Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedures
Open accessAbstract
Despite the advent of non-fluoroscopic technology, fluoroscopy remains the cornerstone of imaging in most interventional electrophysiological procedures, from diagnostic studies over ablation interventions to device implantation. Moreover, many patients receive additional X-ray imaging, such as cardiac computed tomography and others. More and more complex procedures have the risk to increase the radiation exposure, both for the patients and the operators. The professional lifetime attributable excess cancer risk may be around 1 in 100 for the operators, the same as for a patient undergoing repetitive complex procedures. Moreover, recent reports have also hinted at an excess risk of brain tumours among interventional cardiologists. Apart from evaluating the need for and justifying the use of radiation to assist their procedures, physicians have to continuously explore ways to reduce the radiation exposure. After an introduction on how to quantify the radiation exposure and defining its current magnitude in electrophysiology compared with the other sources of radiation, this position paper wants to offer some very practical advice on how to reduce exposure to patients and staff. The text describes how customization of the X-ray system, workflow adaptations, and shielding measures can be implemented in the cath lab. The potential and the pitfalls of different non-fluoroscopic guiding technologies are discussed. Finally, we suggest further improvements that can be implemented by both the physicians and the industry in the future. We are confident that these suggestions are able to reduce patient and operator exposure by more than an order of magnitude, and therefore think that these recommendations are worth reading and implementing by any electrophysiological operator in the field.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Health Risks from Exposure to Low Levels of Ionizing Radiation:: BEIR VII Phase 2
Earth, D. O. (2006) · CERN Document Server (European Organization for Nuclear Research) · Cited by 2,190
Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog
Mettler, F. A., Huda, W., Yoshizumi, T. T., & Mahesh, M. (2008) · Radiology · doi:10.1148/radiol.2481071451 · Cited by 2,008
ESC Guidelines on the management of cardiovascular diseases during pregnancy
Endorsed by the European Society of Gynecology (ESG), T. A. F. E. P. C. (., Authors/Task Force Members, Regitz-Zagrosek, V., Blomstrom Lundqvist, C., Borghi, C., Cifkova, R., Ferreira, R., Foidart, J.-M., Gibbs, J. S. R., Gohlke-Baerwolf, C., et al. (2011) · European Heart Journal · doi:10.1093/eurheartj/ehr218 · Cited by 1,924
Recommendations of the International Commission on Radiological Protection
(1959) · Health Physics · doi:10.1097/00004032-195901000-00001 · Cited by 1,703
Stewart, F. A., Akleyev, A. V., Hauer-Jensen, M., Hendry, J. H., Kleiman, N. J., MacVittie, T. J., Aleman, B. M., Edgar, A. B., Mabuchi, K., Muirhead, C. R., et al. (2012) · Annals of the ICRP · doi:10.1016/j.icrp.2012.02.001 · Cited by 1,604
Benefits and costs, an eternal balance
Dixon, A. K. (2007) · Annals of the ICRP · doi:10.1016/j.icrp.2007.09.001 · Cited by 1,355
Valentin, J. (2000) · Annals of the ICRP · doi:10.1016/s0146-6453(01)00004-5 · Cited by 964
Mettler, F. A., Bhargavan, M., Faulkner, K., Gilley, D. B., Gray, J. E., Ibbott, G. S., Lipoti, J. A., Mahesh, M., McCrohan, J. L., Stabin, M. G., et al. (2009) · Radiology · doi:10.1148/radiol.2532082010 · Cited by 853
Hausleiter, J., Meyer, T., Hadamitzky, M., Huber, E., Zankl, M., Martinoff, S., Kastrati, A., & Schömig, A. (2006) · Circulation · doi:10.1161/circulationaha.105.602490 · Cited by 705
Goldstein, J. A., Gallagher, M. J., O’Neill, W. W., Ross, M. A., O’Neil, B. J., & Raff, G. L. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.08.064 · Cited by 643
Cited by
Tzeis, S., Gerstenfeld, E. P., Kalman, J., Saad, E. B., Shamloo, A. S., Andrade, J. G., Barbhaiya, C. R., Baykaner, T., Boveda, S., Calkins, H., et al. (2024) · Heart Rhythm · doi:10.1016/j.hrthm.2024.03.017 · Cited by 298
Self-citation
Metadata
- Authors
- Heidbuchel, H., Wittkampf, F. H. M., Vano, E., Ernst, S., Schilling, R., Picano, E., Mont, L., ESC Scientific Document Group, Jais, P., de Bono, J., Piorkowski, C., Saad, E., & Femenia, F.
- Venue
- EP Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2014-05-02
- Volume
- 16
- Issue
- 7
- Pages
- 946-964
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W1988444321
- Cited by
- 311
- References
- 67
- FWCI
- 18.76
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedures. https://app.cochat.ai/papers/10.1093/europace/eut409
- The paper
- Heidbuchel, H., Wittkampf, F. H. M., Vano, E., Ernst, S., Schilling, R., Picano, E., Mont, L., ESC Scientific Document Group, Jais, P., de Bono, J., Piorkowski, C., Saad, E., & Femenia, F. (2014). Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedures. EP Europace. https://doi.org/10.1093/europace/eut409
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.