Validation of the mapping accuracy of a novel non-invasive epicardial and endocardial electrophysiology system
Open accessThe DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
AIMS: Use of a non-invasive electrocardiographic mapping system may aid in rapid diagnosis of atrial or ventricular arrhythmias or the detection of ventricular dyssynchrony. The aim of the present study was to validate the mapping accuracy of a novel non-invasive epi- and endocardial electrophysiology system (NEEES). METHODS AND RESULTS: Patients underwent pre-procedural computed tomography or magnetic resonance imaging of the heart and torso. Radiographic data were merged with the data obtained from the NEEES during pacing from implanted pacemaker leads or pacing from endocardial sites using an electroanatomical mapping system (CARTO 3, Biosense Webster). The earliest activation as denoted on the NEEES three-dimensional heart model was compared with the true anatomic location of the tip of the pacemaker lead or the annotated pacing site on the CARTO 3 map. Twenty-nine patients [mean age: 62 ± 11 years, 6/29 (11%) female, 21/29 (72%) with ischaemic cardiomyopathy] were enrolled into the pacemaker verification group. The mean distance from the non-invasively predicted pacing site to the anatomic reference site was 10.8 ± 5.4 mm for the right atrium, 7.7 ± 5.8 mm for the right ventricle, and 7.9 ± 5.7 mm for the left ventricle activated via the coronary sinus lead. Five patients [mean age 65 ± 4 years, 2 (33%) females] underwent CARTO 3 verification study. The mean distance between non-invasively reconstructed pacing site and the reference pacing site was 7.4 ± 2.7 mm for the right atrium, 6.9 ± 2.3 mm for the left atrium, 6.5 ± 2.1 mm for the right ventricle, and 6.4 ± 2.2 for the left ventricle, respectively. CONCLUSION: The novel NEEES was able to correctly identify the site of pacing from various endo- and epicardial sites with high accuracy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“BSPM has nowadays got a real clinical breakthrough as a method for non-invasive 3D EAI, with the competitive concept of “ECG imaging” (ECGi) (239–241), a technology that has rapidly evolved in clinically available devices (242), increasingly used for non-invasive pre-interventional assessment of patients undergoing catheter ablation procedures (243, 244).”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [241]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Marching cubes: A high resolution 3D surface construction algorithm
Lorensen, W. E., & Cline, H. E. (1987) · Proceedings of the 14th annual conference on Computer graphics and interactive techniques · doi:10.1145/37401.37422 · Cited by 10,104 · cited 1×
“The surface triangulation procedure included construction of the primary surface mesh using the marching cubes method.(15) Smoothing and attenuation was performed using the boundary correction method.”
MethodMarching cubes: A high resolution 3D surface construction algorithm
Lorensen, W. E., & Cline, H. E. (1987) · ACM SIGGRAPH Computer Graphics · doi:10.1145/37402.37422 · Cited by 8,647
Automatic three‐dimensional mesh generation by the finite octree technique
Shephard, M. S., & Georges, M. K. (1991) · International Journal for Numerical Methods in Engineering · doi:10.1002/nme.1620320406 · Cited by 732
Cuculich, P. S., Wang, Y., Lindsay, B. D., Faddis, M. N., Schuessler, R. B., Damiano, R. J., Li, L., & Rudy, Y. (2010) · Circulation · doi:10.1161/circulationaha.110.945709 · Cited by 310
Oster, H. S., Taccardi, B., Lux, R. L., Ershler, P. R., & Rudy, Y. (1997) · Circulation · doi:10.1161/01.cir.96.3.1012 · Cited by 256
Noninvasive Panoramic Mapping of Human Atrial Fibrillation Mechanisms: A Feasibility Report
HAISSAGUERRE, M., HOCINI, M., SHAH, A. J., DERVAL, N., SACHER, F., JAIS, P., & DUBOIS, R. (2013) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.12075 · Cited by 221
Jia, P., Ramanathan, C., Ghanem, R. N., Ryu, K., Varma, N., & Rudy, Y. (2006) · Heart Rhythm · doi:10.1016/j.hrthm.2005.11.025 · Cited by 177
Volume discretization into tetrahedra—II. 3D triangulation by advancing front approach
Lo, S. H. (1991) · Computers & Structures · doi:10.1016/0045-7949(91)90059-u · Cited by 127 · cited 1×
“The three-dimensional tetrahedral mesh was constructed using the advancing front approach.(17)”
MethodA Noninvasive Imaging Modality for Cardiac Arrhythmias
Burnes, J. E., Taccardi, B., & Rudy, Y. (2000) · Circulation · doi:10.1161/01.cir.102.17.2152 · Cited by 124
Model-based imaging of cardiac electrical excitation in humans
TILG, B., Fischer, G., Modre, R., Hanser, F., Messnarz, B., Schocke, M. F. H., Kremser, C., Berger, T., Hintringer, F., & Roithinger, F. X. (2002) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2002.804438 · Cited by 106 · cited 1×
“A tailored approach to three-dimensional heart and torso reconstruction has been described previously.(13,14) In the current study, a similar technique was used except for thedifference in custom-built bodysurfacemapping electrodes utilized.”
