Continuous Glucose Monitoring Profiles in Healthy Nondiabetic Participants: A Multicenter Prospective Study
Open accessAbstract
Context: Use of continuous glucose monitoring (CGM) is increasing for insulin-requiring patients with diabetes. Although data on glycemic profiles of healthy, nondiabetic individuals exist for older sensors, assessment of glycemic metrics with new-generation CGM devices is lacking. Objective: To establish reference sensor glucose ranges in healthy, nondiabetic individuals across different age groups using a current generation CGM sensor. Design: Multicenter, prospective study. Setting: Twelve centers within the T1D Exchange Clinic Network. Patients or Participants: Nonpregnant, healthy, nondiabetic children and adults (age ≥6 years) with nonobese body mass index. Intervention: Each participant wore a blinded Dexcom G6 CGM, with once-daily calibration, for up to 10 days. Main Outcome Measures: CGM metrics of mean glucose, hyperglycemia, hypoglycemia, and glycemic variability. Results: A total of 153 participants (age 7 to 80 years) were included in the analyses. Mean average glucose was 98 to 99 mg/dL (5.4 to 5.5 mmol/L) for all age groups except those over 60 years, in whom mean average glucose was 104 mg/dL (5.8 mmol/L). The median time between 70 to 140 mg/dL (3.9 to 7.8 mmol/L) was 96% (interquartile range, 93 to 98). Mean within-individual coefficient of variation was 17 ± 3%. Median time spent with glucose levels >140 mg/dL was 2.1% (30 min/d), and median time spent with glucose levels <70 mg/dL (3.9 mmol/L) was 1.1% (15 min/d). Conclusion: By assessing across age groups in a healthy, nondiabetic population, normative sensor glucose data have been derived and will be useful as a benchmark for future research studies.
AI snapshot
Written from the paper’s full text.
TL;DR
In 153 healthy nondiabetic children and adults aged 7 to 80 years, blinded Dexcom G6 CGM showed a mean 24-hour glucose of 99 ± 7 mg/dL and median 96% time in 70 to 140 mg/dL. Adults ≥60 years had higher mean glucose and less time in range than younger participants.
Key findings
- Overall mean 24-hour CGM glucose was 99 ± 7 mg/dL, mean within-individual coefficient of variation was 17 ± 3%, and median time between 70 and 140 mg/dL was 96% (IQR 93–98).
- Mean average glucose was 98–99 mg/dL for all age groups except those ≥60 years, in whom it was 104 mg/dL (P < 0.001); median time in 70–140 mg/dL was lowest at 93% in adults ≥60 years (P < 0.001).
- Median time >140 mg/dL was 2.1% (30 min/d) and <70 mg/dL was 1.1% (15 min/d); values >180 mg/dL and <54 mg/dL were uncommon, but 28% of participants had at least one hypoglycemic event overall and 14% had an overnight event.
- The correlation between mean glucose and HbA1c was weak (r = 0.27; 95% CI, 0.12–0.41).
Methods
- Multicenter, prospective study at 12 T1D Exchange Clinic Network centers; participants were healthy, nondiabetic, nonpregnant children and adults age ≥6 years with nonobese BMI, HbA1c <5.7%, and no islet autoantibodies. Of 201 screened, 153 were analyzed (age 7–80 years; mean 31 years).
- Participants wore a blinded Dexcom G6 CGM for up to 10 days with once-daily calibration; ≥72 total hours of CGM data with ≥24 overnight hours were required. Mean CGM use was 192 hours (8 days), and MARD versus blood glucose meter was 6.0% (n = 1555 paired measurements).
- Glucose outcomes were summarized overall and by age group and by daytime/nighttime; analyses included t tests for ≥60 vs <60 years, Pearson correlation for mean glucose and HbA1c, and sensitivity analyses removing implausible values.
Limitations
- The paper states that 93% of participants were white, so it could not evaluate whether results were consistent across races; possible sensor pressure during sleep may have caused inaccurate CGM readings, and the study could not determine how many nocturnal hypoglycemic events were due to pressure on the sensor.
- The paper states that after removing implausible data, mean glucose, time in 70–140 mg/dL, and CV were unchanged, but the percentage of participants with a hypoglycemic event was much lower; results may not be generalizable to capillary or laboratory-measured glucose profiles.
- The paper notes the weak HbA1c–glucose correlation may relate to HbA1c being obtained before CGM data collection and to having at most 10 days of CGM data. Evident from the text: participants were primarily recruited from family, friends, and neighbors of patients seen in diabetes clinics, which may limit generalizability to the broader nondiabetic population.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
13 of 17 cited sources checked · 7 supported · 0 flagged · accuracy not yet measured
- Interpretationnone given
The overall results of this study are in agreement with previous studies despite the different generations of CGM systems used in most of the studies.
“The overall results of our study are in agreement with previous studies (3)(4)(5)(6)(7)17) despite the different generations of CGM systems used in most of the studies.”
Quote in the paperNot a substantial claim3 of 3 sources checked- cites [3] Borg, 2010, Real-life glycaemic profiles in non-diabetic individuals with low fasting glucose and normal HbA1c: the A1C-Derived Average Glucose (ADAG) study, Diabetologia source
The Discussion says previous smaller CGM studies found similar glucose patterns, supporting the claim of overall agreement. The cited paper does not compare generations of CGM systems, so it does not support that qualification. (LLM, conf=0.83)
Experimental — accuracy not yet measured - cites [4] Freckmann, 2007, Continuous glucose profiles in healthy subjects under everyday life conditions and after different meals, J Diabetes Sci Technol source
- cites [5] Fox, 2010, Variation of interstitial glucose measurements assessed by continuous glucose monitors in healthy, nondiabetic individuals, Diabetes Care source
- Interpretationcitation
The correlation between HbA1c and glucose is lower than what others have found.
“However, the correlation between HbA1c and glucose is lower than what others have found (15).”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [15] Nathan, 2008, Translating the A1C assay into estimated average glucose values, Diabetes Care source
- Findingown data
Overall, median time spent with glucose levels >140 mg/dL was 2.1% (30 min/d).
“Overall, median time spent with glucose levels .140mg/dL was 2.1% (30 min/d)”
Quote in the paperNot a substantial claim - Findingown data
Overall, median time spent with glucose levels <70 mg/dL was 1.1% (15 min/d).
“and ,70 mg/dL was 1.1% (15 min/d).”
Quote in the paperNot a substantial claim - Findingown data
Median sensor glucose between 70 and 140 mg/dL during nighttime was 99% (IQR, 95 to 100).
“and 99% (IQR, 95 to 100), respectively”
Quote in the paperNot a substantial claim
- Findingown data
The mean within-individual coefficient of variation (CV) was 17 ± 3%.
“the mean of the within-individual coefficient of variation (CV), a metric of glucose variability, was 17 6 3% (Table 2).”
Quote in the paperNot a substantial claim - Findingown data
Only 1% of participants spent ≥2% of time <54 mg/dL.
“but only 1% of participants spent $2% of time ,54 mg/dL.”
Quote in the paperNot a substantial claim - Findingown data
14% of participants had a hypoglycemic event overnight.
“Yet, 14% of participants had a hypoglycemic event overnight (Table 3).”
Quote in the paperNot a substantial claim - Findingown data
The percentage of time between 70 to 120 mg/dL was quite different for daytime vs nighttime in the ≥60 age group (78% vs 92%, respectively).
“The percentage of time between 70 to 120 mg/dL was quite different for daytime vs nighttime in the 601 age group (78% vs 92%, respectively).”
Quote in the paperNot a substantial claim - Findingown data
Overall, the correlation was weak (r = 0.27; 95% CI, 0.12 to 0.41) between mean glucose concentration and HbA1c levels.
“Overall, the correlation was weak (r 5 0.27; 95% confidence interval, 0.12 to 0.41) between mean glucose concentration and HbA1c levels”
Quote in the paperNot a substantial claim - Findingown data
28% of participants had at least one hypoglycemic event.
“However, 28% of participants had at least one hypoglycemic event.”
Quote in the paperNot a substantial claim - Findingown data
In this healthy population, even at the same HbA1c, mean glucose was variable.
