Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor

Wright, James M., Mattu, Gurdial S., Perry, Thomas L., Gelferc, Mark E., Strange, Kevin D., Zorn, Anton, Chen, Yunquan · 2001 · Blood Pressure Monitoring
Cited by 115

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

BACKGROUND: To test the accuracy of a new algorithm for the BPM-100, an automated oscillometric blood pressure (BP) monitor, using stored data from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer. DESIGN: Raw pulse wave and cuff pressure data were stored electronically using embedded software in the BPM-100(Beta), during the validation trial. The 391 sets of measurements were separated objectively into two subsets. A subset of 136 measurements was used to develop a new algorithm to enhance the accuracy of the device when reading higher systolic pressures. The larger subset of 255 measurements (three readings for 85 subjects) was used as test data to validate the accuracy of the new algorithm. METHODS: Differences between the new algorithm BPM-100 and the reference (mean of two observers) were determined and expressed as the mean difference +/- SD, plus the percentage of measurements within 5, 10, and 15 mmHg. RESULTS: The mean difference between the BPM-100 and reference systolic BP was -0.16 +/- 5.13 mmHg, with 73.7% < or = 5 mmHg, 94.9% < or = 10 mmHg and 98.8% < or = 15 mmHg. The mean difference between the BPM-100 and reference diastolic BP was -1.41 +/- 4.67 mmHg, with 78.4% < or = 5 mmHg, 92.5% < or = 10 mmHg, and 99.2% < or = 15 mmHg. These data improve upon that of the BPM-100(Beta) and pass the AAMI standard, and 'A' grade BHS protocol. CONCLUSION: This study illustrates a new method for developing and testing a change in an algorithm for an oscillometric BP monitor utilizing collected and stored electronic data and demonstrates that the new algorithm meets the AAMI standard and BHS protocol.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200106000-00008

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 115

Metadata

Authors
Wright, J. M., Mattu, G. S., Perry, T. L., Gelferc, M. E., Strange, K. D., Zorn, A., & Chen, Y.
Venue
Blood Pressure Monitoring
Published
2001-06
Volume
6
Issue
3
Pages
161-165
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2162576402
Cited by
115
References
1
FWCI
2.27
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. https://app.cochat.ai/papers/10.1097/00126097-200106000-00008
The paper
Wright, J. M., Mattu, G. S., Perry, T. L., Gelferc, M. E., Strange, K. D., Zorn, A., & Chen, Y. (2001). Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200106000-00008

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.