Method
Cited by
In Vivo Validation of Electrocardiographic Imaging
Cluitmans, M. J. M., Bonizzi, P., Karel, J. M. H., Das, M., Kietselaer, B., de Jong, M. M. J., Prinzen, F. W., Peeters, R., Westra, R. L., & Volders, P. G. A. (2017) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2016.11.012 · Cited by 128
How Accurate Is Inverse Electrocardiographic Mapping? A Systematic In Vivo Evaluation
Bear, L. R., LeGrice, I. J., Sands, G. B., Lever, N. A., Loiselle, D. S., Paterson, D. J., Cheng, L. K., & Smaill, B. H. (2018) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.117.006108 · Cited by 92
Carbucicchio, C., Andreini, D., Piperno, G., Catto, V., Conte, E., Cattani, F., Bonomi, A., Rondi, E., Piccolo, C., Vigorito, S., et al. (2021) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-021-01060-5 · Cited by 77
“Non-invasive electrocardiographic imaging [18] (ECGI, EP Solutions SA) was achieved in selected cases.”
MethodClinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Zweerink, A., Владимирович, З. С., Bakelants, E., Potyagaylo, D., Stettler, C., Chmelevsky, M. P., Dupuis‐Lozeron, E., Hachulla, A.-L., Vallée, J., & Burri, H. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2020.11.029 · Cited by 73
Self-citationGraham, A. J., Orini, M., Zacur, E., Dhillon, G. S., Daw, H., Srinivasan, N. T., Lane, J. D., Cambridge, A., Garcia, J. V., O’Reilly, N., et al. (2019) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.118.007120 · Cited by 62
Evaluation of ECG Imaging to Map Hemodynamically Stable and Unstable Ventricular Arrhythmias
Graham, A. J., Orini, M., Zacur, E., Dhillon, G. S., Daw, H., Srinivasan, N. T., Martin, C., Lane, J. D., Mansell, J., Cambridge, A., et al. (2020) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.119.007377 · Cited by 58
Pezzuto, S., Prinzen, F. W., Potse, M., Maffessanti, F., Regoli, F. D., Caputo, M. L., Conte, G., Krause, R., & Auricchio, A. (2020) · EP Europace · doi:10.1093/europace/euaa330 · Cited by 58
“1 and 6 ± 5 mm).(12,13) For those methods only relying on the 12-lead ECG, localization error during pacing was significantly higher (>20 mm) but predicted the correct segment.”
MethodZeppenfeld, K., Rademaker, R., Al‐Ahmad, A., Carbucicchio, C., de Chillou, C., Cvek, J., Ebert, M., Ho, G., Kautzner, J., Lambiase, P. D., et al. (2024) · EP Europace · doi:10.1093/europace/euae214 · Cited by 60
“Initial comparisons with epicardial recordings indicated accuracy within 1 cm 188–190 and compared well with sites of successful RFCA.”
Noninvasive epicardial and endocardial mapping of premature ventricular contractions
Wissner, E., Revishvili, A., Metzner, A., Tsyganov, A., Kalinin, V., Lemes, C., Saguner, A. M., Maurer, T., Deiss, S., Sopov, O., et al. (2016) · Europace · doi:10.1093/europace/euw103 · Cited by 52
“Non-invasive epicardial and endocardial electrophysiology system The methodology of non-invasive imaging using the NEEES has been previously described in detail.(2) Briefly, up to 28 custom-made vertical arrays incorporating 8 unipolar EKG electrodes each (for a maximum of 224 MRI or CT compatible electrodes) were pla…”
Self-citationElectrocardiographic imaging for cardiac arrhythmias and resynchronization therapy
Pereira, H., Niederer, S., & Rinaldi, C. A. (2020) · EP Europace · doi:10.1093/europace/euaa165 · Cited by 67
“As an example, in patients with already-implanted CRTs, ECGi may aid in finding the pacing site with high accuracy, as demonstrated in a study of 29 patients by Revishvili et al.(41) Ploux et al.”
