Performance Evaluation of the Positron Scanner ECAT EXACT
The DOI matches its Crossref record.
Abstract
The Cologne Special is a prototype of the ECAT EXACT (model 921), a new generation of Siemens-CTI PET scanners. It consists of three rings of 48 BGO block detectors each, covering an axial field of view of 16.2 cm with a patient port of 56.2 cm diameter. This results in a total of 24 rings with 384 crystals each, giving 47 contiguous image planes in two-dimensional (2D) mode. Total system sensitivity is 216 kcps/microCi/ml for a 20 cm cylinder phantom in 2D. This increases to 1.5 Mcps/microCi/ml in 3D. Data are acquired in the stationary mode only (no wobble motion), resulting in a transaxial spatial resolution of better than 6 mm full width at half-maximum (FWHM) at the center, which degrades to 7.5 mm tangentially and 9.6 mm radially at a radius of 20 cm. Average axial resolution changes from 5.0 mm FWHM at the center to 8.1 mm at R = 20 cm. Count rate performance was investigated at different low energy discriminator settings and found to be linear up to 2.5 microCi/ml with a 20 cm phantom. The magnitude and distribution of scatter were evaluated for both septa-extended and septa-retracted conditions for a range of energy thresholds. Brain, heart, and whole-body studies with the new tomograph demonstrate the versatility of its applications without compromising on physical performance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
White matter lesions impair frontal lobe function regardless of their location
Tullberg, M., Fletcher, E., DeCarli, C., Mungas, D., Reed, B. R., Harvey, D. J., Weiner, M. W., Chui, H. C., & Jagust, W. J. (2004) · Neurology · doi:10.1212/01.wnl.0000130530.55104.b5 · Cited by 530
Mawlawi, O., Martinez, D., Slifstein, M., Broft, A., Chatterjee, R., Hwang, D.-R., Huang, Y., Simpson, N., Ngo, K., Van Heertum, R., et al. (2001) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1097/00004647-200109000-00002 · Cited by 525
“…47 (three-dimensional in plane and axial resolutions of 6 mm and 4.6 mm FWHM, respectively, at the center of the FOV, slice spacing of 3.375 mm, (Wienhard et al., 1992);(2) total duration of the experiment was 120 minutes (bolus volume of 25.6 mL over 3 minutes; infusion rate of 0.25 mL min−1…”
BackgroundInfluentialWesterterp, M., Pruim, J., Oyen, W., Hoekstra, O., Paans, A., Visser, E., van Lanschot, J., Sloof, G., & Boellaard, R. (2007) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-006-0224-1 · Cited by 294
Hilker, R., Schweitzer, K., Coburger, S., Ghaemi, M., Weisenbach, S., Jacobs, A. H., Rudolf, J., Herholz, K., & Heiss, W.-D. (2005) · Archives of Neurology · doi:10.1001/archneur.62.3.378 · Cited by 293
Self-citationDelineation of Brain Tumor Extent with [11C]l-Methionine Positron Emission Tomography
Kracht, L. W., Miletic, H., Busch, S., Jacobs, A. H., Voges, J., Hoevels, M., Klein, J. C., Herholz, K., & Heiss, W.-D. (2004) · Clinical Cancer Research · doi:10.1158/1078-0432.ccr-04-0262 · Cited by 292
Self-citationGrosu, A.-L., Astner, S. T., Riedel, E., Nieder, C., Wiedenmann, N., Heinemann, F., Schwaiger, M., Molls, M., Wester, H.-J., & Weber, W. A. (2011) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2010.07.002 · Cited by 253
From PET detectors to PET scanners
Humm, J. L., Rosenfeld, A., & Del Guerra, A. (2003) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-003-1266-2 · Cited by 245
Statistical Image Reconstruction Methods for Transmission Tomography
Fessler, J. (2010) · Handbook of Medical Imaging, Volume 2. Medical Image Processing and Analysis · doi:10.1117/3.831079.ch1 · Cited by 240
Fehr, C., Yakushev, I., Hohmann, N., Buchholz, H.-G., Landvogt, C., Deckers, H., Eberhardt, A., Kläger, M., Smolka, M. N., Scheurich, A., et al. (2008) · American Journal of Psychiatry · doi:10.1176/appi.ajp.2007.07020352 · Cited by 226
“) operating in the three-dimensional mode (22).”
