CoChat

Assessment of blood lactate: practical evaluation of the Biosen 5030 lactate analyzer

DAVISON, R. C. RICHARD, COLEMAN, DAMIAN, BALMER, JAMES, NUNN, MAXWELL, THEAKSTON, SIMON, BURROWS, MELONIE, BIRD, STEVE · 2000 · Medicine & Science in Sports & Exercise
Cited by 43

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

PURPOSE: The aim of this study was to assess the validity and reliability of the Biosen 5030 lactate analyzer compared with a YSI 2300 lactate analyzer and a Kodak Ektachem DTII in a practical laboratory study context. METHODS: To assess validity, 144 triplicate capillarized blood samples, across a range of values, were analyzed using the three analyzers. To assess reliability a further 665 samples were repeat analyzed. Temporal stability was determined by the reanalysis of resting and maximal exercise blood samples, after a period of storage ranging from 7 to 20 h, at room temperature. To measure inter- and intra-investigator reliability, 20 resting samples were taken from three different subjects by different investigators and a coefficient of variation was determined. RESULTS: There were strong relationships between the Biosen, the YSI (r2 = 0.97), and the Kodak Ektachem (r2 = 0.91). An analysis of Biosen compared with YSI revealed a positive bias of 0.37 mmol x L(-1) (95% limits of agreement, -0.85 to 1.59 mmol x L(-1)). The test-retest reliability correlation was significant (r2 = 0.99, P < 0.05), but a paired t-test revealed a small (0.03 mmol x L(-1), P < 0.05) significant difference. The coefficient of variation from the three investigators across the 20 samples ranged from 1.3 to 3%. Blood lactate concentration in resting blood samples did significantly increase in value (0.2 mmol x L(-1), P < 0.05) after 7-h exposure to the air, whereas there was no change in maximal exercise blood lactate values after 20-h exposure to the air. CONCLUSIONS: In a practical context, the Biosen 5030 lactate analyzer was comparable to the other analyzers giving fast reliable measures of blood lactate concentrations over the full range of values, which remained stable over extended periods at room temperature.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00005768-200001000-00036

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 43

Metadata

Authors
DAVISON, R. C. R., COLEMAN, D., BALMER, J., NUNN, M., THEAKSTON, S., BURROWS, M., & BIRD, S.
Venue
Medicine & Science in Sports & Exercise
Published
2000-01
Volume
32
Issue
1
Pages
243
ISSN
0195-9131 (ISSN-L)
OpenAlex
W2021738837
Cited by
43
References
2
FWCI
0.76
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Assessment of blood lactate: practical evaluation of the Biosen 5030 lactate analyzer. https://app.cochat.ai/papers/10.1097/00005768-200001000-00036
The paper
DAVISON, R. C. R., COLEMAN, D., BALMER, J., NUNN, M., THEAKSTON, S., BURROWS, M., & BIRD, S. (2000). Assessment of blood lactate: practical evaluation of the Biosen 5030 lactate analyzer. Medicine & Science in Sports & Exercise. https://doi.org/10.1097/00005768-200001000-00036

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.