The Mechanism of Loosening of Cemented Acetabular Components in Total Hip Arthroplasty: Analysis of Specimens Retrieved at Autopsy
Abstract
Late aseptic loosening of cemented acetabular components is governed by the progressive, three-dimensional resorption of the bone immediately adjacent to the cement mantle. This process begins circumferentially at the intraarticular margin and progresses toward the dome of the implant. Evidence of bone resorption at the cement-bone interface was present even in the most well-fixed implants before the appearance of lucent lines on standard roentgenographic views. The mechanical stability of the implant was determined by the three-dimensional extent of bone resorption and membrane formation at the cement-bone interface. The leading edge of the membrane is a transition zone from regions of membrane interposition between the cement and the bone to regions of intimate cement-bone contact. Histologic analysis revealed that progressive bone resorption is fueled by small particles of high density polyethylene (HDP) migrating along the cement-bone interface and bone resorption occurs as a result of the macrophage inflammatory response to the particulate HDP. Evidence in support of a mechanical basis for failure of fixation was lacking. The mechanism of late aseptic loosening of a cemented acetabular component is therefore biologic in nature, not mechanical. This is exactly opposite to the mechanism of loosening on the femoral side of a cemented total hip replacement, which is mechanical in nature.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Kurtz, S. (1999) · Biomaterials · doi:10.1016/s0142-9612(99)00053-8 · Cited by 847
Aseptic loosening, not only a question of wear: A review of different theories
Sundfeldt, M., V Carlsson, L., B Johansson, C., Thomsen, P., & Gretzer, C. (2006) · Acta Orthopaedica · doi:10.1080/17453670610045902 · Cited by 622
Current Concepts Review - Wear in Total Hip and Knee Replacements*
SCHMALZRIED, T. P., & CALLAGHAN, J. J. (1999) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199901000-00016 · Cited by 395
Self-citationWear particles, periprosthetic osteolysis and the immune system
Goodman, S. B. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.06.035 · Cited by 340
Maloney, W. J., Smith, R. L., Schmalzried, T. P., Chiba, J., Huene, D., & Rubash, H. (1995) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199509000-00002 · Cited by 307
Self-citation
Metadata
- Authors
- SCHMALZRIED, T. P., KWONG, L. M., JASTY, M., SEDLACEK, R. C., HAIRE, T. C., O'CONNOR, D. O., BRAGDON, C. R., KABO, J. M., MALCOLM, A. J., & HARRIS, W. H.
- Venue
- Clinical Orthopaedics and Related Research
- Published
- 1992-01
- Volume
- 274
- Issue
- 274
- Pages
- 60???78
- ISSN
- 0009-921X (ISSN-L)
- OpenAlex
- W2091162166
- Cited by
- 443
- References
- 0
- FWCI
- 19.36
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Mechanism of Loosening of Cemented Acetabular Components in Total Hip Arthroplasty: Analysis of Specimens Retrieved at Autopsy. https://app.cochat.ai/papers/10.1097/00003086-199201000-00009
- The paper
- SCHMALZRIED, T. P., KWONG, L. M., JASTY, M., SEDLACEK, R. C., HAIRE, T. C., O'CONNOR, D. O., BRAGDON, C. R., KABO, J. M., MALCOLM, A. J., & HARRIS, W. H. (1992). The Mechanism of Loosening of Cemented Acetabular Components in Total Hip Arthroplasty: Analysis of Specimens Retrieved at Autopsy. Clinical Orthopaedics and Related Research. https://doi.org/10.1097/00003086-199201000-00009
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.