Ghaemi, M. (2002) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp.73.5.517 · Cited by 224
Grouped-coordinate ascent algorithms for penalized-likelihood transmission image reconstruction
Fessler, J. A., Ficaro, E. P., Clinthorne, N. H., & Lange, K. (1997) · IEEE Transactions on Medical Imaging · doi:10.1109/42.563662 · Cited by 222
Dehdashti, F., Mortimer, J. E., Trinkaus, K., Naughton, M. J., Ellis, M., Katzenellenbogen, J. A., Welch, M. J., & Siegel, B. A. (2009) · Breast Cancer Research and Treatment · doi:10.1007/s10549-008-9953-0 · Cited by 213
A comparison of resting-state brain activity in humans and chimpanzees
Rilling, J. K., Barks, S. K., Parr, L. A., Preuss, T. M., Faber, T. L., Pagnoni, G., Bremner, J. D., & Votaw, J. R. (2007) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0705132104 · Cited by 199
Mood, cognition and serotonin transporter availability in current and former ecstasy (MDMA) users
Thomasius, R., Petersen, K., Buchert, R., Andresen, B., Zapletalova, P., Wartberg, L., Nebeling, B., & Schmoldt, A. (2003) · Psychopharmacology · doi:10.1007/s00213-002-1383-9 · Cited by 190
“Imaging was performed on a full-ring whole-body system ECAT EXACT 921/47 (Siemens/CTI, Knoxville, Tenn., USA; Wienhard et al. 1992) in 2D mode.”
MethodPET: Physics, Instrumentation, and Scanners
Cherry, S. R., & Dahlbom, M. (2006) · PET · doi:10.1007/0-387-34946-4_1 · Cited by 181
Pardo, J. V., Lee, J. T., Sheikh, S. A., Surerus-Johnson, C., Shah, H., Munch, K. R., Carlis, J. V., Lewis, S. M., Kuskowski, M. A., & Dysken, M. W. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2006.12.044 · Cited by 176
High opiate receptor binding potential in the human lateral pain system
Baumgärtner, U., Buchholz, H.-G., Bellosevich, A., Magerl, W., Siessmeier, T., Rolke, R., Höhnemann, S., Piel, M., Rösch, F., Wester, H.-J., et al. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.10.033 · Cited by 175
“The camera has a field-of-view of 16.2 cm in 47 planes with a plane spacing of 3.375 mm, an axial resolution of 6.0 mm FWHM in 3D mode (Wienhard et al., 1992) and an in-plane resolution of 6.0 mm (resolution in center with scanner in 3D mode).”
InfluentialHilker, R. (2005) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp.2004.057893 · Cited by 173
Yap, C. S., Czernin, J., Fishbein, M. C., Cameron, R. B., Schiepers, C., Phelps, M. E., & Weber, W. A. (2006) · Chest · doi:10.1378/chest.129.2.393 · Cited by 166
Thomasius, R., Zapletalova, P., Petersen, K., Buchert, R., Andresen, B., Wartberg, L., Nebeling, B., & Schmoldt, A. (2006) · Journal of Psychopharmacology · doi:10.1177/0269881106059486 · Cited by 158
“Imaging was performed on a fullring whole-body system ECAT EXACT 921/47 (Siemens/CTI, Knoxville, TN, USA; Wienhard et al., 1992) in 2d-mode.”
Method(1998) · PubMed · Cited by 61
Mueller, C., Klega, A., Buchholz, H.-G., Rolke, R., Magerl, W., Schirrmacher, R., Schirrmacher, E., Birklein, F., Treede, R.-D., & Schreckenberger, M. (2010) · NeuroImage · doi:10.1016/j.neuroimage.2009.08.033 · Cited by 42
“0 mm FWHM (full width at half maximum) in 3Dmode (Wienhard et al., 1992), and an inplane resolution of 6.”
Influential
Metadata
- Authors
- Wienhard, K., Eriksson, L., Grootoonk, S., Casey, M., Pietrzyk, U., & Heiss, W.-D.
- Venue
- Journal of Computer Assisted Tomography
- Published
- 1992-09
- Volume
- 16
- Issue
- 5
- Pages
- 804-813
- ISSN
- 0363-8715 (ISSN-L)
- OpenAlex
- W1987444921
- Cited by
- 241
- References
- 0
- FWCI
- 8.59
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.