The Mechanism of Loosening of Cemented Acetabular Components in Total Hip Arthroplasty: Analysis of Specimens Retrieved at Autopsy
The DOI matches its Crossref record.
Abstract
Late aseptic loosening of cemented acetabular components is governed by the progressive, three-dimensional resorption of the bone immediately adjacent to the cement mantle. This process begins circumferentially at the intraarticular margin and progresses toward the dome of the implant. Evidence of bone resorption at the cement-bone interface was present even in the most well-fixed implants before the appearance of lucent lines on standard roentgenographic views. The mechanical stability of the implant was determined by the three-dimensional extent of bone resorption and membrane formation at the cement-bone interface. The leading edge of the membrane is a transition zone from regions of membrane interposition between the cement and the bone to regions of intimate cement-bone contact. Histologic analysis revealed that progressive bone resorption is fueled by small particles of high density polyethylene (HDP) migrating along the cement-bone interface and bone resorption occurs as a result of the macrophage inflammatory response to the particulate HDP. Evidence in support of a mechanical basis for failure of fixation was lacking. The mechanism of late aseptic loosening of a cemented acetabular component is therefore biologic in nature, not mechanical. This is exactly opposite to the mechanism of loosening on the femoral side of a cemented total hip replacement, which is mechanical in nature.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Kurtz, S. (1999) · Biomaterials · doi:10.1016/s0142-9612(99)00053-8 · Cited by 847
De Aza, A. H., Chevalier, J., Fantozzi, G., Schehl, M., & Torrecillas, R. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00206-x · Cited by 625
Aseptic loosening, not only a question of wear: A review of different theories
Sundfeldt, M., V Carlsson, L., B Johansson, C., Thomsen, P., & Gretzer, C. (2006) · Acta Orthopaedica · doi:10.1080/17453670610045902 · Cited by 622
Effect of surface topology on the osseointegration of implant materials in trabecular bone
Wong, M., Eulenberger, J., Schenk, R., & Hunziker, E. (1995) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820291213 · Cited by 449
Jiranek, W. A., Machado, M., Jasty, M., Jevsevar, D., Wolfe, H. J., Goldring, S. R., Goldberg, M. J., & Harris, W. H. (1993) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199306000-00007 · Cited by 436
Self-citationQuantitative Assessment of Walking Activity after Total Hip or Knee Replacement*
SCHMALZRIED, T. P., SZUSZCZEWICZ, E. S., NORTHFIELD, M. R., AKIZUKI, K. H., FRANKEL, R. E., BELCHER, G., & AMSTUTZ, H. C. (1998) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199801000-00010 · Cited by 434
Self-citationCurrent Concepts Review - Wear in Total Hip and Knee Replacements*
SCHMALZRIED, T. P., & CALLAGHAN, J. J. (1999) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199901000-00016 · Cited by 395
Self-citationWear particles, periprosthetic osteolysis and the immune system
Goodman, S. B. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.06.035 · Cited by 340
“This concept, later coined “the effective joint space” explains the appearance of osteolytic lesions at distant points around well-fixed and loose implants [3].”
BackgroundOsteolysis and particle disease in hip replacement: A review
Harris, W. H. (1994) · Acta Orthopaedica Scandinavica · doi:10.3109/17453679408993734 · Cited by 319
Self-citationMaloney, W. J., Smith, R. L., Schmalzried, T. P., Chiba, J., Huene, D., & Rubash, H. (1995) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199509000-00002 · Cited by 307
Self-citationUltra-high molecular weight polyethylene. The material and its use in total joint implants.
Li, S., & Burstein, A. H. (1994) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199407000-00020 · Cited by 291
Schmalzried, T. P., Guttmann, D., Grecula, M., & Amstutz, H. C. (1994) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199405000-00008 · Cited by 289
Self-citationA unified theory of wear for ultra-high molecular weight polyethylene in multi-directional sliding
Wang, A. (2001) · Wear · doi:10.1016/s0043-1648(00)00522-6 · Cited by 245
Periprosthetic Osteolysis: Mechanisms, Prevention and Treatment
Goodman, S. B., & Gallo, J. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8122091 · Cited by 238
“Well-fixed cementless implants may demonstrate proximal femoral osteolysis due to the biological reaction to polyethylene debris in the effective joint space [138].”
BackgroundP Value and the Theory of Hypothesis Testing: An Explanation for New Researchers
Biau, D. J., Jolles, B. M., & Porcher, R. (2010) · Clinical Orthopaedics & Related Research · doi:10.1007/s11999-009-1164-4 · Cited by 225
“For instance, with respect to the first example given, we may hypothesize that the presence of a radiolucent line observed in Zone 1 on the postoperative radiograph is a sign of a gap between cement and bone that will favor micromotion and facilitate the passage of polyethylene wear particles, both of which will favor…”
BackgroundGoodman, S. B., Huie, P., Song, Y., Schurman, D. J., Maloney, W. J., Woolson, S. T., & Sibley, R. K. (1998) · Journal of Bone and Joint Surgery - British Volume · doi:10.1302/0301-620x.80b3.8158 · Cited by 217
The role of surface free energy in osteoblast–biomaterial interactions
Gentleman, M. M., & Gentleman, E. (2014) · International Materials Reviews · doi:10.1179/1743280414y.0000000038 · Cited by 216
Kobayashi, A., Freeman, M. A. R., Bonfield, W., Kadoya, Y., Yamac, T., Al-Saffar, N., Scott, G., & Revell, P. A. (1997) · Journal of Bone and Joint Surgery - British Volume · doi:10.1302/0301-620x.79b5.7602 · Cited by 195
Biological reaction to debris in relation to joint prostheses
Revell, P. A., AL-Saffar, N., & Kobayashi, A. (1997) · Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · doi:10.1243/0954411971534304 · Cited by 195
Step activity monitor: Increased accuracy in quantifying ambulatory activity
Shepherd, E. F., Toloza, E., McClung, C. D., & Schmalzried, T. P. (1999) · Journal of Orthopaedic Research · doi:10.1002/jor.1100170512 · Cited by 182
Self-citationBalla, V. K., DeVasConCellos, P. D., Xue, W., Bose, S., & Bandyopadhyay, A. (2009) · Acta Biomaterialia · doi:10.1016/j.actbio.2009.01.011 · Cited by 170
Metadata
- Authors
- SCHMALZRIED, T. P., KWONG, L. M., JASTY, M., SEDLACEK, R. C., HAIRE, T. C., O'CONNOR, D. O., BRAGDON, C. R., KABO, J. M., MALCOLM, A. J., & HARRIS, W. H.
- Venue
- Clinical Orthopaedics and Related Research
- Published
- 1992-01
- Volume
- 274
- Issue
- 274
- Pages
- 60???78
- ISSN
- 0009-921X (ISSN-L)
- OpenAlex
- W2091162166
- Cited by
- 443
- References
- 0
- FWCI
- 19.36
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.