Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Abstract
Objective The purposes of the study were to: (1) describe the aggregate strength of the relationship of arterial oxygen saturation as measured by pulse oximetry with the standard of arterial blood gas analysis as measured by co-oximetry, (2) examine how various factors affect this relationship, and (3) describe an aggregate estimate of the bias and precision between oxygen saturation as measured by pulse oximetry and the standard in vitro measures.Design A meta-analysis was conducted.Sample Seventy-four studies from 1976 to 1994 met the inclusion criteria of: (1) adult study population, (2) quantitative analysis of empirical data, and (3) bivariate correlations or bias and precision estimates between pulse oximeter and co-oximeter values.Results There were a total of 169 oximeter trials on 41 oximeter models from 25 different manufacturers. Studies were conducted in various settings with a variety of subjects, with most being healthy adult volunteers. The weighted mean r, based on 39 studies (62 oximeter trials) for which the r statistic and number of data points were available, was 0.895 (var [r] = 0.014). Based on 23 studies (82 oximeter trials) for which bias and precision estimates and number of data points were available, the mean absolute bias and precision were 1.999 and 0.233, respectively. Several factors were found to affect the accuracy of pulse oximetry.Conclusion Pulse oximeters were found to be accurate within 2% (+/- 1 SD) or 5% (+/- 2 SD) of in vitro oximetry in the range of 70% to 100% Sao2. In comparing ear and finger probes, readings from finger probes were more accurate. Pulse oximeters may fail to record accurately the true Sao2 during severe or rapid desaturation, hypotension, hypothermia, dyshemoglobinemia, and low perfusion states.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The specified values of mean bias (SD for precision) (2% (± 1%)) in current standards are consistent with the outdated 1995 Jensen review results.
“The current standards did not point out evidence sources used to underpin such requirements, but the specified values of mean bias (SD for precision) (2% (± 1%)) are consistent with the outdated 1995 Jensen review results [8].”
From The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: a systematic review and meta-analysis (2022) · cites this paper as [8]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Hunter, J., & Schmidt, F. (2004) · doi:10.4135/9781412985031 · Cited by 3,023 · see citing works
Yelderman, M., & New, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 488 · see citing works
Randomized Evaluation of Pulse Oximetry in 20,802 Patients; II
Møller, J. T., JOHANNESSEN, N. W., Espersen, K., et al. (1993) · Anesthesiology · doi:10.1097/00000542-199303000-00007 · Cited by 444 · see citing works
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 438 · see citing works
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 337 · see citing works
Accuracy of Response of Six Pulse Oximeters to Profound Hypoxia
Severinghaus, J. W., & Naifeh, K. H. (1987) · Anesthesiology · doi:10.1097/00000542-198710000-00017 · Cited by 314 · see citing works
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617328 · Cited by 305 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Spectrophotometric monitoring of arterial oxygen saturation in the fingertip
Yoshiya, I., Shimada, Y., & Tanaka, K. (1980) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442476 · Cited by 262 · see citing works
Randomized Evaluation of Pulse Oximetry in 20,802 Patients; I
Møller, J. T., Pedersen, T., Rasmussen, L. S., Jensen, P. F., Pedersen, B. D., Ravlo, O., Rasmussen, N. H., Espersen, K., JOHANNESSEN, N. W., Cooper, J. B., et al. (1993) · Anesthesiology · doi:10.1097/00000542-199303000-00006 · Cited by 222 · see citing works
Errors in 14 pulse oximeters during profound hypoxia
Severinghaus, J. W., Naifeh, K. H., & Koh, S. O. (1989) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617877 · Cited by 218 · see citing works
Potential errors in pulse oximetry
Webb, R. K., Ralston, A., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09411.x · Cited by 214 · see citing works
Kelleher, J. F. (1989) · The Journal of Clinical Monitoring · doi:10.1007/bf01618369 · Cited by 205 · see citing works
Wynne, J. W., Block, A. J., Hemenway, J., Hunt, L. A., & Flick, M. R. (1979) · The American Journal of Medicine · doi:10.1016/0002-9343(79)91166-5 · Cited by 204 · see citing works
The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
Barker, S. J., & Shah, N. K. (1997) · Anesthesiology · doi:10.1097/00000542-199701000-00014 · Cited by 197 · see citing works
Effects of Intravenously Administered Dyes on Pulse Oximetry Readings
Scheller, M. S., UNGER, R. J., & Kelner, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 194 · see citing works
Schnapp, L. M., & Cohen, N. H. (1990) · CHEST Journal · doi:10.1378/chest.98.5.1244 · Cited by 194 · see citing works
Ralston, A., Webb, R. K., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 190 · see citing works
History of blood gas analysis.
Jw, S., & Astrup, P. (1987) · PubMed · Cited by 187 · see citing works
History of blood gas analysis. VII. Pulse oximetry
Severinghaus, J. W., & Honda, Y. (1987) · The Journal of Clinical Monitoring · doi:10.1007/bf00858362 · Cited by 180 · see citing works
Principles of pulse oximetry: theoretical and practical considerations.