“In this healthy population, even at the same HbA1c, mean glucose was variable.”
Quote in the paperSubstantial claim - Findingown data
Overnight median percentage of glucose values <70 mg/dL was 0.4% (IQR, 0.0 to 2.5).
“and 0.4% (IQR, 0.0 to 2.5), respectively.”
Quote in the paperNot a substantial claim - Findingown data
Mean average glucose during nighttime was 98 ± 9 mg/dL (5.4 ± 0.5 mmol/L).
“and mean average glucose during nighttime was 98 6 9 mg/ dL (5.4 6 0.5 mmol/L).”
Quote in the paperNot a substantial claim - Findingown data
Overall, 51% of participants spent ≥2% of time with sensor glucose >140 mg/dL.
“Overall, 51% of participants spent $2% of time with sensor glucose .140mg/dL”
Quote in the paperNot a substantial claim - Findingown data
Overall, the mean of the individual average 24-hour glucose was 99 ± 7 mg/dL (5.5 ± 0.4 mmol/L).
“Overall, the mean of the individual average 24-hour glucose was 99 6 7 mg/dL (5.5 6 0.4 mmol/L)”
Quote in the paperSubstantial claim - Findingown data
Overall, 35% of participants spent ≥2% of time with sensor glucose <70 mg/dL (almost 30 min/d).
“Overall, 35% of participants spent $2% of time with sensor glucose ,70 mg/dL (almost 30 min/d)”
Quote in the paperSubstantial claim - Findingown data
Mean HbA1c ranged from 4.9% to 5.1% for all other age groups.
“and ranged from a mean of 4.9% to a mean of 5.1% for all other age groups.”
Quote in the paperNot a substantial claim - Findingown data
In the sensitivity analysis after removing implausible data, the percentage of participants who had a hypoglycemic event was lower than in the main analysis.
“however, the percentage of participants who had a hypoglycemic event was lower after removing implausible data (10).”
Quote in the paperNot a substantial claim - Findingown data
<1% of participants spent ≥2% of time >180 mg/dL.
“but ,1% of participants spent $2% of time .180mg/dL (10).”
Quote in the paperNot a substantial claim - Findingown data
Further narrowing the time in the target range of 70 to 120 mg/dL, there was a lower frequency of sensor glucose values that fell within this range moving across the age cohorts.
“Further narrowing the time in the target range of 70 to 120 mg/dL, there was a lower frequency of sensor glucose values that fell within this range moving across the age cohorts”
Quote in the paperNot a substantial claim - Findingown data
More hypoglycemia was seen during the first 48 hours of sensor wear than on subsequent days.
“When data were assessed by days of sensor wear, more hypoglycemia was seen during the first 48 hours of sensor wear than on subsequent days (data not shown).”
Quote in the paperNot a substantial claim - Findingown data
Overall, median percentage of time spent between 70 and 140 mg/dL was 96% (IQR, 93 to 98).
“Overall, median percentage of time spent between 70 and 140 mg/dL (3.9 to 7.8 mmol/L) was 96% (IQR, 93 to 98).”
- Also stated in: § Discussion · plain
“These individuals spent a median of 96% of the time with sensor glucose values between 70 and 140 mg/dL.”
- Also stated in: § Abstract · plain
“The median time between 70 to 140 mg/dL (3.9 to 7.8 mmol/L) was 96% (interquartile range, 93 to 98).”
Quote in the paperSubstantial claim - Also stated in: § Discussion · plain
- Findingown data
Mean glycemic variability measured by CV during nighttime was 13% ± 4%.
“and 13% 6 4%.”
Quote in the paperNot a substantial claim - Findingown data
There was a wide spread of HbA1c levels for a given mean glucose concentration.
“and there was a wide spread of HbA1c levels for a given mean glucose concentration (10).”
Quote in the paperNot a substantial claim - Findingown data
Overnight median percentage of glucose values >140 mg/dL was 0% (IQR, 0.0 to 1.0).
“Glucose concentrations .140mg/dL and ,70 mg/dL were uncommon overnight, with median percentages of values of 0% (IQR, 0.0 to 1.0)”
Quote in the paperNot a substantial claim - Findingown data
Median time spent in the glucose range of 70 to 140 mg/dL was lowest at 93% in adults ≥60 years of age (P < 0.001).
“Similarly, median time spent in the glucose range of 70 to 140 mg/ dL was lowest at 93% in adults $60 years of age (P , 0.001).”
Quote in the paperNot a substantial claim - Findingown data
Overall, mean average glucose during the daytime was 100 ± 7 mg/dL (5.6 ± 0.4 mmol/L).
“Overall, mean average glucose during the daytime was 100 6 7 mg/dL (5.6 6 0.4 mmol/L)”
Quote in the paperNot a substantial claim - Findingown data
Mean HbA1c was 5.3 ± 0.2% for participants aged ≥60 years.
“Mean HbA1c was 5.3 6 0.2% for participants aged $60 years”
Quote in the paperNot a substantial claim - Study design and conductown data
Because the majority of participants were white (93%), the study was unable to evaluate whether results were consistent across races.
“Because the majority of our participants were white (93%), we were unable to evaluate whether results were consistent across races.”
Quote in the paperSubstantial claim - Study design and conductnone given
The study was unable to determine how many nocturnal hypoglycemic events were due to pressure on the sensor because of movement or position during sleep.
“we are unable to determine how many of the nocturnal hypoglycemic events were due to pressure on the sensor because of movement or position of the participant during sleep.”
Quote in the paperNot a substantial claim - Interpretationreasoning
The results may not be generalizable for capillary glucose or laboratory measured glucose profiles.
“Furthermore, this study defined CGM-based glucose profile in healthy nondiabetic participants, and therefore these results may not be generalizable for capillary glucose or laboratory measured glucose profiles.”
Quote in the paperNot a substantial claim - Premisereasoning
There are limited data on CGM-measured glucose concentrations in individuals without diabetes.
“However, there are limited data on CGM-measured glucose concentrations in individuals without diabetes.”
Quote in the paperSubstantial claim - Study design and conductown data
The mean absolute relative difference (MARD) comparing CGM and blood glucose meter measurements was 6.0% based on 1555 paired measurements.
“MARD comparing the CGM and blood glucose meter glucose measurements was 6.0% (n 5 1555 paired measurements).”
Quote in the paperSubstantial claim - Interpretationown data
This study provides data for how often the target range and other ranges are achieved in individuals without diabetes.
“Our study provides data for how often this target range and other ranges are achieved in individuals without diabetes.”
Quote in the paperSubstantial claim - Interpretationown data
This study reports CGM profiles in a large cohort of healthy, nondiabetic individuals with negative islet autoantibodies and normal HbA1c across different age groups.
“Our study reports CGM profiles in a large cohort of healthy, nondiabetic individuals with negative islet autoantibodies and normal HbA1c across different age groups.”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 21
- May combine two claims. 16
- The quote only partly states this claim. 3
Classified as probably not a substantial claim.
Interpretationown dataThe findings support the recent consensus conference recommendation that sensor glucose values <54 mg/dL represent a more meaningful cut-off than 70 mg/dL to define clinically important hypoglycemia.
“Our findings support the recent consensus conference recommendation (12, 13) that sensor glucose values ,54 mg/dL represent a more meaningful cut-off than 70 mg/dL to define clinically important hypoglycemia because CGM glucose levels below 54 mg/dL were uncommon, whereas most participants had some glucose levels of 55 to 69 mg/dL.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [12] Danne, 2017, International consensus on use of continuous glucose monitoring, Diabetes Care source
- cites [13] Maahs, 2016, Outcome measures for artificial pancreas clinical trials: a consensus report, Diabetes Care source
In Additional Recommendations and Limitations (page unavailable), the consensus report suggests reporting CGM excursions below either 70 or 55 mg/dL as possible measures of hypoglycemia. It does not recommend a cutoff below 54 mg/dL or say that cutoff is more clinically meaningful than 70 mg/dL. (LLM, conf=0.85)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationMost studies of CGM profiles in healthy, nondiabetic individuals have had small sample sizes or used early-generation CGM systems, which were less accurate than current devices.
