Validation of a Portable Monitoring System for the Diagnosis of Obstructive Sleep Apnea Syndrome
Open accessAbstract
STUDY OBJECTIVE: To evaluate if a portable monitor could accurately measure the apnea-hypopnea index (AHI) in patients with a suspicion of obstructive sleep apnea (OSA). DESIGN: Prospective and randomized. SETTING: Sleep laboratory. PARTICIPANTS: 80 participants: 70 patients with clinical OSA suspicion and 10 subjects without suspicion of OSA. INTERVENTIONS: N/A. MEASUREMENTS AND RESULTS: Three-order randomized evaluations were performed: (1) STD (Stardust II) used at the participants' home (STD home), (2) STD used simultaneously with PSG in the sleep lab (STD+PSG lab), and (3) PSG performed without the STD (PSG lab). Four AHI values were generated and analyzed: (a) STD home; (b) STD from STD+PSG lab; (c) PSG from STD+PSG (named PSG+STD lab); and (d) PSG lab. Two technicians, blinded to study details, performed the analyses of all evaluations. There was a strong correlation between AHI from the STD and PSG recordings for all 4 AHI values (all correlations above 0.87). Sensitivity, specificity, and positive and negative predictive values at AHI cut-off values of 5, 15, and 30 events/hour were calculated. AHI values from the PSG lab and PSG+STD lab were compared to STD home and STD+PSG lab and showed the best results when STD and PSG were performed simultaneously. In all analyses, the area under ROC curve was at least 0.90. With multiple comparisons, diagnostic agreement was between 91% and 75%. The Bland Altman analyses showed strong agreement between AHI values from the STD and PSG recordings, especially when comparing the AHI from simultaneous STD and PSG recordings. CONCLUSION: These data suggest that the STD is accurate in confirming the diagnosis of OSA where there is a suspicion of the disorder. Better agreement occurred during simultaneous recordings.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
When using an AHI ≥ 5 cutoff, accuracy of Type 3 HSAT devices in a high-risk population (assuming a prevalence of 87%) ranged from 84% to 91%, whereas in a low-risk population (assuming a prevalence of 55%) accuracy ranged from 70% to 78%.
“When using an AHI ≥ 5 cutoff, accuracy in a high-risk population (assuming a prevalence of 87%) ranged from 84% to 91%, whereas in a low-risk population (assuming a prevalence of 55%) accuracy ranged from 70% to 78% based on the seven studies”
From Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline (2017) · cites this paper as [90]
The Statistical Analysis and Results sections report testing one portable monitor at an AHI cutoff of 5 and comparing its results with polysomnography. They do not report accuracy ranges of 84%–91% and 70%–78% for populations with assumed prevalences of 87% and 55%, or an analysis based on seven studies. Page numbers are unavailable in our copy. (LLM, conf=0.84)
Experimental — accuracy not yet measuredUsing a cutoff of AHI ≥ 15, the accuracy of Type 3 HSAT devices in a high-risk population ranged from 65% to 91%, based on six studies.
“Using a cutoff of AHI ≥ 15, the accuracy of these devices in a high-risk population ranged from 65% to 91%, based on six studies”
From Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline (2017) · cites this paper as [90]
The Results section reports evaluation of a portable monitor at an AHI cutoff of 15, and the Discussion describes its performance in patients from a sleep outpatient setting. This is a validation of the Stardust II, not a six-study estimate of Type 3 device accuracy; the cited passages do not report a 65%–91% range at that cutoff. (LLM, conf=0.86)
Experimental — accuracy not yet measuredSeven studies evaluated the performance of Type 3 HSAT devices against PSG, with AHI cutoffs varying across studies.
“Seven studies evaluated the performance of Type 3 HSAT devices against PSG, but the AHI cutoffs employed varied across studies, resulting in sub-grouping by AHI cutoffs for our analyses.”
From Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline (2017) · cites this paper as [90]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A New Method for Measuring Daytime Sleepiness: The Epworth Sleepiness Scale
Johns, M. W. (1991) · Sleep · doi:10.1093/sleep/14.6.540 · Cited by 17,364 · see citing works
Flagged first, then by citations
Research works printed but not linked: 30
- in detection of obstructive sleep apnea. Otolaryngol Head Neck Surg 2001;125:227-30.
- Chesson Jr AL, Berry RB, Pack A. Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults. A joint project sponsored by the American Academy of Sleep Medicine, the American Thoracic Society, and the American College of Chest Physicians. Sleep 2003;26:907-13
- Flemons WW, Littner MR, Rowley JA, et al. Home diagnosis of sleep apnea: a systematic review of the literature. An evidence re- view cosponsored by the American Academy of Sleep Medicine, the American College of Chest Physicians, and the American Tho racic Society. Chest 2003;124:1543-79.
- Netzer NC, Stoohs RA, Netzer CM, Clark K, Strohl KP. Using the Berlin Questionnaire to identify patients at risk for the sleep apnea syndrome. Am Intern Med 1999;131:485-91.
