Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
AIMS: Implantable cardioverter defibrillator (ICD) technology has become more complex, particularly with respect to biventricular resynchronization devices. The incidence of hardware-related complications in single (SC)-, dual (DC)-, and triple (BiV)-chamber devices requiring surgical revision has not been investigated systematically. METHODS AND RESULTS: We analysed data from consecutive ICD recipients implanted between January 2000 and December 2007 with respect to the need of surgical re-intervention for device- or lead-related complications. Generator exchanges due to normal battery depletion were not considered. From 816 patients (81% male, 69% ischaemic cardiomyopathy, 48% secondary prevention ICDs) followed for 31 +/- 24 months (2118 cumulative patient-years), 98 patients underwent 110 revisions (5.2% per patient-year). Complications included lead-related revision procedures in 81 cases and generator-related problems in 29 cases. The annual incidence of surgical revision due to complications was 11.8% in BiV compared with 4.9% in SC and 4.1% in DC patients (P = 0.002). This higher revision rate was mainly caused by lead-related complications. Implantation of a BiV system was an independent risk factor of the need for surgical revision (relative risk 2.37, 95% confidence interval 1.38-4.04). CONCLUSION: Even with long-lasting operator experience, complications requiring surgical revision remain a clinically important problem of ICD therapy. The incidence of complications is significantly higher in BiV resynchronization devices than in SC and DC systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bristow, M. R., Saxon, L. A., Boehmer, J., Krueger, S., Kass, D. A., De Marco, T., Carson, P., DiCarlo, L., DeMets, D., White, B. G., et al. (2004) · New England Journal of Medicine · doi:10.1056/nejmoa032423 · Cited by 5,812
Cardiac Resynchronization in Chronic Heart Failure
Abraham, W. T., Fisher, W. G., Smith, A. L., Delurgio, D. B., Leon, A. R., Loh, E., Kocovic, D. Z., Packer, M., Clavell, A. L., Hayes, D. L., et al. (2002) · New England Journal of Medicine · doi:10.1056/nejmoa013168 · Cited by 4,648
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546
WRITING COMMITTEE MEMBERS, Hunt, S. A., Abraham, W. T., Chin, M. H., Feldman, A. M., Francis, G. S., Ganiats, T. G., Jessup, M., Konstam, M. A., Mancini, D. M., et al. (2005) · Circulation · doi:10.1161/circulationaha.105.167586 · Cited by 2,251
Cited by
Cardiac-Resynchronization Therapy for Mild-to-Moderate Heart Failure
Tang, A. S. L., Wells, G. A., Talajic, M., Arnold, M. O., Sheldon, R., Connolly, S., Hohnloser, S. H., Nichol, G., Birnie, D. H., Sapp, J. L., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1009540 · Cited by 1,859
Self-citationKirkfeldt, R. E., Johansen, J. B., Nohr, E. A., Jorgensen, O. D., & Nielsen, J. C. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht511 · Cited by 871
Poole, J. E., Gleva, M. J., Mela, T., Chung, M. K., Uslan, D. Z., Borge, R., Gottipaty, V., Shinn, T., Dan, D., Feldman, L. A., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.976076 · Cited by 790
An Entirely Subcutaneous Implantable Cardioverter–Defibrillator
Bardy, G. H., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R. E., Wright, D. J., Connelly, D. T., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa0909545 · Cited by 780
Daubert, J.-C., Saxon, L., Adamson, P. B., Auricchio, A., Berger, R. D., Beshai, J. F., Breithard, O., Brignole, M., Cleland, J., DeLurgio, D. B., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.07.025 · Cited by 467
Self-citation
Metadata
- Authors
- Duray, G. Z., Schmitt, J., Cicek-Hartvig, S., Hohnloser, S. H., & Israel, C. W.
- Venue
- EP Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-12-02
- Volume
- 11
- Issue
- 3
- Pages
- 297-302
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W2038284667
- Cited by
- 66
- References
- 4
- FWCI
- 3.30
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.