Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT

El Assaad, Mohamed A., Topouchian, Jirar A., Asmar, Roland G. · 2003 · Blood Pressure Monitoring

Abstract

Background Two devices for self-measurement of blood pressure at the brachial artery—the Omron M5-I and the Omron 705IT—were evaluated according to the international protocol of the European Society of Hypertension. Design The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 supplementary subjects are included making a total number of 33 subjects on which the final validation is performed. Methods For each subject, four blood pressure (BP) measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with three measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 differences were classified into categories (≤5, ≤10, ≤15 mmHg). The number of differences in each category was compared to the number required by the international protocol. An individual analysis was then done to determine for each subject the number of comparisons ≤5 mmHg. At least 22 of the 33 subjects should have two of their three comparisons ≤5 mmHg. Results The two tested devices passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were −0.9±5.8 and −0.8±4.8 mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP) respectively for the Omron M5-I device and −0.2±4.5 and −2.0±4.8 mmHg for the Omron 705IT device. Readings for the two devices differing by less than 5, 10 and 15 mmHg for systolic and diastolic values fulfil the recommendation criteria of the international protocol as well as the individual analysis. Conclusions The Omron M-5I and the Omron 705IT devices pass the validation recommendations of the international protocol.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200306000-00006

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 224

Metadata

Authors
El Assaad, M. A., Topouchian, J. A., & Asmar, R. G.
Venue
Blood Pressure Monitoring
Published
2003-06
Volume
8
Issue
3
Pages
127-133
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2099393780
Cited by
224
References
2
FWCI
4.82
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. https://app.cochat.ai/papers/10.1097/00126097-200306000-00006
The paper
El Assaad, M. A., Topouchian, J. A., & Asmar, R. G. (2003). Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200306000-00006

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.