The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
Abstract
Background: Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy and data dropout rates of three current pulse oximeters during standardized motion in healthy volunteers. Methods: Ten healthy volunteers were monitored by three different pulse oximeters: Nellcor N-200, Nellcor N-3000, and Masimo SET (prototype). Sensors were placed on digits 2, 3, and 4 of the test hand, which was strapped to a mechanical motion table. The opposite hand was used as a stationary control and was monitored with the same pulse oximeters and an arterial cannula. Arterial oxygen saturation was varied from 100% to 75% by changing the inspired oxygen concentration. While SpO2 was both constant and changing, the oximeter sensors were connected before and during motion. Oximeter errors and dropout rates were digitally recorded continuously during each experiment. Results: If the oximeter was functioning before motion began, the following are the percentages of time when the instrument displayed an SpO2 value within 7% of control: N-200 = 76%, N-3000 = 87%, and Masimo = 99%. When the oximeter sensor was connected after the beginning of motion, the values were N-200 = 68%, N-3000 = 47%, and Masimo = 97%. If the alarm threshold was chosen SpO2 less than 90%, then the positive predictive values (true alarms/total alarms) are N-200 = 73%, N-3000 = 81%, and Masimo = 100%. In general, N-200 had the greatest SpO2 errors and N-3000 had the highest dropout rates. Conclusions: The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Lawless (1994) · doi:10.1097/00003246-199406000-00017 · Cited by 406
Cited by
Coté, C. J., Wilson, S., AMERICAN ACADEMY OF PEDIATRICS, & AMERICAN ACADEMY OF PEDIATRIC DENTISTRY (2019) · Pediatrics · doi:10.1542/peds.2019-1000 · Cited by 1,406
American Academy of Pediatrics, American Academy of Pediatric Dentistry, Coté, C. J., Wilson, S., & the Work Group on Sedation (2006) · Pediatrics · doi:10.1542/peds.2006-2780 · Cited by 555
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 491
Jubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 401
Lu, G., Yang, F., Taylor, J. A., & Stein, J. F. (2009) · Journal of Medical Engineering & Technology · doi:10.3109/03091900903150998 · Cited by 324
Metadata
- Authors
- Barker, S. J., & Shah, N. K.
- Venue
- Anesthesiology
- Published
- 1997-01-01
- Volume
- 86
- Issue
- 1
- Pages
- 101-108
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2131516330
- Cited by
- 195
- References
- 1
- FWCI
- 12.73
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication). https://app.cochat.ai/papers/10.1097/00000542-199701000-00014
- The paper
- Barker, S. J., & Shah, N. K. (1997). The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication). Anesthesiology. https://doi.org/10.1097/00000542-199701000-00014
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.