Validation of the Omron HEM-907 device for blood pressure measurement
Abstract
BACKGROUND: The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN: The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed. METHODS: For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. RESULTS: The difference between the two observers was -1 +/- 2 mmHg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were -1 +/- 7 and -5 +/- 6 mmHg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmHg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmHg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmHg for 94 of the systolic readings and 96 of the diastolic readings. CONCLUSIONS: The Omron HEM-907 device passes the two phases of the international validation protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
O??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
White, W. B., Berson, A. S., Robbins, C., Jamieson, M. J., Prisant, L. M., Roccella, E., & Sheps, S. G. (1993) · Hypertension · doi:10.1161/01.hyp.21.4.504 · Cited by 385
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
White, W. B., & Anwar, Y. A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200104000-00007 · Cited by 213
Brien, E. O., & Atkins, N. (1994) · Journal of Hypertension · doi:10.1097/00004872-199409000-00015 · Cited by 51
Proceedings from a Consensus Conference on Self-Blood Pressure Measurement. Devices and validation.
O’Brien, E., Gaudemaris, R. D., Bobrie, G., Rosei, E. A., & Vaisse, B. (2000) · PubMed · Cited by 33
O'Brien, E. T. (2000) · Blood Pressure Monitoring · doi:10.1097/00126097-200002000-00010 · Cited by 23
Cited by
Firmann, M., Mayor, V., Vidal, P. M., Bochud, M., Pécoud, A., Hayoz, D., Paccaud, F., Preisig, M., Song, K. S., Yuan, X., et al. (2008) · BMC Cardiovascular Disorders · doi:10.1186/1471-2261-8-6 · Cited by 721
Kallioinen, N., Hill, A., Horswill, M. S., Ward, H. E., & Watson, M. O. (2017) · Journal of Hypertension · doi:10.1097/hjh.0000000000001197 · Cited by 357
Low socioeconomic status is a risk factor for preeclampsia: the Generation R Study
Silva, L. M., Coolman, M., Steegers, E. A., Jaddoe, V. W., Moll, H. A., Hofman, A., Mackenbach, J. P., & Raat, H. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e3282fcc36e · Cited by 250
Agarwal, R. (2017) · Journal of the American Heart Association · doi:10.1161/jaha.116.004536 · Cited by 144
Guessous, I., Dobrinas, M., Kutalik, Z., Pruijm, M., Ehret, G., Maillard, M., Bergmann, S., Beckmann, J. S., Cusi, D., Rizzi, F., et al. (2012) · Human Molecular Genetics · doi:10.1093/hmg/dds137 · Cited by 70
Metadata
- Authors
- El Assaad, M. A., Topouchian, J. A., Darné, B. M., & Asmar, R. G.
- Venue
- Blood Pressure Monitoring
- Published
- 2002-08
- Volume
- 7
- Issue
- 4
- Pages
- 237-241
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W1775135678
- Cited by
- 279
- References
- 8
- FWCI
- 1.87
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of the Omron HEM-907 device for blood pressure measurement. https://app.cochat.ai/papers/10.1097/00126097-200208000-00006
- The paper
- El Assaad, M. A., Topouchian, J. A., Darné, B. M., & Asmar, R. G. (2002). Validation of the Omron HEM-907 device for blood pressure measurement. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200208000-00006
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.