Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
OBJECT: Self-measurement of blood pressure (BP) might offer some advantages in diagnosis and therapeutic evaluation and in patient management of hypertension. Recently, wrist-cuff devices for self-measurement of BP have gained more than one-third of the world market share. In the present study, we validated wrist-cuff devices and compared the results between wrist- and arm-cuff devices. The factors affecting the accuracy of wrist-cuff devices were also studied. METHOD: The research group to assess the validity of automated blood pressure measuring device consisted of 13 institutes in Japan, which validated two wrist-cuff devices (WC-1 and WC-2) and two arm-cuff devices (AC-1 and AC-2). They used a crossover method, where the comparison was done between auscultation, by two observers by means of a double stethoscope on one arm and the device on the opposite arm or wrist. RESULTS: There was good inter-observer agreement for the auscultation method in each institute (systolic blood pressure (SBP), -0.1 +/- 2.8 mmHg; diastolic blood pressure (DBP), -0.1 +/- 2.6 mmHg, n = 498). The mean difference between auscultation and the device was minimal both in arm-cuff devices (mean difference for AC-1, 2.2/1.9 mmHg, n = 97 and for AC-2, 5.1/2.9 mmHg, n = 136, SBP/DBP) and wrist-cuff devices (mean difference for WC-1, -2.1/1.2 mmHg, n = 173 mmHg and for WC-2, -2.3/-5.6 mmHg, n = 92). The standard deviation of the difference (SDD) in wrist-cuff devices, however (SDD for WC-1, 9.7/7.3 mmHg and for WC-2, 10.2/8.6 mmHg), was larger than that of the arm-cuff devices (SDD for AC-1, 5.6/6.6 mmHg and for AC-2, 6.3/5.1 mmHg). Grading of AC-1 and AC-2 based on criteria of British Hypertension Society was A/A and B/A, respectively, while that of WC-1 and WC-2 was C/B and D/B, respectively. Using the same validation protocol, the results of validation for one device were divergent in each institute. In wrist-cuff devices, the BP value obtained in palmar flexion was significantly higher and that obtained in palmar dorsiflexion was significantly lower than that in palmar extension. In some cases, finger plethysmogram did not disappear during maximum inflation of the wrist-cuff (congruent with 250 mmHg), even in palmar extension and especially in palmar flexion, suggesting incomplete obstruction of radial and/or ulnar arteries during inflation. CONCLUSION: The results suggest that wrist-cuff devices in the present form are inadequate for self-measurement of blood pressure and, thus, are inadequate for general use or clinical and practical use. However, there is much possibility in wrist-cuff device and the accuracy and reliability of wrist-cuff device are warranted by an improvement of technology.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Flynn, J. T., Kaelber, D. C., Baker-Smith, C. M., Blowey, D., Carroll, A. E., Daniels, S. R., de Ferranti, S. D., Dionne, J. M., Falkner, B., Flinn, S. K., et al. (2017) · Pediatrics · doi:10.1542/peds.2017-1904 · Cited by 3,651
The Japanese Society of Hypertension Guidelines for the Management of Hypertension (JSH 2019)
Umemura, S., Arima, H., Arima, S., Asayama, K., Dohi, Y., Hirooka, Y., Horio, T., Hoshide, S., Ikeda, S., Ishimitsu, T., et al. (2019) · Hypertension Research · doi:10.1038/s41440-019-0284-9 · Cited by 1,880
Uhlig, K., Patel, K., Ip, S., Kitsios, G. D., & Balk, E. M. (2013) · Annals of Internal Medicine · doi:10.7326/0003-4819-159-3-201308060-00008 · Cited by 368
Jananese Society of Hypertension (JSH) Guidelines for Self-Monitoring of Blood Pressure at Home
IMAI, Y., OTSUKA, K., KAWANO, Y., SHIMADA, K., HAYASHI, H., TOCHIKUBO, O., MIYAKAWA, M., & FUKIYAMA, K. (2003) · Hypertension Research · doi:10.1291/hypres.26.771 · Cited by 309
Self-citationKuwabara, M., Harada, K., Hishiki, Y., & Kario, K. (2019) · The Journal of Clinical Hypertension · doi:10.1111/jch.13499 · Cited by 140
Metadata
- Authors
- Kikuya, M., Chonan, K., Imai, Y., Goto, E., & Ishii, M.
- Venue
- Journal of Hypertension
- Published
- 2002-04
- Volume
- 20
- Issue
- 4
- Pages
- 629-638
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2010020249
- Cited by
- 91
- References
- 1
- FWCI
- 3.11
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure. https://app.cochat.ai/papers/10.1097/00004872-200204000-00019
- The paper
- Kikuya, M., Chonan, K., Imai, Y., Goto, E., & Ishii, M. (2002). Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure. Journal of Hypertension. https://doi.org/10.1097/00004872-200204000-00019
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.