A 3-year national DRL for CT in hybrid imaging study in Kuwait health environment—impact and implementation
Open accessThe DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
Objective: Diagnostic reference levels (DRLs) for CT in PET-CT are limited, and published DRLs from other countries may not be directly applicable to the State of Kuwait (KW). The authors aimed to carry out the final phase of a 3-year study on DRLs in KW, supporting optimization and dose reduction as imaging technology advances. Methods: In this cohort study, 400 adult oncology patients from 8 PET-CT centres were included, following the same procedures as in the first (2018) and second (2020) years, in accordance with the MOH-KW Ethical Committee’s recommendations. The CT dose index (CTDIvol), dose-length product (DLP), and scan length were recorded, and the median, mean, standard deviation, as well as the 75th and 25th percentiles, along with the whole-body (WB) effective dose (ED), were calculated. Comparative studies were conducted to track implementation and identify any shortfalls. Results: In this study, half-body (HB) and WB scans accounted for 66% and 34% of the total 400 cases, respectively. The proposed local DRL practice among the 8 centres in the 2022 study exhibited a maximum variation of 25%, showing a 30% improvement over 2020. The achievable local DRL remained consistent with 2020 levels. Comparative results of the third quartile DLP (476 mGy cm) and CTDIvol (4 mGy) values for 2022 indicated lower values for the third phase (400 entries) compared to 2020, with a 1.5-fold variation in DLP. The calculated ED for WB scans ranged from 2.6 to 7.1 mSv, with mean values of 4.7 ± 1.25 mSv, using a conversion factor (k = 0.0093 mSv/mGy/cm). The 2022 proposed national diagnostic reference levels (NDRLs) for HB (469 mGy cm, 4.0 mGy) were lower than the Swiss National Data (620 mGy cm, 6.0 mGy) and France (628 mGy cm, 6.6 mGy), but slightly higher than those of the United Kingdom (400 mGy cm, 4.3 mGy), despite the Swiss having about 5000 entries, France 1000 entries, and the United Kingdom 370 HB entries. Conclusions: There was a 11.1% continuous improvement in NDRL for 2022 compared to 9.1% in 2020 and 13% in 2018, demonstrating a trend of enhanced optimization. Advances in knowledge: The data established a trend of NDRL for WBCT (PET-CT) that can serve as a national databank for ongoing optimization. This promotes improvements in patient protection and quality care within the clinical environment of the State of Kuwait, aligning with the strategic goals of Kuwait Vision-2035.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures







Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The United Kingdom possesses a developed DRL system overseen by the UK Health Security Agency (UKHSA).109,110”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [110]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Recommendations of the International Commission on Radiological Protection, ICRP Publication 26
Icrp (1977) · Annals of the ICRP · Cited by 1,780
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
Vañó, E., Miller, D. L., Martin, C. J., Rehani, M. M., Kang, K., Rosenstein, M., Ortiz-López, P., Mattsson, S., Padovani, R., & Rogers, A. (2017) · Annals of the ICRP · doi:10.1177/0146645317717209 · Cited by 1,043
1990 Recommendations of the International Commission on Radiological Protection.
(1991) · PubMed · Cited by 908
Realistic reference phantoms: An ICRP/ICRU joint effort
(2009) · Annals of the ICRP · doi:10.1016/j.icrp.2009.09.001 · Cited by 496
Christner, J. A., Kofler, J. M., & McCollough, C. H. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.09.3462 · Cited by 492
Radiological protection and safety in medicine
(1996) · Annals of the ICRP · doi:10.1016/s0146-6453(00)89195-2 · Cited by 438
Shrimpton, P. C., Jansen, J. T. M., & Harrison, J. D. (2015) · British Journal of Radiology · doi:10.1259/bjr.20150346 · Cited by 134
(1996) · PubMed · Cited by 127
Diagnostic reference levels from the ACR CT accreditation program
McCollough, C., Branham, T., Herlihy, V., Bhargavan, M., Robbins, L., Bush, K., McNitt-Gray, M., Payne, J. T., Ruckdeschel, T., Pfeiffer, D., et al. (2011) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2011.03.014 · Cited by 98
A national survey of computed tomography doses in hybrid PET-CT and SPECT-CT examinations in the UK
Iball, G. R., Bebbington, N. A., Burniston, M. T., Edyvean, S., Fraser, L., Julyan, P. J., Parkar, N., & Wood, T. J. (2017) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000000672 · Cited by 72
Research works printed but not linked: 4
- IAEA, 2016, Radiation Protection of Patients (RPOP)—Diagnostic Reference Levels (DRLs) in Medical Imaging
- National Council on Radiation Protection and Measurements, 2012, Reference Levels and Achievable Doses in Medical and Dental Imaging: Recommendations for the United States
- International Commission on Radiological Protection, 2009, Adult reference computational phantoms. ICRP Publication 110, Ann ICRP, 39, 1
- Vano, 2015, Annals of the ICRP, ICRP publication, Ann ICRP, 44, 135
Cited by
Oglat, A., Shdaifat, S., Sayah, M., Abukonna, A., & Khogali, W. (2026) · Drug, Healthcare and Patient Safety · doi:10.2147/dhps.s595721
Metadata
- Authors
- Masoomi, M., Al-Kandari, L., Al-Shammeri, I., Elrahman, H., & Al-Shammeri, J.
- Affiliations
- Adan Hospital (KW)
- Kuwait University (KW)
- Corresponding author
- Masoomi, Michael
- Venue
- BJR|Open
- Publisher
- Oxford University Press (OUP)
- Published
- 2023-12-12
- Volume
- 6
- Issue
- 1
- Pages
- tzae032-tzae032
- ISSN
- 2513-9878 (ISSN-L)
- PMC
- PMC11495866
- OpenAlex
- W4403129092
- Cited by
- 1
- References
- 19
- Citations by year
- 2026: 1
- FWCI
- 0.16
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- BJR|Open (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Kuwait Foundation for the Advancement of Sciences
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- diagnostic reference levels, CT radiation dose, PET/CT, hybrid imaging, radiation dose optimization, dose reduction, CT dose index, dose-length product, effective dose, cancer patients, Kuwait, radiation protection, conversion factors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A 3-year national DRL for CT in hybrid imaging study in Kuwait health environment—impact and implementation. https://app.cochat.ai/papers/10.1093/bjro/tzae032
- The paper
- Masoomi, M., Al-Kandari, L., Al-Shammeri, I., Elrahman, H., & Al-Shammeri, J. (2023). A 3-year national DRL for CT in hybrid imaging study in Kuwait health environment—impact and implementation. BJR|Open. https://doi.org/10.1093/bjro/tzae032
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.