Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C
The DOI matches its Crossref record.
Abstract
BACKGROUND: The validation of self-measurement devices for clinical use by elderly patients has been recommended. The Omron HEM-722C device has recently been validated according to the British Hypertension Society (BHS) protocol for use for general populations and the Omron HEM-735C is a new fully automatic device with a high capacity for storage of measurements that is integrated with a personal computer. OBJECTIVE: To perform a clinical validation for use by elderly people of the Omron HEM-722C and HEM-735C devices according to the revised protocol of the BHS and the criteria of the Association for the Advancement of Medical Instrumentation (AAMI). METHODS: We carried out a main validation test according to the revised BHS protocol for validation procedures for special groups on two groups of 30 subjects aged more than 65 years (29 men and 31 women), 11 of 30 with systolic blood pressures (SBP) <110mmHg, 10 of 30 with SBP >200 mmHg, 15 of 30 with diastolic blood pressures (DBP( <70 mmHg and 10 of 30 with DBP >110 mmHg. The results were graded according to the BHS system from A to D. RESULTS: The Omron HEM 722C achieved an overall A/A grading and satisfied the AAMI criteria for accuracy whereas the Omron HEM-735C achieved an overall B/A grading and satisfied the AAMI criteria for accuracy. The sphygmomanometer measurements were 147 +/- 31/79 +/- 15 and 144 +/- 30/78 +/- 15 mmHg (means+/-SD) respectively, for the models 722C and 735C. The average differences between mercury sphygmomanometer and HEM-722C readings for SBP and DBP were, respectively, 0.76+/-5 and 0.41+/-8 mmHg; those for HEM-735C were, respectively, 0.24+/-8 and 0.9+/-8 mmHg. Readings of the HEM-722C device differed by less than 5 mmHg for 76% of systolic readings and 96% of the readings differed by less than 10 mmHg. Diastolic measurements differed by less than 5 mmHg for 71% and less than 10 mmHg for 71 and 87% of all readings. Readings of the HEM-735C device differed by less than 5 mmHg for 68% of systolic readings and 74% of the readings differed by less than 10 mmHg. Diastolic measurements differed by less than 5 mmHg in 74% and less than 10 mmHg in 88 and 87% of all readings. CONCLUSIONS: On the basis of these results, for elderly subjects both self-measurement devices (Omron HEM-722C and HEM-735C) satisfied the validation criteria of the BHS and therefore can be recommended for the clinical measurement of blood pressure in elderly patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 903
Taubert, D., Roesen, R., Lehmann, C., Jung, N., & Schömig, E. (2007) · JAMA · doi:10.1001/jama.298.1.49 · Cited by 476
Agarwal, R., Bills, J. E., Hecht, T. J. W., & Light, R. P. (2011) · Hypertension · doi:10.1161/hypertensionaha.110.160911 · Cited by 430
Niiranen, T. J., Hänninen, M.-R., Johansson, J., Reunanen, A., & Jula, A. M. (2010) · Hypertension · doi:10.1161/hypertensionaha.109.149336 · Cited by 390
Johansson, J. K., Niiranen, T. J., Puukka, P. J., & Jula, A. M. (2012) · Hypertension · doi:10.1161/hypertensionaha.111.178657 · Cited by 240
Metadata
- Authors
- Bortolotto, L. A., Henry, O., Hanon, O., Sikias, P., Mourad, J.-J., & Girerd, X.
- Venue
- Blood Pressure Monitoring
- Published
- 1999
- Volume
- 4
- Issue
- 1
- Pages
- 21-25
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W1996834120
- Cited by
- 73
- References
- 0
- FWCI
- 2.94
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. https://app.cochat.ai/papers/10.1097/00126097-199904010-00004
- The paper
- Bortolotto, L. A., Henry, O., Hanon, O., Sikias, P., Mourad, J.-J., & Girerd, X. (1999). Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-199904010-00004
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.