Alexander, C. M., Teller, L. E., & Gross, J. B. (1989) · PubMed · Cited by 175 · see citing works
The Effect of Nail Polish on Pulse Oximetry
COT, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 171 · see citing works
Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09410.x · Cited by 166 · see citing works
Alexander, C. M., Teller, L. E., & Gross, J. B. (1989) · Anesthesia & Analgesia · doi:10.1213/00000539-198903000-00035 · Cited by 157 · see citing works
History of blood gas analysis. VI. Oximetry
Severinghaus, J. W., & Astrup, P. B. (1986) · The Journal of Clinical Monitoring · doi:10.1007/bf02851177 · Cited by 153 · see citing works
Pulse Oximeter Failure Thresholds in Hypotension and Vasoconstriction
Severinghaus, J. W., & Spellman, M. J. (1990) · Anesthesiology · doi:10.1097/00000542-199009000-00025 · Cited by 146 · see citing works
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
Noninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry Device
Mihm, F. G., & Halperin, B. D. (1985) · Anesthesiology · doi:10.1097/00000542-198501000-00020 · Cited by 142 · see citing works
Tyler, I. L., Tantisira, B., WINTER, P. M., et al. (1985) · Anesthesia & Analgesia · doi:10.1213/00000539-198511000-00013 · Cited by 138 · see citing works
Blood Flow Limits and Pulse Oximeter Signal Detection
Lawson, D., Norley, I., Korbon, G., Loeb, R. G., & Ellis, J. B. (1987) · Anesthesiology · doi:10.1097/00000542-198710000-00032 · Cited by 137 · see citing works
Continuous In-Vivo Monitoring of Arterial Oxygenation in Chronic Obstructive Lung Disease
Flick, M. R., & BLOCK, A. J. (1977) · Annals of Internal Medicine · doi:10.7326/0003-4819-86-6-725 · Cited by 132 · see citing works
Tyler, I. L., Tantisira, B., WINTER, P. M., et al. (1986) · Survey of Anesthesiology · doi:10.1097/00132586-198608000-00043 · Cited by 132 · see citing works
Effect of anemia on pulse oximeter accuracy at low saturation
Severinghaus, J. W., & Koh, S. O. (1990) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02828282 · Cited by 129 · see citing works
Bias and Precision of Pulse Oximeters and Arterial Oximeters
Nickerson, B. G., Sarkisian, C. A., & Tremper, K. K. (1988) · CHEST Journal · doi:10.1378/chest.93.3.515 · Cited by 124 · see citing works
The Pulse Oximeter: Applications and Limitations—An Analysis of 2000 Incident Reports
Runciman, W. B., Webb, R. K., Barker, L., et al. (1993) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9302100509 · Cited by 123 · see citing works
The current status of pulse oximetry
Taylor, M. B., & WHITWAM, J. G. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12922.x · Cited by 117 · see citing works
Use of Pulse Oximetry as a Noninvasive Indicator of Intravascular Volume Status
Partridge, B. L. (1987) · The Journal of Clinical Monitoring · doi:10.1007/bf03337381 · Cited by 115 · see citing works
Pulse Oximetry: Technical Aspects of Machine Design
Pologe, J. A. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00009 · Cited by 110 · see citing works
A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
Clayton, D. G., Webb, R. K., Ralston, A., et al. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09303.x · Cited by 109 · see citing works
Evaluation of the Ohmeda 3700 Pulse Oximeter
Kagle, D. M., Alexander, C. M., Berko, R. S., et al. (1987) · Anesthesiology · doi:10.1097/00000542-198703000-00018 · Cited by 109 · see citing works
Hemoglobin desaturation in highly trained athletes during heavy exercise
Williams, J. H., Powers, S. K., & Stuart, M. K. (1986) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-198604000-00004 · Cited by 107 · see citing works
Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985) · PubMed · doi:10.1164/arrd.1985.132.3.685 · Cited by 106 · see citing works
Pulse Oximeter Desaturation Due to Methemoglobinemia
Eisenkraft, J. B. (1988) · Anesthesiology · doi:10.1097/00000542-198802000-00018 · Cited by 106 · see citing works
Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation.