- American Academy of Sleep Medicine Task Force. Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep 1999;22:667-89.
- American Sleep Disorders Association. EEG arousals: scoring rules and examples -Atlas Task Force. Sleep 1992;15:174-82.
- Yin M, Miyazaki S, Itasaka Y, et al. A preliminary study on ap- plication of portable monitoring for diagnosis of obstructive sleep apnea. Auris Nasus Larynx 2005;32:151-6.
- Metersky ML, Castriotta RJ. The effect of polysomnography on sleep position: possible implications on the diagnosis of position- al obstructive sleep apnea. Respiration 1996;63:283-7.
- Bittencourt LR, Suchecki D, Tufik S, et al. The variability of the apnoea-hypopnoea index. J Sleep Res 2001;10:245-51.
- American Academy of Sleep Medicine. Portable monitoring in the diagnosis of obstructive sleep apnea. J Clin Sleep Med 2006;2:274.
- Littner MR. Portable monitoring in the diagnosis of the ob- structive sleep apnea syndrome. Semin Respir Crit Care Med 2005;26:56-67.
- Alonso Alvarez Mde L, Terán Santos J, Cordero Guevara J, et al. Reliability of home respiratory polygraphy for the diagnosis of sleep apnea-hypopnea syndrome: analysis of costs. Arch Bron- coneumol 2008;44:22-8.
- Mulgrew AT, Fox N, Ayas NT, Ryan CF. Diagnosis and initial man- agement of obstructive sleep apnea without polysomnography: a randomized validation study. Ann Intern Med 2007;146:157-66. Monitoring Task Force of the American Academy of Sleep Medi- cine. J Clin Sleep Med 2007;3:737-47.
- American Sleep Disorders Association. Practice parameters for use of portable recording in the assessment of obstructive sleep apnea. Sleep 1994;17:372-77.
- Schäfer H, Ewig S, Hasper E, Lüderitz B. Predictive diagnostic value of clinical assessment and non laboratory monitoring sys- tem recordings in patients with symptoms suggestive of obstruc- tive sleep apnea syndrome. Respiration 1997;64:194-9.
- Westbrook PR, Levendowski DJ, Cvetinovic M, et al. Descrip- tion and validation of the apnea risk evaluation system: a novel method to diagnose sleep apnea-hypopnea in the home. Chest 2005;128:2166-75.
- Parra O, García-Esclasans N, Montserrat JM, et al. Should pa- tients with sleep apnoea/hypopnoea syndrome be diagnosed and managed on the basis of home sleep studies? Eur Respir J 1997;10:1720-4.
- Yin M, Miyazaki S, Ishikawa K. Evaluation of type 3 portable monitoring in unattended home setting for suspected sleep apnea: factors that may affect its accuracy. Otolaryngol Head Neck Surg 2006;134:204-9.
- Zou D, Grote L, Peker Y, Lindblad U, Hedner J. Validation a por- table monitoring device for sleep apnea diagnosis in a population based cohort using synchronized home polysomnography. Sleep 2006;29:367-74.
- Ayappa I, Norman RG, Seelall V, Rapoport DM. Validation of a self-applied unattended monitor for sleep disordered breathing. J Clin Sleep Med 2008;4:26-37.
- Penzel T, Kesper K, Pinnow I, Becker HF, Vogelmeter C. Periph- eral arterial tonometry oximetry and actigraphy for ambulatory recording of sleep apnea. Physiol Meas 2004;25:1025-36.
- Øverland B, Bruskeland G, Akre H, Skatvedt O. Evaluation of a portable recording device (Reggie) with actimeter and na- sopharyngeal/esophagus catheter incorporated. Respiration 2005;72:600-5.
- Man GC, Kang BV. Validation of a portable sleep apnea monitor- ing device. Chest 1995;108:388-93.
- White DP, Gibb TJ, Wall JM, Westbrook PR. Assessment of ac- curacy and analysis time of a novel device to monitor sleep and breathing in the home. Sleep 1995;18:115-26.
- Dingli K, Coleman EL, Vennelle M, et al. Evaluation of a porta- ble device for diagnosing the sleep apnoea/hypopnoea syndrome. Eur Respir J 2003;21:253-9.
- Ballester E, Solans M, Vila X, et al. Evaluation of a portable respiratory recording device for detecting apnoeas and hypo- pnoeas in participants from a general population. Eur Respir J 2000;16:123-7.
- Candela A, Hernández L, Asensio S, et al. Validation of a respira- tory polygraphy system in the diagnosis of sleep apnea syndrome. Arch Bronconeumol 2005;41:71-7.
- Calleja JM, Esnaola S, Rubio R, Durán J. Comparison of a car- diorespiratory device versus polysomnography for diagnosis of sleep apnoea. Eur Respir J 2002;20:1505-10.
- Verse T, Pirsig W, Junge-Hülsing B, Kroker B. Validation of the POLY-MESAM seven-channel ambulatory recording unit. Chest 2000;117:1613-8.
