Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis
Open accessThe DOI matches its Crossref record; 72 of the 72 references it cites were found in a registry or an index.
Abstract
Infectious complications after cardiac implantable electronic device (CIED) implantation are increasing over time and are associated with substantial mortality and healthcare costs. The aim of this study was to systematically summarize the literature on risk factors for infection after pacemaker, implantable cardioverter-defibrillator, and cardiac resynchronization therapy device implantation. Electronic searches (up to January 2014) were performed in PubMed, Scopus, and Web of Science databases. Sixty studies (21 prospective, 9 case-control, and 30 retrospective cohort studies) met the inclusion criteria. The average device infection rate was 1-1.3%. In the meta-analysis, significant host-related risk factors for infection included diabetes mellitus (odds ratio (OR) [95% confidence interval] = 2.08 [1.62-2.67]), end-stage renal disease (OR = 8.73 [3.42-22.31]), chronic obstructive pulmonary disease (OR = 2.95 [1.78-4.90]), corticosteroid use (OR = 3.44 [1.62-7.32]), history of the previous device infection (OR = 7.84 [1.94-31.60]), renal insufficiency (OR = 3.02 [1.38-6.64]), malignancy (OR = 2.23 [1.26-3.95]), heart failure (OR = 1.65 [1.14-2.39]), pre-procedural fever (OR = 4.27 [1.13-16.12]), anticoagulant drug use (OR = 1.59 [1.01-2.48]), and skin disorders (OR = 2.46 [1.04-5.80]). Regarding procedure-related factors, post-operative haematoma (OR = 8.46 [4.01-17.86]), reintervention for lead dislodgement (OR = 6.37 [2.93-13.82]), device replacement/revision (OR = 1.98 [1.46-2.70]), lack of antibiotic prophylaxis (OR = 0.32 [0.18-0.55]), temporary pacing (OR = 2.31 [1.36-3.92]), inexperienced operator (OR = 2.85 [1.23-6.58]), and procedure duration (weighted mean difference = 9.89 [0.52-19.25]) were all predictors of CIED infection. Among device-related characteristics, abdominal pocket (OR = 4.01 [2.48-6.49]), epicardial leads (OR = 8.09 [3.46-18.92]), positioning of two or more leads (OR = 2.02 [1.11-3.69]), and dual-chamber systems (OR = 1.45 [1.02-2.05]) predisposed to device infection. This systematic review on risk factors for CIED infection may contribute to developing better infection control strategies for high-risk patients and can also help risk assessment in the management of device revisions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring inconsistency in meta-analyses
Higgins, J. P. T., Thompson, S. G., Deeks, J. J., & Altman, D. G. (2003) · BMJ · doi:10.1136/bmj.327.7414.557 · Cited by 63,139
Bias in meta-analysis detected by a simple, graphical test
Egger, M., Smith, G. D., Schneider, M., & Minder, C. (1997) · BMJ · doi:10.1136/bmj.315.7109.629 · Cited by 56,722
Meta-analysis in clinical trials
DerSimonian, R., & Laird, N. (1986) · Controlled Clinical Trials · doi:10.1016/0197-2456(86)90046-2 · Cited by 39,552
Duval, S., & Tweedie, R. (2000) · Biometrics · doi:10.1111/j.0006-341x.2000.00455.x · Cited by 14,099
Durack, D. T., Lukes, A. S., Bright, D. K., & Service, D. E. (1994) · The American Journal of Medicine · doi:10.1016/0002-9343(94)90143-0 · Cited by 2,572
Perioperative Management of Antithrombotic Therapy
Douketis, J. D., Spyropoulos, A. C., Spencer, F. A., Mayr, M., Jaffer, A. K., Eckman, M. H., Dunn, A. S., & Kunz, R. (2012) · Chest · doi:10.1378/chest.11-2298 · Cited by 1,641
Baddour, L. M., Epstein, A. E., Erickson, C. C., Knight, B. P., Levison, M. E., Lockhart, P. B., Masoudi, F. A., Okum, E. J., Wilson, W. R., Beerman, L. B., et al. (2010) · Circulation · doi:10.1161/circulationaha.109.192665 · Cited by 1,030
Greenspon, A. J., Patel, J. D., Lau, E., Ochoa, J. A., Frisch, D. R., Ho, R. T., Pavri, B. B., & Kurtz, S. M. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.04.033 · Cited by 780
Management and Outcome of Permanent Pacemaker and Implantable Cardioverter-Defibrillator Infections
Sohail, M. R., Uslan, D. Z., Khan, A. H., Friedman, P. A., Hayes, D. L., Wilson, W. R., Steckelberg, J. M., Stoner, S., & Baddour, L. M. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.01.072 · Cited by 751