Saunders, N. A., Powles, A. C. P., & Rebuck, A. S. (1976) · PubMed · doi:10.1164/arrd.1976.113.6.745 · Cited by 105 · see citing works
Pulse Oximetry: Applications and Limitations
Barker, S. J., & Tremper, K. K. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00010 · Cited by 104 · see citing works
Mengelkoch, L. J., MARTIN, D. S., & Lawler, J. M. (1994) · Physical Therapy · doi:10.1093/ptj/74.1.40 · Cited by 97 · see citing works
Clayton, D. G., Webb, R. K., Ralston, A., DUTHIE, D., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11492.x · Cited by 90 · see citing works
Spurious Pulse Oximeter Desaturation with Methylene Blue Injection
Kessler, M., EIDE, T., HUMAYUN, B., et al. (1986) · Anesthesiology · doi:10.1097/00000542-198610000-00017 · Cited by 89 · see citing works
Kim, J. M., Arakawa, K., Benson, K. T., & Fox, D. K. (1986) · Anesthesia & Analgesia · doi:10.1213/00000539-198612000-00015 · Cited by 86 · see citing works
Efficacy of pulse oximetry and capnometry in postoperative ventilatory weaning
Niehoff, J., DelGuercio, C., LaMorte, W. W., HUGHES-GRASBERGER, S. L., Heard, S. O., Dennis, R. C., & Yeston, N. S. (1988) · Critical Care Medicine · doi:10.1097/00003246-198807000-00011 · Cited by 85 · see citing works
Validity of pulse oximetry during exercise in elite endurance athletes
Martin, D., Powers, S. K., Cicale, M. J., Collop, N. A., Huang, D. T.-C., & Criswell, D. S. (1992) · Journal of Applied Physiology · doi:10.1152/jappl.1992.72.2.455 · Cited by 84 · see citing works
Tremper, K. K. (1989) · CHEST Journal · doi:10.1378/chest.95.4.713 · Cited by 84 · see citing works
The Effect of Sensor Malpositioning on Pulse Oximeter Accuracy during Hypoxemia
Barker, S. J., Hyatt, J., Shah, N. K., et al. (1993) · Anesthesiology · doi:10.1097/00000542-199308000-00009 · Cited by 82 · see citing works
Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room
Dumas, C., Wahr, J. A., & Tremper, K. K. (1996) · Anesthesia & Analgesia · doi:10.1097/00000539-199608000-00012 · Cited by 81 · see citing works
Ear oximetry during combined hypoxia and exercise
Smyth, R. J., D'Urzo, A. D., Slutsky, A. S., Galko, B. M., & Rebuck, A. S. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.2.716 · Cited by 80 · see citing works
Postoperative monitoring of conjunctival oxygen tension and temperature
Fink, S., Abraham, E., & Ehrlich, H. (1988) · International journal of clinical monitoring and computing · doi:10.1007/bf01739231 · Cited by 77 · see citing works
The accuracy of pulse oximeters
Taylor, M. B., & WHITWAM, J. G. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb05549.x · Cited by 77 · see citing works
Shimada, Y., Yoshiya, I., Oka, N., et al. (1984) · Medical & Biological Engineering & Computing · doi:10.1007/bf02447712 · Cited by 76 · see citing works
Reliability of Pulse Oximetry During Exercise in Pulmonary Patients
Escourrou, P., Delaperche, M. F., & Visseaux, A. (1990) · CHEST Journal · doi:10.1378/chest.97.3.635 · Cited by 74 · see citing works
Validity of Ear Oximetry in Clinical Exercise Testing
Hansen, J. E., & Casaburi, R. H. (1987) · CHEST Journal · doi:10.1378/chest.91.3.333 · Cited by 74 · see citing works
Evaluation of the Ohmeda 3700 pulse oximeter.
Warley, A., Mitchell, J. H., & Stradling, J. R. (1987) · Thorax · doi:10.1136/thx.42.11.892 · Cited by 74 · see citing works
Douglas, N. J., Brash, H. M., WRAITH, P. K., et al. (1979) · PubMed · doi:10.1164/arrd.1979.119.2.311 · Cited by 69 · see citing works
Accuracy and precision of fourteen pulse oximeters
Hannhart, B., Haberer, J. P., Saunier, C. F., et al. (1991) · European Respiratory Journal · doi:10.1183/09031936.93.04010115 · Cited by 69 · see citing works
Hyperbilirubinemia Does Not Interfere with Hemoglobin Saturation Measured by Pulse Oximetry
Veyckemans, F., Baele, P., Guillaume, J., et al. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00022 · Cited by 64 · see citing works
Design and evaluation of a new reflectance pulse oximeter sensor.
Mendelson, Y., Kent, J. C., Yocum, B. L., et al. (1988) · PubMed · Cited by 63 · see citing works
Ambient Light Affects Pulse Oximeters
Hanowell, L. H., Eisele, J. H., & DOWNS, D. (1987) · Anesthesiology · doi:10.1097/00000542-198711000-00057 · Cited by 61 · see citing works
Range of Accuracy of Two Wavelength Oximetry
Chapman, K. R., LIU, F. L. W., Watson, R. M., et al. (1986) · CHEST Journal · doi:10.1378/chest.89.4.540 · Cited by 61 · see citing works
Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room
Dumas, C., Wahr, J. A., & Tremper, K. K. (1996) · Anesthesia & Analgesia · doi:10.1213/00000539-199608000-00012 · Cited by 61 · see citing works
CARBOXYHAEMOGLOBINAEMIA AND PULSE OXIMETRY
Vegfors, M., & Lennmarken, C. (1991) · British Journal of Anaesthesia · doi:10.1093/bja/66.5.625 · Cited by 60 · see citing works
GABRIELCZYK, M. R., & BUIST, R. J. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb09025.x · Cited by 59 · see citing works
Pulse Oximetry during One-lung Ventilation
BRODSKY, J. B., SHULMAN, M. S., SWAN, M., et al. (1985) · Anesthesiology · doi:10.1097/00000542-198508000-00021 · Cited by 58 · see citing works
Evaluation of Bedside Testing Options for the Critical Care Unit
Zaloga, G. P. (1990) · CHEST Journal · doi:10.1378/chest.97.5_supplement.185s · Cited by 58 · see citing works
Morris, R., Nairn, M., & Torda, T. A. (1989) · Anaesthesia and Intensive Care · doi:10.1177/0310057x8901700113 · Cited by 57 · see citing works
Infrared Heat Lamps Interfere with Pulse Oximeters
Brooks, T. D., Paulus, D. A., & Winkle, W. E. (1984) · Anesthesiology · doi:10.1097/00000542-198411000-00042 · Cited by 57 · see citing works
West, P., George, C. F., & Kryger, M. H. (1987) · SLEEP · doi:10.1093/sleep/10.3.263 · Cited by 56 · see citing works
Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation
STEWART, K. G., & Rowbottom, S. J. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09720.x · Cited by 55 · see citing works
McGovern, J., Sasse, S. A., Stansbury, D. W., Causing, L. A., & Light, R. W. (1996) · CHEST Journal · doi:10.1378/chest.109.5.1151 · Cited by 55 · see citing works
COMPARISON OF PULSE OXIMETERS: EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENT
Wilkins, C. J., MOORES, M., & Hanning, C. D. (1989) · British Journal of Anaesthesia · doi:10.1093/bja/62.4.439 · Cited by 55 · see citing works
Pulse oximetry versus measurement of transcutaneous oxygen
New, W. (1985) · The Journal of Clinical Monitoring · doi:10.1007/bf02832200 · Cited by 54 · see citing works
Accuracy of pulse oximetry to estimate HbO2 fraction of total Hb during exercise
Powers, S. K., DODD, S. L., Freeman, J., et al. (1989) · Journal of Applied Physiology · doi:10.1152/jappl.1989.67.1.300 · Cited by 53 · see citing works
History and recent developments in pulse oximetry.
