Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution

Goedje, Oliver, Hoeke, Kerstin, Lichtwarck-Aschoff, Michael, Faltchauser, A., Lamm, Peter, Reichart, Bruno·1999·Critical Care Medicine

Abstract

OBJECTIVE: To compare two thermodilution methods for the determination of cardiac output (CO)-thermodilution in the pulmonary artery (COpa) and thermodilution in the femoral artery (COa)-with each other and with CO determined by continuous pulse contour analysis (COpc) in terms of reproducibility, bias, and correlation among the different methods. Good agreement between the methods would indicate the potential of pulse contour analysis to monitor CO continuously and at reduced invasiveness. DESIGN: Prospective criterion standard study. SETTING: Cardiac surgical intensive care unit in a university hospital. PATIENTS: Twenty-four postoperative cardiac surgery patients. INTERVENTIONS: Without interfering with standard hospital cardiac recovery procedures, changes in CO as a result of the postsurgical course, administration of vasoactive substances, and/or fluid administration were recorded. CO was first recorded after a 1-hr stabilization period in the intensive care unit and hourly thereafter for 6 hrs, and by subsequent determinations at 9, 12, and 24 hrs. MEASUREMENTS AND MAIN RESULTS: There were 216 simultaneous determinations of COpa, COa, and COpc. COpc was initially calibrated using COa, and no further recalibration of COpc was performed. COpa ranged from 3.0 to 11.8 L/min, and systemic vascular resistance ranged from 252 to 2434 dyne x sec/cm5. The mean difference (bias) +/-2 SD of differences (limits of agreement) was -0.29+/-1.31 L/min for COpa vs. COa, 0.07+/-1.4 L/min for COpc vs. COpa, and -0.22+/-1.58 L/min for COpc vs. COa. In all but four patients COpc correlated with COa after the initial calibration. Correlation and precision of COpc vs. COa was stable for 24 hrs. CONCLUSIONS: Femoral artery pulse contour CO correlates well with both COpa and COa even during substantial variations in vascular tone and hemodynamics. Additionally, CO determined by arterial thermodilution correlates well with COpa. Thus, COa can be used to calibrate COpc.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003246-199911000-00014

Cites

9 of 9 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 9 references

Cited by

See all 576

Metadata

Authors
Goedje, O., Hoeke, K., Lichtwarck-Aschoff, M., Faltchauser, A., Lamm, P., & Reichart, B.
Venue
Critical Care Medicine
Published
1999-11
Volume
27
Issue
11
Pages
2407-2412
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2008261942
Cited by
576
References
9
FWCI
10.06
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution. https://app.cochat.ai/papers/10.1097/00003246-199911000-00014
The paper
Goedje, O., Hoeke, K., Lichtwarck-Aschoff, M., Faltchauser, A., Lamm, P., & Reichart, B. (1999). Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution. Critical Care Medicine. https://doi.org/10.1097/00003246-199911000-00014

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.