Candidate Reference Method for Total Thyroxine in Human Serum
Abstract
Background: There is a need for a critically evaluated reference method for thyroxine to provide an accuracy base to which routine methods can be traceable. We describe a candidate reference method involving isotope-dilution coupled with liquid chromatography/mass spectrometry. Methods: An isotopically labeled internal standard, thyroxine-d5, was added to serum, followed by equilibration, protein precipitation, and ethyl acetate and solid-phase extractions to prepare samples for liquid chromatography-mass spectrometry electrospray ionization (LC/MS-ESI) analysis. For separation, a Zorbax Eclipse XDB-C18 column was used with a mobile phase consisting of 1 mL/L acetic acid in acetonitrile-water (32:68 by volume) for positive ions and a Zorbax Extend-C18 column with a mobile phase consisting of 2 mL/L ammonium hydroxide in methanol-water (32:68 by volume) for negative ions. [M + H]+ ions at m/z 778 and 783 for thyroxine and its labeled internal standard were monitored for positive ions and [M − H]− ions at m/z 776 and 781 for negative ions. Samples of frozen serum pools were prepared and measured in three separate sets. Results: Within-set CVs were 0.2–1.0%. The correlation coefficients of all linear regression lines (measured intensity ratios vs mass ratios) were 0.999–1.000. Positive- and negative-ion measurements agreed with a mean difference of 0.45% at three concentrations (50, 110, and 168 μg/L). The detection limits (at a signal-to-noise ratio of ∼3 to 5) were 30 and 20 pg for positive and negative ions, respectively. The results from the LC/MS-ESI method were within 1 SD of the composite means from many routine clinical methods, although it appears that the clinical method means may be biased high by 4–5 μg/L across the concentrations. Some routine clinical methods may be biased by up to 20% at low concentrations. Conclusions: This well-characterized LC/MS-ESI method for total serum thyroxine with a theoretically sound approach, demonstrated good accuracy and precision, and low susceptibility to interferences qualifies as a candidate reference method. Use of this reference method as an accuracy base may reduce the apparent biases in routine methods along with the high interlaboratory imprecision.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Several laboratories have developed procedures to accurately determine total T4 and total T3 in human plasma or serum and largely agree on concentration ranges, but this consensus has not been achieved for most other TH metabolites.
“While meanwhile several laboratories developed procedures to accurately determine total T4 and total T3 in human plasma or serum and largely agree on concentrations ranges (25, 67–73), this consensus has not been achieved yet for most of the other TH metabolites, which occur at much lower concentrations.”
From 3,5-T2—A Janus-Faced Thyroid Hormone Metabolite Exerts Both Canonical T3-Mimetic Endocrine and Intracrine Hepatic Action (2020) · cites this paper as [72]
Cited by
Sorted by citations
Yue, B., Rockwood, A. L., Sandrock, T., La’ulu, S. L., Kushnir, M. M., & Meikle, A. W. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2007.098293 · Cited by 127 · see citing works
Carvalho, V. M. (2012) · Journal of Chromatography B · doi:10.1016/j.jchromb.2011.08.027 · Cited by 90 · see citing works
Köhrle, J., Lehmphul, I., Pietzner, M., Renko, K., Rijntjes, E., Richards, K., Anselmo, J., Danielsen, M., & Jonklaas, J. (2020) · Frontiers in Endocrinology · doi:10.3389/fendo.2019.00787 · Cited by 23 · see citing works
Köhrle, J., & Richards, K. H. (2020) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/a-1175-4610 · Cited by 7 · see citing works
- Authors
- Tai, S. S.-C., Sniegoski, L. T., & Welch, M. J.
- Affiliations
- National Institute of Standards and Technology (US)
- Corresponding author
- Tai, Susan S-C
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-04-01
- Volume
- 48
- Issue
- 4
- Pages
- 637-642
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W4319054114
- Cited by
- 80
- References
- 0
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 3 · 2023: 2 · 2022: 2 · 2021: 1 · 2020: 9 · 2019: 2 · 2018: 2 · 2017: 2 · 2016: 4 · 2015: 2 · 2014: 3 · 2013: 2 · 2012: 2
- FWCI
- 3.24
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model (vor)
- Locations
- Clinical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- PubMed
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Thyroid Disorders and Treatments
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
- Keywords
- total thyroxine, reference method, isotope dilution, liquid chromatography-mass spectrometry, electrospray ionization, human serum, methodological bias, thyroid function tests, solid-phase extraction, detection limit
- MeSH terms
- Chromatography, Liquid, Humans, Quality Control, Reference Standards, Reference Values, Sensitivity and Specificity, Thyroxine, Linear Models, Spectrometry, Mass, Electrospray Ionization
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review