KR, Laffel LM, Miller KM, Van Name M, Tamborlane WV et al. Benefits and barriers of continuous glucose monitoring in young children with type 1 diabetes. Diabetes Technol Ther. 2019; 21(9): 493– 8. 31287721 PMC6708264
  12. [17]Howard R, Guo J, Hall KD. Imprecision nutrition? Different simultaneous continuous glucose monitors provide discordant meal rankings for incremental postprandial glucose in subjects without diabetes. Am J Clin Nutr. 2020; 112(4): 1114– 19. 32766882 PMC7528568
  13. [18]Asnicar F, Berry SE, Valdes AM, Nguyen LH, Piccinno G, Drew DA, Leeming E, Gibson R, Le Roy C, Al Khatib H et al. Microbiome connections with host metabolism and habitual diet from 1,098 deeply phenotyped individuals. Nat Med. 2021; 27(2): 321– 32. 33432175 PMC8353542
  14. [19]Steineck IIK, Mahmoudi Z, Ranjan A, Schmidt S, Jørgensen JB, Nørgaard K. Comparison of continuous glucose monitoring accuracy between abdominal and upper arm insertion sites. Diabetes Technol Ther. 2019; 21(5): 295– 302. 30994362
  15. [20]Basu A, Dube S, Veettil S, Slama M, Kudva YC, Peyser T, Carter RE, Cobelli C, Basu R. Time lag of glucose from intravascular to interstitial compartment in type 1 diabetes. J Diabetes Sci Technol. 2015; 9(1): 63– 8. 25305282 PMC4495531
  16. [22]Wadwa RP, Laffel LM, Shah VN, Garg SK. Accuracy of a factory-calibrated, real-time continuous glucose monitoring system during 10 days of use in youth and adults with diabetes. Diabetes Technol Ther. 2018; 20(6): 395– 402. 29901421 PMC6110124
  17. [25]Moser O, Riddell MC, Eckstein ML, Adolfsson P, Rabasa-Lhoret R, van den Boom L, Gillard P, Nørgaard K, Oliver NS, Zaharieva DP et al. Glucose management for exercise using continuous glucose monitoring (CGM) and intermittently scanned CGM (isCGM) systems in type 1 diabetes: position statement of the European Association for the Study of Diabetes (EASD) and of the International Society for Pediatric and Adolescent Diabetes (ISPAD) endorsed by JDRF and supported by the American Diabetes Association (ADA). Diabetologia. 2020; 63(12): 2501– 20. 33047169
  18. [26]Zschornack E, Schmid C, Pleus S, Link M, Klötzer HM, Obermaier K, Schoemaker M, Strasser M, Frisch G, Schmelzeisen-Redeker G et al. Evaluation of the performance of a novel system for continuous glucose monitoring. J Diabetes Sci Technol. 2013; 7(4): 815– 23. 23911162 PMC3879745
  19. [27]Pleus S, Schoemaker M, Morgenstern K, Schmelzeisen-Redeker G, Haug C, Link M, Zschornack E, Freckmann G. Rate-of-change dependence of the performance of two CGM systems during induced glucose swings. J Diabetes Sci Technol. 2015; 9(4): 801– 7. 25852074 PMC4525645
  20. [28]Freckmann G, Pleus S, Link M, Zschornack E, Kloẗzer HM, Haug C. Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel. J Diabetes Sci Technol. 2013; 7(4): 842– 53. 23911165 PMC3879748
  21. [30]Herbert M, Pendyal S, Rairikar M, Halaby C, Benjamin RW, Kishnani PS. Role of continuous glucose monitoring in the management of glycogen storage disorders. J Inherit Metab Dis. 2018; 41(6): 917– 27. 29802555
  22. [31]Davis GM, Spanakis EK, Migdal AL, Singh LG, Albury B, Urrutia MA, Zamudio-Coronado KW, Scott WH, Doerfler R, Lizama S et al. Accuracy of Dexcom G6 continuous glucose monitoring in non–critically ill hospitalized patients with diabetes. Diabetes Care. 2021; 44(7): 1641– 6. 34099515 PMC8323182
  23. [32]Gecili E, Huang R, Khoury JC, King E, Altaye M, Bowers K, Szczesniak RD. Functional data analysis and prediction tools for continuous glucose-monitoring studies. J Clin Transl Sci. 2021: 5(1): e51. PMC8057494 33948272

Cited by

Sorted by citations

Authors
Merino, J., Linenberg, I., Bermingham, K. M., Ganesh, S., Bakker, E., Delahanty, L. M., Chan, A. T., Capdevila Pujol, J., Wolf, J., Al Khatib, H., Franks, P. W., Spector, T. D., Ordovas, J. M., Berry, S. E., & Valdes, A. M.
Affiliations
  • Broad Institute (US)
  • ZOE (United Kingdom) (GB)
  • King's College London (GB)
  • Harvard University (US)
  • University of Nottingham (GB)
Corresponding author
Berry, Sarah E
Venue
The American Journal of Clinical Nutrition
Publisher
Elsevier BV
Published
2022-06
Volume
115
Issue
6
Pages
1569-1576
ISSN
0002-9165 (ISSN-L)
OpenAlex
W4212862774
Cited by
46
References
7
Citations by year
2026: 15 · 2025: 9 · 2024: 11 · 2023: 8 · 2022: 3
FWCI
5.50
Type
journal article
Language
en
Open access
Locations
  • American Journal of Clinical Nutrition (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Research Portal (King's College London) (publishedVersion, other-oa)
  • DIGITAL.CSIC (Spanish National Research Council (CSIC)) (submittedVersion, cc-by)
  • Lund University Publications (Lund University) (submittedVersion)
Funders
  • Wellcome Trust
  • Medical Research Council
  • Versus Arthritis
  • Chronic Disease Research Foundation
  • NIHR
  • CRN
  • Biotechnology and Biological Sciences Research Council (BB/NO12739/1)
  • MGH (R35 CA253185)
  • PWF
  • European Research Council (CoG-2015_681742_NASCENT)
  • Swedish Research Council
  • Novo Nordisk Foundation
  • Swedish Foundation for Strategic Research
  • National Institute for Health Research
  • European Union
  • Biomedical Research Centre
  • American Diabetes Association (#7-21-JDFM-005)
  • NIH (P30 DK40561)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Genetics › Nutrition, Genetics, and Disease
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes, Cardiovascular Risks, and Lipoproteins
Keywords
continuous glucose monitoring, postprandial glycemic response, personalized nutrition, glycemic variability, time in range, Dexcom G6, ad libitum meal, device reliability, personalized dietary recommendations
MeSH terms
Blood Glucose, Diet, Glucose, Humans, Blood Glucose Self-Monitoring, Reproducibility of Results, Meals

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review