“Most studies of CGM profiles in healthy, nondiabetic individuals have had small sample sizes or used early-generation CGM systems, which were less accurate than current devices (3)(4)(5)(6)(7).”
Quote in the paperSubstantial claim3 of 3 sources checked- cites [3] Borg, 2010, Real-life glycaemic profiles in non-diabetic individuals with low fasting glucose and normal HbA1c: the A1C-Derived Average Glucose (ADAG) study, Diabetologia source
The Discussion describes earlier CGM studies of 32 and 15 people, supporting the small-sample part of the claim; Methods reports that this study used CGM in 80 nondiabetic participants. Neither section says the systems were early-generation or compares their accuracy with current devices. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [4] Freckmann, 2007, Continuous glucose profiles in healthy subjects under everyday life conditions and after different meals, J Diabetes Sci Technol source
The Methods report CGM profiles from 24 healthy volunteers using SCGM1 devices, supporting this paper as an example of a small study using an older system. The paper reports the SCGM1’s measurement error but does not compare its accuracy with current devices or establish that most such studies share these limitations. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [5] Fox, 2010, Variation of interstitial glucose measurements assessed by continuous glucose monitors in healthy, nondiabetic individuals, Diabetes Care source
The Abstract’s Research Design and Methods describes a CGM study of 74 healthy participants. Research Design and Methods identifies the devices used and reports missing or implausible readings from one sensor, noting that a commercially available version had newer software. The paper does not establish that most such studies had small samples or that earlier devices were less accurate than current devices. (LLM, conf=0.83)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S2 (checkable).
PremisecitationAs assessment of diabetes treatment outcomes moves beyond HbA1c to include sensor glucose profiles, there is an increased need for a repository of normative CGM data with up-to-date and widely used devices.
“As the assessment of outcomes of treatment of T1D and type 2 diabetes in clinical practice and research trials moves beyond the narrow focus of HbA1c levels to include the results of sensor glucose profiles (14), there is an increased need for a repository of”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [14] Beck, 2017, The fallacy of average: how using HbA1c alone to assess glycemic control can be misleading, Diabetes Care source
May combine two claims.
InterpretationcitationThe variability of mean glucose at the same HbA1c is similar to what has been observed in the T1D population and highlights the limitation of HbA1c to reflect true glycemic status for some individuals.
“This is similar to what has been observed in the T1D population and highlights the limitation of HbA1c to reflect true glycemic status for some individuals (14).”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [14] Beck, 2017, The fallacy of average: how using HbA1c alone to assess glycemic control can be misleading, Diabetes Care source
Classified as probably not a substantial claim. May combine two claims.
PremisecitationCurrent consensus conferences have somewhat arbitrarily set time in ranges for CGM sensor glucose levels for patients with T1D, with time in target identified as being between 70 to 180 mg/dL.
“The current consensus conferences have somewhat arbitrarily set time in ranges for CGM sensor glucose levels for patients with T1D, with time in target identified as being between 70 to 180 mg/dL (12,13).”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [12] Danne, 2017, International consensus on use of continuous glucose monitoring, Diabetes Care source
- cites [13] Maahs, 2016, Outcome measures for artificial pancreas clinical trials: a consensus report, Diabetes Care source
The Objective and Rationale section advocates standardized CGM glucose metrics for artificial-pancreas studies and says they are broadly applicable to type 1 diabetes studies. Additional Recommendations and Limitations mentions arbitrary glycemic ranges, but the retrieved passages do not identify 70–180 mg/dL as the target range or say that consensus conferences set that range somewhat arbitrarily. Page numbers were not available in our copy. (LLM, conf=0.80)
Experimental — accuracy not yet measured
Classified as probably not a substantial claim.
Findingown dataA total of 153 participants aged 7 to 80 years were included in the analyses.
“A total of 153 participants (age 7 to 80 years) were included in the analyses.”
- Also stated in: § Results · plain
“Therefore, 153 participants were included in the analyses, ranging from 7 to 80 years (mean, 31 years).”
Quote in the paperSubstantial claim- Also stated in: § Results · plain
The quote only partly states this claim.
Findingown dataMedian sensor glucose between 70 and 140 mg/dL during the daytime was 96% (IQR, 92 to 97%).
“Median sensor glucose between 70 and 140 mg/dL during the daytime and nighttime was 96% (IQR, 92 to 97%)”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSensor glucose values >180 mg/dL and <54 mg/dL were uncommon.
“Sensor glucose values .180mg/dL (10.0 mmol/L) and ,54 mg/dL (3.0 mmol/L) were uncommon.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn a sensitivity analysis after removing participants with first-degree family members with T1D (n = 101), mean CGM glucose, time spent in the 70 to 140 mg/dL range, and glycemic variability as measured by CV were not different from the main analysis.
“In a sensitivity analysis after removing participants with first-degree family members with T1D (n 5 101), mean CGM glucose, time spent in the 70 to 140 mg/dL range, and glycemic variability as measured by CV were not different from the main analysis (data not shown).”
Quote in the paperNot a substantial claimMay combine two claims. The quote only partly states this claim.
Findingown dataMean average glucose was 98 to 99 mg/dL for all age groups except those over 60 years, in whom mean average glucose was 104 mg/dL (P < 0.001 comparing those ≥60 years old with those younger).
“Mean average glucose was 98 to 99 mg/dL (5.4 to 5.5 mmol/L) for all age groups except those over 60 years, in whom mean average glucose was 104 mg/dL (5.8 mmol/L).”
- Also stated in: § Results · plain
“The mean average glucose level was 98 to 99 mg/dL (5.4 to 5.5 mmol/L) for all age groups except those $60 years old in whom mean average glucose a The underweight, normal weight, and overweight BMI categories for participants aged $18 years are ,18.5, 18.5 to ,24.9, and $24.9, respectively.”
Quote in the paperSubstantial claim- Also stated in: § Results · plain
May combine two claims.
Findingown dataIn a sensitivity analysis after removing implausible data, mean glucose, time spent in the 70 to 140 mg/dL range, and CV were not different from the main analysis.
“Similarly, in a sensitivity analysis after removing implausible data, mean glucose, time spent in the 70 to 140 mg/dL range, and CV were not different from the main analysis”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Findingown dataMean glycemic variability measured by CV during the daytime was 17% ± 3%.
“mean glycemic variability measured by CV during the daytime and nighttime was 17% 6 3%”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataMean CGM use by participants was 192 hours (8 days).
“Mean CGM use by the participants was 192 hours (8 days)”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataParticipants 6 to 11 years old had a median of 90% of their readings in the 70 to 120 mg/dL range, as compared with only 81% in the cohort over age 60 years.
“with participants 6 to 11 years old having a median of 90% of their readings in this range, as compared with only 81% in the cohort over age 60 years.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationcitationTwo possible factors may have contributed to the weaker-than-expected correlation: HbA1c was obtained prior to collection of CGM glucose data, and there are at most 10 days of CGM data.
“There are two possible factors that may have contributed to this weaker-than-expected correlation: (1) HbA1c was obtained prior to collection of CGM glucose data and ( 2) there are at most 10 days of CGM data (16).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16] Riddlesworth, 2018, Optimal sampling duration for continuous glucose monitoring to determine long-term glycemic control, Diabetes Technol Ther source
Probably not a substantial claim: may not pass S2 (checkable).
Premisenone givenUnderstanding CGM metrics in healthy participants without diabetes is needed to serve as a benchmark for research studies to define impaired glycemic status and for future technologies aiming to help individuals achieve near-normal glycemic profiles.
“Understanding CGM metrics in healthy participants without diabetes is needed to serve as a benchmark for research studies to define impaired glycemic status and for future technologies aiming to help individuals achieve near-normal glycemic profiles.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenStrengths of the study include the large sample size crossing the lifespan, multiple centers conducting the study, use of the latest generation technology, and up to 10 days of CGM wear time.
“Strengths of our study include the large sample size crossing the lifespan, multiple centers conducting the study, use of the latest generation technology, and up to 10 days of CGM wear time.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
Interpretationnone givenThis study fills the current gap in normative CGM data.