Sohns, C., Lemeš, C., Metzner, A., Fink, T., Chmelevsky, M. P., Maurer, T., Budanova, M. A., Солнцев, В. Н., Schulze, W. H. W., Staab, W., et al. (2017) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.116.004419 · Cited by 43
Self-citationSchuler, S., Pilia, N., Potyagaylo, D., & Loewe, A. (2021) · Medical Image Analysis · doi:10.1016/j.media.2021.102247 · Cited by 40
Rudic, B., Chaykovskaya, M., Tsyganov, A., Kalinin, V., Tülümen, E., Papavassiliu, T., Dösch, C., Liebe, V., Kuschyk, J., Röger, S., et al. (2016) · Journal of the American Heart Association · doi:10.1161/jaha.116.004095 · Cited by 38
“Body surface ECG data were processed by NEEES, using its inverse problem solution software in combination with heart and torso anatomy, allowing for reconstruction of unipolar EGs at approximately 2500 points on epicardium and endocardium.(23,24) A brief description of the mathematical method for epicardial and endoca…”
MethodSelf-citationRodrigo, M., Climent, A. M., Liberos, A., Hernández‐Romero, I., Arenal, Á., Bermejo, J., Fernández‐Avilés, F., Atienza, F., & Guillem, M. S. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2707413 · Cited by 38
“This technique which allows to reconstruct the electrical activity in the whole heart with much higher spatial resolution than routinely used invasive techniques that employ a few recording points, and the potential to simultaneously estimate the electrical signal everywhere in the myocardium, has been exploited in th…”
Solving the Inverse Problem of Electrocardiography on the Endocardium Using a Single Layer Source
Kalinin, A. V., Potyagaylo, D., & Kalinin, V. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00058 · Cited by 35
“Nevertheless, solution of the endo-epicadial inverse problem was employed for detection of origins of focal ventricular tachycardia (Revishvili et al., 2015; Wissner et al., 2016), determination of electrical rotors in atrial fibrillation (Metzner et al.”
MethodSelf-citationParreira, L., Tsyganov, A., Artyukhina, Е. А., Vernooy, K., Tondo, C., Adragão, P., Ascione, C., Carmo, P., Carvalho, M. S., Egger, M., et al. (2023) · EP Europace · doi:10.1093/europace/euad041 · Cited by 36
“7 Electrocardiographic imaging has demonstrated the ability to assess the entire ventricular activation sequence with good correlation to invasive contact mapping technique in native rhythms with wide QRS, bundle branch block, and paced rhythms.”
Self-citationAita, S., Ogata, K., Yoshida, K., Inaba, T., Kosuge, H., Machino, T., Tsumagari, Y., Hattori, A., Ito, Y., Komatsu, Y., et al. (2019) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2019.06.010 · Cited by 29
Tsyganov, A., Wißner, E., Metzner, A., Mironovich, S., Chaykovskaya, M., Kalinin, V., Chmelevsky, M. P., Lemeš, C., & Kück, K. (2017) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2017.07.018 · Cited by 32
Self-citationThe Impact of Torso Signal Processing on Noninvasive Electrocardiographic Imaging Reconstructions
Bear, L. R., Doğrusöz, Y. S., Good, W. W., Švehlíková, J., Coll‐Font, J., van Dam, E., & MacLeod, R. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3003465 · Cited by 31
ECG Adapted Fastest Route Algorithm to Localize the Ectopic Excitation Origin in CRT Patients
Potyagaylo, D., Chmelevsky, M. P., van Dam, P. M., Budanova, M. A., Владимирович, З. С., Treshkur, T., & Lebedev, D. S. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00183 · Cited by 21
“, 2018), isochronal, isopotential, or phase maps (van Dam et al., 2009a; Revishvili et al., 2015; Wang et al., 2016; Rodrigo et al., 2017), and substrate characterization (Rudy, 2013; Sohns et al.”
BackgroundSelf-citationJanssen, A. M., Potyagaylo, D., Dössel, O., & Oostendorp, T. F. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-017-1715-x · Cited by 21
“Despite the progress made in the methodology of ECG imaging over the past decades and the promising application of the technique in clinical studies [3, 39, 48], an accurate estimation of cardiac sources remains a challenging problem.”
