The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
Background: Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy and data dropout rates of three current pulse oximeters during standardized motion in healthy volunteers. Methods: Ten healthy volunteers were monitored by three different pulse oximeters: Nellcor N-200, Nellcor N-3000, and Masimo SET (prototype). Sensors were placed on digits 2, 3, and 4 of the test hand, which was strapped to a mechanical motion table. The opposite hand was used as a stationary control and was monitored with the same pulse oximeters and an arterial cannula. Arterial oxygen saturation was varied from 100% to 75% by changing the inspired oxygen concentration. While SpO2 was both constant and changing, the oximeter sensors were connected before and during motion. Oximeter errors and dropout rates were digitally recorded continuously during each experiment. Results: If the oximeter was functioning before motion began, the following are the percentages of time when the instrument displayed an SpO2 value within 7% of control: N-200 = 76%, N-3000 = 87%, and Masimo = 99%. When the oximeter sensor was connected after the beginning of motion, the values were N-200 = 68%, N-3000 = 47%, and Masimo = 97%. If the alarm threshold was chosen SpO2 less than 90%, then the positive predictive values (true alarms/total alarms) are N-200 = 73%, N-3000 = 81%, and Masimo = 100%. In general, N-200 had the greatest SpO2 errors and N-3000 had the highest dropout rates. Conclusions: The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Moller, J. T., Johannessen, N. W., Espersen, K., Ravlo, O., Pedersen, B. D., Jensen, P. F., Rasmussen, N. H., Rasmussen, L. S., Pedersen, T., Cooper, J. B., et al. (1993) · Anesthesiology · doi:10.1097/00000542-199303000-00007 · Cited by 435
Lawless (1994) · doi:10.1097/00003246-199406000-00017 · Cited by 406
Effects of Methemoglobinemia on Pulse Oximetry and Mixed Venous Oximetry
Barker, S. J., Tremper, K. K., & Hyatt, J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00021 · Cited by 382
Self-citationRecent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 333
Statistics and ethics in medical research
Altman, D. (1981) · BMJ · doi:10.1136/bmj.282.6268.990-a · Cited by 151
EFFECT OF MOTION ARTEFACT ON PULSE OXIMETERS: EVALUATION OF FOUR INSTRUMENTS AND FINGER PROBES
LANGTON, J. A., & Hanning, C. D. (1990) · British Journal of Anaesthesia · doi:10.1093/bja/65.4.564 · Cited by 58
Cited by
Coté, C. J., Wilson, S., AMERICAN ACADEMY OF PEDIATRICS, & AMERICAN ACADEMY OF PEDIATRIC DENTISTRY (2019) · Pediatrics · doi:10.1542/peds.2019-1000 · Cited by 1,406
American Academy of Pediatrics, American Academy of Pediatric Dentistry, Coté, C. J., Wilson, S., & the Work Group on Sedation (2006) · Pediatrics · doi:10.1542/peds.2006-2780 · Cited by 555
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 491
Jubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 401
Lu, G., Yang, F., Taylor, J. A., & Stein, J. F. (2009) · Journal of Medical Engineering & Technology · doi:10.3109/03091900903150998 · Cited by 324
Update on Bispectral Index monitoring
Johansen, J. W. (2006) · Best Practice & Research Clinical Anaesthesiology · doi:10.1016/j.bpa.2005.08.004 · Cited by 313
Overnight Pulse Oximetry for Sleep-Disordered Breathing in Adults
Netzer, N., Eliasson, A. H., Netzer, C., & Kristo, D. A. (2001) · Chest · doi:10.1378/chest.120.2.625 · Cited by 311
Masimo Signal Extraction Pulse Oximetry
Goldman, J. M., Petterson, M. T., Kopotic, R. J., & Barker, S. J. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011493521730 · Cited by 267
Self-citationCoté, C. J., Wilson, S., AMERICAN ACADEMY OF PEDIATRICS, & AMERICAN ACADEMY OF PEDIATRIC DENTISTRY (2016) · Pediatrics · doi:10.1542/peds.2016-1212 · Cited by 260
Romero, L. M., & Remage-Healey, L. (2000) · General and Comparative Endocrinology · doi:10.1006/gcen.2000.7491 · Cited by 259
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243
Effect of Nitrous Oxide on Excitatory and Inhibitory Synaptic Transmission in Hippocampal Cultures
Mennerick, S., Jevtovic-Todorovic, V., Todorovic, S. M., Shen, W., Olney, J. W., & Zorumski, C. F. (1998) · The Journal of Neuroscience · doi:10.1523/jneurosci.18-23-09716.1998 · Cited by 224
(2008) · Pediatric Anesthesia · doi:10.1111/j.1460-9592.2007.02404.x · Cited by 216
“Motion-Resistant” Pulse Oximetry: A Comparison of New and Old Models
Barker, S. J. (2002) · Anesthesia & Analgesia · doi:10.1097/00000539-200210000-00033 · Cited by 204
Self-citationNeonatal and Pediatric Pulse Oximetry
Salyer, J. W. (2003) · Respiratory Care · doi:10.4187/respcare.03480386 · Cited by 183
The Effect of Motion on Pulse Oximetry and Its Clinical Significance
Petterson, M. T., Begnoche, V. L., & Graybeal, J. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000278134.47777.a5 · Cited by 176
Automated Epileptic Seizure Detection Based on Wearable ECG and PPG in a Hospital Environment
Vandecasteele, K., De Cooman, T., Gu, Y., Cleeren, E., Claes, K., Paesschen, W., Huffel, S., & Hunyadi, B. (2017) · Sensors · doi:10.3390/s17102338 · Cited by 174
Larsen, B., Seitz, A., & Larsen, R. (2000) · Anesthesia & Analgesia · doi:10.1097/00000539-200001000-00035 · Cited by 173
Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion
Louie, A., Feiner, J. R., Bickler, P. E., Rhodes, L., Bernstein, M., & Lucero, J. (2018) · Anesthesiology · doi:10.1097/aln.0000000000002002 · Cited by 106
Chong, J. W., Dao, D. K., Salehizadeh, S. M. A., McManus, D. D., Darling, C. E., Chon, K. H., & Mendelson, Y. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1080-y · Cited by 79
Alterations in electrodermal activity and cardiac parasympathetic tone during hypnosis
Kekecs, Z., Szekely, A., & Varga, K. (2016) · Psychophysiology · doi:10.1111/psyp.12570 · Cited by 47
“…the volar surface of the medial phalanges on the index and middle fingers of the non-dominant hand, while the pulse photoplethysmography (PPG) transducer was clipped to the left earlobe to decrease PPG’s vulnerability to movement artifacts (Barker & Shah, 1997; G. Lu, Yang, Taylor, & Stein, 2009).”
MethodBerkenbosch, J. W., & Tobias, J. D. (2006) · Respiratory Care · doi:10.4187/respcare.06510726 · Cited by 38
Metadata
- Authors
- Barker, S. J., & Shah, N. K.
- Venue
- Anesthesiology
- Published
- 1997-01-01
- Volume
- 86
- Issue
- 1
- Pages
- 101-108
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2131516330
- Cited by
- 195
- References
- 6
- FWCI
- 12.73
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.