“Although our study fills the current gap in the data”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe reported mean absolute relative difference (MARD) for the Dexcom G6 in accuracy studies comparing sensor glucose values with reference blood glucose values is 9.0%.
“The reported mean absolute relative difference (MARD) for the Dexcom G6 in accuracy studies comparing sensor glucose values with reference blood glucose values is 9.0% (9).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Shah, 2018, Performance of a factory-calibrated real-time continuous glucose monitoring system utilizing an automated sensor applicator, Diabetes Technol Ther source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenPressure on the sensor during sleep may have resulted in inaccuracy of some CGM readings.
“Although pressure on the sensor during sleep may have resulted in inaccuracy of some CGM readings”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Interpretationnone givenCGM profiles provide the best window available to date to evaluate the frequency of biochemical hypoglycemia in clinical trials of new drugs and devices for the treatment of diabetes.
“CGM profiles also provide the best window available to date to evaluate the frequency of biochemical hypoglycemia in clinical trials of new drugs and devices for the treatment of diabetes.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
PremisereasoningHbA1c does not capture clinically relevant hyperglycemic and hypoglycemic patterns, which limits its utility in personalizing insulin-dosing decisions.
“However, HbA1c does not capture clinically relevant hyperglycemic and hypoglycemic patterns, which limits its utility in personalizing insulin-dosing decisions.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Premisenone givenAlthough data on glycemic profiles of healthy, nondiabetic individuals exist for older sensors, assessment of glycemic metrics with new-generation CGM devices is lacking.
“Although data on glycemic profiles of healthy, nondiabetic individuals exist for older sensors, assessment of glycemic metrics with new-generation CGM devices is lacking.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationHbA1c reflects overall glycemic control from the prior 60-90 days and is a surrogate marker for diabetes microvascular complications risk.
“This is due to the fact that HbA1c is a reflection of overall glycemic control from the prior 60-90 days, as well as clear evidence from the DCCT establishing it as a surrogate marker for diabetes microvascular complications risk (1).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [1] Nathan, 1993, The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus, N Engl J Med source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenUse of continuous glucose monitoring (CGM) is increasing for insulin-requiring patients with diabetes.
“Use of continuous glucose monitoring (CGM) is increasing for insulin-requiring patients with diabetes.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenParticipants were healthy, nondiabetic children and adults, primarily recruited from family, friends, and neighbors of patients seen in the diabetes clinics.
“Participants were healthy, nondiabetic children and adults, primarily recruited from family, friends, and neighbors of patients seen in the diabetes clinics.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenCGM metrics are being used to examine and compare the relative effectiveness and safety of new therapeutic agents and modern technologies in clinical diabetes research.
“Moreover, CGM metrics are being used to examine and compare the relative effectiveness and safety of new therapeutic agents and modern technologies in clinical diabetes research.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenCGM provides information on trends for hypoglycemia, hyperglycemia, and glycemic variability that helps patients and providers optimize glycemic control without increasing the risk for hypoglycemia.
“Continuous glucose monitoring (CGM) provides information on trends for hypoglycemia, hyperglycemia, and glycemic variability that helps patients and providers optimize glycemic control without increasing the risk for hypoglycemia.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenStudy participants provided written informed consent prior to study participation.
“Study participants provided written informed consent prior to study participation.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductnone givenOnly participants with ≥72 total hours of CGM data (with at least 24 hours overnight) were included in the analysis.
“Only participants with $72 total hours of CGM data (with at least 24 hours overnight) were included in the analysis.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenOnly participants with no islet autoantibodies and a central laboratory HbA1c level <5.7% were included in the analysis.
“Only participants with no islet autoantibodies and a central laboratory HbA1c level ,5.7% were included in the analysis.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductcitationThe study was conducted at 12 diabetes centers within the T1D Exchange Clinic Network after approval by institutional review boards.
“The study was conducted at 12 diabetes centers within the T1D Exchange Clinic Network after approval by institutional review boards (8).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [8] Beck, 2012, The T1D Exchange clinic registry, J Clin Endocrinol Metab source
The paper relies on this without support.
The 153 participants are representative of healthy nondiabetic individuals across age groups.
“Participants were healthy, nondiabetic children and adults, primarily recruited from family, friends, and neighbors of patients seen in the diabetes clinics.”
“Because the majority of our participants were white (93%), we were unable to evaluate whether results were consistent across races.”
The paper relies on this without support.
CGM profiles over up to 10 days reflect usual glucose profiles in healthy nondiabetic individuals.
“An early-generation Dexcom G6 CGM (Dexcom, Inc., San Diego, CA) in blinded mode (i.e., the participant was unable to see the glucose values) was worn for up to 10 days with one daily calibration using a study-provided blood glucose meter (Contour Next EZ; Ascensia Diabetes Care, Parsippany, NJ).”
“there are at most 10 days of CGM data (16).”
The authors state this as an assumption.
Low CGM readings <50 mg/dL during sleep are likely due to the participant lying on the CGM sensor.
“any low CGM readings ,50 mg/dL (2.8 mmol/L) during sleep time as defined by logs (in which case the participant was likely to be lying on the CGM sensor)”
The paper relies on this without support.
Participants' daily logs accurately recorded sleep and wake times.
“Participants completed a daily log indicating times and intensity of any exercise, start times of meals and snacks, and sleep and wake times.”
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069 · see citing works
Self-citationTranslating the A1C assay into estimated average glucose values
Nathan, D. M., Kuenen, J., Borg, R., Zheng, H., Schoenfeld, D., Heine, R. J., & for the A1c-Derived Average Glucose (ADAG) Study Group (2008) · Diabetes Care · doi:10.2337/dc08-0545 · Cited by 1,812 · see citing works
The fallacy of average: how using HbA1c alone to assess glycemic control can be misleading
Beck, R. W., Connor, C. G., Mullen, D. M., Wesley, D. M., & Bergenstal, R. M. (2017) · Diabetes Care · doi:10.2337/dc17-0636 · Cited by 477 · see citing works
Self-citationThe T1D exchange clinic registry
Beck, R. W., Tamborlane, W. V., Bergenstal, R. M., Miller, K. M., DuBose, S. N., Hall, C. A., & for the T1D Exchange Clinic Network (2012) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2012-1561 · Cited by 456 · see citing works
Self-citationBergenstal, R. M., Ahmann, A. J., Bailey, T., Beck, R. W., Bissen, J., Buckingham, B., Deeb, L., Dolin, R. H., Garg, S. K., Goland, R., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0051 · Cited by 307 · see citing works
Optimal Sampling Duration for Continuous Glucose Monitoring to Determine Long-Term Glycemic Control
Riddlesworth, T. D., Beck, R. W., Gal, R. L., Connor, C. G., Bergenstal, R. M., Lee, S., & Willi, S. M. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0455 · Cited by 266 · see citing works
Self-citationDeSalvo, D. J., Miller, K. M., Hermann, J. M., Maahs, D. M., Hofer, S. E., Clements, M. A., Lilienthal, E., Sherr, J. L., Tauschmann, M., Holl, R. W., et al. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12711 · Cited by 212 · see citing works
