Comparison of wrist blood pressure measurement with conventional sphygmomanometry at a cardiology outpatient clinic
Abstract
OBJECTIVE: Oscillometric measurement of blood pressure at the wrist is becoming a widely used method for detection of hypertension and its control by treatment. The objective of the present study was to evaluate accuracy and suitability of wrist measurement in a clinical routine setting. PATIENTS AND METHODS: A series of 333 consecutive patients admitted to our cardiology outpatient clinic were included. Blood pressure was measured at both upper arms according to World Health Organization-International Society of Hypertension guidelines. Oscillometric measurement was performed at the contralateral wrist simultaneously. Blood pressure readings were taken by an oscillometric device applied at the wrist ('Klock'; Industrielle Entwicklung Medizintechnik, Stolberg, Germany) and a conventional mercury sphygmomanometer applied at the upper arm. RESULTS: Seventy-eight patients were excluded due to differences in blood pressure > 5 mmHg between both upper arms or due to 'error' messages of the wrist device. The data of the remaining 255 patients (149 males; mean age, 65 +/- 13 years; range, 18-95 years) are presented. Mean conventional blood pressure was significantly lower compared with the wrist device (137 +/- 20/80 +/- 11 mmHg versus 153 +/- 28/87 +/- 18 mmHg; P < 0.001 and P < 0.001). The mean difference was 16 +/- 25/6 +/- 17 mmHg. In clinical terms, differences in blood pressure exceeding +/-20/+/-10 mmHg reflecting classification of hypertension are considered important. Measurements of 101 (40%) patients were within these limits. Systolic readings of 110 (43%) and diastolic readings of 117 patients (46%) were beyond this scope. CONCLUSION: Due to low reliability of wrist blood pressure measurement, it cannot compete with the upper arm standard procedure. If ever, it should only be used if test readings in an individual comparing wrist and upper arm measurement show differences within a range of +/-20/+/-10 mmHg.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
These findings are consistent with previously described limitations of wrist-based cuff blood pressure measurements.
“These findings are consistent with previously described limitations of wrist-based cuff BP measurements26.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [26]
Cites
(1997) · Archives of Internal Medicine · doi:10.1001/archinte.157.21.2413 · Cited by 6,345 · see citing works
Subcommittee, G. (1999) · Journal of Hypertension · doi:10.1097/00004872-199917020-00001 · Cited by 3,895 · see citing works
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343 · see citing works
Ambulatory blood pressure. An independent predictor of prognosis in essential hypertension.
Verdecchia, P., Porcellati, C., Schillaci, G., Borgioni, C., Ciucci, A., Battistelli, M., Guerrieri, M., Gatteschi, C., Zampi, I., Santucci, A., et al. (1994) · Hypertension · doi:10.1161/01.hyp.24.6.793 · Cited by 1,671 · see citing works
Ohkubo, T., Imai, Y., Tsuji, I., Nagai, K., Kato, J., Kikuchi, N., Nishiyama, A., Aihara, A., Sekino, M., Kikuya, M., et al. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816070-00010 · Cited by 782 · see citing works
Brien, E. O., Petrie, J. C., Littler, W. A., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D. G., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 764 · see citing works
Ambulatory blood pressure normality: results from the PAMELA study
Mancia, G., Sega, R., Bravi, C., Vito, G. D., Valagussa, F., Cesana, G., & Zanchetti, A. (1995) · Journal of Hypertension · doi:10.1097/00004872-199512000-00003 · Cited by 619 · see citing works
Mancia, G., Zanchetti, A., Agebiti-Rosei, E., Benemio, G., De Cesaris, R., Fogari, R., Pessino, A., Porcellati, C., Salvetti, A., & Trimarco, B. (1997) · Circulation · doi:10.1161/01.cir.95.6.1464 · Cited by 596 · see citing works
