Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies--with specific reference to the measurement of cardiac output
Open accessAbstract
Bland-Altman analysis is used for assessing agreement between two measurements of the same clinical variable. In the field of cardiac output monitoring, its results, in terms of bias and limits of agreement, are often difficult to interpret, leading clinicians to use a cutoff of 30% in the percentage error in order to decide whether a new technique may be considered a good alternative. This percentage error of +/- 30% arises from the assumption that the commonly used reference technique, intermittent thermodilution, has a precision of +/- 20% or less. The combination of two precisions of +/- 20% equates to a total error of +/- 28.3%, which is commonly rounded up to +/- 30%. Thus, finding a percentage error of less than +/- 30% should equate to the new tested technique having an error similar to the reference, which therefore should be acceptable. In a worked example in this paper, we discuss the limitations of this approach, in particular in regard to the situation in which the reference technique may be either more or less precise than would normally be expected. This can lead to inappropriate conclusions being drawn from data acquired in validation studies of new monitoring technologies. We conclude that it is not acceptable to present comparison studies quoting percentage error as an acceptability criteria without reporting the precision of the reference technique.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Echocardiography shows a precision error of around 30% when compared to gold standard dilution techniques, which falls within the clinically accepted range.
“For instance, echocardiography shows a precision error of around 30% when compared to gold standard dilution techniques, which falls within the clinically accepted range [54,55].”
From Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review (2025) · cites this paper as [55]
The Validation of cardiac output monitoring devices section (page unavailable) describes ±30% as a commonly used percentage-error cutoff for comparing a new cardiac-output technique with intermittent thermodilution. It does not report that echocardiography has a 30% error; it also warns that the cutoff alone cannot establish clinical acceptability without measuring the reference technique’s precision. (LLM, conf=0.88)
Experimental — accuracy not yet measuredSimilar associations between inflammatory mediators and disease severity have been demonstrated in the context of influenza virus, coronavirus, and parainfluenza virus infections.
“Similar associations have been demonstrated in the context of influenza virus (21, 43, 206, 283), coronavirus (328), and parainfluenza virus (58) infections.”
From The airway epithelium: soldier in the fight against respiratory viruses (2011) · cites this paper as [21]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
1 retracted
Mayer, J., Boldt, J., Schöllhorn, T., Röhm, K. D., Mengistu, A. M., & Suttner, S. (2007) · British Journal of Anaesthesia · doi:10.1093/bja/ael341 · Cited by 146 · see citing works
RetractedSTATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619 · see citing works
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343 · see citing works
Critchley, L. A. H., & Critchley, J. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009982611386 · Cited by 1,328 · see citing works
Pearse, R. M., Dawson, D., Fawcett, J., Rhodes, A., Grounds, R. M., & Bennett, E. (2005) · Critical Care · doi:10.1186/cc3887 · Cited by 698 · see citing works
Self-citationGoedje, O., Hoeke, K., Lichtwarck‐Aschoff, M., et al. (1999) · Critical Care Medicine · doi:10.1097/00003246-199911000-00014 · Cited by 579 · see citing works
Rocca, C. D., Costa, M. G., Pompei, L., Coccia, C., & Pietropaoli, P. (2002) · British Journal of Anaesthesia · doi:10.1093/bja/88.3.350 · Cited by 239 · see citing works
Early Goal-Directed Therapy after Major Surgery Reduces Complications and Duration of Hospital Stay
Elliott, D. (2006) · Journal of the Intensive Care Society · doi:10.1177/175114370600700123 · Cited by 228 · see citing works