Self-citationOutcome measures for artificial pancreas clinical trials: a consensus report
Maahs, D. M., Buckingham, B. A., Castle, J. R., Cinar, A., Damiano, E. R., Dassau, E., DeVries, J. H., Doyle, F. J., Griffen, S. C., Haidar, A., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2716 · Cited by 211 · see citing works
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197 · see citing works
Self-citationMazze, R. S., Strock, E. S., Wesley, D., Borgman, S. K., Morgan, B., Bergenstal, R. M., & Cuddihy, R. M. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0293 · Cited by 183 · see citing works
Self-citationFreckmann, G., Hagenlocher, S., Baumstark, A., Jendrike, N., Gillen, R. C., Rössner, K., & Haug, C. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100513 · Cited by 174 · see citing works
Bergenstal, R. M., Ahmann, A., Bailey, T. S., Beck, R. W., Bissen, J., Buckingham, B. A., Deeb, L. C., Dolin, R. H., Garg, S. K., Goland, R., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700234 · Cited by 167 · see citing works
Self-citationJuvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-1971 · Cited by 138 · see citing works
Reference Values for Continuous Glucose Monitoring in Chinese Subjects
Zhou, J., Li, H., Ran, X., Yang, W., Li, Q., Peng, Y., Li, Y., Gao, X., Luan, X., Wang, W., et al. (2009) · Diabetes Care · doi:10.2337/dc09-0076 · Cited by 138 · see citing works
on behalf of the ADAG Study Group, Borg, R., Kuenen, J. C., Carstensen, B., Zheng, H., Nathan, D. M., Heine, R. J., Nerup, J., Borch-Johnsen, K., & Witte, D. R. (2010) · Diabetologia · doi:10.1007/s00125-010-1741-9 · Cited by 67 · see citing works
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S. K., Heinemann, L., Hirsch, I. B., Amiel, S. A., et al. (2018) · Yearbook of pediatric endocrinology · doi:10.1530/ey.15.10.10 · Cited by 44 · see citing works
Self-citationContinuous Glucose Monitoring in Healthy Children Aged 2–8 Years
Sundberg, F., & Forsander, G. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0270 · Cited by 21 · see citing works
DeSalvo, D. J., Km, M., Hermann, J., Maahs, D. M., Hofer, S., Ma, C., Lilienthal, E., Sherr, J. L., Tauschmann, M., Holl, R. W., et al. (2019) · Yearbook of pediatric endocrinology · doi:10.1530/ey.16.10.9 · Cited by 6 · see citing works
Shah, V. N., DuBose, S. N., Li, Z., et al. (2019) · Zenodo (CERN European Organization for Nuclear Research) · doi:10.5061/dryad.h7d11cd
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Self-citationConsensus report: definition and interpretation of remission in type 2 diabetes
Riddle, M. C., Cefalu, W. T., Evans, P. H., Gerstein, H. C., Nauck, M. A., Oh, W. K., Rothberg, A. E., le Roux, C. W., Rubino, F., Schauer, P., et al. (2021) · Diabetes Care · doi:10.2337/dci21-0034 · Cited by 464 · see citing works
de Bock, M. I., Ethel Codner, D., Craig, M. E., Huynh, T., Maahs, D. M., Mahmud, F. H., Marcovecchio, L., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13455 · Cited by 260 · see citing works
Review of methods for detecting glycemic disorders
Bergman, M., Abdul‐Ghani, M. A., DeFronzo, R. A., Manco, M., Sesti, G., Fiorentino, T. V., Ceriello, A., Rhee, M. K., Phillips, L. S., Chung, S. T., et al. (2020) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2020.108233 · Cited by 202 · see citing works
“7% [39 mmol/mol]) aged 7–80 years [181] wearing the Dexcom G6 system for approximately 10 days on an ambulatory basis, Shah et al established that the average 24-h glucose was 99 ± 7 mg/dl (5.”
MethodSherr, J. L., Schoelwer, M., Dos Santos, T. J., Reddy, L., Biester, T., Galderisi, A., van Dyk, J. C., Hilliard, M. E., Berget, C., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13421 · Cited by 152 · see citing works
“5 mmol/L (98–99 mg/dl) and 96% of time spent in this tighter target range.(264) The true test of new technologies will be to see how”
BackgroundSelf-citationBesser, R. E. J., Bell, K. J., Couper, J., Ziegler, A., Wherrett, D. K., Knip, M., Speake, C., Casteels, K., Driscoll, K. A., Jacobsen, L. M., et al. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13410 · Cited by 147 · see citing works
“Normative data taken from children, adolescents, and adults who are islet autoantibody-negative demonstrate a narrow variability in glucose using continuous glucose monitoring (CGM).(79) CGM provides”
BackgroundConsensus report: definition and interpretation of remission in type 2 diabetes
Riddle, M. C., Cefalu, W. T., Evans, P. H., Gerstein, H. C., Nauck, M. A., Oh, W. K., Rothberg, A. E., le Roux, C. W., Rubino, F., Schauer, P., et al. (2021) · Diabetologia · doi:10.1007/s00125-021-05542-z · Cited by 121 · see citing works
Holzer, R., Bloch, W., & Brinkmann, C. K. (2022) · Sensors · doi:10.3390/s22052030 · Cited by 108 · see citing works
“In a recently published multicenter study, CGM systems were used to collect data from healthy, nondiabetic, normal-weight individuals of different age groups [27].”
MethodCastañeda, J., Arrieta, A., van den Heuvel, T., Battelino, T., & Cohen, O. (2024) · Diabetes Care · doi:10.2337/dc23-1581 · Cited by 107 · see citing works
Zahedani, A. D., McLaughlin, T., Veluvali, A., Aghaeepour, N., Hosseinian, A., Agarwal, S., Ruan, J., Tripathi, S., Woodward, M., Hashemi, N., et al. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00956-y · Cited by 96 · see citing works
Use of Continuous Glucose Monitors by People Without Diabetes: An Idea Whose Time Has Come?
Klonoff, D. C., Nguyen, K. T., Xu, N. Y., Gutierrez, A., Espinoza, J. C., & Vidmar, A. P. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221110830 · Cited by 93 · see citing works
CGMap: Characterizing continuous glucose monitor data in thousands of non-diabetic individuals
Keshet, A., Shilo, S., Godneva, A., Talmor‐Barkan, Y., Aviv, Y., Segal, E., & Rossman, H. (2023) · Cell Metabolism · doi:10.1016/j.cmet.2023.04.002 · Cited by 93 · see citing works
Consensus Report: Definition and Interpretation of Remission in Type 2 Diabetes
Riddle, M. C., Cefalu, W. T., Evans, P. H., Gerstein, H. C., Nauck, M. A., Oh, W. K., Rothberg, A. E., le Roux, C. W., Rubino, F., Schauer, P., et al. (2022) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab585 · Cited by 92 · see citing works
“In cases where the accuracy of HbA1c values is uncertain, CGM can be used to assess the correlation between mean glucose and HbA1c and identify patterns outside the usual range of normal (27,28).”
Jarvis, P. R. E., Cardin, J. L., Nisevich-Bede, P. M., & McCarter, J. P. (2023) · Metabolism · doi:10.1016/j.metabol.2023.155640 · Cited by 86 · see citing works
Lin, Y. K., Fisher, S. J., & Pop‐Busui, R. (2020) · Journal of Diabetes Investigation · doi:10.1111/jdi.13290 · Cited by 82 · see citing works
Dunn, T. C., Ajjan, R., Bergenstal, R. M., & Xu, Y. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0565 · Cited by 83 · see citing works
Self-citationAcute Effects of Coffee Consumption on Health among Ambulatory Adults
Marcus, G. M., Rosenthal, D. G., Nah, G., Vittinghoff, E., Fang, C., Ogomori, K., Joyce, S., Yιlmaz, D., Yang, V. B., Kessedjian, T., et al. (2023) · New England Journal of Medicine · doi:10.1056/nejmoa2204737 · Cited by 76 · see citing works
Meulemeester, J. D., Charleer, S., Visser, M. M., Block, C. E. M. D., Mathieu, C., & Gillard, P. (2024) · Diabetologia · doi:10.1007/s00125-024-06171-y · Cited by 82 · see citing works
“Recently, guidelines have recommended that time in tight range (TITR) should also be reported [13]; this is defined as the percentage of time spent in normoglycaemia (3.9–7.8 mmol/l) [14].”
InfluentialShah, V. N., Kanapka, L. G., Aktürk, H. K., Polsky, S., Forlenza, G. P., Kollman, C., Beck, R. W., & Snell‐Bergeon, J. K. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0486 · Cited by 75 · see citing works
Self-citationPassanisi, S., Piona, C. A., Salzano, G., Marigliano, M., Bombaci, B., Morandi, A., Alibrandi, A., Maffeis, C., & Lombardo, F. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0373 · Cited by 66 · see citing works
Glycaemic variability: The under‐recognized therapeutic target in type 1 diabetes care
Wilmot, E. G., Choudhary, P., Leelarathna, L., & Baxter, M. (2019) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13842 · Cited by 65 · see citing works
“8 mmol/L).(15) Data concerning individuals with T1DM suggest that even those patients with so-called “good variability” demonstrate a CoV of approximately 37%, which is more than twice that of individuals without diabetes.”
