Measurement of gas exchange in intensive care: Laboratory and clinical validation of a new device
The DOI matches its Crossref record.
Abstract
The performance of a new gas exchange monitor was assessed both in laboratory simulation and in ICU patients. Laboratory simulation using N2 and CO2 injections resulted in a mean error of 2 +/- 2% in CO2 production (VCO2) and 4 +/- 4% in oxygen consumption (VO2) in respirator measurements (n = 55) and in a mean error of 3 +/- 2% in VCO2 and 4 +/- 2% in VO2 in canopy measurements (n = 25). The mean error in RQ during ethanol burning was 2 +/- 2% in respirator measurements (n = 45) and 1 +/- 1% in canopy measurements. FIO2 had little effect on the accuracy of VCO2, whereas the accuracy on high rates of VO2 (VO2 = 400 ml/min) was reduced, when FIO2 increased: the error ranged from 1 +/- 1% to 6 +/- 1%, except at VO2 400 ml/min during FIO2 0.8, where the error was 16 +/- 3%. Neither peak airway pressure (+13 to +63 cm H2O) nor PEEP (0 to +20 cm H2O) had an effect on the accuracy. The highest level of minute ventilation studied (22.5 L/min) reduced the accuracy slightly (mean error of VCO2 4 +/- 1% and VO2 7 +/- 2%). In patients during controlled mechanical ventilation, increasing FIO2 from 0.4 to 0.6 had no effect on the results. VO2 was consistently higher by gas exchange than by the Fick principle: 16 +/- 9% during controlled ventilation (n = 20), 21 +/- 8% on synchronized intermittent mandatory ventilation (n = 10) and 25 +/- 8% during spontaneous breathing. We conclude that the device proved to be accurate for gas exchange measurements in the ICU.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Indirect Calorimetry: A Practical Guide for Clinicians
Haugen, H. A., Chan, L., & Li, F. (2007) · Nutrition in Clinical Practice · doi:10.1177/0115426507022004377 · Cited by 512
Singer, P., Theilla, M., Fisher, H., Gibstein, L., Grozovski, E., & Cohen, J. (2006) · Critical Care Medicine · doi:10.1097/01.ccm.0000206111.23629.0a · Cited by 393
Preoperative oral carbohydrate administration reduces postoperative insulin resistance
Nygren, J., Soop, M., Thorell, A., Efendic, S., Nair, K. S., & Ljungqvist, O. (1998) · Clinical Nutrition · doi:10.1016/s0261-5614(98)80307-5 · Cited by 313
Indirect calorimetry in nutritional therapy. A position paper by the ICALIC study group
Oshima, T., Berger, M. M., De Waele, E., Guttormsen, A. B., Heidegger, C.-P., Hiesmayr, M., Singer, P., Wernerman, J., & Pichard, C. (2017) · Clinical Nutrition · doi:10.1016/j.clnu.2016.06.010 · Cited by 283
“However, the use of the Haldane transformation formula introduces a mathematical limitation, especially in case of O2 enrichment higher than 60% as the inaccuracy of the analyzers will be enhanced by the calculation [28,30,34].”
BackgroundMethodInfluentialPlank, L. D., Connolly, A. B., & Hill, G. L. (1998) · Annals of Surgery · doi:10.1097/00000658-199808000-00002 · Cited by 238
Citation classics in critical care medicine
Baltussen, A., & Kindler, C. H. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2195-7 · Cited by 207
Asymptomatic atherosclerosis and insulin resistance.
Laakso, M., Sarlund, H., Salonen, R., Suhonen, M., Pyörälä, K., Salonen, J. T., & Karhapää, P. (1991) · Arteriosclerosis and Thrombosis: A Journal of Vascular Biology · doi:10.1161/01.atv.11.4.1068 · Cited by 206
Invited Review: Use of Indirect Calorimetry in Clinical Nutrition
McClave, S. A., & Snider, H. L. (1992) · Nutrition in Clinical Practice · doi:10.1177/0115426592007005207 · Cited by 195
Tibby, S. M., Hatherill, M., Marsh, M. J., Morrison, G., Anderson, D., & Murdoch, I. A. (1997) · Intensive Care Medicine · doi:10.1007/s001340050443 · Cited by 178
“Sources of error in calculation of CO from the Fick principle were kept to a minimum by avoiding air leaks through the use of cuffed endotracheal tubes where necessary [15], avoidance of VO2 measurement at higher FIO2 levels [16, 17] and use of co-oximetry to measure oxygen saturations.”
Indirect calorimetry: methodology, instruments and clinical application
da Rocha, E. E. M., Alves, V. G. F., & da Fonseca, R. B. V. (2006) · Current Opinion in Clinical Nutrition and Metabolic Care · doi:10.1097/01.mco.0000222107.15548.f5 · Cited by 177
The Measurement of Energy Expenditure
Branson, R. D., & Johannigman, J. A. (2004) · Nutrition in Clinical Practice · doi:10.1177/0115426504019006622 · Cited by 111
Can calculation of energy expenditure based on CO2 measurements replace indirect calorimetry?
Oshima, T., Graf, S., Heidegger, C.-P., Genton, L., Pugin, J., & Pichard, C. (2017) · Critical Care · doi:10.1186/s13054-016-1595-8 · Cited by 48
Taittonen, M. T., Kirvela, O. A., Aantaa, R., & Kanto, J. H. (1998) · Anesthesia & Analgesia · doi:10.1213/00000539-199807000-00034 · Cited by 41
Metadata
- Authors
- TAKALA, J., KEINÄNEN, O., VÄISÄNEN, P., & KARI, A.
- Venue
- Critical Care Medicine
- Published
- 1989-10
- Volume
- 17
- Issue
- 10
- Pages
- 1041-1047
- ISSN
- 0090-3493 (ISSN-L)
- OpenAlex
- W2026326574
- Cited by
- 303
- References
- 0
- FWCI
- 15.64
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.