Validation of the mapping accuracy of a novel non-invasive epicardial and endocardial electrophysiology system
Open accessThe DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
AIMS: Use of a non-invasive electrocardiographic mapping system may aid in rapid diagnosis of atrial or ventricular arrhythmias or the detection of ventricular dyssynchrony. The aim of the present study was to validate the mapping accuracy of a novel non-invasive epi- and endocardial electrophysiology system (NEEES). METHODS AND RESULTS: Patients underwent pre-procedural computed tomography or magnetic resonance imaging of the heart and torso. Radiographic data were merged with the data obtained from the NEEES during pacing from implanted pacemaker leads or pacing from endocardial sites using an electroanatomical mapping system (CARTO 3, Biosense Webster). The earliest activation as denoted on the NEEES three-dimensional heart model was compared with the true anatomic location of the tip of the pacemaker lead or the annotated pacing site on the CARTO 3 map. Twenty-nine patients [mean age: 62 ± 11 years, 6/29 (11%) female, 21/29 (72%) with ischaemic cardiomyopathy] were enrolled into the pacemaker verification group. The mean distance from the non-invasively predicted pacing site to the anatomic reference site was 10.8 ± 5.4 mm for the right atrium, 7.7 ± 5.8 mm for the right ventricle, and 7.9 ± 5.7 mm for the left ventricle activated via the coronary sinus lead. Five patients [mean age 65 ± 4 years, 2 (33%) females] underwent CARTO 3 verification study. The mean distance between non-invasively reconstructed pacing site and the reference pacing site was 7.4 ± 2.7 mm for the right atrium, 6.9 ± 2.3 mm for the left atrium, 6.5 ± 2.1 mm for the right ventricle, and 6.4 ± 2.2 for the left ventricle, respectively. CONCLUSION: The novel NEEES was able to correctly identify the site of pacing from various endo- and epicardial sites with high accuracy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“BSPM has nowadays got a real clinical breakthrough as a method for non-invasive 3D EAI, with the competitive concept of “ECG imaging” (ECGi) (239–241), a technology that has rapidly evolved in clinically available devices (242), increasingly used for non-invasive pre-interventional assessment of patients undergoing catheter ablation procedures (243, 244).”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [241]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Marching cubes: A high resolution 3D surface construction algorithm
Lorensen, W. E., & Cline, H. E. (1987) · Proceedings of the 14th annual conference on Computer graphics and interactive techniques · doi:10.1145/37401.37422 · Cited by 10,104 · cited 1×
“The surface triangulation procedure included construction of the primary surface mesh using the marching cubes method.(15) Smoothing and attenuation was performed using the boundary correction method.”
MethodMarching cubes: A high resolution 3D surface construction algorithm
Lorensen, W. E., & Cline, H. E. (1987) · ACM SIGGRAPH Computer Graphics · doi:10.1145/37402.37422 · Cited by 8,647
Automatic three‐dimensional mesh generation by the finite octree technique
Shephard, M. S., & Georges, M. K. (1991) · International Journal for Numerical Methods in Engineering · doi:10.1002/nme.1620320406 · Cited by 732
Cuculich, P. S., Wang, Y., Lindsay, B. D., Faddis, M. N., Schuessler, R. B., Damiano, R. J., Li, L., & Rudy, Y. (2010) · Circulation · doi:10.1161/circulationaha.110.945709 · Cited by 310
Oster, H. S., Taccardi, B., Lux, R. L., Ershler, P. R., & Rudy, Y. (1997) · Circulation · doi:10.1161/01.cir.96.3.1012 · Cited by 256
Noninvasive Panoramic Mapping of Human Atrial Fibrillation Mechanisms: A Feasibility Report
HAISSAGUERRE, M., HOCINI, M., SHAH, A. J., DERVAL, N., SACHER, F., JAIS, P., & DUBOIS, R. (2013) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.12075 · Cited by 221
Jia, P., Ramanathan, C., Ghanem, R. N., Ryu, K., Varma, N., & Rudy, Y. (2006) · Heart Rhythm · doi:10.1016/j.hrthm.2005.11.025 · Cited by 177
Volume discretization into tetrahedra—II. 3D triangulation by advancing front approach
Lo, S. H. (1991) · Computers & Structures · doi:10.1016/0045-7949(91)90059-u · Cited by 127 · cited 1×
“The three-dimensional tetrahedral mesh was constructed using the advancing front approach.(17)”
MethodA Noninvasive Imaging Modality for Cardiac Arrhythmias
Burnes, J. E., Taccardi, B., & Rudy, Y. (2000) · Circulation · doi:10.1161/01.cir.102.17.2152 · Cited by 124
Model-based imaging of cardiac electrical excitation in humans
TILG, B., Fischer, G., Modre, R., Hanser, F., Messnarz, B., Schocke, M. F. H., Kremser, C., Berger, T., Hintringer, F., & Roithinger, F. X. (2002) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2002.804438 · Cited by 106 · cited 1×
“A tailored approach to three-dimensional heart and torso reconstruction has been described previously.(13,14) In the current study, a similar technique was used except for thedifference in custom-built bodysurfacemapping electrodes utilized.”
