Continuous Cardiac Output in Septic Shock by Simulating a Model of the Aortic Input Impedance: A Comparison with Bolus Injection Thermodilution

Open access
Jellema, Wilbert T., Wesseling, Karel H., Groeneveld, Johan A. B., Stoutenbeek, Chris P., Thijs, Lambertus G., van Lieshout, Johannes J. · 1999 · Anesthesiology

Abstract

Background: To compare continuous cardiac output obtained by simulation of an aortic input impedance model to bolus injection thermodilution (TDCO) in critically ill patients with septic shock. Methods: In an open study, mechanically ventilated patients with septic shock were monitored for 1 (32 patients), 2 (15 patients), or 3 (5 patients) days. The hemodynamic state was altered by varying the dosages of dopamine, norepinephrine, or dobutamine. TDCO was estimated 189 times as the series average of four automated phase-controlled injections of iced 5% glucose, spread equally over the ventilatory cycle. Continuous model-simulated cardiac output (MCO) was computed from radial or femoral artery pressure. On each day, the first TDCO value was used to calibrate the model. Results: TDCO ranged from 4.1 to 18.2 l/min. The bias (mean difference between MCO and TDCO) on the first day before calibration was -1.92 +/- 2.3 l/min (mean +/- SD; n = 32; 95% limits of agreement, -6.5 to 2.6 l/min). The bias increased at higher levels of cardiac output (P < 0.05). In 15 patients studied on two consecutive days, the precalibration ratio TDCO:MCO on day 1 was 1.39 +/- 0.28 (mean +/- SD) and did not change on day 2 (1.39 +/- 0.34). After calibration, the bias was -0.1 +/- 0.8 l/min with 82% of the comparisons (n = 112) < 1 l/min and 58% (n = 79) < 0.5 l/min, and independent of the level of cardiac output. Conclusions: In mechanically ventilated patients with septic shock, changes in bolus TDCO are reflected by calibrated MCO over a range of cardiac output values. A single calibration of the model appears sufficient to monitor continuous cardiac output over a 2-day period with a bias of -0.1 +/- 0.8 l/min.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00000542-199905000-00016

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 213

Metadata

Authors
Jellema, W. T., Wesseling, K. H., Groeneveld, J. A. B., Stoutenbeek, C. P., Thijs, L. G., & van Lieshout, J. J.
Venue
Anesthesiology
Published
1999-05-01
Volume
90
Issue
5
Pages
1317-1328
ISSN
0003-3022 (ISSN-L)
OpenAlex
W2168586605
Cited by
213
References
1
FWCI
10.38
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Continuous Cardiac Output in Septic Shock by Simulating a Model of the Aortic Input Impedance: A Comparison with Bolus Injection Thermodilution. https://app.cochat.ai/papers/10.1097/00000542-199905000-00016
The paper
Jellema, W. T., Wesseling, K. H., Groeneveld, J. A. B., Stoutenbeek, C. P., Thijs, L. G., & van Lieshout, J. J. (1999). Continuous Cardiac Output in Septic Shock by Simulating a Model of the Aortic Input Impedance: A Comparison with Bolus Injection Thermodilution. Anesthesiology. https://doi.org/10.1097/00000542-199905000-00016

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.