Parreira, L., Carmo, P., Adragão, P., Nunes, S., Soares, A., Marinheiro, R., Budanova, M. A., Владимирович, З. С., Chmelevsky, M. P., Pinho, J., et al. (2020) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2020.07.004 · Cited by 19
Self-citationPilia, N., Schuler, S., Rees, M., Moik, G., Potyagaylo, D., Dössel, O., & Loewe, A. (2023) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2023.102619 · Cited by 16
Metzner, A., Wißner, E., Tsyganov, A., Kalinin, V., Schlüter, M., Lemeš, C., Mathew, S., Maurer, T., Heeger, C.-H. H., Reißmann, B., et al. (2017) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12527 · Cited by 15
“Based on body- surface ECG mapping, the ECGI procedure enables reconstruction of unipolar EGs of thousands of points on the myocardium (Revishvili et al., 2015; Rudic et al., 2016).”
MethodSelf-citationHis-Optimized and Left Bundle Branch-Optimized Cardiac Resynchronization Therapy
Zweerink, A., & Burri, H. (2022) · Cardiac Electrophysiology Clinics · doi:10.1016/j.ccep.2021.12.006 · Cited by 14
Electrocardiographic imaging in the atria
Hernández‐Romero, I., Molero, R., Santos, C. F., Herrero-Martín, C., Climent, A. M., & Guillem, M. S. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02709-7 · Cited by 24
Wißner, E., Saguner, A. M., Metzner, A., Chmelesky, M., Tsyganov, A., Deiß, S., Maurer, T., & Kück, K. (2016) · HeartRhythm Case Reports · doi:10.1016/j.hrcr.2016.02.004 · Cited by 11
Self-citationNew approaches to predicting the risk of sudden death
Behr, E. R., & Ensam, B. (2016) · Clinical Medicine · doi:10.7861/clinmedicine.16-3-283 · Cited by 9
Zhou, S., Sapp, J. L., Šťovíček, P., & Horáček, B. M. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2887041 · Cited by 11
Parreira, L., Carmo, P., Adragão, P., Pinho, J., Budanova, M. A., Владимирович, З. С., Cavaco, D., Marinheiro, R., Carmo, J., Costa, F. M., et al. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.08.046 · Cited by 10
Bhagirath, P., van der Graaf, M., van Dongen, E., de Hooge, J., van Driel, V., Ramanna, H., de Groot, N. M. S., & Götte, M. (2015) · Journal of the American Heart Association · doi:10.1161/jaha.115.002222 · Cited by 9
Parreira, L., Carmo, P., Marinheiro, R., Mesquita, D., Chmelevsky, M. P., Ferreira, A. M., Marques, L., Pinho, J., Chambel, D., Nunes, S., et al. (2022) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2022.05.005 · Cited by 8
Self-citationZhou, Z., Jin, Q., Yu, L. H., Wu, L., & He, B. (2016) · PLoS ONE · doi:10.1371/journal.pone.0163445 · Cited by 10
Atrial location optimization by electrical measures for Electrocardiographic Imaging
Gisbert, V., Jiménez-Serrano, S., Roses-Albert, E., & Rodrigo, M. (2020) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2020.104031 · Cited by 8
Graham, A. J., & SCHILLING, R. J. (2021) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2021.27 · Cited by 10
van der Graaf, A. W. M., Bhagirath, P., de Hooge, J., Ramanna, H., van Driel, V. J. H. M., de Groot, N. M. S., & Götte, M. (2015) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-015-0054-9 · Cited by 7
“[18] demonstrated the accuracy of a novel non-invasive epi- and endocardial electrophysiology system (NEEES) in 29 patients with implanted devices (26 CRT devices in three dual-chamber pacemakers).”
Namdar, M., Chmelevsky, M. P., Hachulla, A.-L., & Shah, D. C. (2016) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.116.003949 · Cited by 7
“The main advantages of NEEES are its noninvasive nature, simultaneous mapping of all 4 chambers of the heart in only 1 cardiac cycle and of epicardial and endocardial surfaces.(11) Such information may be important to guide the identification and ablation of challenging arrhythmias and their sources and potentially of…”
BackgroundSelf-citationParameter variations in personalized electrophysiological models of human heart ventricles
Ushenin, K. S., Kalinin, V., Gitinova, S., Sopov, O., & Solovyova, O. E. (2021) · PLoS ONE · doi:10.1371/journal.pone.0249062 · Cited by 10
“Body-surface electrode positioning was detected by an expert using CT data and Amycard 01 K software [26].”
MethodSelf-citationZhou, S., Sapp, J. L., Dawoud, F., & Horáček, B. M. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2872983 · Cited by 7
Localization of the ventricular pacing site from BSPM and standard 12-lead ECG: a comparison study
Sedova, K. A., van Dam, P. M., Blahova, M., Necasova, L., & Kautzner, J. (2023) · Scientific Reports · doi:10.1038/s41598-023-36768-z · Cited by 6
“That is a localization error of the same range as reported in studies with 184–224 body surface leads 30,34,37 .”