MethodShah, A. J., Hocini, M., Xhaet, O., Pascale, P., Roten, L., Wilton, S. B., Linton, N., Scherr, D., Miyazaki, S., Jadidi, A. S., et al. (2013) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2013.03.082 · Cited by 94
Intini, A., Goldstein, R. N., Jia, P., Ramanathan, C., Ryu, K., Giannattasio, B., Gilkeson, R., Stambler, B. S., Brugada, P., Stevenson, W. G., et al. (2005) · Heart Rhythm · doi:10.1016/j.hrthm.2005.08.019 · Cited by 88
Atrial Noninvasive Activation Mapping of Paced Rhythm Data
Modre, R., Tilg, B., Fischer, G., Hanser, F., Messnarz, B., Seger, M., Schocke, M. F. H., Berger, T., Hintringer, F., & Roithinger, F. X. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.02558.x · Cited by 57 · cited 1×
“A tailored approach to three-dimensional heart and torso reconstruction has been described previously.(13,14) In the current study, a similar technique was used except for thedifference in custom-built bodysurfacemapping electrodes utilized.”
MethodNoninvasive electrocardiographic imaging (ECGI) of scar-related atypical atrial flutter
Wang, Y., Schuessler, R. B., Damiano, R. J., Woodard, P. K., & Rudy, Y. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.08.019 · Cited by 45
Lundqvist, C. B., Auricchio, A., Brugada, J., Boriani, G., Bremerich, J., Cabrera, J. Á., Frank, H., Gutberlet, M., Heidbüchel, H., Kück, K., et al. (2013) · EP Europace · doi:10.1093/europace/eut084 · Cited by 43
Bokeriya, L. A., Revishvili, A. S., Kalinin, A. V., Kalinin, V. V., Lyadzhina, O. A., & Фетисова, Е. А. (2008) · Biomedical Engineering · doi:10.1007/s10527-009-9065-0 · Cited by 19
Self-citationBokeriia, L. A., Revishvili, A. S., Kalinin, A. V., Kalinin, V. V., Oa, L., & Fetisova, E. (2008) · PubMed · Cited by 14
Self-citationДенисов, А. М., Захаров, Е. В., Kalinin, A. V., & Kalinin, V. V. (2010) · Computational Mathematics and Mathematical Physics · doi:10.1134/s0965542510070067 · Cited by 12
Self-citationKalinin, A. V. (2011) · Computational Mathematics and Modeling · doi:10.1007/s10598-011-9086-x · Cited by 10
Self-citation
Cited by
In Vivo Validation of Electrocardiographic Imaging
Cluitmans, M. J. M., Bonizzi, P., Karel, J. M. H., Das, M., Kietselaer, B., de Jong, M. M. J., Prinzen, F. W., Peeters, R., Westra, R. L., & Volders, P. G. A. (2017) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2016.11.012 · Cited by 128
How Accurate Is Inverse Electrocardiographic Mapping? A Systematic In Vivo Evaluation
Bear, L. R., LeGrice, I. J., Sands, G. B., Lever, N. A., Loiselle, D. S., Paterson, D. J., Cheng, L. K., & Smaill, B. H. (2018) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.117.006108 · Cited by 92
Carbucicchio, C., Andreini, D., Piperno, G., Catto, V., Conte, E., Cattani, F., Bonomi, A., Rondi, E., Piccolo, C., Vigorito, S., et al. (2021) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-021-01060-5 · Cited by 77
“Non-invasive electrocardiographic imaging [18] (ECGI, EP Solutions SA) was achieved in selected cases.”
MethodClinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Zweerink, A., Владимирович, З. С., Bakelants, E., Potyagaylo, D., Stettler, C., Chmelevsky, M. P., Dupuis‐Lozeron, E., Hachulla, A.-L., Vallée, J., & Burri, H. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2020.11.029 · Cited by 73
Self-citation
Metadata
- Authors
- Revishvili, A. S., Wissner, E., Lebedev, D. S., Lemes, C., Deiss, S., Metzner, A., Kalinin, V. V., Sopov, O. V., Labartkava, E. Z., Kalinin, A. V., Chmelevsky, M., Zubarev, S. V., Chaykovskaya, M. K., Tsiklauri, M. G., & Kuck, K.-H.
- Affiliations
- Russian Academy of Sciences (RU)
- Asklepios Klinik St. Georg (DE)
- Federal Almazov North-West Medical Research Centre (RU)
- Corresponding author
- Wissner, Erik
- Venue
- Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2015-02-02
- Volume
- 17
- Issue
- 8
- Pages
- 1282-1288
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- PMC
- PMC4535554
- OpenAlex
- W2056403421
- Cited by
- 89
- References
- 19
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 4 · 2023: 8 · 2022: 4 · 2021: 14 · 2020: 10 · 2019: 13 · 2018: 8 · 2017: 11 · 2016: 8 · 2015: 6
- FWCI
- 8.01
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc
- Locations
- EP Europace (publishedVersion, cc-by-nc)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Funders
- Russian Science Foundation
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Arrhythmias and Treatments
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- epicardial mapping, endocardial mapping, map accuracy, electroanatomical mapping, CARTO 3, ischemic cardiomyopathy, ventricular dyssynchrony, atrial arrhythmias, ventricular arrhythmias, three-dimensional heart model, cardiac CT, cardiac magnetic resonance imaging
- MeSH terms
- Cardiac Pacing, Artificial, Endocardium, Equipment Design, Female, Humans, Male, Middle Aged, Pericardium, Sensitivity and Specificity, Ventricular Fibrillation, Reproducibility of Results, Tachycardia, Ventricular, Body Surface Potential Mapping, Equipment Failure Analysis, Imaging, Three-Dimensional
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.