CoChat

Fully automated assay of glycohemoglobin with the Abbott IMx analyzer: novel approaches for separation and detection

Wilson, D H, Bogacz, J P, Forsythe, C M, Turk, P J, Lane, T L, Gates, R C, Brandt, D R · 1993 · Clinical Chemistry
Cited by 59

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

We describe a novel assay for measuring glycohemoglobin directly from anticoagulated whole blood with the Abbott IMx analyzer. The glycohemoglobin is labeled with a soluble polyanionic affinity reagent and the anionic complex is then captured with a cationic solid-phase matrix. Glycohemoglobin is quantified by measuring the quenching by heme of the static fluorescence from an added fluorophore. The assay is standardized to report both percent total glycohemoglobin (%GHb) and percent hemoglobin A1c (%HbA1c). Glucose, bilirubin, triglycerides, labile fraction, and hemoglobin variants do not interfere in the assay. Within- and between-run CVs are approximately 4-5%, with total CVs of approximately 6.5%. Highly significant linear correlations (r > 0.97) were obtained in comparison studies with two major assay methodologies. The time to obtain one result is approximately 10 min (including assay of a control), 56 min for 22 results. We describe the development, standardization, and validation of this new method.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/clinchem/39.10.2090

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 59

Metadata

Authors
Wilson, D. H., Bogacz, J. P., Forsythe, C. M., Turk, P. J., Lane, T. L., Gates, R. C., & Brandt, D. R.
Venue
Clinical Chemistry
Published
1993-10-01
Volume
39
Issue
10
Pages
2090-2097
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W1891741539
Cited by
59
References
1
FWCI
1.46
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Fully automated assay of glycohemoglobin with the Abbott IMx analyzer: novel approaches for separation and detection. https://app.cochat.ai/papers/10.1093/clinchem/39.10.2090
The paper
Wilson, D. H., Bogacz, J. P., Forsythe, C. M., Turk, P. J., Lane, T. L., Gates, R. C., & Brandt, D. R. (1993). Fully automated assay of glycohemoglobin with the Abbott IMx analyzer: novel approaches for separation and detection. Clinical Chemistry. https://doi.org/10.1093/clinchem/39.10.2090

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.