The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)

Barker, Steven J., Shah, Nitin K.·1997·Anesthesiology

Abstract

Background: Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy and data dropout rates of three current pulse oximeters during standardized motion in healthy volunteers. Methods: Ten healthy volunteers were monitored by three different pulse oximeters: Nellcor N-200, Nellcor N-3000, and Masimo SET (prototype). Sensors were placed on digits 2, 3, and 4 of the test hand, which was strapped to a mechanical motion table. The opposite hand was used as a stationary control and was monitored with the same pulse oximeters and an arterial cannula. Arterial oxygen saturation was varied from 100% to 75% by changing the inspired oxygen concentration. While SpO2 was both constant and changing, the oximeter sensors were connected before and during motion. Oximeter errors and dropout rates were digitally recorded continuously during each experiment. Results: If the oximeter was functioning before motion began, the following are the percentages of time when the instrument displayed an SpO2 value within 7% of control: N-200 = 76%, N-3000 = 87%, and Masimo = 99%. When the oximeter sensor was connected after the beginning of motion, the values were N-200 = 68%, N-3000 = 47%, and Masimo = 97%. If the alarm threshold was chosen SpO2 less than 90%, then the positive predictive values (true alarms/total alarms) are N-200 = 73%, N-3000 = 81%, and Masimo = 100%. In general, N-200 had the greatest SpO2 errors and N-3000 had the highest dropout rates. Conclusions: The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00000542-199701000-00014

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 195

Metadata

Authors
Barker, S. J., & Shah, N. K.
Venue
Anesthesiology
Published
1997-01-01
Volume
86
Issue
1
Pages
101-108
ISSN
0003-3022 (ISSN-L)
OpenAlex
W2131516330
Cited by
195
References
1
FWCI
12.73
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication). https://app.cochat.ai/papers/10.1097/00000542-199701000-00014
The paper
Barker, S. J., & Shah, N. K. (1997). The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication). Anesthesiology. https://doi.org/10.1097/00000542-199701000-00014

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.