Severinghaus, J. W. (1993) · PubMed · Cited by 53 · see citing works
Ear oximetry in clinical practice.
Chaudhary, B. A., & Burki, N. K. (1978) · PubMed · doi:10.1164/arrd.1978.117.1.173 · Cited by 52 · see citing works
Evaluation of a pulse oximeter in the prehospital setting
MCGUIRE, T. J., & Pointer, J. E. (1988) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(88)80444-x · Cited by 52 · see citing works
A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters
Cecil, W. T., Thorpe, K. J., Fibuch, E. E., & Tuohy, G. F. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01618105 · Cited by 51 · see citing works
Accuracy of pulse oximetry in patients with low systemic vascular resistance
Secker, C., & Spiers, P. (1997) · Anaesthesia · doi:10.1111/j.1365-2044.1997.32-az0062.x · Cited by 51 · see citing works
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50 · see citing works
Limitations of pulse oximetry.
Mardirossian, G., & Schneider, R. E. (1992) · PubMed · Cited by 50 · see citing works
The prevalence of hypoxemia detected by pulse oximetry during recovery from anesthesia
Morris, R. W., Buschman, A., Warren, D. L., et al. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01618102 · Cited by 50 · see citing works
Pulse oximetry in sickle cell disease.
Pianosi, P. T., Charge, T. D., Esseltine, D. W., & Coates, A. L. (1993) · Archives of Disease in Childhood · doi:10.1136/adc.68.6.735 · Cited by 50 · see citing works
Accuracy of Pulse Oximetry During Hypoxemia
Thrush, D., & HODGES, M. R. (1994) · Southern Medical Journal · doi:10.1097/00007611-199404000-00019 · Cited by 49 · see citing works
Pulse oximetry ‐ clinical implications and recent technical developments
Lindberg, L., Lennmarken, C., & Vegfors, M. (1995) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1995.tb04062.x · Cited by 48 · see citing works
Agreement between Noninvasive Oximetric Values for Oxygen Saturation
Cahan, C., Decker, M. J., Hoekje, P. L., & Strohl, K. P. (1990) · CHEST Journal · doi:10.1378/chest.97.4.814 · Cited by 47 · see citing works
Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery
PALVE, H., & Vuori, A. (1989) · Critical Care Medicine · doi:10.1097/00003246-198901000-00014 · Cited by 47 · see citing works
Inman, K. J., Sibbald, W. J., Rutledge, F. S., et al. (1993) · CHEST Journal · doi:10.1378/chest.104.2.542 · Cited by 44 · see citing works
Pulse oximetry: Gold standard or false sense of security?
Grace, R. F. (1994) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1994.tb125873.x · Cited by 43 · see citing works
Jones, J. S., Heiselman, D., Cannon, L., et al. (1988) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(88)80237-3 · Cited by 42 · see citing works
Reliability of Six Pulse Oximeters in Chronic Obstructive Pulmonary Disease
Hannhart, B., Michalski, H., Delorme, N., Chapparo, G., & Polu, J. M. (1991) · CHEST Journal · doi:10.1378/chest.99.4.842 · Cited by 39 · see citing works
Quantification of oxygen saturation during episodic hypoxemia.
Slutsky, A. S., & Strohl, K. P. (1980) · PubMed · doi:10.1164/arrd.1980.121.5.893 · Cited by 38 · see citing works
Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter
Mackenzie, N. (1985) · The Journal of Clinical Monitoring · doi:10.1007/bf02832255 · Cited by 37 · see citing works
ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULT
Tremper, K. K., Hufstedler, S. M., Barker, S. J., Adams, A. L., Wong, D. H. W., Zaccari, J., Benik, K., & Lemons, V. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00175 · Cited by 37 · see citing works
DiBenedetto, R. J., Graves, S. A., Gravenstein, N. L., et al. (1994) · Anesthesia & Analgesia · doi:10.1213/00000539-199402000-00028 · Cited by 37 · see citing works
Evaluation of pulse oximetry during cardiopulmonary resuscitation
SPITTAL, M. J. (1993) · Anaesthesia · doi:10.1111/j.1365-2044.1993.tb07185.x · Cited by 37 · see citing works
The accuracy of the biox 3700 pulse oximeter in patients receiving vasoactive therapy
Ibáñez, J., Velasco, J., & Raurich, J. M. (1991) · Intensive Care Medicine · doi:10.1007/bf01690773 · Cited by 37 · see citing works
In vitro investigation of the factors affecting pulse oximetry
de Kock, J. P., & Tarassenko, L. (1991) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(91)90046-a · Cited by 36 · see citing works
Assessment of two noninvasive monitors of arterial oxygenation in anesthetized man.