O'Brien, E. T., Mee, F. I., Atkins, N., & Thomas (1996) · PubMed · Cited by 409 · see citing works
British Hypertension Society guidelines for hypertension management 1999: summary
Ramsay, L. E., Williams, B., Johnston, G., MacGregor, G. A., Poston, L., POTTER, J. F., Poulter, N. R., & Russell, G. I. (1999) · BMJ · doi:10.1136/bmj.319.7210.630 · Cited by 373 · see citing works
Prognostic Significance of the White Coat Effect
Verdecchia, P., Schillaci, G., Borgioni, C., Ciucci, A., & Porcellati, C. (1997) · Hypertension · doi:10.1161/01.hyp.29.6.1218 · Cited by 135 · see citing works
Sympathetic Neural Mechanisms of Cyclosporine-Induced Hypertension*
Sander, M., Lyson, T., Thomas, G. D., & Victor, R. G. (1996) · American Journal of Hypertension · doi:10.1016/0895-7061(96)00288-9 · Cited by 116 · see citing works
WATSON, S., WENZEL, R., DIMATTEO, C., Meier, B., & Lüscher, T. F. (1998) · American Journal of Hypertension · doi:10.1016/s0895-7061(98)00188-5 · Cited by 55 · see citing works
O’Rourke, M. F. (1990) · American Journal of Hypertension · doi:10.1093/ajh/3.10.803 · Cited by 42 · see citing works
Task force VI: Self-monitoring of the blood pressure
White, W. B., Asmar, R., Lmai, Y., et al. (1999) · Blood Pressure Monitoring · doi:10.1097/00126097-199912000-00007 · Cited by 37 · see citing works
Accuracy of a New Wrist Cuff Oscillometric Blood Pressure Device
Watson, S., Wenzel, R. R., Matteo, C. D., et al. (1998) · doi:10.7892/boris.115289 · Cited by 31 · see citing works
Factors influencing validation of ambulatory blood pressure measuring devices
Brien, E. O., Atkins, N., & Staessen, J. A. (1995) · Journal of Hypertension · doi:10.1097/00004872-199511000-00003 · Cited by 27 · see citing works
Rogers, P., Burke, V., Stroud, P., & Puddey, I. B. (1999) · Clinical and Experimental Pharmacology and Physiology · doi:10.1046/j.1440-1681.1999.03059.x · Cited by 26 · see citing works
Eckert, S., Gleichmann, U., Zagorski, O., & Klapp, A. (1997) · PubMed · Cited by 25 · see citing works
Clinical test of a blood pressure measurement device with a wrist cuff.
Wonka, F., Thümmler, M., & Schöppe, A. (1996) · PubMed · Cited by 23 · see citing works
Wrist Measurement of Blood Pressure: Some Critical Remarks to Oscillometry
Weber, F. P., Erbel, R. M., Schäfers, R., & Philipp, T. (1999) · Kidney & Blood Pressure Research · doi:10.1159/000025923 · Cited by 10 · see citing works
Saul, F., Aristidou, Y., Klaus, D., et al. (1995) · PubMed · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Flynn, J. T., Kaelber, D. C., Baker-Smith, C. M., Blowey, D., Carroll, A. E., Daniels, S. R., de Ferranti, S. D., Dionne, J. M., Falkner, B., Flinn, S. K., et al. (2017) · Pediatrics · doi:10.1542/peds.2017-1904 · Cited by 3,651 · see citing works
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
- Authors
- Zweiker, R., Schumacher, M., Fruhwald, F. M., Watzinger, N., & Klein, W.
- Affiliations
- University of Graz (AT)
- Corresponding author
- Zweiker, Robert
- ORCID
- Zweiker, Robert
- Venue
- Journal of Hypertension
- Published
- 2000-08
- Volume
- 18
- Issue
- 8
- Pages
- 1013-1018
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2001639725
- Cited by
- 29
- References
- 22
- Citations by year
- 2026: 1 · 2025: 1 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 1 · 2018: 1 · 2017: 1 · 2013: 1
- FWCI
- 0.74
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- oscillometric blood pressure, hypertension diagnosis, measurement agreement, hypertension classification, cardiology outpatient clinic
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Blood Pressure, Blood Pressure Determination, Female, Humans, Male, Middle Aged, Nurses, Outpatient Clinics, Hospital, Wrist, Cardiology Service, Hospital
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review