- Reichert JA, Bloch DA, Cundiff E, Votteri BA. Comparison of the NovaSom QSG, a new sleep apnea home-diagnostic system, and polysomnography. Sleep Med 2003;4:213-8.
Cited by
Sorted by citations
Kapur, V. K., Auckley, D. H., Chowdhuri, S., Kuhlmann, D. C., Mehra, R., Ramar, K., & Harrod, C. G. (2017) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.6506 · Cited by 3,253 · see citing works
High STOP-Bang score indicates a high probability of obstructive sleep apnoea
Chung, F., Subramanyam, R., Liao, P., Sasaki, E., Shapiro, C., & Sun, Y. (2012) · British Journal of Anaesthesia · doi:10.1093/bja/aes022 · Cited by 988 · see citing works
Obstructive Sleep Apnea Devices for Out-Of-Center (OOC) Testing: Technology Evaluation
Collop, N. A., Tracy, S. L., Kapur, V., Mehra, R., Kuhlmann, D., Fleishman, S. A., & Ojile, J. M. (2011) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.1328 · Cited by 427 · see citing works
El Shayeb, M., Topfer, L.-A., Stafinski, T., Pawluk, L., & Menon, D. (2014) · Canadian Medical Association Journal · doi:10.1503/cmaj.130952 · Cited by 237 · see citing works
Masa, J. F., Corral, J., Pereira, R., Duran-Cantolla, J., Cabello, M., Hernandez-Blasco, L., Monasterio, C., Alonso, A., Chiner, E., Rubio, M., et al. (2011) · Thorax · doi:10.1136/thx.2010.152272 · Cited by 161 · see citing works
Prevalence, risk factors, and type of sleep apnea in patients with paroxysmal atrial fibrillation
Traaen, G. M., Øverland, B., Aakerøy, L., Hunt, T. E., Bendz, C., Sande, L., Aakhus, S., Zaré, H., Steinshamn, S., Anfinsen, O. G., et al. (2020) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2019.100447 · Cited by 71 · see citing works
Guerrero, A., Embid, C., Isetta, V., Farre, R., Duran-Cantolla, J., Parra, O., Barbé, F., Montserrat, J. M., & Masa, J. F. (2014) · Sleep · doi:10.5665/sleep.3932 · Cited by 62 · see citing works
Severity of individual obstruction events is gender dependent in sleep apnea
Leppänen, T., Kulkas, A., Duce, B., Mervaala, E., & Töyräs, J. (2017) · Sleep and Breathing · doi:10.1007/s11325-016-1430-0 · Cited by 45 · see citing works
Chernyshev, O. Y., McCarty, D. E., Moul, D. E., Liendo, C., Caldito, G. C., Munjampalli, S. K., Kelley, R., & Chesson Jr, A. L. (2015) · Nature and Science of Sleep · doi:10.2147/nss.s85780 · Cited by 23 · see citing works
Kawakami, H., Nagai, T., Fujii, A., Uetani, T., Nishimura, K., Inoue, K., Suzuki, J., Oka, Y., Okura, T., Higaki, J., et al. (2016) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-016-0148-z · Cited by 23 · see citing works
- Authors
- Santos-Silva, R., Sartori, D. E., Truksinas, V., Truksinas, E., Alonso, F. F. F. D., Tufik, S., & Bittencourt, L. R. A.
- Affiliations
- Universidade Federal de São Paulo (BR)
- Corresponding author
- Bittencourt, Lia R. A.
- Venue
- Sleep
- Publisher
- Oxford University Press (OUP)
- Published
- 2009-05
- Volume
- 32
- Issue
- 5
- Pages
- 629-636
- ISSN
- 0161-8105 (ISSN-L)
- 1550-9109
- PMC
- PMC2675898
- OpenAlex
- W154426793
- Cited by
- 142
- References
- 1
- Citations by year
- 2026: 4 · 2025: 2 · 2024: 4 · 2023: 5 · 2022: 6 · 2021: 9 · 2020: 11 · 2019: 9 · 2018: 9 · 2017: 8 · 2016: 9 · 2015: 9 · 2014: 11 · 2013: 10 · 2012: 14
- FWCI
- 7.57
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- SLEEP (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Americanae (AECID Library) (submittedVersion, other-oa)
- LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) (submittedVersion, other-oa)
- Funders
- Associação Fundo de Incentivo à Pesquisa
- Topics
- Health Sciences › Medicine › Physiology › Obstructive Sleep Apnea Research
- Social Sciences › Psychology › Experimental and Cognitive Psychology › Sleep and related disorders
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- obstructive sleep apnea, apnea-hypopnea index, home sleep monitoring, home sleep testing, polysomnography, diagnostic accuracy, OSA diagnosis
- MeSH terms
- Adult, Brazil, Equipment Design, Female, Humans, Male, Middle Aged, Prospective Studies, Sensitivity and Specificity, Polysomnography, Monitoring, Ambulatory, Sleep Apnea, Obstructive
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review