BackgroundAdvances in Continuous Glucose Monitoring: Clinical Applications
Kwon, S. Y., & Moon, J. S. (2025) · Endocrinology and Metabolism · doi:10.3803/enm.2025.2370 · Cited by 67 · see citing works
Continuous glucose monitoring and intrapersonal variability in fasting glucose
Shilo, S., Keshet, A., Rossman, H., Godneva, A., Talmor‐Barkan, Y., Aviv, Y., & Segal, E. (2024) · Nature Medicine · doi:10.1038/s41591-024-02908-9 · Cited by 61 · see citing works
Scully, K. J., Sherwood, J. S., Martin, K., Ruazol, M. P., Marchetti, P., Larkin, M. E., Zheng, H., Sawicki, G. S., Uluer, A., Neuringer, I. P., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab857 · Cited by 52 · see citing works
Glycemic Patterns Are Distinct in Post-Bariatric Hypoglycemia After Gastric Bypass (PBH-RYGB)
Lee, D., Dreyfuss, J. M., Sheehan, A., Puleio, A., Mulla, C. M., & Patti, M. E. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab323 · Cited by 45 · see citing works
Time-in-range for monitoring glucose control: Is it time for a change?
Bellido, V., Pinés-Corrales, P. J., Villar‐Taibo, R., & Ampudia‐Blasco, F. J. (2021) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2021.108917 · Cited by 46 · see citing works
Mo, Y., Lu, J., & Zhou, J. (2023) · Journal of Diabetes Investigation · doi:10.1111/jdi.14112 · Cited by 48 · see citing works
“83 in the majority of healthy participants of European ethnicity.”
BackgroundMerino, J., Linenberg, I., Bermingham, K. M., Ganesh, S., Bakker, E., Delahanty, L. M., Chan, A. T., Capdevila Pujol, J., Wolf, J., Al Khatib, H., et al. (2022) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqac026 · Cited by 46 · see citing works
Ylescupidez, A., Speake, C., Pietropaolo, S. L., Wilson, D. M., Steck, A. K., Sherr, J. L., Gaglia, J. L., Bender, C., Lord, S. M., & Greenbaum, C. J. (2023) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgad472 · Cited by 41 · see citing works
Self-citationContinuous Glucose Monitoring for Prediabetes: What Are the Best Metrics?
Zahalka, S. J., Galindo, R. J., Shah, V. N., & Wang, C. C. L. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241242487 · Cited by 45 · see citing works
“24,25,41 Therefore, the goal for prediabetes TIR (54-140 mg/dl) that we propose is > 95%.”
Self-citationLupoli, R., Lembo, E., Rainone, C., Schiavo, L., Iannelli, A., Minno, M. N. D. D., & Capaldo, B. (2021) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2021.08.047 · Cited by 42 · see citing works
Montaser, E., Farhy, L. S., & Kovatchev, B. P. (2024) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgae379 · Cited by 39 · see citing works
Spartano, N. L., Sultana, N., Lin, H., Cheng, H., Lu, S., Fei, D., Murabito, J. M., Walker, M. E., Wolpert, H. A., & Steenkamp, D. (2024) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgae626 · Cited by 40 · see citing works
Lin, Y. K., Groat, D., Chan, O. T. M., Hung, M., Sharma, A. R., Varner, M. W., Gouripeddi, R., Facelli, J. C., & Fisher, S. J. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0016 · Cited by 38 · see citing works
Layne, J. E., Jepson, L. H., Carite, A. M., Parkin, C. G., & Bergenstal, R. M. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0197 · Cited by 37 · see citing works
Self-citationHazlehurst, J., Khoo, B. C. E., Lobato, C. B., Ilesanmi, I., Abbott, S., Chan, T. L., Pillai, S., Maslin, K., Purkayastha, S., McGowan, B. M., et al. (2024) · Endocrine Connections · doi:10.1530/ec-23-0285 · Cited by 43 · see citing works
Divilly, P., Martine‐Edith, G., Zaremba, N., Søholm, U., MAHMOUDI, Z., Cigler, M., Ali, N., Abbink, E. J., Brøsen, J. M. B., de Galan, B. E., et al. (2024) · Diabetes Care · doi:10.2337/dc23-2332 · Cited by 42 · see citing works
Lu, J., Home, P. D., & Zhou, J. (2020) · Diabetes Care · doi:10.2337/dc20-0561 · Cited by 36 · see citing works
Sofizadeh, S., Pehrsson, A., Ólafsdóttir, A. F., & Lind, M. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820965599 · Cited by 36 · see citing works
Cichosz, S. L., Kronborg, T., Laugesen, E., Hangaard, S., Fleischer, J., Hansen, T. K., Jensen, M. H., Poulsen, P. L., & Vestergaard, P. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0226 · Cited by 36 · see citing works
DuBose, S. N., Li, Z., Sherr, J. L., Beck, R. W., Tamborlane, W. V., & Shah, V. N. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820905904 · Cited by 35 · see citing works
“25f Overnightd nadir glucose, mean (SD)—mg/dL 83 (12) 82 (9) 76 (14) .”
InfluentialSelf-citationLandstra, C. P., Andrès, A., Chetboun, M., Conte, C., Kelly, Y. M., Berney, T., de Koning, E. J. P., Piemonti, L., Stock, P. G., Pattou, F. N., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab386 · Cited by 34 · see citing works
A Data-Driven Approach to Classifying Daily Continuous Glucose Monitoring (CGM) Time Series
Lobo, B. J., Farhy, L. S., Shafiei, M., & Kovatchev, B. P. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3103127 · Cited by 36 · see citing works
Zhang, Z., Wang, Y., Lu, J., & Zhou, J. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16033 · Cited by 36 · see citing works
Leahy, J. L., Aleppo, G., Fonseca, V. A., Garg, S. K., Hirsch, I. B., McCall, A. L., McGill, J. B., & Polonsky, W. H. (2019) · Journal of the Endocrine Society · doi:10.1210/js.2019-00222 · Cited by 38 · see citing works
Barua, S., Sabharwal, A., GLANTZ, N. M., Conneely, C., LAREZ, A. J., Bevier, W. C., & Kerr, D. E. (2021) · EClinicalMedicine · doi:10.1016/j.eclinm.2021.100853 · Cited by 34 · see citing works
“CGMmeasures separated by time of day akin to ours have been computed in healthy, non-diabetic individuals elsewhere [9].”
