Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room

Dumas, Constantine, Wahr, Joyce A., Tremper, Kevin K. · 1996 · Anesthesia & Analgesia
Cited by 89

The DOI matches its Crossref record.

Abstract

The frequency and nature of spurious pulse oximetry readings were compared using both a conventional pulse oximeter (CPO) and a prototype Masimo signal extraction technology pulse oximeter (Masimo SET). At a university hospital, 50 ASA physical status I-IV adult patients who underwent general or spinal-epidural anesthesia were selected from a group of 250 patients on the basis of high-alarm generation with routine postoperative pulse oximetry. Pulse oximetry data were recorded simultaneously from both devices with a computer. Overall, the CPO alarm frequency (i.e., oxygen saturation < 90%, or complete signal loss) was once every 13 min, and 87% of these alarms were considered false. Alarms were considered false based on reference electrocardiographs (16 patients), arterial blood gases (7 patients), and clinical assessment. The prototype Masimo SET device alarm frequency was once every 30 min, and 59% of these were considered false. During arm motion with 15 patients, the CPO device produced spurious signals on 54 occasions compared with five for the prototype Masimo SET. The incidence of artifactual pulse oximetry events during patient motion appear to be substantially reduced with the prototype Masimo SET device, relative to a CPO device.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00000539-199608000-00012

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 89

Metadata

Authors
Dumas, C., Wahr, J. A., & Tremper, K. K.
Venue
Anesthesia & Analgesia
Published
1996-08
Volume
83
Issue
2
Pages
269-272
ISSN
0003-2999 (ISSN-L)
OpenAlex
W2004663224
Cited by
89
References
0
FWCI
7.31
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room. https://app.cochat.ai/papers/10.1097/00000539-199608000-00012
The paper
Dumas, C., Wahr, J. A., & Tremper, K. K. (1996). Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room. Anesthesia & Analgesia. https://doi.org/10.1097/00000539-199608000-00012

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.