Klug, D., Balde, M., Pavin, D., Hidden-Lucet, F., Clementy, J., Sadoul, N., Rey, J. L., Lande, G., Lazarus, A., Victor, J., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.678664 · Cited by 700
Cited by
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
2023 ESC Guidelines for the management of endocarditis
Delgado, V., Ajmone Marsan, N., de Waha, S., Bonaros, N., Brida, M., Burri, H., Caselli, S., Doenst, T., Ederhy, S., Erba, P. A., et al. (2023) · European Heart Journal · doi:10.1093/eurheartj/ehad193 · Cited by 1,754
Kusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R., Cha, Y.-M., Clancy, J., Deharo, J.-C., Ellenbogen, K. A., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.09.001 · Cited by 1,199
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.044 · Cited by 1,008
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Challenges in Infective Endocarditis
Cahill, T. J., Baddour, L. M., Habib, G., Hoen, B., Salaun, E., Pettersson, G. B., Schäfers, H. J., & Prendergast, B. D. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.10.066 · Cited by 715
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Circulation · doi:10.1161/cir.0000000000000628 · Cited by 603
Blomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · EP Europace · doi:10.1093/europace/euz246 · Cited by 423
Antibacterial Envelope to Prevent Cardiac Implantable Device Infection
Tarakji, K. G., Mittal, S., Kennergren, C., Corey, R., Poole, J. E., Schloss, E., Gallastegui, J., Pickett, R. A., Evonich, R., Philippon, F., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1901111 · Cited by 405
Management Considerations in Infective Endocarditis: A Review
Wang, A., Gaca, J. G., & Chu, V. H. (2018) · JAMA · doi:10.1001/jama.2018.7596 · Cited by 368
Blomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · European Journal of Cardio-Thoracic Surgery · doi:10.1093/ejcts/ezz296 · Cited by 318
Cardiac Pacemakers: Function, Troubleshooting, and Management
Mulpuru, S. K., Madhavan, M., McLeod, C. J., Cha, Y.-M., & Friedman, P. A. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.10.061 · Cited by 303
Clinically Significant Pocket Hematoma Increases Long-Term Risk of Device Infection
Essebag, V., Verma, A., Healey, J. S., Krahn, A. D., Kalfon, E., Coutu, B., Ayala-Paredes, F., Tang, A. S., Sapp, J., Sturmer, M., et al. (2016) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.01.009 · Cited by 201
Boriani, G., Savelieva, I., Dan, G.-A., Deharo, J. C., Ferro, C., Israel, C. W., Lane, D. A., La Manna, G., Morton, J., Mitjans, A. M., et al. (2015) · EP Europace · doi:10.1093/europace/euv202 · Cited by 182
Deer, T. R., Pope, J. E., Hayek, S. M., Lamer, T. J., Veizi, I. E., Erdek, M., Wallace, M. S., Grider, J. S., Levy, R. M., Prager, J., et al. (2017) · Neuromodulation: Technology at the Neural Interface · doi:10.1111/ner.12579 · Cited by 174
“Factors associated with an increased risk of SSI include anemia, smoking (134,135), diabetes mellitus (136,137), cancer, malnutrition, obesity, cardiovascular disease (other than hypertension), and active drug and alcohol abuse (138).”
BackgroundJuneau, D., Golfam, M., Hazra, S., Zuckier, L. S., Garas, S., Redpath, C., Bernick, J., Leung, E., Chih, S., Wells, G., et al. (2017) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.116.005772 · Cited by 160
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Heart Rhythm · doi:10.1016/j.hrthm.2018.10.037 · Cited by 131
Endovascular neuromodulation: safety profile and future directions
Raza, S. A., Opie, N. L., Morokoff, A., Sharma, R. P., Mitchell, P. J., & Oxley, T. J. (2020) · Frontiers in Neurology · doi:10.3389/fneur.2020.00351 · Cited by 29
Targeting Cardiovascular Implant Infection: Multimodality and Molecular Imaging
Sohns, J. M., Bavendiek, U., Ross, T. L., & Bengel, F. M. (2017) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.117.005376 · Cited by 24
Metadata
- Authors
- Polyzos, K. A., Konstantelias, A. A., & Falagas, M. E.
- Venue
- EP Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2015-04-29
- Volume
- 17
- Issue
- 5
- Pages
- 767-777
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W2153172728
- Cited by
- 493
- References
- 72
- FWCI
- 6.02
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.