Rl, K., Jl, C., Kieraszewicz, H., et al. (1982) · PubMed · Cited by 36 · see citing works
Utility of the Pulse Oximeter during Cardiopulmonary Resuscitation
Narang, V. (1986) · Anesthesiology · doi:10.1097/00000542-198608000-00039 · Cited by 36 · see citing works
Reflection and transmission pulse oximetry during compromised peripheral perfusion
Pälve, H. (1992) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01618081 · Cited by 34 · see citing works
Accuracy of pulse oximetry during moderate exercise: a comparative study
Barthélemy, J., GEYSSANT, A., Riffat, J., et al. (1990) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519009089168 · Cited by 34 · see citing works
Ambulatory Monitoring of Arterial Oxygen Saturation
Decker, M. J., Hoekje, P. L., & Strohl, K. P. (1989) · CHEST Journal · doi:10.1378/chest.95.4.717 · Cited by 34 · see citing works
Evaluation of Biox IIA ear oximeter.
Tweeddale, P. M., & Douglas, N. J. (1985) · Thorax · doi:10.1136/thx.40.11.825 · Cited by 33 · see citing works
An evaluation of pulse oximetry in prehospital care
Aughey, K., HESS, D. R., Eitel, D. R., Bleecher, K., Cooley, M., Ogden, C. S., & Sabulsky, N. K. (1991) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(05)81432-5 · Cited by 33 · see citing works
ACCURACY OF A NEW NON-INVASIVE OXYGEN SATURATION MONITOR
Sarnquist, F. H., Todd, C., & Whitcher, C. E. (1980) · Anesthesiology · doi:10.1097/00000542-198009001-00163 · Cited by 33 · see citing works
History and recent developments in pulse oximetry
Severinghaus, J. W. (1993) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519309090686 · Cited by 33 · see citing works
Accuracy of Pulse Oximetry during Exercise Stress Testing
Norton, L., Squires, B., Craig, N., et al. (1992) · International Journal of Sports Medicine · doi:10.1055/s-2007-1021310 · Cited by 32 · see citing works
The Nellcor N‐101 pulse oximeter
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32 · see citing works
Evaluation of the influence of movement on saturation readings from pulse oximeters
Plummer, J. L., ZAKARIA, A. Z., Ilsley, A. H., et al. (1995) · Anaesthesia · doi:10.1111/j.1365-2044.1995.tb05998.x · Cited by 32 · see citing works
Møller, J. T. (1994) · PubMed · Cited by 31 · see citing works
Weaning from ventilation after cardiopulmonary bypass: evaluation of a non-invasive technique
Withington, D. E., Ramsay, J. G., Saoud, A. T., et al. (1991) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03009157 · Cited by 31 · see citing works
Pulse oximetry in sickle cell disease
Smith, S. G. W., Glass, U. H., ACHARYA, J., et al. (1989) · Clinical & Laboratory Haematology · doi:10.1111/j.1365-2257.1989.tb00207.x · Cited by 30 · see citing works
The influence of changes in blood flow on the accuracy of pulse oximetry in humans
Vegfors, M., Lindberg, L., & Lennmarken, C. (1992) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1992.tb03479.x · Cited by 30 · see citing works
Pulse oximetry for hypoxemia: a warning to users and manufacturers
Fanconi, S. (1989) · Intensive Care Medicine · doi:10.1007/bf00273571 · Cited by 30 · see citing works
Roizen, M. F., Schreider, B. D., Austin, W., et al. (1993) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02886693 · Cited by 30 · see citing works
Clinical evaluation of a new ear oximeter.