Button, D., Weibel, L., Reuthebuch, O. T., Genoni, M., Zollinger, A., & Hofer, C. K. (2007) · British Journal of Anaesthesia · doi:10.1093/bja/aem188 · Cited by 173 · see citing works
Pittman, J., Bar-Yosef, S., Sum-Ping, J., Sherwood, M. W., & Mark, J. B. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000179021.36805.1f · Cited by 135 · see citing works
Hamilton, T. T., Huber, L. M., & Jessen, M. E. (2002) · The Annals of Thoracic Surgery · doi:10.1016/s0003-4975(02)04059-6 · Cited by 135 · see citing works
Sander, M., Spies, C., Grubitzsch, H., Foer, A., Müller, M. C., & von Heymann, C. (2006) · Critical Care · doi:10.1186/cc5103 · Cited by 121 · see citing works
Bein, B., Worthmann, F., Tonner, P. H., Paris, A., Steinfath, M., Hedderich, J., & Scholz, J. (2004) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2004.01.025 · Cited by 98 · see citing works
Change in stroke volume in response to fluid challenge: assessment using esophageal Doppler
Roeck, M., Jakob, S. M., Boehlen, T., et al. (2003) · Intensive Care Medicine · doi:10.1007/s00134-003-1720-4 · Cited by 62 · see citing works
Evaluation of a thoracic bioimpedance cardiac output monitor during cardiac catheterization
Barin, E. S., Haryadi, D. G., Schookin, S. I., et al. (2000) · Critical Care Medicine · doi:10.1097/00003246-200003000-00016 · Cited by 49 · see citing works
Continuous cardiac output measurements in the perioperative period
Jakobsen, C.-J., Melsen, N. C., & Andresen, E. B. (1995) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1995.tb04104.x · Cited by 48 · see citing works
Munro, H. M., Wood, C. E., Taylor, B. L., et al. (1994) · PubMed · Cited by 24 · see citing works
Testing the safety of baxter continuous cardiac output monitoring system
Lichtenthal, P. R., & Gordan, D. (1996) · The Journal of Clinical Monitoring · doi:10.1007/bf00857646 · Cited by 9 · see citing works
Tzenkov, I. G., & Peña, J. (2003) · British Journal of Anaesthesia · doi:10.1093/bja/aeg521 · Cited by 8 · see citing works
Nevière, R., Mathieu, D., Riou, Y., et al. (1994) · PubMed · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
The airway epithelium: soldier in the fight against respiratory viruses
Vareille, M., Kieninger, E., Edwards, M. R., & Regamey, N. (2011) · Clinical Microbiology Reviews · doi:10.1128/cmr.00014-10 · Cited by 655 · see citing works
Noori, S., Drabu, B., Soleymani, S., & Seri, I. (2012) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/fetalneonatal-2011-301090 · Cited by 146 · see citing works
Nedelcu, C., Ionescu, N. S., Bizubac, A. M., Filip, C., & Cirstoveanu, C. (2025) · Life · doi:10.3390/life15101621 · Cited by 2 · see citing works
“For instance, echocardiography shows a precision error of around 30% when compared to gold standard dilution techniques, which falls within the clinically accepted range [54,55].”
- Authors
- Cecconi, M., Rhodes, A., Poloniecki, J., Della Rocca, G., & Grounds, R. M.
- Affiliations
- University of Udine (IT)
- St George's Hospital (GB)
- St George's, University of London (GB)
- Corresponding author
- Cecconi, Maurizio
- Venue
- Critical Care
- Published
- 2009
- Volume
- 13
- Issue
- 1
- Pages
- 201
- ISSN
- 1364-8535 (ISSN-L)
- PMC
- PMC2688094
- OpenAlex
- W2155688319
- Cited by
- 351
- References
- 24
- Citations by year
- 2026: 16 · 2025: 16 · 2024: 18 · 2023: 20 · 2022: 15 · 2021: 24 · 2020: 17 · 2019: 20 · 2018: 18 · 2017: 23 · 2016: 29 · 2015: 42 · 2014: 24 · 2013: 20 · 2012: 15
- FWCI
- 16.56
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Critical Care (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- St George's Online Research Archive (St George's University of London) (acceptedVersion)
- Institutional Research Information System (University of Udine) (submittedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- cardiac output monitoring, method comparison, Bland-Altman analysis, measurement agreement, percentage error, limits of agreement, bias
- MeSH terms
- Cardiac Output, Humans, Monitoring, Physiologic, Reference Standards, Reproducibility of Results, Point-of-Care Systems, Biomedical Research
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review