Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults

White, William B., Anwar, Yusra Anis·2001·Blood Pressure Monitoring

Abstract

BACKGROUND: Non-invasive self blood pressure monitoring has become increasingly popular. To assure the accuracy of devices used for this purpose, all need to be validated independently prior to marketing. The objective of this study was to assess the accuracy of the HEM-907, a new semi-automatic, non-invasive, oscillometric blood pressure monitoring device specifically designed to be used in the clinic or physician's office setting. METHODS: Blood pressure measurements taken employing this device were compared with the results obtained by two experienced observers using a mercury sphygmomanometer on 100 subjects and patients (384 measurements). The limits of agreement were calculated for the device compared with the results of the two observers according to the standards of the Association for the Advancement of Medical Instrumentation (AAMI). RESULTS: The agreement between the two observers was -0.36+/-2.32mmHg for systolic blood pressure and 0.02+/-2.42mmHg for diastolic blood pressure. The agreement between the device and the observers was 1.56+/-4.42mmHg and 3.49+/-4.61mmHg for systolic and diastolic blood pressure respectively. CONCLUSIONS: The Omron HEM-907 satisfied the AAMI criteria for accuracy for a non-invasive blood pressure monitoring device.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200104000-00007

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 213

Metadata

Authors
White, W. B., & Anwar, Y. A.
Venue
Blood Pressure Monitoring
Published
2001-04
Volume
6
Issue
2
Pages
107-110
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2047667671
Cited by
213
References
2
FWCI
1.36
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults. https://app.cochat.ai/papers/10.1097/00126097-200104000-00007
The paper
White, W. B., & Anwar, Y. A. (2001). Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200104000-00007

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.