Scoggin, C., Nett, L. M., & Tl, P. (1977) · PubMed · Cited by 30 · see citing works
Ries, A. L. (1987) · CHEST Journal · doi:10.1378/chest.91.3.316 · Cited by 28 · see citing works
Pulse oximetry and capnography in anaesthetic practice: an epidemiological appraisal
Duncan, P. G., & Cohen, M. M. (1991) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03008199 · Cited by 27 · see citing works
Comparison of recorded values from six pulse oximeters
Choe, H., Tashiro, C., Fukumitsu, K., et al. (1989) · Critical Care Medicine · doi:10.1097/00003246-198907000-00016 · Cited by 27 · see citing works
Respiratory Monitoring During Mechanical Ventilation
Tobin, M. J. (1990) · Critical Care Clinics · doi:10.1016/s0749-0704(18)30361-0 · Cited by 26 · see citing works
History of blood gas analysis. V. Oxygen measurement
Severinghaus, J. W., & Astrup, P. B. (1986) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01620550 · Cited by 26 · see citing works
Assessment of Two Noninvasive Monitors of Arterial Oxygenation in Anesthetized Man
Knill, R. L., Clement, J. L., Kieraszewicz, H., et al. (1982) · Anesthesia & Analgesia · doi:10.1213/00000539-198207000-00006 · Cited by 26 · see citing works
A model for technology assessment applied to pulse oximetry
(1993) · Critical Care Medicine · doi:10.1097/00003246-199304000-00024 · Cited by 25 · see citing works
Pulse oximetry versus arterial blood gas specimens in long-term oxygen therapy
Pierson, D. J. (1990) · Lung · doi:10.1007/bf02718208 · Cited by 24 · see citing works
The Oxygen Analyser: Applications and Limitations – An Analysis of 2000 Incident Reports
Barker, L., Webb, R. K., Runciman, W. B., et al. (1993) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9302100513 · Cited by 24 · see citing works
An Unexpected Finding with Pulse Oximetry in a Patient with Hemoglobin Köln
Gottschalk, A., & Silverberg, M. B. (1994) · Anesthesiology · doi:10.1097/00000542-199402000-00030 · Cited by 23 · see citing works
Validation of pulse oximetry for monitoring of hypoxaemic episodes in the late postoperative period
Stausholm, K., Rosenberg-Adamsen, S., Edvardsen, L., et al. (1997) · British Journal of Anaesthesia · doi:10.1093/bja/78.1.86 · Cited by 22 · see citing works
Brown, M., & Vender, J. S. (1988) · Critical Care Clinics · doi:10.1016/s0749-0704(18)30478-0 · Cited by 21 · see citing works
History, Status and Future of Pulse Oximetry
Severinghaus, J. W. (1987) · Springer eBooks · doi:10.1007/978-1-4613-1927-6_1 · Cited by 21 · see citing works
Clinical evaluation of a new finger oximeter
Kim, S. K., BAIDWAN, B. S., & Petty, T. L. (1984) · Critical Care Medicine · doi:10.1097/00003246-198410000-00016 · Cited by 21 · see citing works
Accuracy of three pulse oximeters at low cardiac index and peripheral temperature
Pälve, H., & Vuori, A. (1991) · Critical Care Medicine · doi:10.1097/00003246-199104000-00018 · Cited by 20 · see citing works
Pälve, H., & Vuori, A. (1991) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/1053-0770(91)90154-l · Cited by 19 · see citing works
Use of pulse oximetry in critically ill adults.
Szaflarski, N. L., & Nh, C. (1989) · PubMed · Cited by 18 · see citing works
Misiano, D. R., Meyerhoff, M. E., & Collison, M. E. (1990) · CHEST Journal · doi:10.1378/chest.97.5_supplement.204s · Cited by 16 · see citing works
Pulse Oximetry Correction for Smoking Exposure
Glass, K. L., Dillard, T. A., Phillips, Y. Y., et al. (1996) · Military Medicine · doi:10.1093/milmed/161.5.273 · Cited by 15 · see citing works
Evaluation of the Datascope ACCUSAT pulse oximeter in healthy adults
Mendelson, Y., Kent, J. C., SHAHNARIAN, A., et al. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01618108 · Cited by 15 · see citing works
Pulse Oximetry and Patient Positioning
Hovagim, A. R., Backus, W. W., Manecke, G. R., et al. (1989) · Anesthesiology · doi:10.1097/00000542-198909000-00026 · Cited by 15 · see citing works
A clinical evaluation of pulse oximetry during thoracic surgery
Desiderio, D. P., Wong, G. Y., Shah, N. K., Liu, J., Loughlin, C. J., & Bedford, R. F. (1990) · Journal of Cardiothoracic Anesthesia · doi:10.1016/0888-6296(90)90443-j · Cited by 15 · see citing works
Noll, M. L., & Byers, J. F. (1995) · Heart & Lung · doi:10.1016/s0147-9563(05)80040-8 · Cited by 15 · see citing works
Gaskin, L., & Thomas, J. (1995) · Physiotherapy · doi:10.1016/s0031-9406(05)66819-8 · Cited by 14 · see citing works
Pulse oximetry in the emergency department
Galdun, J. P., Paris, P. M., & Stewart, R. D. (1989) · The American Journal of Emergency Medicine · doi:10.1016/0735-6757(89)90054-5 · Cited by 14 · see citing works
ACCURACY OF PULSE OIIMETRY IN CYANOTIC CONGENITAL HEART DISEASE
Lazzell, V. A., & Jopling, M. W. (1987) · Anesthesiology · doi:10.1097/00000542-198709001-00169 · Cited by 14 · see citing works
VENOUS CONGESTION AFFECTS ARTERIAL HEMOGLOBIN SATURATION MEASURED BY THE PULSE OXIMETER
Kim, J.-M., & Mathewson, H. S. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00174 · Cited by 14 · see citing works
Pulse Meters and Pulse Oximeters
Griffiths, D. M., Ilsley, A. H., & Runciman, W. B. (1988) · Anaesthesia and Intensive Care · doi:10.1177/0310057x8801600118 · Cited by 14 · see citing works
Poppius, H., & Viljanen, A. A. (1977) · PubMed · Cited by 13 · see citing works
Broome, I. J., MILLS, G. H., SPIERS, P., et al. (1993) · Anaesthesia · doi:10.1111/j.1365-2044.1993.tb07017.x · Cited by 13 · see citing works
Continuous in vivo measurement of arterial oxygen saturation by oximetry.