Discordance Between Glucose Management Indicator and Glycated Hemoglobin in People Without Diabetes
Shah, V. N., Vigers, T., Pyle, L. L., Calhoun, P. M., & Bergenstal, R. M. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0544 · Cited by 34 · see citing works
Self-citationClinical Application of Time in Range and Other Metrics
Aleppo, G. (2021) · Diabetes Spectrum · doi:10.2337/ds20-0093 · Cited by 31 · see citing works
Key indices of glycaemic variability for application in diabetes clinical practice
Monnier, L., Bonnet, F., Colette, C., Renard, É., & Owens, D. R. (2023) · Diabetes & Metabolism · doi:10.1016/j.diabet.2023.101488 · Cited by 34 · see citing works
Glucose levels measured with continuous glucose monitoring in uncomplicated pregnancies
Carlson, A. L., Beck, R. W., Li, Z., Norton, E., Bergenstal, R. M., Johnson, M. L., DUNNIGAN, S. M., Banfield, M., KRUMWIEDE, K. J., Sibayan, J., et al. (2024) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2023-003989 · Cited by 32 · see citing works
Self-citationPitfalls of HbA1c in the Diagnosis of Diabetes
Bergman, M., Abdul‐Ghani, M. A., Neves, J. S., Monteiro, M. P., Medina, J. L., Dorcely, B., & Buysschaert, M. (2020) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgaa372 · Cited by 31 · see citing works
Gubitosi-Klug, R. A., Braffett, B. H., Bebu, I., Johnson, M. L., Farrell, K., Kenny, D., Trapani, V. R., Meadema-Mayer, L., Soliman, E. Z., Pop-Busui, R., et al. (2022) · Diabetes Care · doi:10.2337/dc21-0629 · Cited by 31 · see citing works
Self-citationRossi, A., Venema, A., Haarsma, P., Feldbrugge, L., Burghard, R., Rodriguez‐Buritica, D., Parenti, G., Oosterveer, M. H., & Derks, T. G. J. (2022) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgac411 · Cited by 31 · see citing works
Ajjan, R. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16098 · Cited by 35 · see citing works
Digital biomarkers of ageing for monitoring physiological systems in community-dwelling adults
Lu, J. K., Wang, W., Mahadzir, M. D. A., Poganik, J. R., Moqri, M., Herzog, C. M. S., Verdin, E. M., Sebastiano, V., Gladyshev, V. N., Maier, A. B., et al. (2025) · The Lancet Healthy Longevity · doi:10.1016/j.lanhl.2025.100725 · Cited by 30 · see citing works
Consensus report: Definition and interpretation of remission in type 2 diabetes
Riddle, M. C., Cefalu, W. T., Evans, P. H., Gerstein, H. C., Nauck, M. A., Oh, W. K., Rothberg, A. E., le Roux, C. W., Rubino, F., Schauer, P. R., et al. (2021) · Diabetic Medicine · doi:10.1111/dme.14669 · Cited by 30 · see citing works
Reframing prediabetes: A call for better risk stratification and intervention
Kim, S. H. (2024) · Journal of Internal Medicine · doi:10.1111/joim.13786 · Cited by 30 · see citing works
Wang, Y., Lu, J., Yu, J., Ni, J., Wang, M., Lu, W., Zhu, W. M., Guo, J., Bao, Y., & Zhou, J. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16143 · Cited by 29 · see citing works
Battelino, T., Lalic, N. M., Hussain, S., Ceriello, A., Majanović, S. K., Davies, S., Topsever, P., Heverly, J., Ulivi, F., Brady, K., et al. (2025) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2025.112111 · Cited by 34 · see citing works
Cummings, C., Jiang, A., Sheehan, A., Ferraz‐Bannitz, R., Puleio, A., Simonson, D. C., Dreyfuss, J. M., & Patti, M. E. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15096 · Cited by 27 · see citing works
Continuous Glucose Monitoring Profiles in Healthy, Nondiabetic Young Children
DuBose, S. N., Kanapka, L. G., Bradfield, B., Sooy, M., Beck, R. W., & Steck, A. K. (2022) · Journal of the Endocrine Society · doi:10.1210/jendso/bvac060 · Cited by 27 · see citing works
Self-citationLim, B. S. Y., Yang, Q., Choolani, M. A., Gardner, D. S.-L., Chong, Y. S., Zhang, C., Chan, S., & Li, L. (2024) · Diabetes Care · doi:10.2337/dc24-0944 · Cited by 30 · see citing works
Oser, T. K., Cucuzzella, M. T., Stasinopoulos, M., MONCRIEF, M. A., McCall, A. L., & Cox, D. J. (2022) · JMIR Diabetes · doi:10.2196/34465 · Cited by 27 · see citing works
Continuous glucose monitoring in individuals undergoing gestational diabetes screening
Bartal, M. F., Cornthwaite, J. A. A., Ghafir, D., Ward, C., Nazeer, S. A., Blackwell, S. C., Pedroza, C., Chauhan, S. P., & Sibai, B. M. (2023) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2023.04.021 · Cited by 27 · see citing works
Matabuena, M., Pazos-Couselo, M., Alonso‐Sampedro, M., Fernández‐Merino, C., González‐Quintela, A., & Gudé, F. (2023) · Scientific Reports · doi:10.1038/s41598-023-40949-1 · Cited by 26 · see citing works
“The use of continuous glucose monitoring in healthy populations 11, 35–39 is topic of growing interest due to its multiple applications in epidemiological studies 17, 38 or diet optimization 40, 41 .”
MethodAktürk, H. K., Battelino, T., Castañeda, J., Arrieta, A., van den Heuvel, T. R. A., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0432 · Cited by 26 · see citing works
Cai, J., Liu, J., Lu, J., Ni, J., Wang, C., Chen, L., Lu, W., Zhu, W., Xia, T., & Zhou, J. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16212 · Cited by 27 · see citing works
Sundberg, F., Nåtman, J., Franzén, S., Åkesson, K., & Särnblad, S. (2021) · Pediatric Diabetes · doi:10.1111/pedi.13211 · Cited by 24 · see citing works
“1%.(29) Whether this observed difference in time below target in the youngest age group is constant and clinically significant remains to be further investigated.”
BackgroundJansen, L. T., Yang, N., Wong, J. M., Mehta, T. S., Allison, D. B., Ludwig, D. S., & Ebbeling, C. B. (2022) · Diabetes Care · doi:10.2337/dc21-1970 · Cited by 25 · see citing works
Brummer, J., Glasbrenner, C., Figueroa, S. H., Koehler, K., & Höchsmann, C. (2024) · Frontiers in Nutrition · doi:10.3389/fnut.2023.1308348 · Cited by 26 · see citing works
“This is particularly important as the generalizability of previous findings to non-diabetic individuals is likely limited, for instance, due to the usually far lower variations in blood glucose levels in persons without diabetes (77) as compared to persons with diabetes (78).”
Tattoo Inks for Optical Biosensing in Interstitial Fluid
Pazos, M. D., Hu, Y., Elani, Y., Browning, K. L., Jiang, N., & Yetisen, A. K. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202101238 · Cited by 23 · see citing works
Real World Interstitial Glucose Profiles of a Large Cohort of Physically Active Men and Women
Skroče, K., Zignoli, A., Fontana, F. Y., Maturana, F. M., Lipman, D., Tryfonos, A., Riddell, M. C., & Zisser, H. (2024) · Sensors · doi:10.3390/s24030744 · Cited by 23 · see citing works
“This finding is in line with recent observations of Shah et al. [26] in 153 apparently healthy (i.e., non-diabetic subjects) adults and youth where the mean 24 h glucose was 98–99 mg/dL for those less than 60 years of age.”
BackgroundResultEviz, E., Kıllı, N. E., Karakus, K. E., Can, E., Gökçe, T., Mutlu, G. Y., & Hatun, Ş. (2024) · Diabetic Medicine · doi:10.1111/dme.15333 · Cited by 23 · see citing works
“10,11 With the use of AID systems, a new era in diabetes management has begun, and it has become possible to not only meet the targets defined in the CGM consensus, but also to go beyond them without any increase in the risk of hypoglycaemic events.”
The Use of Continuous Glucose Monitoring to Diagnose Stage 2 Type 1 Diabetes
Mader, J. K., Wong, J. C., Freckmann, G., Garcia‐Tirado, J., Hirsch, I. B., Johnson, S. B., Kerr, D. E., Kim, S. H., Lal, R., Montaser, E., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251333441 · Cited by 26 · see citing works
Self-citationDadlani, V., Kaur, R. J., Stegall, M. D., Xyda, S., Kumari, K., Bonner, K., Smith, B. H., Thapa, P. B., Dean, P. G., & Kudva, Y. C. (2019) · Clinical Transplantation · doi:10.1111/ctr.13719 · Cited by 22 · see citing works
Clavero‐Jimeno, A., Dote‐Montero, M., Migueles, J. H., Camacho-Cardeñosa, A., Osés, M., Medina, J. E., Alcántara, J. M. A., Muñoz‐Torres, M., Labayen, I., & Ruiz, J. R. (2024) · Obesity · doi:10.1002/oby.24063 · Cited by 25 · see citing works
Addala, A., Zaharieva, D. P., Gu, A., Prahalad, P., Scheinker, D., Buckingham, B. A., Hood, K. K., & Maahs, D. M. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab522 · Cited by 21 · see citing works
Schiaffini, R., Lumaca, A., Martino, M., Rapini, N., Deodati, A., Amodeo, M. E., Ciampalini, P., Matteoli, M. C., Pampanini, V., & Cianfarani, S. (2024) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3826 · Cited by 21 · see citing works
Is It Time to Replace Time-in-Range with Time-in-Tight-Range? Maybe Not
Beck, R. W. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0602 · Cited by 20 · see citing works
Self-citationFreckmann, G., Schauer, S., Beltzer, A., Waldenmaier, D., Buck, S., Baumstark, A., Jendrike, N., Link, M., Zschornack, E., Haug, C., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221113341 · Cited by 17 · see citing works
“11 In general, it should not be neglected that glucose excursions always depend on the composition of the meal (eg carbohydrate/fat/protein content, glycemic index) and may differ between studies according to eating patterns in the respective country.”
