Accuracy of AccuChek glucose measurement in intensive care patients *

Meynaar, Iwan A., van Spreuwel, Margot, Tangkau, Peter L., Dawson, Lilian, Visser, Steven Sleeswijk, Rijks, Lode, Vlieland, Thea Vliet · 2009 · Critical Care Medicine
Cited by 42

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: To evaluate the accuracy of the AccuChek Inform point-of-care glucose measurement device as compared with central laboratory glucose measurement. DESIGN: Prospective, observational study. SETTING: A ten-bed mixed closed format intensive care unit ina 500-bed general hospital. The unit has a computerized insulin protocol aiming for 81 to 135 mg/dL. PATIENTS: All intensive care unit patients were eligible. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Paired samples (AccuChek glucose in whole blood calibrated to give whole blood results and central laboratory glucose in serum) were taken simultaneously. In 32 critically ill patients, we obtained the following information: mean +/- standard deviation age 71.6 +/- 11.9 yrs; mean Acute Physiology and Chronic Health Evaluation II score at admission 17.8 +/- 6.7; 239 paired samples were taken from arterial catheters. Mean AccuChek whole blood glucose was 126 +/- 36 mg/dL (7.0 +/- 2.0 mmol/L); mean central laboratory serum glucose was 137 +/- 38 mg/dL (7.6 +/- 2.1 mmol/L). Mean difference was 11 mg/dL (0.61 mmol/L) (8%) (95% Confidence Interval 9-13 mg/dL, p < .001). ISO 15197 guideline requires 95% of point-of-care measurements to be within 15 mg/dL margins with reference <75 mg/dL or within 20% if reference is higher. In total, 216 (90.4%) of AccuChek measurements were within ISO 15197 margins. Because AccuChek was calibrated to give whole blood results, we calculated a correction factor of 1.086 from the two mean values to correct whole blood AccuChek into serum-like results. This is almost the same as the correction factor of 1.080 given by Roche Diagnostics. By multiplying AccuChek whole blood results with 1.086, 225 (94.1%) of results were within the ISO 15197 margins. Hematocrit did not influence AccuChek results in the 0.20 to 0.44 range. Beyond this range, there were not enough data to draw conclusions. CONCLUSIONS: In critically ill patients, the accuracy of AccuChek glucose measurement calibrated to give serum-like results with blood samples derived from arterial catheters is acceptable but falls short by about 1% of complying with the ISO 15197 guideline.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003246-200910000-00002

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 42

Metadata

Authors
Meynaar, I. A., van Spreuwel, M., Tangkau, P. L., Dawson, L., Visser, S. S., Rijks, L., & Vlieland, T. V.
Venue
Critical Care Medicine
Published
2009-10
Volume
37
Issue
10
Pages
2691-2696
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2061434420
Cited by
42
References
3
FWCI
5.44
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accuracy of AccuChek glucose measurement in intensive care patients *. https://app.cochat.ai/papers/10.1097/00003246-200910000-00002
The paper
Meynaar, I. A., van Spreuwel, M., Tangkau, P. L., Dawson, L., Visser, S. S., Rijks, L., & Vlieland, T. V. (2009). Accuracy of AccuChek glucose measurement in intensive care patients *. Critical Care Medicine. https://doi.org/10.1097/00003246-200910000-00002

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.