Flick, M. R., & Block, A. J. (1978) · PubMed · Cited by 13 · see citing works
Emerging technology in critical care: continuous intra-arterial blood gas monitoring
Szaflarski, N. L. (1996) · American Journal of Critical Care · doi:10.4037/ajcc1996.5.1.55 · Cited by 12 · see citing works
An Evaluation of Prolonged Oximetric Data Acquisition
Jones, R. D. M., Lawson, A., Gunawardene, W. M. S., et al. (1992) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9202000304 · Cited by 12 · see citing works
Derived Oxygen Saturations are not Clinically Useful for the Calculation of Oxygen Consumption
Myburgh, J. (1992) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9202000411 · Cited by 12 · see citing works
A Comparison between Direct and Calculated Oxygen Saturation in Intensive Care
Johnson, P. A., Bihari, S., Raper, R. F., Haughton, M. A., Fisher, M. M., & HERKES, R. G. (1993) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9302100117 · Cited by 12 · see citing works
Accuracy of Pulse Oximetry in Patients with Hyperbilirubinemia
Chelluri, L., Snyder, J. V., & Bird, J. R. (1991) · Respiratory Care · doi:10.1177/194336549103601204 · Cited by 12 · see citing works
Respiratory Artifact during Pulse Oximetry in Critically III Patients
Scheller, J., & Loeb, R. G. (1988) · Anesthesiology · doi:10.1097/00000542-198810000-00025 · Cited by 11 · see citing works
Principles and Application of Oximetry
Rutherford, K. A. (1989) · Critical Care Nursing Clinics of North America · doi:10.1016/s0899-5885(18)30854-2 · Cited by 11 · see citing works
Changing perspectives in the assessment of oxygenation
Ahrens, T. (1993) · Critical Care Nurse · doi:10.4037/ccn1993.13.4.78 · Cited by 10 · see citing works
Accuracy and Response Time of a Portable Pulse Oximeter
Modica, R., & Rizzo, A. (2009) · Respiration · doi:10.1159/000195917 · Cited by 10 · see citing works
Blood Flow Limits and Signal Detection Comparing Five Different Models of Pulse Oximeters
Greenblott, G., Gerschultz, S., & Tremper, K. K. (1989) · Anesthesiology · doi:10.1097/00000542-198902000-00037 · Cited by 10 · see citing works
Comparison of three transmittance oximeters.
Strohl, K., House, P. M., Holic, J. F., et al. (1986) · PubMed · Cited by 10 · see citing works
Evaluation of pulse oximetry in anemia from hemoglobin-H disease
Jay, G. D., & Renzi, F. P. (1992) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(05)82528-4 · Cited by 9 · see citing works
Effects of carboxyhemoglobin on pulse oximetry in humans
Tashiro, C., Koo, Y. H., Fukumitsu, K., et al. (1988) · Journal of Anesthesia · doi:10.1007/s0054080020036 · Cited by 9 · see citing works
Intravenous Conscious Sedation Use in Endoscopy
HINZMANN, C. A., BUDDEN, P. M., & OLSON, J. (1992) · Gastroenterology Nursing · doi:10.1097/00001610-199208000-00003 · Cited by 9 · see citing works
Case studies in pulse oximetry monitoring
Grossbach, I. (1993) · Critical Care Nurse · doi:10.4037/ccn1993.13.4.63 · Cited by 9 · see citing works
Misiano, D. R., Meyerhoff, M. E., & Collison, M. E. (1990) · CHEST Journal · doi:10.1378/chest.97.5.204s · Cited by 9 · see citing works
Pulse oximeter signal at various blood flow conditions in anin vitro model
Lindberg, L., Vegfors, M., Lennmarken, C., et al. (1995) · Medical & Biological Engineering & Computing · doi:10.1007/bf02522952 · Cited by 8 · see citing works
SWEENEY, C. (1991) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(05)81444-1 · Cited by 8 · see citing works
Viitanen, A., Salmenperä, M., Heinonen, J., et al. (1987) · The Journal of Clinical Monitoring · doi:10.1007/bf00858356 · Cited by 8 · see citing works
Noninvasive Respiratory Monitoring During Ventilatory Support
HESS, D. R. (1991) · Critical Care Nursing Clinics of North America · doi:10.1016/s0899-5885(18)30683-x · Cited by 8 · see citing works
Noninvasive monitoring of oxygenation with pulse oximetry.
Gilboy, N. S., & McGaffigan, P. (1989) · PubMed · Cited by 7 · see citing works
Increasing Clinical Use of Pulse Oximetry
Peters, K., Caulfield, A., Schultz, P., et al. (1990) · Dimensions of Critical Care Nursing · doi:10.1097/00003465-199003000-00012 · Cited by 6 · see citing works
Noninvasive monitoring of the mechanically ventilated patient
McCauley, M. D., & Rueden, K. T. V. (1988) · Critical Care Nursing Quarterly · doi:10.1097/00002727-198812000-00006 · Cited by 6 · see citing works
Failure rate of pulse oximetry in the postanesthesia care unit
Gillies, B. S., Posner, K. L., Freund, P. R., & Cheney, F. W. (1993) · The Journal of Clinical Monitoring · doi:10.1007/bf01618673 · Cited by 5 · see citing works
Noninvasive Assessment of Gas Exchange in the Critically Ill Patient
Rueden, K. T. V. (1990) · AACN Advanced Critical Care · doi:10.4037/15597768-1990-2002 · Cited by 5 · see citing works
Accuracy of pulse oximetry in aortomyoplasty and balloon counterpulsation.
Visram, A., & Marshall, C. (1997) · PubMed · Cited by 5 · see citing works
Pulse oximetry during open heart surgery.