Chmelevsky, M. P., Владимирович, З. С., Budanova, M. A., Treshkur, T. V., & Lebedev, D. S. (2019) · Journal of Arrhythmology · doi:10.35336/va-2019-2-19-28 · Cited by 5
Self-citationParreira, L., Carmo, P., Adragão, P., Marinheiro, R., Budanova, M. A., Cardim, N. M., & Gonçalves, P. D. A. (2020) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2020.07.003 · Cited by 5
Stable Identification of Sources Located on Interface of Nonhomogeneous Media
Mones, J. J. C., Aguayo, E. R. E., Oliveros, J. J. O., Hernández, C., & Castillo, M. M. M. (2021) · Mathematics · doi:10.3390/math9161932 · Cited by 5
The use of noninvasive ECG imaging for examination of a patient with Brugada syndrome
Chaykovskaya, M. K., Rudic, B., Tsyganov, A., Zaklyazminskaya, E. V., Yakovleva, M., & Borggrefe, M. M. (2015) · HeartRhythm Case Reports · doi:10.1016/j.hrcr.2015.04.009 · Cited by 4
“ECGI methodology was extensively validated in a series of animal and human studies.(3,4,14) The combination of optical mapping and phase analysis of optical action potentials has thus become a well-established technique for this kind of experimental investigation of reentrant arrhythmias.”
MethodSelf-citationRole of Myocardial Properties and Pacing Lead Location on ECG in Personalized Paced Heart Models
Ushenin, K. S., Dokuchaev, A., Magomedova, S. M., Sopov, O. V., Kalinin, V., & Solovyova, O. E. (2017) · Computing in cardiology · doi:10.22489/cinc.2017.227-432 · Cited by 4
Self-citationYashkov, М. V. (2018) · Translational Medicine · doi:10.18705/2311-4495-2018-5-2-22-29 · Cited by 4
Bessière, F., Zorgani, A., Robert, J., Daunizeau, L., Cao, E., Vaillant, F., Abell, E., Quesson, B., Catheline, S., Chevalier, P., et al. (2019) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2019.10.017 · Cited by 6
Noninvasive arrhythmia mapping and ablation - myth or reality?
Revishvili, A. S. (2020) · Journal of Arrhythmology · doi:10.35336/va-2020-3-5-8 · Cited by 4
“Параллельно, в 2008-2010 годах нами были опубликованы первые клинические результаты применения оригинальной системы неинвазивного многоканального компьютерного картирования (система Amycard, РФ) для эндо- и эпикардиальной топической диагностики предсердных и желудочковых аритмий [6].”
Self-citationFruelund, P. Z., van Dam, P. M., Melgaard, J., Sommer, A., Lundbye‐Christensen, S., Søgaard, P., Zaremba, T., Graff, C., & Riahi, S. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1087568 · Cited by 3
Kashtanova, N. Y., Gruzdev, I. S., Кондратьев, Е. В., Artyukhina, Е. А., Yashkov, М. V., & Revishvili, A. S. (2018) · Medical Visualization · doi:10.24835/1607-0763-2018-3-33-48 · Cited by 4
Self-citationPulmonalvenenisolation bei Vorhofflimmern: Erkenntnisse aus klinischen Langzeitdaten
Heeger, C.-H. H., Kück, K., & Ouyang, F. O. (2017) · Herz · doi:10.1007/s00059-017-4565-2 · Cited by 4
Noninvasive Cardiac Electrical Activity Mapping Systems: Current Available Options
Ballatore, A., Saglietto, A., Cavallone, E., Ferrari, G. M. D., Anselmino, M., & Dusi, V. (2025) · Reviews in Cardiovascular Medicine · doi:10.31083/rcm44335 · Cited by 3
“Similarly, in patients undergoing AF catheter ablation, it identified pacing sites in different atrial regions with an average error of ~7 millimeters (7.4 mm in the right atrium and 6.9 mm in the left) [54].”
