Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometer

Mattu, Gurdial S., Perry, Thomas L., Wright, James M. · 2001 · Blood Pressure Monitoring
Cited by 82

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

BACKGROUND: To compare directly the accuracy of the BPM-100(Beta) monitor (an automated oscillometric blood pressure device) with standard auscultatory mercury sphygmomanometry. DESIGN: The BPM-100(Beta) was connected in parallel via a T-tube to a mercury sphygmomanometer. The BPM-100(Beta) and two trained observers (blinded from each other and from the BPM-100(Beta)) measured the sitting blood pressure simultaneously. METHODS: Means, standard deviations and ranges were calculated for all the demographic data: age, arm size, heart rate and blood pressure. The agreement between the BPM-100(Beta) and the mean of two observers (the reference) was determined and expressed as the mean +/- SD, as well as the percentage of differences falling within 5, 10 and 15 mmHg. RESULTS: Of the 92 subjects recruited, 85 (92.4%) met the inclusion criteria, and 391 sets of sitting blood pressure and heart rate measurements were available for analysis. The mean difference between the BPM-100(Beta) monitor and the reference was -0.62 +/- 6.96 mmHg for systolic blood pressure, -1.48 +/- 4.80 mmHg for diastolic blood pressure and 0.14 +/- 1.86 beats/min for heart rate. The only limitation of the device was its tendency to underestimate higher systolic blood pressures. This problem has been addressed by a minor change in the algorithm (see the companion publication, Blood Press Monit, 6, 161-165, 2001). CONCLUSION: The BPM-100(Beta) is an accurate blood pressure monitor for the office setting, meeting all requirements of the Association for the Advancement of Medical Instrumentation and achieving an 'A' grade according to the British Hypertension Society protocol.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200106000-00007

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 82

Metadata

Authors
Mattu, G. S., Perry, T. L., & Wright, J. M.
Venue
Blood Pressure Monitoring
Published
2001-06
Volume
6
Issue
3
Pages
153-159
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2047906635
Cited by
82
References
2
FWCI
2.04
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometer. https://app.cochat.ai/papers/10.1097/00126097-200106000-00007
The paper
Mattu, G. S., Perry, T. L., & Wright, J. M. (2001). Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometer. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200106000-00007

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.