Polychromatic Streak Artifacts in Computed Tomography Images

Duerinckx, André J., Macovski, Albert·1978·Journal of Computer Assisted Tomography

Abstract

A series of artifacts in computed tomography (CT) images, consisting of streaks coming off edges and flares surrounding the object being scanned, is shown to be due to polychromaticity of the X-ray beam. These artifacts are seldom seen on conventional CT brain scans. They are helpful in understanding artifacts such as those seen around the petrous bone of the skull. Following this, a comparison is made between "single energy" and "dual energy" correction schemes for poly-artifacts. The theoretical assumptions underlying either method are analyzed, and the importance of these concepts in dual energy reconstructions is stressed. A new test phantom is suggested to evaluate different poly-correction schemes. All the simulated projection data in this study are noise free and are reconstructed using an industrial convolution-back projection technique. In addition, reconstructions are shown of the phantom using real data taken with and without a water bag path length compensator.

Integrity signals

Registry

Crossref · 10.1097/00004728-197809000-00020

Cites

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • The large extracranial metal components and dental cement of the implant can cause beam hardening that can appear as cupping, streaks, dark bands or flare in the μCT.

    “the large extracranial metal components and dental cement of the implant can cause beam hardening that can appear as cupping, streaks, dark bands or flare in the µCT38–40.”

Cited by

Sorted by citations

Authors
Duerinckx, A. J., & Macovski, A.
Affiliations
Stanford University (US)
Venue
Journal of Computer Assisted Tomography
Published
1978-09
Volume
2
Issue
4
Pages
481-487
ISSN
0363-8715 (ISSN-L)
OpenAlex
W2049442067
Cited by
72
References
0
Citations by year
2025: 1 · 2023: 2 · 2022: 3 · 2021: 1 · 2020: 2 · 2019: 2 · 2018: 1 · 2017: 1 · 2016: 2 · 2015: 2 · 2014: 2 · 2013: 2 · 2012: 3
FWCI
2.78
Type
journal article
Language
en
Open access
closed
Locations
  • Journal of Computer Assisted Tomography (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
Keywords
streak artifacts, polychromatic X-ray beam, dual-energy CT, single-energy CT, artifact removal, petrous bone, test phantom, convolution backprojection
MeSH terms
Mathematics, Models, Structural, Tomography, X-Ray Computed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review