Kurki, T. S. O., Smith, N. T., Sanford, T. J., et al. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198802001-00123 · Cited by 4 · see citing works
Getting the most from pulse oximetry.
Durren, M. (1992) · PubMed · Cited by 4 · see citing works
Monitoring the respiratory system of the mechanically ventilated patient.
Jd, T., & Df, R. (1994) · PubMed · Cited by 4 · see citing works
Pulse oximetry: pearls and pitfalls.
Harrahill, M. A. (1991) · PubMed · Cited by 4 · see citing works
Evaluation of ear oximetry in the measurement of arterial oxygen saturation
Cable, J. A. (1981) · The Journal of the American Osteopathic Association · doi:10.1515/jom-1981-811015 · Cited by 4 · see citing works
NON- INVASIVE PULSE-OXIMETRY DURING ONE-LUNG VENTILATION
Shulman, M. S., Brodsky, J. B., Mark, J. B. D., et al. (1984) · Anesthesiology · doi:10.1097/00000542-198409001-00098 · Cited by 3 · see citing works
Experimental and predicted dual oximetry variability
Wong, D. H. W., Weir, P. D., Mahutte, C. K., ONISHI, R., Daves, S., Foran, W., Ferraro, L., & Barker, S. J. (1993) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02886697 · Cited by 3 · see citing works
VALIDITY OF PULSE OXIMETRY DURING MAXIMAL TREADMILL EXERCISE
Mengelkoch, L. J., MARTIN, D. S., Cicale, M. J., et al. (1992) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199205001-00954 · Cited by 3 · see citing works
Clinical use of non-invasive method for determining oxygen saturation: ear oximetry.
Ishikawa, S., Linzmayer, I., & Segal, M. S. (1978) · PubMed · Cited by 3 · see citing works
323 VALIDITY OP PULSE EMI OXIMETRY FOR DETERMINING OXYGEN HEMOGLOBIN SATURATION DURING EXERCISE
Brown, D. D., KNOWLTON, R. G., Szurgot, B., et al. (1990) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199004000-00323 · Cited by 2 · see citing works
Evaluation of ear oximetry in the measurement of arterial oxygen saturation.
Cable, J. (1981) · PubMed · Cited by 2 · see citing works
Innovations in technology: continuous intra-arterial blood gas monitoring
Dirks, J. (1995) · Critical Care Nurse · doi:10.4037/ccn1995.15.2.19 · Cited by 2 · see citing works
Lavies, N. G. (1997) · Anaesthesia · doi:10.1111/j.1365-2044.1997.00036.x · Cited by 2 · see citing works
The current practice of pulse oximetry and capnometry/capnography in the prehospital setting
Kissinger, D. P., Hamilton, I., & Rozycki, G. S. (1991) · Cited by 2 · see citing works
Sao2 monitoring in the postanesthesia care unit.
Berels, D. J., & Marz, M. S. (1991) · PubMed · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
ATS Statement: Guidelines for the Six-Minute Walk Test
(2002) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.166.1.at1102 · Cited by 11,195 · see citing works
Uronis, H. E., Currow, D. C., McCrory, D. C., Samsa, G. P., & Abernethy, A. P. (2008) · British Journal of Cancer · doi:10.1038/sj.bjc.6604161 · Cited by 165 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Psychological testing and psychological assessment: A review of evidence and issues.
Meyer, G. J., Finn, S. E., Eyde, L. D., Kay, G. G., Moreland, K. L., Dies, R. R., Eisman, E. J., Kubiszyn, T. W., & Reed, G. M. (2001) · American Psychologist · doi:10.1037//0003-066x.56.2.128 · Cited by 1,469 · see citing works
Wilson, M. J., Lopez, M., Vargas, M., Julian, C., Tellez, W., Rodriguez, A., Bigham, A., Armaza, J. F., Niermeyer, S., Shriver, M., et al. (2007) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00806.2006 · Cited by 101 · see citing works
Impact of skin tone on photoacoustic oximetry and tools to minimize bias
Mantri, Y., & Jokerst, J. V. (2022) · Biomedical Optics Express · doi:10.1364/boe.450224 · Cited by 75 · see citing works
- Authors
- Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N.
- Affiliations
- University of Alberta (CA)
- ORCID
- Jensen, Louise A.
- Venue
- Heart & Lung
- Publisher
- Elsevier BV
- Published
- 1998-11
- Volume
- 27
- Issue
- 6
- Pages
- 387-408
- ISSN
- 0147-9563 (ISSN-L)
- OpenAlex
- W2112793666
- Cited by
- 243
- References
- 198
- Citations by year
- 2026: 6 · 2025: 10 · 2024: 8 · 2023: 8 · 2022: 11 · 2021: 12 · 2020: 9 · 2019: 12 · 2018: 9 · 2017: 9 · 2016: 6 · 2015: 13 · 2014: 7 · 2013: 9 · 2012: 12
- FWCI
- 1.34
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Heart & Lung (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Keywords
- pulse oximetry, arterial oxygen saturation, co-oximetry, arterial blood gases, measurement accuracy, bias and precision, finger probe, desaturation, hypotension, hypothermia, low perfusion, adults
- MeSH terms
- Adult, Clinical Trials as Topic, Female, Humans, Male, Monitoring, Physiologic, Oximetry, Oxygen, Oxygen Consumption, Reference Values, Sensitivity and Specificity, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review