InfluentialGagyi, R. B., Hoogendijk, M. G., Yap, S., & Szili‐Törok, T. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.2012449 · Cited by 2
Thind, M., Lou, Q., Zado, E. S., Markman, T. M., Schaller, R. D., Nazarian, S., Frankel, D. S., Hyman, M. C., Tschabrunn, C. M., & Marchlinski, F. E. (2024) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.16347 · Cited by 2
“…was 10.8 mm, and also from catheter ‐ based pacing during PVI reported a mean accuracy of 7.4 mm in the right atrium (RA) and 6.9 mm in the LA. 11 Hohmann et al. 12 performed a porcine study in which 63 pacing events in the atria of nine subjects resulted in a median ECGi accuracy of 21 mm…”
InfluentialNon-invasive electrophysiological cardiac mapping from development to practice
Artyukhina, Е. А., Yashkov, М. V., & Revishvili, A. S. (2020) · Patologiya krovoobrashcheniya i kardiokhirurgiya · doi:10.21688/1681-3472-2020-4-33-41 · Cited by 3
Self-citationComparison of His-Purkinje and Biventricular pacing in patient-specific computer models
Dokuchaev, A., Bazhutina, A., Khamzin, S., Владимирович, З. С., & Solovyova, O. E. (2021) · IEEE Ural-Siberian Conference on Computational Technologies in Cognitive Science, Genomics and Biomedicine (CSGB) · doi:10.1109/csgb53040.2021.9496032 · Cited by 2
History of development of ventricular arrhythmias diagnostics
Taymasova, I. A., Yashkov, M. V., Dedukh, E. V., Artyukhina, E. A., & Revishvili, A. S. (2021) · Kardiologiia · doi:10.18087/cardio.2021.12.n1469 · Cited by 2
Self-citationBudanova, M. A., Chmelevsky, M. P., Владимирович, З. С., Potyagaylo, D., Rudic, B., Tülümen, E., & Borggrefe, M. (2019) · Computing in cardiology · doi:10.22489/cinc.2019.399 · Cited by 2
Self-citationNon-invasive identification of atrial fibrillation drivers.
Rodrigo, M. (2016) · RiuNet (Universitat Politècnica de València) · doi:10.4995/thesis/10251/75346 · Cited by 4
Non‐invasive Body Surface Potential Mapping of Reentrant Drivers in Human Atrial Fibrillation
Martin, R. I. R., Hocini, M., Dubois, R., Derval, N., Jaïs, P., & Haïssaguerre, M. (2019) · Wiley eBooks · doi:10.1002/9781119152637.ch17 · Cited by 2
Magomedova, S. M., Sopov, O. V., Lomidze, N., Проничева, И. В., Kotanova, E. S., & Serguladze, S. Y. (2017) · Annaly aritmologii · doi:10.15275/annaritmol.2017.4.3 · Cited by 1
Self-citationNoninvasive Phase Mapping of Atrial Flutter in Humans – Comparison with Invasive Mapping
Wißner, E., Asklepios Klinik St Georg, H., Kalinin, V., EP Solutions SA, Y.-L.-B., Kalinin, A. V., EP Solutions SA, Y.-L.-B., Sopov, O., Chmelevsky, M. P., EP Solutions SA, Y.-L.-B., Metzner, A., et al. (2018) · European Journal of Arrhythmia & Electrophysiology · doi:10.17925/ejae.2018.04.01.15 · Cited by 1
“The three-dimensional models of the atrium generated by CARTO 3 and NEEES were merged utilising a customized software.(4) The position of the individual poles of the HALO catheter were projected onto the NEEES atrial model (Figure 1).”
MethodSelf-citationВладимирович, З. С., Chmelevsky, M. P., Potyagaylo, D., Budanova, M. A., Дмитриевич, Р. С., Рыжков, А. В., & Lebedev, D. S. (2018) · Computing in cardiology · doi:10.22489/cinc.2018.347 · Cited by 1
“The diagnostic accuracy of the ECGI application in combination with the CT scans was previously reported [5].”
