The ECAT EXACT HR: Performance of a New High Resolution Positron Scanner

Wienhard, Klaus, Dahlbom, Magnus, Eriksson, Lars, Michel, Christian, Bruckbauer, Thomas, Pietrzyk, Uwe, Heiss, Wolf-Dieter·1994·Journal of Computer Assisted Tomography

Abstract

OBJECTIVE: The ECAT EXACT HR is a newly designed CTI-Siemens PET scanner with high spatial resolution. Its physical performance with respect to resolution, count rate efficiency, and scatter was investigated and evaluated with phantom studies. MATERIALS AND METHODS: The new tomograph consists of three rings of 112 BGO block detectors (50 mm x 23 mm x 30 mm deep) each, covering an axial field of view of 15 cm with a patient port of 56 cm diameter. Each block is sawed into an 8 x 7 matrix giving 24 detector rings with 784 crystals each. RESULTS: Total sensitivity for a 20 cm cylinder phantom is 177 kcps/microCi/ml in two-dimensional (2D) mode and increases to 1.46 Mcps/microCi/ml in 3D mode. Count rate performance was investigated for different low energy discriminator thresholds. Smaller detector blocks improve noise equivalent counts by approximately 50% compared with the EXACT system both in 2D and in 3D mode. Scatter fractions vary in 2D from 0.09 to 0.13 for energy thresholds from 450 to 250 keV for line sources in a 20 cm diameter phantom. In 3D mode an increase of scatter by a factor of 3 is observed. Transaxial spatial resolution varies from 3.6 mm full width at half-maximal (FWHM) at the center to 4.5 mm FWHM tangentially and 7.4 mm FWHM radially at R = 20 cm. Average axial resolution changes from 4.0 mm FWHM at center to 6.7 mm FWHM at R = 20 cm. CONCLUSION: Due to its special properties, the EXACT HR can be equally applied to routine clinical brain and whole-body imaging and to noninvasive experimental studies of regional tracer concentrations in medium-sized animals.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00004728-199401000-00023

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 561

Metadata

Authors
Wienhard, K., Dahlbom, M., Eriksson, L., Michel, C., Bruckbauer, T., Pietrzyk, U., & Heiss, W.-D.
Venue
Journal of Computer Assisted Tomography
Published
1994-01
Volume
18
Issue
1
Pages
110-118
ISSN
0363-8715 (ISSN-L)
OpenAlex
W2089686687
Cited by
561
References
0
FWCI
24.88
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The ECAT EXACT HR: Performance of a New High Resolution Positron Scanner. https://app.cochat.ai/papers/10.1097/00004728-199401000-00023
The paper
Wienhard, K., Dahlbom, M., Eriksson, L., Michel, C., Bruckbauer, T., Pietrzyk, U., & Heiss, W.-D. (1994). The ECAT EXACT HR: Performance of a New High Resolution Positron Scanner. Journal of Computer Assisted Tomography. https://doi.org/10.1097/00004728-199401000-00023

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.