BackgroundTime above range for predicting the development of type 2 diabetes
Marco, A., Pazos-Couselo, M., Moreno-Fernández, J., Díez‐Fernández, A., Alonso‐Sampedro, M., Fernández‐Merino, C., González‐Quintela, A., & Gudé, F. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.1005513 · Cited by 21 · see citing works
Bahillo-Curieses, P., Fernández-Velasco, P., Pérez-López, P., Martínez, A. M. V., de la Marca, M. D. L. O. N., & Díaz-Soto, G. (2024) · Endocrine · doi:10.1007/s12020-024-03881-6 · Cited by 22 · see citing works
Time in range—A new gold standard in type 2 diabetes research?
Goshrani, A., Lin, R., O'Neal, D. N., & Ekinci, E. I. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16279 · Cited by 23 · see citing works
Allo Beta Cell transplantation: specific features, unanswered questions, and immunological challenge
Caldara, R., Tomajer, V., Monti, P., Sordi, V., Citro, A., Chimienti, R., Gremizzi, C., Catarinella, D., Tentori, S., Paloschi, V., et al. (2023) · Frontiers in Immunology · doi:10.3389/fimmu.2023.1323439 · Cited by 19 · see citing works
Felix dies natalis, insulin… ceterum autem censeo “beta is better”
Piemonti, L. (2021) · Acta Diabetologica · doi:10.1007/s00592-021-01737-3 · Cited by 18 · see citing works
Exploring Minimally Invasive Approach to Define Stages of Type 1 Diabetes Remotely
Kontola, H., Alanko, I., Koskenniemi, J. J., Löyttyniemi, E., Itoshima, S., Knip, M., Veijola, R., Toppari, J., & Kero, J. T. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0554 · Cited by 17 · see citing works
Oganesova, Z., Pembérton, J. H., & Brown, A. C. (2024) · Diabetic Medicine · doi:10.1111/dme.15369 · Cited by 22 · see citing works
Glucose Patterns in Very Old Adults: A Pilot Study in a Community-Based Population
Selvin, E., Wang, D., Tang, O., Minotti, M., Echouffo-Tcheugui, J. B., & Coresh, J. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0156 · Cited by 18 · see citing works
Analytical Performance of the FreeStyle Libre 2 Glucose Sensor in Healthy Male Adults
Fellinger, E., Brandt, T., Creutzburg, J., Rommerskirchen, T., & Schmidt, A. (2024) · Sensors · doi:10.3390/s24175769 · Cited by 18 · see citing works
“Additionally, more and more studies focus on non-diabetic cohorts to gain insight into the glycemic pro fi les of healthy individuals [6–8].”
BackgroundSundberg, F., Smart, C. E., Samuelsson, J., Åkesson, K., & Krogvold, L. (2024) · Diabetes Care · doi:10.2337/dci24-0058 · Cited by 16 · see citing works
Bermingham, K. M., Smith, H., Duncan, E. L., Gonzalez, J. T., Valdes, A. M., Franks, P. W., Delahanty, L. M., Dashti, H. S., Davies, R., Hadjigeorgiou, G. M., et al. (2026) · Nature Communications · doi:10.1038/s41467-026-70308-3 · Cited by 15 · see citing works
Discordance between postprandial plasma glucose measurement and continuous glucose monitoring
Barua, S., Wierzchowska-McNew, R. A., Deutz, N. E. P., & Sabharwal, A. (2022) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqac181 · Cited by 16 · see citing works
A Conditional Generative Adversarial Network for Synthesis of Continuous Glucose Monitoring Signals
Cichosz, S. L., & Xylander, A. A. P. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211014255 · Cited by 17 · see citing works
Chekima, K., Wong, B. T., Noor, M. I., Ooi, Y. B. H., Yan, S. W., & Chekima, B. (2022) · Foods · doi:10.3390/foods11070983 · Cited by 15 · see citing works
Patel, N. J., Tozzo, V., Higgins, J. M., & Stone, J. H. (2022) · Arthritis & Rheumatology · doi:10.1002/art.42405 · Cited by 15 · see citing works
Fløde, M., Hermann, M., Haugstvedt, A., Søfteland, E., Igland, J., Åsberg, A., Jenssen, T. G., & Graue, M. (2023) · BMC Endocrine Disorders · doi:10.1186/s12902-023-01472-6 · Cited by 15 · see citing works
Pathophysiology, Diagnostic Criteria, and Approaches to Type 2 Diabetes Remission
Khamis, A. M. (2023) · Cureus · doi:10.7759/cureus.33908 · Cited by 18 · see citing works
“This standard would apply to all glucose-lowering medications, even those with other effects [42,43].”
BackgroundThe GO MOMs Study Group (2024) · BMJ Open · doi:10.1136/bmjopen-2024-084216 · Cited by 13 · see citing works
Wang, F., Wang, D., Xi-Ling, L. U., Xiao-Ming, S. U. N., & Duan, B. (2022) · Minerva Endocrinology · doi:10.23736/s2724-6507.20.03284-8 · Cited by 14 · see citing works
Standardized Hybrid Closed-Loop System Reporting
Shah, V. N., & Garg, S. K. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0622 · Cited by 13 · see citing works
Self-citation
- Authors
- Shah, V. N., DuBose, S. N., Li, Z., Beck, R. W., Peters, A. L., Weinstock, R. S., Kruger, D., Tansey, M., Sparling, D., Woerner, S., Vendrame, F., Bergenstal, R., Tamborlane, W. V., Watson, S. E., & Sherr, J.
- Affiliations
- Barbara Davis Center for Diabetes, Aurora, Colorado
- Jaeb Center for Health Research (US)
- University of Southern California (US)
- SUNY Upstate Medical University (US)
- Henry Ford Health (US)
- University of Iowa (US)
- University of Oklahoma Health Sciences Center (US)
- Indiana University School of Medicine
- University of Miami (US)
- Park Nicollet Clinic (US)
- Yale University (US)
- University of Louisville (US)
- Corresponding author
- DuBose, Stephanie N
- ORCID
- Venue
- The Journal of Clinical Endocrinology & Metabolism
- Publisher
- The Endocrine Society
- Published
- 2019-04-25
- Volume
- 104
- Issue
- 10
- Pages
- 4356-4364
- ISSN
- 0021-972X (ISSN-L)
- 1945-7197
- PMC
- PMC7296129
- OpenAlex
- W2941246007
- Semantic Scholar
- 22170f815dcf3fd4385ff1485a7902d3413b5ae7
- Cited by
- 315
- References
- 17
- Citations by year
- 2026: 57 · 2025: 71 · 2024: 63 · 2023: 41 · 2022: 38 · 2021: 26 · 2020: 15 · 2019: 4
- FWCI
- 14.29
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- The Journal of Clinical Endocrinology & Metabolism (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion)
- Funders
- Leona M. and Harry B. Helmsley Charitable Trust
- National Institutes of Health
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes, Cardiovascular Risks, and Lipoproteins
- Keywords
- continuous glucose monitoring, glycemic variability, hypoglycemia, hyperglycemia, Dexcom G6, reference values
- MeSH terms
- Adolescent, Adult, Age Factors, Aged, Blood Glucose, Child, Circadian Rhythm, Female, Humans, Hyperglycemia, Hypoglycemia, Male, Middle Aged, Monitoring, Physiologic, Prospective Studies, Reference Values, Sex Factors, Blood Glucose Self-Monitoring, Benchmarking, Young Adult, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-14.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-14.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review