MethodSelf-citationBudanova, M. A., Chmelevsky, M. P., Владимирович, З. С., Potyagaylo, D., Rudic, B., Tülümen, E., & Borggrefe, M. M. (2018) · Computing in cardiology · doi:10.22489/cinc.2018.384 · Cited by 1
Self-citationResynchronization Therapy for Chronic Heart Failure: Diagnostic and Therapeutic Approaches
Е.Т., С., Dusypov, A. A., Abdrakhmanov, A. S., Orekhov, A. Y., & Turubaev, E. M. (2019) · Kardiologiia · doi:10.18087/cardio.2019.12.n391 · Cited by 4
Artyukhina, Е. А., Викторовна, Д. Е., Yashkov, М. V., & Revishvili, A. S. (2021) · Bulletin of Siberian Medicine · doi:10.20538/1682-0363-2021-3-213-218 · Cited by 1
Self-citationMuça, M., Владимирович, З. С., Bastian, D., Walaschek, J., Buia, V., Rittger, H., Dokuchaev, A., Bayer, T., & Serdoz, L. V. (2024) · Diagnostics · doi:10.3390/diagnostics14060622 · Cited by 1
“It 3-dimensional (3D) CARTO mapping system [7] and with ventricular 18]. implies the combined integration of non-invasive electrocardiographic recordings and cardiac computer tomographic (cCT) or magnetic resonance imaging (cMRI) data. three-case series, we compare the non-invasive and invasive 3D…”
BackgroundMethodVaidya, V. R., Sugure, A., & Asirvatham, S. J. (2017) · Journal of Innovations in Cardiac Rhythm Management · doi:10.19102/icrm.2017.081206 · Cited by 3
Multimodal high-resolution mapping of contracting intact Langendorff-perfused hearts
Schröder-Schetelig, J. (2021) · eDiss (Georg-August-Universität Göttingen) · doi:10.53846/goediss-8250 · Cited by 1
Chronic effects of right ventricular pacing: The impact of right ventricular lead position
Fruelund, P. Z. (2023) · VBN Forskningsportal (Aalborg Universitet) · doi:10.54337/aau534290121 · Cited by 1
Innovative Verfahren der Vorhofflimmertherapie
Metzner, A., Wißner, E., Fink, T., Ouyang, F. O., & Kück, K. (2015) · Der Kardiologe · doi:10.1007/s12181-015-0032-6
Self-citationWhat is Established and What is New in Ablation of Persistent Atrial Fibrillation?
Kück, K., & Metzner, A. (2015) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2015.04.02.80
Cluitmans, M. J. M. (2016) · Research Publications (Maastricht University) · doi:10.26481/dis.20160929mc · Cited by 5
Владимирович, З. С., Chmelevsky, M. P., Budanova, M. A., Ryzhkov, A. V., Trukshina, M. A., Lebedeva, V. K., Sitnikova, M. Y., & Lebedev, D. S. (2017) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2017-7-93-98
“In all patients, there was complete His left bundle branch block (LBBB) with QRS width 205 (190; 215) ms. Non-invasive electrophysiological mapping (NEM) was done with the “Amycard 01C EP LAB” (EP Solutions SA, Switzerland), and magnetic-resonance imaging (MRI).”
InfluentialSelf-citationModels of human heart and torso electrophysiology verified against clinical data
Ushenin, K. S., Dokuchaev, A., Magomedova, S. M., Sopov, O. V., Kalinin, V., & Solovyova, O. E. (2018) · Doklady Meždunarodnoj konferencii "Matematičeskaâ biologiâ i bioinformatika" · doi:10.17537/icmbb18.41
Self-citationDEVELOPMENT OF NONINVASIVE TOPICAL DIAGNOSTICS OF CARDIAC ARRHYTHMIAS
Хлынин, М. С., Batalov, R. E., Kiselev, N. V., Nam, I., Silivanov, V. V., & Spisivtsev, S. A. (2019) · Сибирский журнал клинической и экспериментальной медицины · doi:10.29001/2073-8552-2019-34-2-9-20 · Cited by 1
Phase Mapping for Cardiac Unipolar Electrograms with Neural Network Instead of Phase Transformation
Ushenin, K. S., Nesterova, T., Smarko, D., & Sholokhov, V. (2020) · Ural Symposium on Biomedical Engineering, Radioelectronics and Information Technology (USBEREIT) · doi:10.1109/usbereit48449.2020.9117627
“It may be applied to the processing of unipolar electrograms from the unipolar catheter, multi-leads catheter, and balloons, microelectrode arrays, invasive and non-invasive systems of cardiac mapping [15], [16].”
Budanova, M. A., Chmelevsky, M. P., Владимирович, З. С., Potyagaylo, D., Rudic, B., Tueluemen, E., & Borggrefe, M. (2019) · doi:10.23919/cinc49843.2019.9005738
“The results of recent validation studies showed high precision of non-invasive electrocardiographic imaging (ECGI) for focal arrhythmia localization [1-5].”
ResultSelf-citationKashtanova, N. Y., Kondratyev, E. V., Karmazanovsky, G. G., Gruzdev, I. S., Artyukhina, Е. А., Yashkov, М. V., & Revishvili, A. S. (2021) · Journal of Arrhythmology · doi:10.35336/va-2021-e-25-31
Self-citationАзатовна, Т. И., Yashkov, М. V., Artyukhina, Е. А., & Revishvili, A. S. (2021) · Journal of Arrhythmology · doi:10.35336/va-2021-4-57-61
“Все это позволяет определить вид и тактику оперативного вмешательства на основании дифференциальной оценки зоны интереса по локализации в эпи- и эндокарде [7].”
Self-citationKashtanova, N. Y., Kondratyev, E. V., Karmazanovsky, G. G., Gruzdev, I. S., Artyukhina, Е. А., Yashkov, М. V., & Revishvili, A. S. (2021) · Journal of Arrhythmology · doi:10.35336/va-2021-1-14-22
Self-citationPreoperative Prediction of Optimal Method and Site of Left Ventricular Electrode Implantation
Stepanova, V. V., Алексеевич, М. В., & Владимирович, З. С. (2022) · Cardiac Arrhythmias · doi:10.17816/cardar108644
The Application of Non-invasive Inverse ECG Techniques in Cardiomyopathy
Boonstra, M. (2023) · Utrecht University Repository (Utrecht University) · doi:10.33540/1807
Reconstruction of data inside myocardium from spatial measurements with neural networks
Ushenin, K. S., Dordiuk, V., & Dzhigil, M. (2023) · Ural Symposium on Biomedical Engineering, Radioelectronics and Information Technology (USBEREIT) · doi:10.1109/usbereit58508.2023.10158895
“Systems for non-invasive cardiac mapping reconstruct extracellular potential on the myocardial surface from potentials on the human torso surface [2], [3].”
BackgroundStepanova, V. V., Владимирович, З. С., Алексеевич, М. В., Savelyeva, M. A., & Lebedev, D. S. (2024) · Eurasian heart journal · doi:10.38109/2225-1685-2024-4-82-89
Self-citationMalishevskii, L., Владимирович, З. С., Bazhutina, A., Марков, Н. С., Mikhaylov, E. N., Chumarnaya, T. V., Stepanova, V. V., Лебедева, В. К., Solovyova, O. E., & Lebedev, D. S. (2026) · Diagnostics · doi:10.3390/diagnostics16111724
“Validation of the Amycard electrophysiology system for noninvasive epi-and endocardial mapping has been reported previously [14–16].”
InfluentialSelf-citationThengilan, S., Venuturumilli, R., Sen, N., Solanki, O. B., Sambhav, D. K., & Mohsin, N. (2026) · Cureus · doi:10.7759/cureus.111780
“Three-dimensional electroanatomical mapping systems and high-density mapping systems allow the visualization of the arrhythmic circuits in a more detailed manner, as well as identification of arrhythmogenic substrates at a more detailed level [30,31].”
Innovative Verfahren der Vorhofflimmertherapie
Metzner, A., Wißner, E., Fink, T., Ouyang, F. O., & Kück, K. (2015) · Herz · doi:10.1007/s00059-014-4194-y
Self-citation
Metadata
- Authors
- Revishvili, A. S., Wissner, E., Lebedev, D. S., Lemes, C., Deiss, S., Metzner, A., Kalinin, V. V., Sopov, O. V., Labartkava, E. Z., Kalinin, A. V., Chmelevsky, M., Zubarev, S. V., Chaykovskaya, M. K., Tsiklauri, M. G., & Kuck, K.-H.
- Affiliations
- Russian Academy of Sciences (RU)
- Asklepios Klinik St. Georg (DE)
- Federal Almazov North-West Medical Research Centre (RU)
- Corresponding author
- Wissner, Erik
- Venue
- Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2015-02-02
- Volume
- 17
- Issue
- 8
- Pages
- 1282-1288
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- PMC
- PMC4535554
- OpenAlex
- W2056403421
- Cited by
- 89
- References
- 19
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 4 · 2023: 8 · 2022: 4 · 2021: 14 · 2020: 10 · 2019: 13 · 2018: 8 · 2017: 11 · 2016: 8 · 2015: 6
- FWCI
- 8.01
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc
- Locations
- EP Europace (publishedVersion, cc-by-nc)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Funders
- Russian Science Foundation
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Arrhythmias and Treatments
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- epicardial mapping, endocardial mapping, map accuracy, electroanatomical mapping, CARTO 3, ischemic cardiomyopathy, ventricular dyssynchrony, atrial arrhythmias, ventricular arrhythmias, three-dimensional heart model, cardiac CT, cardiac magnetic resonance imaging
- MeSH terms
- Cardiac Pacing, Artificial, Endocardium, Equipment Design, Female, Humans, Male, Middle Aged, Pericardium, Sensitivity and Specificity, Ventricular Fibrillation, Reproducibility of Results, Tachycardia, Ventricular, Body Surface Potential Mapping, Equipment Failure Analysis, Imaging, Three-Dimensional
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.